The SureClip Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult patients only via a straight or side viewing flexible endoscope for the purpose of : (1) Endoscopic marking; (2) Hemostasis for (a) Mucosal / sub-mucosal defects < 3cm, (b) Bleeding ulcers, (c) Polyps < 1.5cm in diameter, (d) Diverticula in the colon, (e) Arteries < 2 mm (f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection; (3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively.
Device Story
SureClip Repositionable Hemostasis Clip is a sterile, single-use endoscopic device; consists of delivery system and clip assembly. Used by physicians during endoscopic procedures via straight or side-viewing flexible endoscopes. Device allows for clip placement to mark tissue, achieve hemostasis in bleeding sites or defects, and close small GI perforations. Repositionable design allows clinician to open/close clip multiple times before final deployment. Output is physical mechanical closure of tissue. Benefits include effective bleeding control and defect closure, reducing need for more invasive interventions.
Single-use, sterile, ethylene oxide (EO) sterilized. Comprises delivery system and clip assembly. Available in 8mm, 11mm, 16mm open widths; 1650mm and 2350mm working lengths. Compatible with 2.8mm minimum working channel endoscopes. Mechanical actuation principle. Complies with ISO 10993-1 (biocompatibility), ISO 11135 (EO sterilization), and ISO 10993-7 (EO residuals).
Indications for Use
Indicated for adult patients requiring endoscopic clip placement in the GI tract for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, polyps <1.5cm, colonic diverticula, arteries <2mm, post-resection prophylaxis), or closure of GI luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2018
Micro-Tech (Nanjing) Co., Ltd. Becky Li Quality and Regulatory Affairs Director NO. 10 Gaoke Third Road Nanjing, Jiangsu 210032 China
Re: K182556
> Trade/Device Name: SureClip Repositionable Hemostasis Clip Regulation Number: 21 CFR§ 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: II Product Code: PKL Dated: September 13, 2018 Received: September 17, 2018
Dear Becky Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel G. Walter Jr -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K182556
#### Device Name
SureClip Repositionable Hemostasis Clip
Indications for Use (Describe)
The SureClip Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult patients only via a straight or side viewing flexible endoscope for the purpose of :
(1) Endoscopic marking;
(2) Hemostasis for
(a) Mucosal / sub-mucosal defects < 3cm,
- (b) Bleeding ulcers,
(c) Polyps < 1.5cm in diameter,
(d) Diverticula in the colon,
(e) Arteries < 2 mm
(f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection;
(3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional font. The 'M' is larger and positioned slightly behind the 'T', creating a sense of depth. To the right of the logo, there is some text, but it is too blurry to read.
510K Summary
# 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K182556
# 1. Date of Preparation: 2018-12-05
# 2. Sponsor Identification
# Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone,
Nanjing, Jiangsu Province, PRC
# Establishment Registration Number: 3004837686
Contact Person: Becky Li
Position: Quality and Regulatory Affairs Director
Tel: +86-25-58646378
Fax: +86-25-58744269
Email: Beckyli mt@outlook.com
# 3. Identification of Proposed Device
Trade Name: SureClip™ Repositionable Hemostasis Clip
Common Name: Hemostasis Clipping Device
# Regulatory Information
Classification Name: Hemorrhoidal Ligator Classification: II Product Code: PKL Requlation Number: 876.4400 Review Panel: Gastroenterology/Urology
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for MICRO-TECH (Nanjing) Co., Ltd. The logo consists of the company name in English, followed by the company name in Chinese characters. The English name is in a blue sans-serif font, and the Chinese name is in a similar font and color. There is a blue graphic to the left of the company name.
510K Summary
#### 4. Identification of Predicate Device
#### Predicate Device
510(k) Number: K151802
Product Name: Resolution 360™ Clip
Manufacturer: Boston Scientific Corporation
### Reference Device
510(k) Number: K152001
Product Name: Sterile Repositionable Hemostasis Clipping Device
Manufacturer: Micro-Tech (Nanjing) Co., Ltd.
#### 5. Indications for Use
The SureClip™ Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult patients only via a straight or side viewing flexible endoscope for the purpose of :
- (1) Endoscopic marking;
- (2) Hemostasis for
- (a) Mucosal / sub-mucosal defects < 3cm,
- (b) Bleeding ulcers,
- (c) Polyps < 1.5cm in diameter,
- (d) Diverticula in the colon,
- (e) Arteries < 2 mm,
(f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion
resection;
(3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively.
#### 6. Device Description
The proposed device SureClip™ Repositionable Hemostasis Clip is a sterile,
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a blue and white logo with the letters "MT" stacked on top of each other. The letters are in a bold, sans-serif font and have a 3D effect. The "M" is on top and the "T" is on the bottom. The logo is simple and modern.
510K Summary
single-use endoscopic clipping device in adult patients only via a straight or side viewing flexible endoscope, intended to be used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract. It consists of two main components, delivery system and clip assembly. And it is offered in different dimensions.
# 7. Comparison of Technological Characteristics
The SureClip™ Repositionable Hemostasis Clip incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device Resolution 360™ Clip.
### Comparison to predicate Devices:
| Item | Proposed Device<br>SureClip™ Repositionable<br>Hemostasis Clip | Predicate Device<br>Resolution 360™ Clip | Substantial<br>Equivalence |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Product Code | PKL | PKL | Same |
| Regulation No. | 876.4400 | 876.4400 | Same |
| Class | II | II | Same |
| Supplied in<br>Sterile | Yes | Yes | Same |
| Configuration | Delivery system and clip<br>assembly | Delivery system and clip assembly | Same |
| Open width | 8mm, 11mm and 16mm | 11mm | Similar |
| Minimal<br>working<br>channel | 2.8mm | 2.8mm | Same |
| Working<br>Length | 1650mm, 2350mm | 2350mm | Similar |
| Indications for<br>Use | The SureClip™ Repositionable<br>Hemostasis Clip is indicated<br>for Endoscopic clip placement<br>within the Gastrointestinal tract<br>in adult patients only via a<br>straight or side viewing flexible | The Resolution 360™ Clip is indicated for<br>clip placement within the gastrointestinal<br>(GI) tract for the purpose of:<br>1. Endoscopic marking<br>2. Hemostasis for<br>● Mucosal/sub-mucosal defects < 3 | Similar |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a blue logo with the letters 'M' and 'T' stacked on top of each other. The letters are in a bold, sans-serif font and have a 3D effect, with the left and top sides of the letters appearing to be slightly raised. The logo is simple and modern, and the blue color gives it a professional and trustworthy feel.
| Item | Proposed Device<br>SureClip™ Repositionable<br>Hemostasis Clip | 510K Summary<br>Predicate Device<br>Resolution 360™ Clip | Substantial<br>Equivalence |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| | endoscope for the purpose of :<br>(1) Endoscopic marking;<br>(2) Hemostasis for<br>(a) mucosal / sub-mucosal<br>defects < 3cm,<br>(b) bleeding ulcers,<br>(c) polyps < 1.5cm in diameter,<br>(d) diverticula in the colon,<br>(e) Arteries < 2 mm,<br>(f) Prophylactic clipping to<br>reduce the risk of delayed<br>bleeding post lesion resection;<br>(3) As a supplementary<br>method, closure of GI tract<br>luminal perforations <20mm<br>that can be treated<br>conservatively. | cm<br>● Bleeding ulcers<br>● Arteries < 2 mm`<br>● Polyps < 1.5 cm in diameter<br>● Diverticula in the colon<br>● Prophylactic clipping to reduce the<br>risk of delayed bleeding post<br>lesion resection<br>3. Anchoring to affix jejunal feeding tubes<br>to the wall of the small bowel; and<br>Anchoring to affix fully covered<br>esophageal self-expanding metal<br>stents to the wall of the esophagus<br>4. As a supplemental closure method of<br>luminal perforations < 20 mm that can<br>be treated conservatively | |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO sterilized<br>pouch with one device per<br>pouch | Single-use EO sterilized pouch with one<br>device per pouch | Same |
| Shelf Life | Three years | Three years | Same |
# 8. Performance Data
The proposed device the SureClip™ Repositionable Hemostasis Clip meets the
requirements of ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1:
Evaluation and Testing Within A Risk Management Process", ISO 11135
"Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for
Development, Validation, and Routine Control of Sterilization processes for
Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part
7: ethylene oxide sterilization residuals",
The following bench tests were performed on the SureClip™ Repositionable
### Hemostasis Clip
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional design. The 'M' is formed by two angled lines that meet at the top, and the 'T' is a simple vertical line. The logo has a blocky, geometric appearance, and the blue color is consistent throughout the design.
- > Dimension;
- Release Force; >
- > Clamping Strength;
- > Mechanical integrity Of Clip Assembly;
- > Tensile Strength;
- Clip Assembly Repeated Open/Close; >
- > Clip Open And Close Force
- > Rotation;
- > Scope Compatibility/Usability;
- > Endoscope Damage;
- > Torque;
- > Biopsy Valve Compatibility;
- > Clip Approach;
- > Coil to Handle Tensile
The testing performed demonstrated that the proposed device and predicate device are equivalent.
### 9. Clinical Test Conclusion
No clinical study is included in this submission.
#### 10. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the SureClip™ Repositionable Hemostasis Clip has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Resolution 360™ Clip (K151802).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.