← Product Code [ONJ](/submissions/GU/subpart-e%E2%80%94surgical-devices/ONJ) · K090430

# BACKSTOP AND BACKSTOP INJECTOR (K090430)

_Pluromed, Inc. · ONJ · Jun 2, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/ONJ/K090430

## Device Facts

- **Applicant:** Pluromed, Inc.
- **Product Code:** [ONJ](/submissions/GU/subpart-e%E2%80%94surgical-devices/ONJ.md)
- **Decision Date:** Jun 2, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4680
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

BackStop is intended for use in the ureter to prevent migration of renal calculi during intracorporeal lithotrypsy procedures.

## Device Story

BackStop is a self-forming, reverse thermosensitive polymeric plug used during intracorporeal lithotripsy. The device is supplied as a viscous liquid at room temperature and is manually injected into the ureter via an accessory injector. Upon contact with body temperature, the material increases in viscosity to form a mechanical plug that occludes the ureter, preventing renal calculi or fragments from migrating (retropulsion) during stone fragmentation. The plug is water-soluble and is removed by irrigation with saline at the conclusion of the procedure. Used in clinical settings by urologists, the device provides a temporary barrier to improve stone management efficiency. It benefits patients by reducing the risk of stone migration, potentially minimizing the need for additional procedures to retrieve escaped fragments.

## Clinical Evidence

Prospective, randomized, single-blind, multi-site clinical study (n=68). Subjects with single proximal ureteral stone randomized to BackStop (n=34) or control (n=34). Primary endpoint: retropulsion rate. Results: BackStop group showed significantly lower retropulsion rate (8.8%, n=3) vs. control (52.9%, n=18) (p=0.0002). No significant difference in stone-free rates or need for additional procedures. No adverse events or post-lithotripsy ureteral occlusion reported in the BackStop group.

## Technological Characteristics

Rapid transition polymer (reverse thermosensitive) and saline. Form factor: injectable viscous liquid that transitions to a gel plug in situ. Mechanism: mechanical occlusion via ureteral conformability. Removal: saline dissolution. Single-use, sterile.

## Regulatory Identification

A ureteral stone dislodger is a device that consists of a bougie or a catheter with an expandable wire basket near the tip, a special flexible tip, or other special construction. It is inserted through a cystoscope and used to entrap and remove stones from the ureter. This generic type of device includes the metal basket and the flexible ureteral stone dislodger.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Cook Urological N-Trap™ ([K863081](/device/K863081.md))
- Boston Scientific Stone Cone™ ([K970121](/device/K970121.md))
- Boston Scientific Temporary Indwelling Ureteral Catheter / Ureteral Stent ([K013784](/device/K013784.md))
- Boston Scientific Temp-Tip Drainage Catheter ([K924608](/device/K924608.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary

# JUN - 2 2009

### Safety and Effectiveness Data Summary

Prepared By:

Telephone Number: Fax Number:

Contact Person:

Proprietary Name:

Classification Name: Common Name:

Classification: Regulation Number: Product Code:

Performance Standards:

Predicates:

Description of Device:

Pluromed, Inc. 25-H Olympia Avenue Woburn; MA 01801 781.932.0574 419.828.6350

James Wilkie

BackStop™ and BackStop™ Injector (accessory to BackStop) Ureteral Stone Dislodger Urological stone entrapment device

Class II 21 CFR 876.4680 FFL

No Performance Standards have been established for the device described in this Premarket Notification under section 514 of the FD&C Act.

- 1. Cook Urological N-Trap™ (K863081)
- 2. Boston Scientific Stone Cone™ (K970121)
- 3. Boston Scientific Temporary Indwelling Ureteral Catheter / Ureteral Stent (K013784)
- 4. Boston Scientific Temp-Tip Drainage Catheter (K924608)

BackStop™ is a self-forming polymeric plug with reverse thermosensitive properties. At room temperature it is viscous but injectable. When the temperature of the plug increases , its viscosity increases and it forms a plug that is intended for use in the ureter to prevent migration of renal calculi during intracorporeal lithotripsy procedures. The material is water soluble and is removed by irrigating the plug with saline.

Substantial Equivalence:

Below is a comparison of predicate devices to BackStop™ .

{1}------------------------------------------------

|                | Pluromed<br>BackStop<br>current 510k                                                                                                          | Cook Urological<br>N-Trap<br>510(k)<br>K863081                                        | Boston Scientific<br>Stone Cone<br>510(k)<br>K970121                                | Boston Scientific<br>Temporary Indwelling<br>Ureteral Stent<br>510(k) K013784     | Boston Scientific<br>Temp-Tip drainage<br>catheter<br>510(k) K924608 |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Intended Use   | BackStop is intended<br>for use in the ureter to<br>prevent migration of<br>renal calculi during<br>intracorporeal<br>lithotripsy procedures. | To prevent retrograde<br>migration of stone<br>fragments during<br>lithotripsy        | To entrap stone<br>fragments during<br>lithotripsy                                  | To facilitate the<br>passage of urine from<br>the kidney to the<br>bladder.       | To facilitate drainage from<br>the kidney to the bladder.            |
| Mode of Action | Mechanically occludes<br>the ureter. Stones or<br>fragments are trapped<br>by the plug.                                                       | Mechanically occludes<br>the ureter. Stones or<br>fragments are trapped<br>in basket. | Mechanically occludes<br>the ureter. Stones or<br>fragments are trapped<br>in cone. | Facilitates ureteral<br>drainage                                                  | Facilitates ureteral<br>drainage                                     |
| Removal        | Dissolved with saline,<br>excreted                                                                                                            | Manually removed                                                                      | Manually removed                                                                    | Dissolution and<br>excretion in urine                                             | Dissolution of PVA distal<br>tip and excretion in urine              |
| Application    | Manual insertion into<br>the ureter                                                                                                           | Manual insertion into<br>ureter                                                       | Manual insertion into<br>ureter                                                     | Manual insertion into<br>the ureter                                               | Manual insertion into the<br>ureter                                  |
| Materials      | Rapid Transition<br>Polymer, saline                                                                                                           | Nitinol                                                                               | Nitinol                                                                             | Reversible cross linked<br>alginate polymer with<br>incorporated<br>radiopacifier | PVA distal tip                                                       |
| Shape          | Gel plug takes the<br>shape of the ureter                                                                                                     | Round mesh basket<br>with tether for removal<br>from ureter                           | Conical tipped basket<br>with tether for removal<br>from ureter                     | Extended polymeric<br>tube with open ended<br>coil at each end                    | Pigtail shape                                                        |
| Stone Removal  | Conventional stone<br>removal                                                                                                                 | Physically captured in<br>mesh basket and<br>manually withdrawn                       | Physically captured in<br>conical basket and<br>manually withdrawn                  | N/A                                                                               | N/A                                                                  |
| How Supplied   | Sterile package                                                                                                                               | Sterile package                                                                       | Sterile package                                                                     | Sterile package                                                                   | Sterile package                                                      |
| Use            | Single Use                                                                                                                                    | Single Use                                                                            | Single Use                                                                          | Single Use                                                                        | Single Use                                                           |
| Sterility      | Sterile                                                                                                                                       | Sterile                                                                               | Sterile                                                                             | Sterile                                                                           | Sterile                                                              |

·

. :

:

: 上一篇:

.

: 上一篇:

:

1

K096430 Page 2/3

1

.

. .

:

·

100 - 100 -

. .

: 上一篇:

. :

{2}------------------------------------------------

### Clinical Summary:

BackStop™ was evaluated in a prospective, randomized, single-blind, controlled multi-site clinical study. A total of sixty-eight (68) subjects with a single stone in the proximal ureter and an indication for ureteroscopic lithotripsy were enrolled and each subject was randomly assigned to either the experimental group (BackStop™, n=34) or the control group (standard of care, no anti-retropulsion device, n=34). Measured endpoints included the presence of retropulsion, the need for subsequent procedures, stone-free rate at follow-up, the occurrence of adverse events and the ability to dissolve BackStop™ and reverse the ureteral plug at the completion of lithotripsy. Results: There were no adverse events, nor any incidents of an occluded ureter postlithotripsy in the BackStop™ group. The test of the primary study hypothesis revealed a statistically significant lower rate of retropulsion {p=0.0002} in the BackStop™ group {8.8%, n = 3) compared to the control group (52.9%, n = 18). There were no significant differences between the two groups with respect to the secondary efficacy endpoints of stone-free rate at follow-up or need for additional procedures. The investigators concluded that BackStop™ is a novel and easy to use method of preventing stone fragment retropulsion during lithotripsy for the management of ureteral stones. Pluromed feels the study supports the conclusion that BackStop™ is a safe and effective method of preventing retropulsion during lithotripsy. BackStop™ prevents retropulsion of all size fragments and is easily removed after the procedure.

{3}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of a circular emblem that contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 2 2009

Mr. James Wilkie Vice President, Operations Pluromed. Inc. · 25-H Olympia Avenue WOBURN MA 01801

Re: K090430

Trade/Device Name: BackStop™M Regulation Number: 21 CFR 876.4680 Regulation Name: Ureteral stone dislodger Regulatory Class: II Product Code: ONJ Dated: May 7, 2009 Received: May 11, 2009

Dear Mr. Wilkie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 884.xxx (Obstetrics/Gynecology) (240) 276-0115 21 CFR 892.xxx (Radiology) (240) 276-0120 Other (240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Irine M. Morris

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K090430

Device Name:

BackStop

## Indications For Use:

BackStop is intended for use in the ureter to prevent migration of renal calculi during intracorporeal lithotripsy procedures.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hogu M. Whang

(Division Sigh-Off)
Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number

Page 1 of

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/ONJ/K090430](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/ONJ/K090430)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
