The reprocessed hot biopsy forceps are intended to be used for endoscopic tissue sample acquisition, controlled coagulation/cautery, and sessile polyp removal in the gastrointestinal tract.
Device Story
Reprocessed hot biopsy forceps; used for endoscopic tissue sampling, coagulation, and sessile polyp removal in GI tract. Device comprises proximal handle with electrosurgical plug, connecting rod/coil mechanism, and distal biopsy cup or alligator forceps. Length 160-240 cm; diameter 2.2-2.5 mm. Jaws may include serrations, fenestrations, needles, or biting teeth. Sheathed in electrical insulation. Operated by physician during endoscopic procedures; connects to external electrosurgical unit (not included). Provides tissue samples for biopsy and cauterization for hemostasis or polyp removal. Benefits include clinical utility of reprocessed single-use devices.
Clinical Evidence
Bench testing only. Includes functional testing of representative samples and process validation for cleaning, sterilization, and packaging. Manufacturing process includes visual and functional inspection of all units.
Technological Characteristics
Electrosurgical biopsy forceps; 160-240 cm length; 2.2-2.5 mm diameter. Materials include electrical insulation sheath, metal biopsy cups/jaws (cup, alligator, serrated, fenestrated, needle, or biting tooth options). Connects to external electrosurgical unit via proximal plug. Manual operation.
Indications for Use
Indicated for endoscopic tissue sample acquisition, controlled coagulation/cautery, and sessile polyp removal in the gastrointestinal tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
# JUL 2 0 2005
Image /page/0/Picture/1 description: The image shows the text "K050136" at the top. Below that, the text "PAGE 1 OF 2" is visible. The text appears to be handwritten in black ink on a white background. The image seems to be a page number or identification marker.
#### SUMMARY AND CERTIFICATION SECTION 2.
# 2.A. 510(K) SUMMARY
.
| Submitter: | SterilMed, Inc. |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dr. Bruce R. Lester<br>SterilMed, Inc.<br>11400 73rd Avenue North<br>Minneapolis, MN 55369<br>Ph: 763-488-3400<br>Fax: 763-488-3350 |
| Date Prepared: | January 19, 2005 |
| Trade Name: | Reprocessed Hot Biopsy Forceps |
| Classification Name:<br>and Number: | Electric Biopsy Forceps<br>Class II, 21 CFR 876.4300 |
| Product Code: | NLU |
| Predicate Device(s): | The reprocessed hot biopsy forceps are substantially<br>equivalent to: "Thermal Option" Hot Biopsy Forceps<br>(K932790), manufactured by Ballard (formerly Cox); "The<br>Shark" Hot Biopsy Forceps manufactured by Wilson-Cook<br>(K923470); "Radial Jaw" Hot Biopsy Forceps (K860366),<br>manufactured by Microvasive; and "Precisor" Hot Biopsy<br>Forceps (K905088), manufactured by Bard (formerly<br>ESCO). |
| Device Description: | The device consists of a proximal handle containing an<br>electrosurgical plug which is connected, via a coil, spring<br>or rod mechanism, to the biopsy cup or alligator forceps at<br>the distal tip. The biopsy forceps range from 160 to 240<br>cm in length and 2.2-2.5 mm in diameter. The jaws range<br>from 2.2 to 2.5 mm in diameter. The cup forceps may or<br>may not have serrated edges, a fenestration, needle, or<br>distal biting tooth. |
| | The forceps are covered with a sheath of electrical<br>insulation. The electrosurgical plug on the handle used for<br>connection to the appropriate electrosurgical unit. This<br>submission does not include the electrosurgical box that the<br>forceps connect to or its associated connecting cable. |
| Intended Use: | The reprocessed hot biopsy forceps are intended to be used<br>for endoscopic tissue sample acquisition, controlled<br>coagulation/cautery, and sessile polyp removal in the<br>gastrointestinal tract. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed hot biopsy forceps<br>underwent bench testing to demonstrate appropriate<br>functional characteristics. Process validation testing was<br>done to validate the cleaning and sterilization procedures as<br>well as the device's packaging. In addition, the<br>manufacturing process includes visual and functional<br>testing of all products produced. |
| Conclusion: | The hot biopsy forceps reprocessed by SterilMed are<br>substantially equivalent to the following specific predicate<br>devices: "Thermal Option" Hot Biopsy Forceps (K932790),<br>manufactured by Ballard (formerly Cox); "The Shark" Hot<br>Biopsy Forceps manufactured by Wilson-Cook (K923470);<br>"Radial Jaw" Hot Biopsy Forceps (K860366),<br>manufactured by Microvasive; and "Precisor" Hot Biopsy<br>Forceps (K905088), manufactured by Bard (formerly<br>ESCO). This conclusion is based upon the devices'<br>similarities in functional design, materials, indications for<br>use and methods of construction. |
{1}------------------------------------------------
・
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a logo with a stylized bird figure. The bird is depicted with three curved lines forming its body and wings. The logo also includes text arranged in a circular pattern around the bird, although the specific text is not clear due to the image quality. The overall design is simple and abstract.
JUL 2 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bruce R. Lester, Ph.D. Vice President of R&D SterilMed, Inc. 11400 73rd Avenue North MAPLE GROVE MN 55369
Re: K050136
Trade/Device Name: SterilMed Reprocessed Hot Biopsy Forceps (see enclosure) Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: NLU Dated: April 27, 2005 Received: April 28, 2005
Dear Dr. Lester:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have route determined the device is substantially equivalent (for the indications for use stated in above and have as a legally marketed predicate devices marketed in interstate commerce prior to the cholors, the enactment date of the Medical Device Amendments, or to devices that have been May 20, 1710, and classified in accordans of the Federal Food, Drug, and Cosmetic Act (Act) that tecassinod in accordatios approval application (PMA). You may, therefore, market the do not require approvations of the Act. The general controls provisions of the Act. device, subjoc to annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your device is older to such additional controls. Existing major regulations affecting your Approval), a the you be boor of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I caral statues and rogalding, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (2) CFR Part 807); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter with anow you to ogin manenes of your device of your device to a legally premarked nothleation: "The PDF mailig sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific auvice tor your acrees on the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, " have the on your responsibilities under the Act from the 807.97). Tou Inay Outain other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Attachment - Reprocessed Hot Biopsy Forceps (K050136)
| Bard |
|-------------|
| 000852 |
| 000853 |
| 000854 |
| 000856 |
| Ballard |
| 60101 |
| 60102 |
| 60103 |
| 60104 |
| 60107 |
| 60108 |
| 60109 |
| 60110 |
| 60113 |
| 60114 |
| 60115 |
| 60116 |
| 60119 |
| 60120 |
| 60121 |
| 60122 |
| Microvasive |
1 250
### Wilson-Cook
HS-2.5-230 HDS-2.5-230
'문관 1
:
.
{5}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K050136 |
|---------------------------|--------------------------------|
| Device Name: | Reprocessed Hot Biopsy Forceps |
| Indications For Use: | |
The Reprocessed Hot Biopsy Forceps are intended to be used for endoscopic tissue sample acquisition, controlled coagulation/cautery, and sessile polyp removal in the gastrointestinal tract.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Darrey C. hagdon
(Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 0501-510(k) Number _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.