← Product Code [NLR](/submissions/GU/subpart-e%E2%80%94surgical-devices/NLR) · K030919

# SISS, INC. (D.B.A.MEDISISS) REPROCESSD ELECTROSURGICAL INSTRUMENTS AND ACCESSORIES (K030919)

_Medisiss · NLR · Jul 3, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/NLR/K030919

## Device Facts

- **Applicant:** Medisiss
- **Product Code:** [NLR](/submissions/GU/subpart-e%E2%80%94surgical-devices/NLR.md)
- **Decision Date:** Jul 3, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Intended Use

The SISS Inc. (d.b.a. MediSISS) Reprocessed electrosurgical instruments have applications in a variety of minimally invasive surgical procedures to manipulate and manage internal soft tissue by grasping, cutting, and/or to facilitate coagulation, transection, resection, mobilization, and dissections of tissue.

## Device Story

Reprocessed electrosurgical instruments; hand-manipulated devices with electrocautery capability; used in minimally invasive surgery. Input: manual operation by surgeon. Transformation: mechanical transmission of handle movement to distal end-effector (dissectors, graspers, scissors, shears, forceps) via insulated shaft; electrical energy delivery for monopolar, bipolar, or tripolar coagulation/cutting. Output: physical tissue manipulation and electrosurgical effect. Features: optional 360-degree shaft rotation; jaw locking mechanisms. Used in OR; operated by surgeons. Benefits: provides reprocessed, cost-effective surgical instrumentation for soft tissue management.

## Clinical Evidence

Bench testing only. Evidence includes validation of cleaning, sterilization (ANSI/AAMI/ISO 11135-1994), and packaging processes. Biocompatibility testing (ANSI/AAMI/ISO 10993-1:1997) and mechanical/functional testing performed on representative samples to confirm performance characteristics remain equivalent to original devices.

## Technological Characteristics

Hand-manipulated electrosurgical instruments; monopolar, bipolar, or tripolar configurations. Materials and design identical to predicate devices. Features include insulated shafts, rotation knobs, and jaw-locking mechanisms. Sterilization via ethylene oxide per ANSI/AAMI/ISO 11135-1994. Biocompatibility per ANSI/AAMI/ISO 10993-1:1997 and 10993-7:1995.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Circon/ACMI electrosurgical instruments ([K884306](/device/K884306.md), [K932293](/device/K932293.md), [K965176](/device/K965176.md))
- Alliance Corporation reprocessed electrosurgical instruments ([K012608](/device/K012608.md), [K012638](/device/K012638.md), [K012603](/device/K012603.md))
- Adven Medical, Inc. Reprocessed Used Disposable Endoscopic Scissors and Graspers ([K012696](/device/K012696.md))
- SterilMed, Inc. Reprocessed Laparoscopic Electric Instruments ([K012598](/device/K012598.md))
- Vanguard Medical Concepts, Inc. Vanguard Reprocessed Endoscopic Instruments ([K012700](/device/K012700.md))
- SISS, Inc. Electrosurgical Electrodes ([K012669](/device/K012669.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3 2003 JUL

Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The sequence starts with the letter 'K', followed by the numbers '030919'. The characters are written in a casual, cursive style, with varying stroke thicknesses. The image has a simple composition, focusing solely on the legibility of the characters.

510(k): SISS Inc. (d.b.a. MediSISS) Reprocessed Electrosurgical Instruments and Accessories

Page 41 of 44

Summerer of Stiller The Price Differ 201 231 2 2011 10:00 10 10 10 10 10 10 10 10 10 10 1

submitted by

SISS Inc. (d.b.a. MediSISS) P.O. 2060 723 Curtis Court Sisters, OR 97759 Phone: (800)-860-9482 Fax: (541) 549-4527 Email: mabarker@MediSISS.com

Contact Person: Device Trade Names: Common Names: Classification Names:

Mary Ann Barker Reprocessed Electrosurgical Instruments Reprocessed Electrosurgical Instruments Electrosurgical cutting and coagulation device and accessories. CFR §876.4300

#### Identification of a Legally Marketed Predicate Device

The SISS Inc. (d.b.a. MediSISS) Reprocessed Electrosurgical Instruments are substantially equivalent to the electrosurgical instruments manufactured by:

Circon/ACMI Weck (Pilling Surgical)

These devices are legally marketed and distributed pursuant to 510(k)'s K884306, K932293, and K965176 and the counterpart devices from the original manufacturers.

They are also similar to the reprocessed electrosurgical instruments and accessories reprocessed by Alliance Corporation and legally marketed and distributed pursuant to 510(k)'s K012608, K012638, K012603. Likewise, Adven Medical, Inc Reprocessed Used Disposable Endoscopic Scissors and Graspers - 510(k) K012696; SterilMed, Inc. Reprocessed Laparoscopic Electric Instruments: - 510(k) K012598; Vanguard Medical. Concepts, Inc Vanguard Reprocessed Endoscopic Instruments, 510(k) K012700; and SISS, Inc Electrosurgical Electrodes - 510(k) K012669.

{1}------------------------------------------------

#### Device Description

Reprocessed Electrosurgical Instruments may consist of hand-manipulated devices with electrocautery capability and with or without rotation capability.

The handpiece handles are connected to the distal end-effector by a narrow-diameter insulated barrel or shaft. The distal end of the device consists of a variety of distal end configurations including: dissectors (straight or curved), graspers, scissors (curved, hooked, or metzenbaum), shears, and cutting or dissecting forceps. The devices may be monopolar, bipolar, or tripolar. The devices are designed to be inserted through an appropriately sized trocar sleeve or cannula.

If the instrument has a scissor or jaw end-effector these are opened and closed using the handles. The device's insulated shaft may be designed to (depending on the device model and type) be rotated (up to 360°) either direction (using a knob on the handle.) The jaws of some models may be rotated by manipulating controls on the handpiece. Grasper and clamp models may have manipulating jaws operated at the handpiece to lock and hold tissue.

## Intended Use

The SISS Inc. (d.b.a. MediSISS) Reprocessed electrosurgical instruments have applications in a variety of minimally invasive surgical procedures to manipulate and manage internal soft tissue by grasping, cutting, and/or to facilitate coagulation, transection, resection, mobilization, and dissections of tissue.

## Summary of Technological Characteristics

The intended use and technological features of the reprocessed devices do not differ from the legally marketed predicate device(s). Both the reprocessed devices(s) and the predicate device(s) have the same materials and product design. There are no changes to the claims, intended use, clinical applications, patient populations, performance specifications, or methods of operation. The technological characteristics of the reprocessed electrosurgical devices are the same as those of the legally marketed predicate devices. The technological characteristics of the reprocessed electrosurgical device(s) are the same as those of the legally marketed predicate devices. In addition the MediSISS™ manufacturing process includes 100 % visual and mechanical testing of all products prior to packaging, labeling, and sterilization.

## Summary of Performance Data

The SISS Inc. (d.b.a. MediSISS) Reprocessed Electrosurgical Instruments comply with the following standards, practices, and guidance's:

- ANSI/AAMI/ISO 11135-1994, Medical Devices-Validation and Routine ● Control of Ethylene Oxide Sterilization
- ANSI/AAMI/ISO 10993-7:1995, Biological Evaluation of Medical ● Devices-Part 7: Ethylene oxide sterilization residuals

{2}------------------------------------------------

- ANSVAAMI/ISO 10993-1: 1997 (1999 Edition). Biological Evaluation of ● Medical Devices Part 1: Evaluation and Testing.
Cleaning, sterilization, packaging validations, and visual/mechanical testing demonstrate that the devices are equivalent and continue to be safe and effective for their intended use.

MediSISS™ Reprocessed Electrosurgical Instruments undergo mechanical testing to demonstrate that the parts do not change in function. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device's packaging.

Cleaning, sterilization, and packaging validations, functional/performance testing, (product certification,) and (biocompatibility) testing (or certification of the replacement insulation material) demonstrate that the devices are equivalent and continue to be safe and effective for their intended use.

Representative samples of reprocessed electrosurgical instruments underwent bench testing to demonstrate appropriate functional characteristics. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device's packaging. In addition, the manufacturing process includes visual and functional testing of all products produced.

The SISS Inc. (d.b.a. MediSISS) Reprocessed Electrosurgical Instruments are substantially equivalent to the identified predicate devices. This has been demonstrated through bench testing and comparative analysis of features.

{3}------------------------------------------------

# Conclusion

Since the Reprocessed Electrosurgical Instruments meet the requirements of the stated standards and embody technological characteristics identical to the predicate device, we believe the device is safe and effective and performs as well as or better than the predicate device. The Reprocessed Electrosurgical Instruments will be reprocessed per specifications and good manufacturing practices that ensure the device is safe and effective for its intended use.

In accordance with the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this Premarket notification, SISS Inc. (d.b.a. MediSISS) concludes that the modified devices (Reprocessed Electrosurgical Instruments) are safe, effective and substantially equivalent to the predicate devices described herein.

Based on the information provided herein, the 510(k) "Substantial Equivalence" Decision Making Process Chart, and the FDA - "510(k) Guidance Document for General Surgical Electrosurgical Devices" 5/10/95, we conclude that the SISS Inc. (d.b.a. MediSISS) reprocessed electrosurgical instruments are substantially equivalent to the predicate devices under the Federal Food and Drug, and Cosmetic Act.

This conclusion is based upon the devices' similarities in functional design, materials, indications for use, and methods of construction.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge of the circle. In the center of the circle is a stylized image of an eagle's head with three lines representing the feathers.

Public Health Service

3 2003 JUL

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siss. Inc. (d.b.a. MediSiSS") Mr. Marc M. Mouser Project Engineer, Medical Devices FDA Office Coordinator UL Conformity Assessment Services Underwriters Laboratory, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542

Re: K030919

Trade/Device Name: See Attachment Regulation Number: 876.4300 Regulation Name: Reprocessed Endoscopic Electrosurgical Unit (with or without Accessories) Regulatory Class: II Product Code: NLR Dated: June 16, 2003 Received: June 20, 2003

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure ,

l'age 2

{6}------------------------------------------------

| Device                                                         | Manufacturer            | Model No. |
|----------------------------------------------------------------|-------------------------|-----------|
| Dolphin Nose Dissector                                         | Weck (Pilling Surgical) | 722420a   |
| Maryland Dissector                                             | Weck (Pilling Surgical) | 722425a   |
| Straight Dissector                                             | Weck (Pilling Surgical) | 725401a   |
| Fenestrated Dissector                                          | Weck (Pilling Surgical) | 725430a   |
| Mixter Dissector 90° angle                                     | Weck (Pilling Surgical) | 727210a   |
| Mixter Dissector 45° angle                                     | Weck (Pilling Surgical) | 727211a   |
| Maryland Dissector                                             | Weck (Pilling Surgical) | 727425a   |
| Cone Tip Dissector with spoon                                  | Weck (Pilling Surgical) | 725405a   |
| Dolphin Nose Dissector                                         | Weck (Pilling Surgical) | 725420a   |
| Maryland Dissector                                             | Weck (Pilling Surgical) | 725425a   |
| Tri-polar Cutting Forceps,<br>10 mm, 32 cm                     | Circon ACMI             | 006689-90 |
| Tri-polar Cutting Forceps<br>w/rotation, 10 mm, 32 cm          | Circon ACMI             | 006689-90 |
| Tri-polar Cutting Forceps<br>for open procedures, 10 mm, 15 cm | Circon ACMI             | 008140-90 |
| Tri-polar Cutting Forceps,<br>5 mm, 32 cm                      | Circon ACMI             | 008593-90 |

:

:

 1

:

K030919 Premarket Notification of MediSISS Reprocessed Electrosurgical Devices

{7}------------------------------------------------

Page 5 of 44

Indications for Usa

|                           | Page 1 of 1                                                                          |
|---------------------------|--------------------------------------------------------------------------------------|
| 510(k) Number (if known): |                                                                                      |
| Device Name:              | SISS Inc. (d.b.a. MediSISS) Reprocessed Electrosurgical Instruments and Accessories. |
| Indications for Use:      |                                                                                      |

The SISS Inc. (d.b.a. MediSISS) Reprocessed electrosurgical instruments have applications in a variety of minimally invasive surgical procedures to manipulate and manage internal soft tissue by grasping, cutting, and/or to facilitate coagulation, transection, resection, mobilization, and dissections of tissue.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use (Per 21 CFR 801.109) | OR (Division Sign-Off) | Over-The-Counter Use (Optional Format 1-2-96) |
|---------------------------------------|------------------------|-----------------------------------------------|
|---------------------------------------|------------------------|-----------------------------------------------|

Nancy c Brogdon

(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices K030919

510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/NLR/K030919](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/NLR/K030919)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
