The InScope Multi-Clip Applier is indicated for use with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of: - Endoscopic marking, . - Hemostasis for: ● - o Mucosal and submucosal defects < 3 cm - Bleeding ulcers o - o Arteries < 2 mm - o Polyps < 1.5 cm in diameter - o Diverticula in the colon - Anchoring to affix jejunal feeding tubes to the wall of the small bowel, . - Closure of GI tract luminal perforations < 20 mm in patients where an alternative . to surgery is required.
Device Story
InScope Multi-Clip Applier is a sterile, single-use, disposable endoscopic instrument; designed for use with flexible endoscopes (min 3.2 mm working channel). Device features two opposing jaws controlled by a proximal handle lever; manual firing mechanism deploys preloaded titanium clips. Used by physicians in clinical settings to achieve hemostasis, mark tissue, anchor feeding tubes, or close GI perforations. Device provides mechanical tissue approximation; benefits include minimally invasive alternative to surgery for specific GI defects.
Clinical Evidence
Bench and preclinical testing only. Results demonstrate performance equivalence to predicates in hemostasis, clip retention, and tissue viability. Literature review supports use for perforation closure, fistulas, anastomotic leaks, and diverticula bleeding without increased safety risks.
Technological Characteristics
Sterile, single-use, disposable endoscopic instrument. Materials include titanium clips. Sensing/actuation: manual lever-actuated mechanical grasping and firing mechanism. Dimensions: compatible with flexible endoscopes having minimum 3.2 mm working channel. No electronic components or software.
Indications for Use
Indicated for patients requiring endoscopic marking, hemostasis (mucosal/submucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), jejunal feeding tube anchoring, or closure of GI luminal perforations <20mm where surgery is not an option.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center - WO66-G609
Silver Spring, MD 20993-0002
Ethicon Endo-Surgery, Inc.
Kimberly Shoemaker, RAC
Manager, Regulatory Affairs
4545 Creek Road
Cincinnati, OH 45242-2839
JUL 27 2015
Re: K051950
Trade/Device Name: InScope™ Multi-Clip Applier
Regulation Number: 21 CFR § 876.4400
Regulation Name: Hemorrhoidal ligator
Regulatory Class: II
Product Code: MND, FHN, OCW, HBT and MCH
Dated (Date on orig SE ltr): September 23, 2005
Received (Date on orig SE ltr): September 26, 2005
Dear Ms. Shoemaker,
This letter corrects our substantially equivalent letter of October 7, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the *Federal Register*.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D.
Director
Division of Reproductive, Gastro-Renal, and Urological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
K05/1950
# Indications for Use
510(k) Number (if known): K051950
Device Name: InScope™ Multi-Clip Applier
Indications for Use:
The InScope Multi-Clip Applier is indicated for use with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of:
- Endoscopic marking,
- Hemostasis for:
- Mucosal and submucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Anchoring to affix jejunal feeding tubes to the wall of the small bowel,
- Closure of GI tract luminal perforations < 20 mm in patients where an alternative to surgery is required.
Prescription Use ☑
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Sypem
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K051950
Page 1 of 1
{3}
OCT 7 - 2005
K051950
Pg1 of 2
# InScope™ Multi-Clip Applier
510(k) Summary of Safety and Effectiveness
## Company
INSCOPE
A Division of Ethicon Endo-Surgery, Inc.
4545 Creek Rd.
Cincinnati, OH 45242
## Contact
Kimberly Shoemaker, RAC
Manager, Regulatory Affairs
## Date Prepared:
July 15, 2005
## Name of Device
Trade Name: InScope™ Multi-Clip Applier
Classification Name: Clip Applier/Endoscopic Clipping Device
## Predicate Devices:
Trade Name:
Syntheon, LLC: Endoscopic Multi-Fire Clip Applier
Olympus Optical Co., Ltd.: Rotatable Clip Fixing Device HX-5/6-1
## Device Description
The InScope™ Multi-Clip Applier is a sterile, single patient use, disposable instrument capable of attaching clips to the mucosal lining of the gastrointestinal (GI) tract. The clips are used to create hemostasis and/or tissue approximation. The distal, flexible portion of the applier is designed to work with a flexible endoscope having a minimum working channel diameter of 3.2 mm. Two opposing jaws grasp tissue via a manually activated lever on a proximal handle. Once the closure lever is latched, a firing mechanism is manually activated, which deploys a titanium clip over the target tissue. The device is supplied preloaded with four (4) titanium clips.
{4}
K051950
Pg 2 of 2
# Indications for Use
The InScope Multi-Clip Applier is indicated for use with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of:
- Endoscopic marking,
- Hemostasis for:
- Mucosal and submucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Anchoring to affix jejunal feeding tubes to the wall of the small bowel,
- Closure of GI tract luminal perforations < 20 mm in patients where an alternative to surgery is required.
# Technological Characteristics
The InScope™ Multi-Clip Applier does not have the identical indication statements as the predicate devices, but the differences noted do not alter the intended therapeutic/diagnostic effect.
# Performance Data
Results of bench and preclinical testing demonstrate that performance of the InScope Multi-Fire Clip Applier is substantially equivalent to the predicate devices in regards to hemostasis, clip retention and lack of tissue damage upon clip placement (tissue remains viable).
A literature review in combination with the preclinical data demonstrates the use of the InScope Multi-Clip Applier for closure of gastrointestinal perforations, fistulas, anastomotic leaks and for control of bleeding diverticula is not likely to raise any new or increased risks to safety and effectiveness compared to risks experienced with similar marketed devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.