← Product Code [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS) · K241271

# Aqua Medical RF Vapor Ablation System (K241271)

_Aqua Medical, Inc. · KNS · Dec 12, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K241271

## Device Facts

- **Applicant:** Aqua Medical, Inc.
- **Product Code:** [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS.md)
- **Decision Date:** Dec 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.

## Device Story

Catheter-based system for GI tissue coagulation and ablation; delivers heated water vapor to tissue surface. System components: software-controlled RF generator, disposable RF vapor catheters, saline delivery tubing/syringe. Generator uses RF energy to heat saline within catheter tip, creating vapor; vapor applied to target tissue. Operated via GUI, foot pedal, or catheter-integrated button. Used in clinical settings by physicians during endoscopic procedures (via 3.7mm working channel). Output is thermal energy for tissue coagulation/ablation. Benefits include non-contact tissue heating, avoiding direct RF application to tissue.

## Clinical Evidence

Bench testing only. Hardware verification included fatigue cycle testing, catheter joint forces, and sliding forces. Lean beef testing compared ablation depths between previous time-based treatment and modified energy-level-based treatment, demonstrating equivalent heat penetration.

## Technological Characteristics

Catheter-based RF vapor ablation system. Components: RF generator, 7F disposable catheters (10.5F sheath), saline delivery system. Energy source: Bipolar RF energy used to generate water vapor. Connectivity: GUI-controlled generator. Software-controlled. Single-use, sterile catheters.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Aqua Medical RF Vapor System ([K213627](/device/K213627.md), [K211282](/device/K211282.md), [K183595](/device/K183595.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 12, 2024

Aqua Medical, Inc. % Bosmat Friedman-Cox Regulatory Consultant ProMedoss, Inc. 6026 Beech Cove Ln. Charlotte, North Carolina 28269

Re: K241271

Trade/Device Name: Aqua Medical RF Vapor Ablation System Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit And Accessories Regulatory Class: Class II Product Code: KNS. GEI Dated: May 6, 2024 Received: November 8, 2024

Dear Bosmat Friedman-Cox:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Anthony Lee -S" in a large, bold font. The text is black against a white background. The letters are clear and easy to read. There is a watermark in the background.

Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K241271

Device Name

Aqua Medical RF Vapor Ablation System

Indications for Use (Describe)

The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.

Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| Image: AQUA MEDICAL logo    | Aqua Medical RF Vapor Ablation System |
|-----------------------------|---------------------------------------|
| Traditional 510(k); K241271 | 510(k) Summary                        |

### SUBMITTER 1.

## Applicant's Name:

Aqua Medical, Inc. 6634 Owens Drive Pleasanton, CA 94588 Phone: 949-233-5172

## Primary Contact:

Bosmat Friedman-Cox Regulatory Affairs Consultant 6026 Beech Cove Ln. Charlotte, NC 28269 Phone: 980-308-1636 bosmat.f(@promedoss.com

## Date Prepared:

November 26, 2024

### 2. DEVICE

Trade Name: Aqua Medical RF Vapor Ablation System

| Classification Name: | Endoscopic electrosurgical unit and accessories                |
|----------------------|----------------------------------------------------------------|
| Product Code:        | KNS                                                            |
| Regulation No:       | 21 CFR 876.4300                                                |
| Class:               | 2                                                              |
| Review Panel:        | Gastroenterology/Urology                                       |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Product Code:        | GEI                                                            |
| Regulation No:       | 21 CFR 878.4400                                                |
| Class:               | 2                                                              |
| Review Panel:        | General & Plastic Surgery                                      |

#### 3. PREDICATE DEVICES

## Primary Predicate:

Aqua Medical RF Vapor System, by Aqua Medical Inc., Product code KNS, GEI cleared Under: K213627, K211282 and K183595.

### 4. DEVICE DESCRIPTION

The Aqua RF Vapor Ablation System uses a catheter to deliver heated water vapor to the tissue surface undergoing treatment. The Aqua RF Vapor Ablation System is a catheter-based system for heating and coagulating tissue. The heat is delivered to the target tissue in the form of water

{5}------------------------------------------------

| Image: AQUA MEDICAL logo<br>AQUA MEDICAL | Aqua Medical RF Vapor Ablation System |  |
|------------------------------------------|---------------------------------------|--|
| Traditional 510(k); K241271              | 510(k) Summary                        |  |

vapor that is generated within the catheter tip using radiofrequency (RF) energy. Application of heat to the tissue results in coagulation and ablation of the tissues being treated. This technique avoids the need for direct RF energy application to heat the tissue. The Aqua Medical RF VaporAblation System consists of the following components:

- Aqua RF Vapor Generator: A software-controlled RF generator is operated through a graphical user interface (GUI) and incorporates a syringe pump that delivers saline to the catheter. Delivery of bipolar RF energy to the catheter is controlled either by an accessory foot pedal or a button incorporated in the disposable catheter.
- Aqua RF Vapor Catheters: disposable, sterile, single-use catheters with a diameter of 7F -(2.3 mm) with an outer catheter sheath that is 10.5F and a length between 145 cm and 210 cm. There are 3 types of RF Vapor catheters available. Catheters are designed to be used through an endoscope working channel diameter of 3.7mm.
- Saline Delivery Tubing and Syringe: Tubing and syringe (60cc) provide a means of delivering saline to the catheter during treatment.

### 5. INDICATIONS FOR USE

The Aqua Medical RF Vapor System is indicated for use in the coagulation and ablation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus.

Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP). The device is to be used in adults only.

### SUBSTANTIAL EQUIVALENCE 6.

The reason for this submission includes minor software and device modifications, addition of a new circumferential catheter size and to update the previously cleared indication to include the term "ablation". To support the requested modifications the company submitted supporting nonclinical data as well as justification for the addition of "ablation" to the indication. The proposed modifications do not introduce new safety and effectiveness concerns and support the company's substantial equivalency claim.

## 7. PERFORMANCE DATA

In order to support the proposed modifications, the company conducted Hardware Verification Testing, Lean Beef Testing and Software Testing. The proposed modifications did not require new or supplemental biocompatibility, sterilization, packaging or shelf-life testing. Hardware Verification testing included fatigue cycle testing, catheter joint forces and sliding forces. The catheter joint forces and sliding forces passed the specification requirements for the testing. The fatigue cycle testing was completed to validate the physical integrity and durability of the subject device under simulated-use conditions including extreme tortuosity, mechanical cycling, highvapor output and temperature exposure. The Lean Beef testing demonstrated equivalent results between the previous time-based treatment when compared to the modified energy level-based

{6}------------------------------------------------

|                     |                                       |                | Page 3 of 3 |
|---------------------|---------------------------------------|----------------|-------------|
| Image: AQUA MEDICAL | Aqua Medical RF Vapor Ablation System |                |             |
|                     | Traditional 510(k); K241271           | 510(k) Summary |             |

K241271

treatment. The test also demonstrated that the ablation depths for the modified Circumferential Delivery Catheters are substantially equivalent to those produced by the Predicate Circumferential Delivery Catheters through the dose ranges evaluated in the test. Equivalent ablation depths (penetration of heat) were produced by both systems supporting our substantial equivalency claim.

## 8. CONCLUSION

Aqua Medical has demonstrated that the Aqua RF Vapor Ablation System is substantially equivalent to the predicate device. Differences between the proposed Aqua RF Vapor Ablation System and the predicate device do not raise new questions of safety or effectiveness. The requested modifications have been adequately supported by the provided bench testing results.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K241271](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K241271)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
