← Product Code [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS) · K172288

# Fusion OMNI Sphincterotome (K172288)

_Wilson-Cook Medical Inc./Cook Endoscopy · KNS · Apr 17, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K172288

## Device Facts

- **Applicant:** Wilson-Cook Medical Inc./Cook Endoscopy
- **Product Code:** [KNS](/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS.md)
- **Decision Date:** Apr 17, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.

## Device Story

Sterile, single-use endoscopic sphincterotome; consists of long plastic cannula with internal wire; distal wire segment exposed. Operated by physician during ERCP via endoscope accessory channel. High-frequency electrical current applied to wire to perform sphincterotomy (opening papilla roof) for biliary/pancreatic orifice access. Preloaded configuration aids in bridging difficult strictures. Device facilitates selective cannulation and device exchange via distal wire guide port. Benefits include improved access to ductal systems and procedural efficiency.

## Clinical Evidence

Bench testing only. Testing included shelf life, packaging validation, sterile barrier inspection, functional testing, joint strength (drive wire and brass insert), fluoroscopic visibility, and active cord pin detachment. Electrical safety and EMC assessed per IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18. Biocompatibility evaluated per ISO 10993-1.

## Technological Characteristics

Sterile, single-use endoscopic accessory. Materials: plastic cannula, metal wire. Energy: high-frequency electrosurgical current. Features: distal wire guide port, variable catheter/wire guide diameters. Standards: IEC 60601-1 (safety), IEC 60601-2-2 (HF surgical equipment), IEC 60601-2-18 (endoscopic equipment), ISO 10993-1 (biocompatibility).

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Wilson-Cook OMNI™ Sphincterotome ([K052051](/device/K052051.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2018

Wilson-Cook Medical, Inc./ Cook Endoscopy Sierra Lowe Regulatory Affairs Specialist - I 4900 Bethania Station Road Winston Salem. NC 27105

Re: K172288

Trade/Device Name: Fusion® OMNI™ Sphincterotome Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: March 6, 2018 Received: March 7, 2018

Dear Sierra Lowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the text "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and is set against a white background. The letters are clear and easy to read. The text appears to be a name, possibly of a person.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K172288

Device Name

Fusion® OMNITM Sphincterotome

#### Indications for Use (Describe)

This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### K172288 Page 1 of 3

Image /page/3/Picture/1 description: The image shows a close-up of a logo against a red background. The word "COOK" is prominently displayed in white, block letters at the top. Below "COOK", the word "MEDICAL" is written in smaller, white block letters on a dark red or brown rectangular background.

COOK ENDOSCOPY 4900 BETHANIA STATION ROAD WINSTON-SALEM, NC 27105 U.S.A. PHONE: 336.744.0157 TOLL FREE: 800.245.4707 WWW.COOKMEDICAL.COM

## 510(k) Summary

# Fusion® OMNI Sphincterotome

Traditional 510(k) Premarket Notification

March 6, 2018

## Applicant Information

| Applicant: | Wilson-Cook Medical, Inc. /Cook Endoscopy<br>4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
|------------|----------------------------------------------------------------------------------------------------------------|
| Contact:   | Sierra Lowe, Specialist I - Regulatory Affairs                                                                 |
| Phone:     | (336) 744-0157 ext. 396506                                                                                     |
| Fax:       | (336) 201-5994                                                                                                 |

## Device Information

| Trade Name:          | Fusion® OMNI Sphincterotome                     |
|----------------------|-------------------------------------------------|
| Common Names:        | Papillotome, sphincterotome                     |
| Classification Name: | Endoscopic electrosurgical unit and accessories |
| Regulation Number:   | 21 CFR 876.4300                                 |
| Product Code:        | KNS                                             |
| Device Class:        | Class II                                        |
| Review Panel:        | Gastroenterology-Urology                        |

#### Predicate Device

| Name:          | Wilson-Cook OMNI <sup>TM</sup> Sphincterotome |
|----------------|-----------------------------------------------|
| 510(k) Number: | K052051                                       |
| Date:          | Cleared June 18, 1990                         |

## Intended Use

This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.

{4}------------------------------------------------

### Device Description

The Fusion® OMNI Sphincterotome (subject device) is a sterile, single use device compatible with the accessory channel of endoscope. The device consists of a long, thin plastic tube (cannula) with a wire running the length of its interior. A small portion of that wire is exposed at its distal end. The roof of the papilla is opened by passing high- frequency current through the wire, exposing the biliary or pancreatic orifices for selective cannulation.

#### Substantial Equivalence

Minor changes have been made to the predicate sphincterotomes cleared to market via K052051. The most notable changes to the subject devices include: Slightly different catheter diameters and compatible wire guide diameters to accommodate a range of endoscope channels and/or smaller papilla, inclusion of a wire guide port at the distal end (patient contacting end) of the device in order to facilitate an additional option to perform a device exchange, slightly decreased maximum rated input voltage in order to be representative of the electrical testing conducted on the subject devices.

The subject device and predicate device have the same intended use and different technological characteristics. None of the differences in technological characteristics raise different questions of safety and effectiveness. Furthermore, performance data from acceptable scientific testing methods provide evidence that the subject device is substantially equivalent to the predicate device.

#### Performance Data

Performance testing consisting of sterilization, shelf life, biocompatibility, and non-clinical bench testing demonstrate that the Fusion® OMNI Sphincterotome meets the performance requirements to fulfill the intended use of the device.

#### Summary of Non-Clinical Testing

The following non-clinical testing was conducted to demonstrate the performance of the subject device and confirmed that the subject device performs as intended.

- . Shelf Life Testing
- . Packaging Validation
- Non-clinical, bench testing of the subject devices methods included: sterile barrier ● inspection, functional testing, drive wire to handle joint strength, brass insert to handle joint strength, fluoroscopic visibility, active cord pin detachment.
- The electromagnetic compatibility and electrical safety of the subject devices were . assessed in accordance with the following standards: IEC 60601-1: 2005 + A1: 2012 (3.1

{5}------------------------------------------------

edition)- Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, IEC 60601-2-2: 2017 (6.0 edition) - Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories, and IEC 60601-2-18: 2009 (3.0 edition) - Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment

- Biocompatibility testing was conducted in accordance with ISO 10993-1:2009 and FDA . Guidance: Use of International Standard ISO 10993-1:2009 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K172288](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KNS/K172288)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
