Avulsion Forceps

K183428 · STERIS Corporation · KGE · Apr 16, 2019 · Gastroenterology, Urology

Device Facts

Record IDK183428
Device NameAvulsion Forceps
ApplicantSTERIS Corporation
Product CodeKGE · Gastroenterology, Urology
Decision DateApr 16, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2

Indications for Use

Disposable biopsy forceps are used to obtain mucosal tissue samples for pathology using monopolar energy, including the avulsion technique, during endoscopic procedures in the gastrointestinal tract, for example the esophagus, stomach, small intestine and colon. Tissue samples will be submitted for pathologic review. This product is intended only for adult populations.

Device Story

Avulsion Forceps are endoscopic instruments used to obtain mucosal tissue samples. The device features a proximal handle that actuates distal toothed jaws via a single drive cable. It connects to an electrosurgical generator via an Olympus-style active cord connector to deliver monopolar energy. During procedures, the clinician uses manual traction and high-energy cutting current to resect tissue. The device is used in the gastrointestinal tract (esophagus, stomach, small intestine, colon). The output is a tissue sample for pathologic review. The device is disposable and supplied sterile.

Clinical Evidence

Bench testing only. Verification activities included structural integrity, dimensional analysis, functional testing in tortuous conditions, simulation testing for tissue retrieval, electrical safety testing per IEC 60601-2-2 (3rd Edition), and biocompatibility testing per ISO-10993-1. All tests passed.

Technological Characteristics

Monopolar electrosurgical biopsy forceps; 2.6 mm outer tube diameter; 230 cm working length; toothed jaw design; single drive cable mechanism; Olympus-style active cord connector. Materials are biocompatible per ISO-10993-1. Sterile, single-use, 1-year shelf life. Electrical safety per IEC 60601-2-2.

Indications for Use

Indicated for adult patients undergoing endoscopic procedures in the gastrointestinal tract (esophagus, stomach, small intestine, colon) to obtain mucosal tissue samples for pathology using monopolar energy, including avulsion technique.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 16, 2019 STERIS Corporation Eileen McCafferty Senior Regulatory Specialist 5960 Heisley Rd Mentor, OH 44060 Re: K183428 Trade/Device Name: Avulsion Forceps Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KGE Dated: December 10, 2018 Received: December 11, 2018 Dear Eileen McCafferty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls' provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling {1}------------------------------------------------ (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, (for) Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183428 Device Name Avulsion Forceps Indications for Use (Describe) Disposable biopsy forceps are used to obtain mucosal tissue samples for pathology using monopolar energy, including the avulsion technique, during endoscopic procedures in the gastrointestinal tract, for example the esophagus, stomach, small intestine and colon. Tissue samples will be submitted for pathologic review. This product is intended only for adult populations. Type of Use (Select one or both, as applicable) | <div> <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |---------------------------------------------------------------------------------------------------------------| | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the STERIS logo. The logo is in black and white and consists of the word "STERIS" in large, bold letters, with the registered trademark symbol to the right. Below the word is a graphic of several horizontal wavy lines in blue. At the top of the image, the text "K183428" and "1 of 5" are printed. # 510(k) Summary For Avulsion Forceps STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Eileen McCafferty Contact: Senior Regulatory Affairs Specialist Tel: 440-392-7189 Fax: 440-357-9198 Email: eileen_mccafferty@steris.com April 3, 2019 Summary Date: Premarket Notification Number: K183428 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {4}------------------------------------------------ #### 1. Device Name | Trade Name: | Avulsion Forceps | |------------------------|-------------------------------------------------| | Device Classification: | Class II | | Common/usual Name: | Electric Biopsy Forceps, Hot Biopsy Forceps | | Classification Name: | Endoscopic Electrosurgical unit and accessories | | Classification Number: | 21 CFR 876.4300 | | Product Code: | KGE | #### 2. Predicate Device US Endoscopy Disposable Hot Biopsy Forceps K925361 #### 3. Description of Device The Avulsion Forceps is used during endoscopic procedures in the gastrointestinal tract. The avulsion technique in Endoscopy involves using both mechanical forces as well as high energy cutting current to tear the tissue away from the gastrointestinal tract. The Avulsion Forceps consists of a handle on the proximal end that actuates two toothed jaws on the distal end. An Olympus-style active cord connector is included that allows the Avulsion Forceps to be connected to an electrosurgical generator for use with monopolar energy. The device also includes a spring sheath catheter housing the drive cable which transfers the linear motion of the handle slider to open and close the jaws. #### 4. Intended Use/Indications for Use Disposable biopsy forceps are used to obtain mucosal tissue samples for pathology using monopolar energy, including the avulsion technique, during endoscopic procedures in the gastrointestinal tract, for example the esophagus, stomach, small intestine and colon. Tissue samples will be submitted for pathologic review. This product is intended only for adult populations. {5}------------------------------------------------ #### ട്. Description of Technological Similarities and Differences The US Endoscopy (USE) Avulsion Forceps are very similar to the USE Disposable Hot Biopsy Forceps in intended use and mode of operation. The major differences between the USE Avulsion Forceps and the USE Hot Biopsy Forceps include the indication for use in avulsion technique that the predicate did not have. However, this does not affect the intended use of the device as the intended use of each device is to resect tissue for removal and/or further diagnosis. The proposed and predicate devices have the same technological characteristics in that both devices use electrical energy and manual traction (pull force) to resect tissue. Avulsion is a resection technique that uses a different setting on the electrical surgical generator (Cut as opposed to Coagulation) and a different application of that energy (continuous flow as opposed to modulated flow). This change in technique does not affect the fact that the Avulsion forceps and its predicate both use manual traction and electrical energy to resect tissue for the purpose of removal and/or further diagnosis. Also, the Avulsion Forceps use a single drive cable to control the jaws, whereas the predicate hot biopsy forceps use a dual drive cable system, and the Avulsion Forceps has a toothed cup while the predicate had a plain cup type. The difference in cup design has no effect on the intended use of the devices as both cup types can be used to grasp tissue and deliver current. These proposed changes raise no new concerns of safety and effectiveness when compared to the predicate device. ### 6. Description of Safety and Substantial Equivalence Table 5-1 summarizes the difference between the proposed device and the predicate. | Feature | Proposed<br>Avulsion Forceps | Predicate K925361/A<br>USE Disposable Hot<br>Biopsy Forceps | Comparison | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use /<br>Indications for Use | Disposable biopsy forceps<br>are used to obtain mucosal<br>tissue samples for<br>pathology using<br>monopolar energy,<br>including the avulsion<br>technique, during<br>endoscopic procedures in<br>the gastrointestinal tract,<br>for example the<br>esophagus, stomach, small<br>intestine and colon. Tissue<br>samples will be submitted<br>for pathologic review. The | The US Endoscopy<br>Disposable hot biopsy<br>forceps are used to obtain<br>mucosal tissue samples<br>using monopolar energy<br>during endoscopic<br>procedures. | Proposed device<br>includes<br>indications for use<br>in avulsion<br>techniques.<br>Avulsion is a<br>resection<br>technique that uses<br>a different setting<br>on the electrical<br>surgical generator<br>(Cut as opposed to<br>Coagulation) and a<br>different | | Feature | Proposed<br>Avulsion Forceps | Predicate K925361/A<br>USE Disposable Hot<br>Biopsy Forceps | Comparison | | | product is intended only<br>for adult populations. | | application of that<br>energy<br>(continuous flow<br>as opposed to<br>modulated flow).<br>This change in<br>technique does not<br>affect the fact that<br>the Avulsion<br>Forceps and its<br>predicate both use<br>manual traction<br>and electrical<br>energy to resect<br>tissue for the<br>purpose of<br>removal and/or<br>further diagnosis. | | Mode of Operation | Hand Control | Hand Control | Identical | | Jaws Shape | Serrated<br>Image: Serrated Jaws | Oval<br>Image: Oval Jaws | The proposed<br>avulsion forceps<br>jaw design utilizes<br>toothed cups,<br>while the predicate<br>hot biopsy forceps<br>utilizes plain cups.<br>This difference<br>has no effect on<br>the intended use or<br>safety of the<br>devices as both<br>jaws are used to<br>grasp tissue and<br>deliver current. | | Outer Tube<br>Diameter | $Φ$ 2.6 mm | $Φ$ 2.6 mm | Identical | | Minimum Working<br>Channel | $Φ$ 2.8 mm | $Φ$ 2.8 mm | Identical | | Working Length | 230 cm | 230 cm | Identical | | Energy | Yes | Yes | Identical | | Monopolar/Bipolar | Monopolar | Monopolar | Identical | | Biocompatibility | Yes | Yes | Identical | | Feature | Proposed<br>Avulsion Forceps | Predicate K925361/A<br>USE Disposable Hot<br>Biopsy Forceps | Comparison | | Packaging | Sterile Tyvek pouch | Sterile Tyvek pouch | Identical | | Disposable | Yes | Yes | Identical | | Supplied Sterile | Yes | Yes | Identical | | Accessories | None | None | Identical | | Shelf-life | 1 year | 1 year | Identical | Table 5-1. Predicate Device Comparison Table {6}------------------------------------------------ ## STERIS Traditional 510(k) PREMARKET NOTIFICATION AVULSION FORCEPS {7}------------------------------------------------ ### STERIS Traditional 510(k) PREMARKET NOTIFICATION AVULSION FORCEPS Table 5-2 summarizes the verification activities that were performed with respective acceptance criteria to ensure that this modification does not affect the safety or effectiveness of the Avulsion Forceps device. | Table 5-2. Summary of Verification Activities. | | | |------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------| | Testing | Acceptance Criteria | Proposed Avulsion<br>Forceps | | Structural<br>Integrity Testing | All Avulsion Forceps structural components<br>must, minimally, meet or exceed the structural<br>integrity limits by the predicate hot biopsy<br>forceps | PASS | | Dimensional | All Avulsion Forceps dimensional components<br>must meet the dimensional specification for<br>each forceps | PASS | | Functional<br>Testing | The Avulsion Forceps must function as<br>intended in tortuous conditions | PASS | | Simulation<br>Testing | The Avulsion Forceps retrieve tissue samples | PASS | | Electrical<br>Testing | All Avulsion Forceps electrical requirements<br>must meet current industry standards according<br>to IEC 60601-2-2 (3rd Edition) | PASS | | Biocompatibility<br>Testing | The patient contacting components of the<br>device must be biologically safe to use in<br>accordance with ISO-10993-1 | PASS | ## 7. Conclusion Shelf Life for 12 months Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs at least as well as the legally marketed predicate device (K925361), Class II (21 CFR 876.4300), product code KGE. The Avulsion Forceps must maintain device performance through a shelf life of 1 year PASS
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...