← Product Code [KGE](/submissions/GU/subpart-e%E2%80%94surgical-devices/KGE) · K022751

# REPROCESSED USED DISPOSABLE HOT AND COLD BIOPSY FORCEPS (K022751)

_Medical Device Services · KGE · Nov 4, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KGE/K022751

## Device Facts

- **Applicant:** Medical Device Services
- **Product Code:** [KGE](/submissions/GU/subpart-e%E2%80%94surgical-devices/KGE.md)
- **Decision Date:** Nov 4, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

Cold biopsy forceps are intended to be used through an endoscope to remove polyps and/or tissue specimens throughout the alimentary tract. Hot Biopsy Forceps are intended to be used through an endoscope to cauterize and remove polyps and/or tissue specimens throughout the alimentary tract.

## Device Story

Device consists of reprocessed single-use gastrointestinal biopsy forceps; includes both cold and hot biopsy variants. Input: manual operation by clinician via endoscope working channel. Mechanism: mechanical jaw assembly for tissue collection; hot biopsy variants include electrical connectivity for cauterization. Output: excised tissue samples. Used in clinical settings (endoscopy suites) by physicians. Reprocessing involves cleaning and sterilization of used disposable devices to render them patient-ready. Benefit: provides cost-effective access to biopsy instrumentation for diagnostic and therapeutic tissue sampling.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and validation of the reprocessing protocol to ensure the device remains safe and effective for its intended use.

## Technological Characteristics

Mechanical biopsy forceps with jaw mechanism; hot biopsy variants include electrical connectors for cauterization. Dimensions vary by model (1.8mm to 3.3mm jaw O.D.; 100cm to 240cm length). Materials are consistent with original single-use predicate. Sterilization is performed as part of the reprocessing protocol. No software or electronic algorithms involved.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Microvasive Radial Jaw Biopsy Forceps ([K932266](/device/K932266.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Medical Device Services, I

144 West Brigham Road, Bldg. E. St. George, UT 84790

K022751/
Page 1 of 1

Tel: (435) 652-307. Fax: (435) 652-3087

### 510(k) SUMMARY

4 2002 NOV

#### Reference: Medical Device Services, Incorporated Section 510(k) Notification MDS Reprocessed Biopsy Forceps

| Classification name:    | Instrument, Biopsy, Mechanical, Gastrointestinal                                                   |
|-------------------------|----------------------------------------------------------------------------------------------------|
| Common/Usual Name:      | Gastrointestinal Biopsy Forceps                                                                    |
| Proprietary Name:       | MDS Reprocessed Biopsy Forcep                                                                      |
| Establishment Reg. No.: | 1724309                                                                                            |
| Classification:         | The FDA has classified gastrointestinal biopsy forceps as a<br>Class II device in 21 CFR 876.1075. |

MDS intends to market Reprocessed Used Disposable Biopsy Forceps. Reprocessing Biopsy Forceps is performed by MDS to MDS protocol Number 40003.

"Reprocessed," means all operations performed to render a contaminated single-use device patient ready (Enforcement Priorities for Single-Use Devices Reprocessed by Third Party Reprocessors and Hospitals). MDS is a "third party reprocessor" and reprocesses used, single-use medical devices.

MDS believes that Used Disposable Biopsy Forceps can be considered "reusable - by MDS" as defined in the Food and Drug Administration Compliance Policy Guide #7124.16: they are able to withstand the necessary cleaning and sterilization process, the physical characteristics or quality of the device will not be adversely effected, and the device remains safe and effective for its intended use.

Biopsy forceps are long instruments with a small jaw type mechanism on one end and a handle mechanism on the other. Forceps are designed to collect tissue samples.

Medical Device Services, Reprocessed Used Disposable Biopsy Forceps are substantially equivalent to disposable biopsy forceps currently marketed new by Microvasive under 510(k) 932266.

{1}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

| NOV 4 2002                                                                                                                    | Food and Drug Administration<br>9200 Corporate Boulevard<br>Rockville MD 20850     |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Medical Device Services<br>c/o Mr. Mark W. Aldana<br>President<br>Adven Medical, Inc.<br>1001 Slaton Hwy.<br>LUBBOCK TX 79404 | Re: K022751                                                                        |
|                                                                                                                               | Trade/Device Name: Reprocessed Used Disposable<br>Biopsy Forceps (SEE ENCLOSURE 1) |
|                                                                                                                               | Regulation Number: 21 CFR §876.4300                                                |
|                                                                                                                               | Regulation Name: Endoscopic electrosurgical unit<br>and accessories                |
|                                                                                                                               | Regulatory Class: II                                                               |
|                                                                                                                               | Product Code: 78 KGE                                                               |
|                                                                                                                               | Regulation Number: 21 CFR §876.1075                                                |
|                                                                                                                               | Regulation Name: Gastroenterology-urology biopsy instrument                        |
|                                                                                                                               | Regulatory Class: I                                                                |
|                                                                                                                               | Product Code: 78 FCL                                                               |
|                                                                                                                               | Dated: August 9, 2002                                                              |
|                                                                                                                               | Received: August 19, 2002                                                          |

Dear Mr. Aldana:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | : (301) 594-4591 |
|----------------------------------|------------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616   |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616   |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | : (301) 594-4654 |
| Other                            | (301) 594-4692   |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## ENCLOSURE 1 K022751

## CLASS I: FCL 21CFR §876.1075

## Medical Device Services Reprocessed Used Disposable Mechanical and Electric Biopsy Forceps

Manufacturer: MICROVASIVE

Radial Jaw* 3 Max Capacity Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1586                 | 3.3              | 160            | 3.8                     | Yellow        |
| 1587 with needle     | 3.3              | 160            | 3.8                     | Yellow        |
| 1588                 | 3.3              | 240            | 3.3                     | Orange        |
| 1589 with needle     | 3.3              | 240            | 3.8                     | Orange        |

Radial Jaw* II Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1562                 | 2.2              | 160            | 2.8                     | Yellow        |
| 1563 with needle     | 2.2              | 160            | 2.8                     | Yellow        |
| 1564                 | 2.2              | 240            | 2.8                     | Orange        |
| 1565 with needle     | 2.2              | 240            | 2.8                     | Orange        |

Radial Jaw* LC II Large Capacity Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1591                 | 2.2              | 160            | 2.8                     | Yellow        |
| 1592 with needle     | 2.2              | 160            | 2.8                     | Yellow        |
| 1593                 | 2.2              | 240            | 2.8                     | Orange        |
| 1594 with needle     | 2.2              | 240            | 2.8                     | Orange        |

Radial Jaw Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1260                 | 2.2              | 160            | 2.8                     | Yellow        |
| 1263 with needle     | 2.2              | 160            | 2.8                     | Yellow        |
| 1271                 | 2.2              | 240            | 2.8                     | Orange        |
| 1265 with needle     | 2.2              | 240            | 2.8                     | Orange        |

{4}------------------------------------------------

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1260                 | 2.2              | 160            | 2.8                     | Yellow        |
| 1263 with needle     | 2.2              | 160            | 2.8                     | Yellow        |
| 1271                 | 2.2              | 240            | 2.8                     | Orange        |
| 1265 with needle     | 2.2              | 240            | 2.8                     | Orange        |
| 1582                 | 3.3              | 160            | 3.8                     | Yellow        |
| 1583 with needle     | 3.3              | 160            | 3.8                     | Yellow        |
| I 584                | 3.3              | 240            | 3.8                     | Orange        |
| 1585 with needle     | 3.3              | 240            | 3.8                     | Orange        |

Radial Jaw MC 3.3 Single-Use Max Capacity Biopsy Forceps

Radial Jaw LC Large Capacity Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1273                 | 2.2              | 240            | 2.8                     | Orange        |
| 1274 with needle     | 2.2              | 240            | 2.8                     | Orange        |

Radial Jaw GP Gastro-pediatric Single-Use Biopsy Forceps

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1281                 | 1.8              | 160            | 2.0                     | Yellow        |
| 1286 with needle     | 1.8              | 160            | 2.0                     | Yellow        |

Multibite''' Multiple Sample Single-Use Biopsy Forceps

| Manufacturer Numbers | Length<br>(cm) | Working Channel<br>(mm) |
|----------------------|----------------|-------------------------|
| 1010                 | 160            | 2.8                     |
| 1012                 | 240            | 2.8                     |

Radial Jaw 3"' Single-Use Biopsy Forceps

| Manufacturer Numbers     | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|--------------------------|------------------|----------------|-------------------------|---------------|
| 1534 (Box 5)             | 2.2              | 160            | 2.8                     | Yellow        |
| 1535 with needle (Box 5) | 2.2              | 160            | 2.8                     | Yellow        |
| 1536 (Box 5)             | 2.2              | 240            | 2.8                     | Orange        |
| 1537 with needle (Box 5) | 2.2              | 240            | 2.8                     | Orange        |

{5}------------------------------------------------

| Manufacturer Numbers | Jaw O.D.<br>(mm) | Length<br>(cm) | Working Channel<br>(mm) | Color<br>Code |
|----------------------|------------------|----------------|-------------------------|---------------|
| 1281                 | 1.8              | 160            | 2.0                     | Yellow        |
| 1286 with needle     | 1.8              | 160            | 2.0                     | Yellow        |
| 1596                 | 2.2              | 160            | 2.8                     | Yellow        |
| 1597 with needle     | 2.2              | 160            | 2.8                     | Yellow        |
| 1598                 | 2.2              | 240            | 2.8                     | Orange        |
| 1599 with needle     | 2.2              | 240            | 2.8                     | Orange        |

. . . .

## Radial Jaw 3 Large Capacity Single-Use Biopsy Forceps

# CLASS II; KGE; 21CFR §876.4300

Radial Jaw''' 3 Single-Use Hot Biopsy Forceps

| Manufacturer Numbers                  | Jaw O.D.<br>(mm) | Length<br>(cm) |
|---------------------------------------|------------------|----------------|
| 1550 (Box 5) (Olympus® Connector)     | 2.2              | 240            |
| 1551 (Box 5) (Microvasive® Connector) | 2.2              | 2.40           |

. · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

Radial Jaw Hot Biopsy Forceps

| Manufacturer Numbers                  | Jaw O.D.<br>(mm) | Length<br>(cm) |
|---------------------------------------|------------------|----------------|
| 1274 (Box 5) (Microvasive® Connector) | 2.2              | 240            |
| 1277 (Box 5) (Olympus® Connector)     | 2.2              | 2.40           |

{6}------------------------------------------------

| 1266             | 1.8 | 100 | 2.0 | Blue |
|------------------|-----|-----|-----|------|
| 1267             | 2.2 | 100 | 2.8 | Blue |
| 1268 with needle | 2.2 | 100 | 2.8 | Blue |
| 1269 with needle | 1.8 | 100 | 2.0 | Blue |

| 1531 | with needle (Box 5) | 2.2 | 100 | 2.8 | Blue |
|------|---------------------|-----|-----|-----|------|
| 1266 |                     | 1.8 | 100 | 2.0 | Blue |
| 1267 |                     | 2.2 | 100 | 2.8 | Blue |
| 1268 | with needle         | 2.2 | 100 | 2.8 | Blue |
| 1269 | with needle         | 1.8 | 100 | 2.0 | Blue |

100

2.2

2.8

Blue

XX 1260-20 Radial Jaw 20-pack 2.2 160 2.8 Yellow

1530 (Box 5)

XX 1263-20 Radial Jaw 20-pack with needle 2.2 160 2.8 Yellow

XX 1265-20 Radial Jaw 20-pack with needle 2.2 240 2.8 Orange

XX1267-20 Radial Jaw 20-pack 2.2 240 2.8 Orange

{7}------------------------------------------------

Page 1 of

510(k) Number:

K022751

Device Name:

Reprocessed Used Disposable Biopsy Forceps

Medical Device Services (MDS) intends to reprocess used disposable hot and cold biopsy forceps manufactured by Micovasive.

Cold biopsy forceps are intended to be used through an endoscope to remove polyps and/or tissue specimens throughout the alimentary tract

Hot Biopsy Forceps are intended to be used through an endoscope to cauterize and remove polyps and/or tissue specimens throughout the alimentary tract.

MDS reprocessed biopsy forceps are disposable unless reprocessed again by MDS.

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

David A. Szymon

on Sign-Off of Reproductive. Abdominal onical Davies

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KGE/K022751](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/KGE/K022751)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
