K172299 · Ths International, Inc. D/B/A Accordion Medical · FFL · Sep 19, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K172299
Device Name
Accordion Piccolo Stone Management Device
Applicant
Ths International, Inc. D/B/A Accordion Medical
Product Code
FFL · Gastroenterology, Urology
Decision Date
Sep 19, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Accordion Piccolo Stone Management Device is intended to be used endoscopically to entrap and remove calculi and other foreign objects from the urinary tract and to guide instrumentation within the ureteral tract.
Device Story
Device is a ureteral stone dislodger used endoscopically; operated by physicians (urologists) in clinical settings. Input: manual advancement through cystoscope working channel under direct vision or fluoroscopy. Operation: physician pulls handle to advance movable core wire, deploying a film membrane within the ureteral lumen. Output: deployed membrane acts as a physical barrier (backboard) to prevent proximal stone migration during lithotripsy or to sweep stone fragments into the bladder. Benefits: prevents stone migration into the kidney during surgical procedures; facilitates fragment removal.
Clinical Evidence
Bench testing only. Testing included torque handle grip tests, Nitinol core wire floppiness tests, and Nitinol poke strength tests to ensure compatibility and performance of the modified components.
Technological Characteristics
Ureteral stone dislodger consisting of a film membrane, two-part guidewire (inner/outer shaft), and handle. Materials: Polyimide braided outer shaft, Nitinol core wire. Dimensions: 94cm length, 0.032" OD shaft. Deployment: Mechanical, via manual handle actuation. Biocompatible materials. Non-powered, manual device.
Indications for Use
Indicated for endoscopic entrapment and removal of calculi and foreign objects from the urinary tract and to guide instrumentation within the ureteral tract in patients requiring stone management.
Regulatory Classification
Identification
A ureteral stone dislodger is a device that consists of a bougie or a catheter with an expandable wire basket near the tip, a special flexible tip, or other special construction. It is inserted through a cystoscope and used to entrap and remove stones from the ureter. This generic type of device includes the metal basket and the flexible ureteral stone dislodger.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a stacked formation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2017
THS International, Inc. dba Accordion Medical David Ouigley President 9465 Counselors Row, Suite 200 Indianapolis. Indiana 46240
K172299 Re:
Trade/Device Name: Accordion Piccolo Stone Management Device Regulation Number: 21 CFR 876.4680 Regulation Name: Ureteral Stone Dislodger Regulatory Class: Class II Product Code: FFL Dated: September 5, 2017 Received: September 11, 2017
Dear David Quigley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Charles Viviano -S
For
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172299
Device Name
Accordion Piccolo Stone Management Device
Indications for Use (Describe)
The Accordion Piccolo Stone Management Device is intended to be used endoscopically to entrap and remove calculi and other foreign objects from the urinary tract and to guide instrumentation within the ureteral tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for THS International. The logo consists of a blue square with a white abstract design on the left, followed by the letters "THS" in large, bold, blue font. Below the letters, the word "INTERNATIONAL" is written in a smaller, blue font.
# 510(K)SUMMARY
# THS International, Inc. dba Accordion Medical's Accordion Piccolo Stone Management Device
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
THS International, Inc. dba Accordion Medical
9465 Counselors Row Suite 200
Indianapolis, IN 46240
| Phone: | (317) 759-2869 |
|-----------------|----------------|
| Facsimile: | (317) 688-8373 |
| Contact Person: | David Quigley |
| Date Prepared: | June 30, 2017 |
#### Device Information
| Proprietary Name: | Accordion Piccolo Stone Management Device (PA |
|----------------------|-----------------------------------------------|
| | 1205-27-10) |
| Classification Name: | Ureteral Stone Dislodger |
| Product Code: | FFL |
| Device Class: | 2 |
#### Predicate Devices/ Substantial Equivalence
Accordion Stone Management Device (PA 1205-06-10) (K052048) Accordion CoAx Stone Control Device (PA 1305-01-10, PA 1305-01-15) (K082803)
#### Intended Use/Indications for Use
The Accordion Piccolo Stone Management Device is intended to be used endoscopically to entrap and remove calculi and other foreign objects from the urinary tract and to guide instrumentation within the ureteral tract.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for THS International. The logo consists of a square with a white abstract design inside, followed by the letters "THS" in large, bold font. Below the letters "THS" is the word "INTERNATIONAL" in a smaller font. The logo is a dark blue color.
#### Rational for Special 510k
| Accordion Stone<br>Management Device<br>vs Accordion Piccolo | Effect on<br>Intended Use | Alters the fundamental scientific technology of the<br>device |
|-----------------------------------------------------------------|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Length of device:<br>From 150cm to 94cm | No Change to<br>Intended Use | No change to fundamental scientific technology of the device: Device still spans the ureteral length and is longer than average flexible and semi-rigid scopes<br>This length falls within already approved length parameters stated in the Accordion Stone<br>Management Device 510k submission (K050428) |
| Diameter of shaft:<br>From 0.035" OD to<br>0.032"OD | No Change to<br>Intended Use | No change to fundamental scientific technology of the device: still delivered with a cystoscope and can be used to pass other instrumentation<br>This diameter falls within already approved diameter parameters stated in the Accordion Stone<br>Management Device 510k submission (K050428) |
| Material of Shaft:<br>From Pebax to<br>Polyimide | No Change to<br>Intended Use | No change to fundamental scientific technology of the device: Still biocompatible material: same material used in the CoAx Stone Control Device (K082803 on Nov. 13, 2008) with same Intended Use |
| Material of Core<br>Wire:<br>From Stainless steel to<br>Nitinol | No Change to<br>Intended Use | No change to fundamental scientific technology of the device: Same nitinol used in wires for Accordion occlusion film. |
### Device Description
The Accordion Piccolo's film membrane is attached to a two-part (inner and outer shaft) guidewire and deploys within the lumen of the ureter. The device can be advanced via the working channel of a cystoscope, and advanced under direct vision or fluoroscopy until the tip is proximal to the kidney stone or foreign object. To contain the stone during lithotripsy or entrap a foreign object, the movable core of the guidewire component is advanced by pulling on the handle, which causes the film membrane to deploy within the lumen of the ureter and prevent proximal movement of the stone or object during the surgical procedures. The physician may maintain the film membrane in the deployed position and "sweep" the stone fragments down the ureter and into the bladder or use the deployed film as a "backboard" to prevent migration of the stone fragments into the kidney as he/she removes the fragments with a basket or other device.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for THS International. The logo consists of a square with a white design inside, followed by the letters "THS" in large, bold font. Below the letters, the word "INTERNATIONAL" is written in a smaller font size.
#### Technological Characteristics
The Accordion Piccolo is fundamentally a shorter version of the Accordion Stone Management device (K052048). Its core wire, or inner shaft, uses Nitinol rather than stainless steel. The nitinol allows a more flexible tip which some Doctors prefer. Testing of the core wire was conducted to ensure that while the tip is more flexible, the device still performs the same functions. The Accordion Piccolo core wire is the same nitinol found in the occlusion film.
The braided shaft, or outer shaft. is the only material difference between the Accordion Piccolo and the Accordion Stone Management Device. The braided shaft of the Accordion Piccolo has the same polyimide material used in manufacturing of the braided shaft of the CoAx Stone Control Device (K082803). The change in materials allowed us a smaller diameter, preferred by doctors, without changing the functionality of the device, allowing the same structure but with a smaller diameter.
The occlusion film, handle, introducer, and all packaging components are the same as the Accordion Stone Management Device. Testing was conducted to ensure that the change diameter would not impact the functionality of the current Accordion Stone Control handle. The film, introducer and packaging are unchanged by the length or diameter changes.
#### Performance Data
50-0111-01 - Accordion Piccolo Benchtop Report (Full Device)
- Torque Handle grip test (Core wire and handle compatibility)
- Benchtop Test Accordion Niti Floppiness test (Core wire testing)
- Benchtop test Accordion Niti poke strength test (Core wire testing)
#### Conclusion
The Accordion Piccolo device testing demonstrates that the device is as safe, as effective, and performs as well as the Accordion Stone Management Device.
#### Previous Submissions for this device
N/A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.