The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
Device Story
Endoscopic mucosal resection (EMR) device for upper GI tract; kit/pack configuration. Components: ligation handle, loading device, connector tube, cylindrical barrel with 5 bands, deployment cord, hexagonal snare. Operation: manual ligation of tissue followed by diathermic resection using snare connected to electrosurgical generator (e.g., ERBE VIO 300-D, US Endoscopy gi4000). Used in clinical settings by physicians. Universal connector allows attachment to endoscopes lacking luer connections. Band lock secures bands during shipping. Black band indicates proximity to end of supply. Benefits: provides options for treatment area size; facilitates tissue resection; extended shelf life (24 months).
Clinical Evidence
Bench testing only. Includes verification testing (measurements, visual inspection, shipping, packaging), biocompatibility testing (ISO 10993-4, -5, -10, -11; USP38-NF33), and simulated use testing.
Indicated for patients undergoing endoscopic mucosal resection in the upper GI tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2021
Steris Corporation Carroll Martin Regulatory Affairs Director 5976 Heislev Road Mentor, OH 44060
Re: K212860
> Trade/Device Name: SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4), SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4), SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with kit) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical unit and accessories Regulatory Class: Class II Product Code: FDI Dated: September 7, 2021 Received: September 8, 2021
Dear Carroll Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212860
Device Name
SmartBand® EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4) SmartBand® EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4) SmartSnare™ EMR Hexagonal Snare (SS-230-1 or packaged with Kit)
Indications for Use (Describe)
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the first page of a document with the identifier K212860 and page number 1 of 6. The document features the STERIS logo in a large, bold font. Below the company name is a graphic of several horizontal, curved lines, resembling waves.
# 510(k) Summary for the SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4) SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4) SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit)
STERIS Corporation 5976 Heisley Road Mentor, OH 44060
Contact: Carroll Martin Regulatory Affairs Director Tel: 440-358-6259 Email: Carroll Martin@steris.com
Submission Date: September 7, 2021
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
#### 1. Device Name
| Trade Name: | SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4)<br>SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4)<br>SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class: | Class II |
| Regulation Name: | Endoscopic Electrosurgical Unit and Accessories |
| Common/usual Name: | Ligation Device |
| Regulation Number: | 21 CFR 876.4300 |
| Product Code: | FDI |
| Review Panel: | Gastroenterology/Urology |
#### 2. Predicate Device
SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12) SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12) SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit) All cleared under K190512
#### 3. Device Description
The SmartBand EMR Device is an endoscopic mucosal resection device that is provided as a kit or a pack that is designed for ligation assisted endoscopic mucosal resection (EMR) in the upper gastrointestinal (GI) tract. The SmartBand EMR Kit consists of the SmartBand Ligator Kit, supplied non-sterile and the SmartSnare EMR Hexagonal Snare, supplied sterile. The ligator kit contains a ligation handle, loading device, connector tube and a cylindrical barrel with 5 bands separated by a deployment cord. The SmartBand EMR Pack consists of a cylindrical barrel with 5 bands separated by a deployment cord. The SmartBand EMR Pack can be used when more bands are needed to complete a procedure. The SmartBand EMR Hexagonal Snare consists of a handle, finger slide and a hexagonal snare loop on the distal end.
#### 4. Indications for Use
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
#### న్. Technological Characteristics Comparison Table
A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1.
{6}------------------------------------------------
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The SmartBand® EMR Device is<br>intended for endoscopic mucosal<br>resection in the upper GI tract. | The SmartBand® EMR Device is<br>intended for endoscopic mucosal<br>resection in the upper GI tract. | Same |
| Device<br>Components | Ligation Device<br>5 non-latex bands (4 green and 1<br>blue)<br>Trigger cord<br>Loading Device<br>Flush Tube<br>Deployment Handle<br>Pentax Adaptor | Ligation Device<br>5 non-latex bands (4 green and 1<br>black)<br>Trigger cord<br>Loading Device<br>Flush Tube<br>Deployment Handle<br>Pentax Adaptor<br>Universal Connector<br>Band Lock | Similar. The color<br>change of one band<br>and the addition of<br>the Universal<br>Connector and<br>Band Lock do not<br>affect the intended<br>use or fundamental<br>technology of the<br>device. |
| | Snare<br>Stainless steel snare loop<br>Shaft<br>Finger Slide<br>Snare Handle Body | Snare<br>Stainless steel snare loop<br>Shaft<br>Finger Rings<br>Snare Handle Body | Same |
| Materials of<br>Construction | Snare<br>Snare loop - stainless steel<br>(patient contacting)<br>Sheath - Teflon (patient<br>contacting)<br>Core Wire - Nitinol (patient<br>contacting)<br>Stainless steel (snare loop) Strain<br>relief - polyurethane (non-patient<br>contacting)<br>Finger slide - ABS (non-patient<br>contacting)<br>Snare handle body - ABS (non-<br>patient contacting)<br>Ligation Device<br>Bands - Non-latex synthetic<br>polyisoprene (patient contacting)<br>Barrel - Polycarbonate (patient<br>contacting)<br>Trigger cord - Liquid crystal<br>polymer (patient contacting)<br>Loading device - Stainless steel<br>(non-patient contacting)<br>Flush tube - ABS (non-patient<br>contacting)<br>Handle - ABS (non-patient<br>contacting) | Snare<br>Snare loop - stainless steel<br>(patient contacting)<br>Sheath - Teflon (patient<br>contacting)<br>Core Wire - Nitinol (patient<br>contacting)<br>Stainless steel (snare loop) Strain<br>relief - polyurethane (non-<br>patient contacting)<br>Finger slide - ABS (non-patient<br>contacting)<br>Snare handle body - ABS (non-<br>patient contacting)<br>Ligation Device<br>Bands - Non-latex synthetic<br>polyisoprene (patient contacting)<br>Barrel - Polycarbonate (patient<br>contacting)<br>Trigger cord - Liquid crystal<br>polymer (patient contacting)<br>Loading device - Stainless steel<br>(non-patient contacting)<br>Flush tube - ABS (non-patient<br>contacting)<br>Handle - ABS (non-patient<br>contacting)<br>Universal Connector - Silicone<br>(non-patient contacting)<br>Band Lock - Polyurethane (non-<br>patient contacting) | Similar. The<br>materials of the<br>Universal<br>Connector and<br>Band Lock do not<br>affect the intended<br>use or fundamental<br>technology of the<br>device. |
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
| # of bands/color | 5 non-latex bands (4 green and 1<br>blue) | 5 non-latex bands (4 green and 1<br>black) | Similar. This<br>change aids the<br>user in determining<br>when they are<br>getting close to<br>using all of the<br>bands, as the black<br>band is the second<br>to last band in line<br>on the barrel. The<br>change in color<br>does not affect the<br>intended use or<br>fundamental<br>technology of the<br>device. |
| Barrel<br>Dimensions | Inside diameter:<br>10.75mm and 12.0mm<br>Depth: 7.75mm | Inside diameter<br>9.4mm, 10.75mm, and12.0mm<br>Depth 7.75mm (10.75mm and<br>12.0mm sizes)<br>Depth 9.5mm (9.4mm size) | Similar. This<br>change gives the<br>user more options<br>based on the size<br>of the treatment<br>area. |
| Sterile/Non-<br>sterile | Sterile (snare) and Non-sterile<br>(ligation device) | Sterile (snare) and Non-sterile<br>(ligation device) | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single used; snare is used with<br>diathermic energy | Single used; snare is used with<br>diathermic energy | Same |
| Target<br>Population | Patients undergoing an endoscopic<br>mucosal resection procedure | Patients undergoing an<br>endoscopic mucosal resection<br>procedure | Same |
| Energy<br>Used/Delivered | Diathermic energy (snare) | Diathermic energy (snare) | Same |
| Method of<br>Application | Manual | Manual | Same |
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
| Packaging<br>Change | No band lock | Band lock present to secure<br>the bands during shipping | Different. The band lock<br>prevents the bands from<br>dislodging from the barrel<br>during shipping. The<br>additional box is a<br>convenience for the user.<br>These changes have no effect<br>on how the device is used or<br>its fundamental technology. |
| | No stand-alone box for the<br>snare | Addition of stand-alone box<br>for snare | |
| Compatible<br>endoscopes | Pentax, Fujinon and Olympus | Pentax, Fujinon and Olympus | Same<br>Similar. The function of the<br>electrical snare of the<br>proposed device remains the<br>same in that it is attached to<br>the electrical surgical<br>generator to allow the flow of<br>electrical energy to the snare<br>loop for the purpose of<br>resecting tissue. The use of<br>the modified device with the<br>US Endoscopy gi4000<br>Electrosurgical Generator<br>does not change the intended<br>use or fundamental<br>technology of the modified<br>device. |
| Compatible<br>electrical surgical<br>generators | ERBE VIO 300-D | ERBE VIO 300-D<br>US Endoscopy gi4000<br>Electrosurgical Generator | |
| Addition of<br>universal<br>connector | Not present. | Present | Different. The universal<br>connector allows the user to<br>attach the ligation device<br>handle directly to the<br>accessory ports of endoscopes<br>that do not have a luer<br>connection on the accessory<br>port. This change has no<br>effect on how the device is<br>used or its fundamental<br>technology. |
| Expiration<br>Dating | 12 months | 24 months | Different. The extended shelf<br>life of the proposed device<br>provides the user greater<br>flexibility related to device<br>use. It has no effect on how<br>the device is used or its<br>fundamental technology. |
Table 1. Technological Characteristics Comparison Table
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### 6. Summarv of Non-Clinical Performance Testing
Verification testing consisting of measurements, visual inspection, ship testing and packaging testing.
Biocompatibility testing in accordance with ISO 10993-4, -5, -10 and -11 and USP38-NF33.
Simulated use testing.
#### 7. Conclusion
The minor changes that are the subject of this submission have no impact on the fundamental scientific technology of the device or how it is used. STERIS has determined that the modified device is as safe and effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.