The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper G1 tract.
Device Story
Device facilitates ligation-assisted endoscopic mucosal resection (EMR) in upper GI tract. System comprises cylindrical barrel with 5 bands, deployment cord, ligation handle, loading device, and electrosurgical hexagonal snare. Operated by gastroenterologist via endoscope accessory channel. Procedure involves attaching barrel to distal end of endoscope; suctioning mucosal tissue into barrel; deploying band to create pseudopolyp; passing electrosurgical snare through accessory channel to capture pseudopolyp; and performing resection via electrocautery. Device enables tissue removal for diagnostic or therapeutic purposes. Used in clinical settings.
Clinical Evidence
Bench testing only. Testing included system suction delivery efficiency, band ultimate tensile force, and snare loop to pull wire tensile strength. Biocompatibility testing per ISO 10993-1 (Cytotoxicity, Sensitization, Acute Systemic Toxicity, USP Physicochemical) showed acceptable results.
Technological Characteristics
Accessory for endoscopes. Components: cylindrical barrel, bands, deployment cord, ligation handle, electrosurgical snare. Sterilization: ETO (EN ISO 11135-1, EN ISO 11138-2). Biocompatibility: ISO 10993-1 compliant. Mechanical operation via manual handle rotation and suction.
Indications for Use
Indicated for endoscopic mucosal resection in the upper GI tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2019
Intelligent Endoscopy Melissa Clark President 4740 Commercial Park Ct, Suite 1 Clemmons, NC 27012
Re: K190512
Trade/Device Name: SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12) SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12) SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit) Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FDI Dated: May 2, 2019 Received: May 6, 2019
Dear Melissa Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K190512
Device Name
SmartBand® EMR Kit (SB-EMR-K, SB-EMR-K-12) SmartBand® EMR Pack (SB-EMR-P, SB-EMR-P-12) SmartSnare™ EMR Hexagonal Snare (SS-230-1 or packaged with Kit)
Indications for Use (Describe)
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper G1 tract.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CER 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------|
| <div> <span>☐</span> <span>Over-The-Counter Use (21 CER 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | Intelligent Endoscopy, LLC<br>4740 Commercial Park Ct, Suite 1<br>Clemmons, NC 27012<br>Establishment Registration Number: 3011324403 |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Melissa Clark<br>President<br>Intelligent Endoscopy, LLC<br>Office: (336) 608-4375<br>Fax: (336) 893-7268<br>mclark@intelligentendoscopy.com |
| Date Prepared: | February 27, 2019 |
| Trade Name: | SmartBand® EMR Kit (SB-EMR-K, SB-EMR-K-12)<br>SmartBand® EMR Pack (SB-EMR-P, SB-EMR-P-12)<br>SmartSnare™ EMR Hexagonal Snare (SS-230-1 or packaged with Kit) |
| Regulation Number: | 21 CFR 876.4300 |
| Regulation Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | FDI |
| Classification: | Class II |
| Review Panel: | Gastroenterology / Urology |
| Predicate Device: | Captivator™ EMR Device (K140726) |
# Device Description:
The SmartBand® EMR Device is an endoscopic mucosal resection device that is provided as a Kit or a Pack that is designed for ligation assisted endoscopic mucosal resection (EMR) in the upper gastrointestinal (GI) tract. The SmartBand® EMR Kit consists of the SmartBand Ligator Kit, supplied non-sterile and the SmartSnare™ EMR Hexagonal Snare, supplied sterile. The ligator kit contains a ligation handle, loading device, connector tube, and a cylindrical barrel with 5 bands separated by a deployment cord. The SmartBand® EMR Pack consists of a cylindrical barrel with 5 bands separated by a deployment cord, which can be used when more bands are needed to complete the procedure.
# Indications for Use:
The SmartBand® EMR Device is indicated for endoscopic mucosal resection in the upper GI tract.
# Technological Characteristics:
The SmartBand® EMR Device is an accessory to be used with an appropriate endoscope of the gastroenterologist's choice. The deployment cord is passed retrograde through the accessory channel of the endoscope using the loading device and then assembled to the ligation handle. The cylindrical barrel with 5 bands, attaches to the distal end of the endoscope. When the handle is turned, a band is deployed around the mucosal tissue creating a pseudopolyp. The ligation handle allows for the passage of the electrosurgical snare through the accessory channel of the endoscope. The
{4}------------------------------------------------
SmartSnare™ wire loop is placed around the pseudopolyp and the targeted tissue is removed using electrocautery. The SmartBand® EMR Device shares substantially equivalent design features and function with the predicate.
The SmartBand® EMR Device has the same intended use and is placed using the same methodology as the predicate device. The SmartBand® EMR Device and the predicate device use a barrel / cap to suction and capture tissue during an endoscopic mucosal resection procedure, band the tissue into a pseudopolyp, and use an electrosurgical snare to resect the tissue. The SmartBand® EMR Device and the predicate device are both used to perform ligation-assisted endoscopic mucosal resections.
# Performance Testing Summary:
The SmartBand® EMR Kit is provided in a box with two packaged components. The SmartBand® Ligator Kit, provided non-sterile, and the SmartSnare™ EMR Hexagonal Snare, provided sterile. The SmartBand® EMR Pack is provided non-sterile in a pouch. All components have been performance bench tested to ensure safe and effective function of the device. The non-clinical comparative performance testing successfully determined that the SmartBand® EMR Device is substantially equivalent to the predicate. This testing included but was not limited to system suction delivery efficiency with the snare, band ultimate tensile force, and snare loop to pull wire tensile strength.
The SmartSnare™ EMR Hexagonal Snare is provided with the Kit in a separate sterile pouch. Testing according to ENISO 11135-1 and EN ISO 11138-2, sterilization of health care products, verifies that the ETO sterilization method and parameters that are used to sterilize the SmartSnare™ EMR Hexagonal Snare. The predicate device also uses ETO sterilization.
#### Biocompatibility Testing Summary:
Representative production units of The SmartBand® EMR Device were submitted to complete biocompatibility studies in accordance to ISO 10993-1. The patient contacting components of the device (the barrel, bands, cord and snare) were subjected to and showed acceptable results to Cytotoxicity, Sensitization, Acute Systemic Toxicity, and USP Physicochemical testing.
# Conclusion Summary of Substantial Equivalence:
There have been no prior submissions for the subject device. Intelligent Endoscopy has demonstrated in this original Traditional 501(k) that the SmartBand® EMR Device is substantially equivalent to the Captivator™ EMR Device (K140726) that is currently available in the market.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.