The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.
Device Story
ClearGrasp Snare is a monopolar electrosurgical instrument for endoscopic mucosal resection (EMR). Device consists of flexible rope wire loop, catheter tube, and handle with control knobs (handle and slider). Operated by physician during endoscopic procedures; device passes through endoscope working channel. User selects loop shape (oval or crescent), size, and rotational capability based on lesion characteristics. When connected to an electrosurgical generator, loop delivers monopolar current to resect targeted GI tract lesions. Single-patient use; supplied sterile. Benefits include precise grasping and resection of polyps, facilitating removal during standard GI endoscopy.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1, -5, -10, -11. Performance testing per IEC 60601-2-2 and ANSI/AAMI HF18 (appearance, dimension, continuity, tensile strength, withstand voltage, leakage current). Sterilization validation per ISO 11737 and shelf-life testing per ASTM F1980.
Technological Characteristics
Monopolar electrosurgical snare. Materials: SUS 304, PTFE, ABS. Dimensions: 1600-2400mm working length; 6-30mm loop sizes. Connectivity: Connects to external electrosurgical generator. Sterilization: Ethylene Oxide. Design: Flexible catheter with handle-actuated loop extension/rotation. No software/algorithm.
Indications for Use
Indicated for endoscopic removal of diminutive, sessile, and pedunculated polyps and tissue from the GI tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 6, 2019
FINEMEDIX CO., LTD. % April Lee Consultant Withus Group, Inc. 106 Superior Irvine, CA 92620
Re: K183289
> Trade/Device Name: ClearGrasp Snare Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FDI Dated: January 29, 2019 Received: February 5, 2019
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark J. Antonino -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183289
Device Name ClearGrasp Snare
Indications for Use (Describe)
The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, pedunculated polyps and tissue from within the GI tract.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for FINEMEDIX. The logo consists of the word "FINEMEDIX" in a bold, sans-serif font, with an orange circle above the "E" in "FINEMEDIX". The word "FINEMEDIX" is in a dark blue color. The logo is simple and modern.
#### 510(k) Summary
#### Submitter
FINEMEDIX CO., LTD. Hee Kyung Kwon 60, Maeyeo-ro, Dong-gu, Daegu 41065 South Korea Email: hkyung1230@finemedix.com Tel. +82-53-741-8388 Fax. +82-53-741-8168
#### Official Correspondent
Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
#### Device Information
- Trade Name: ClearGrasp Snare
- Common Name: Snare, Flexible
- Classification Name: Endoscopic electrosurgical unit and accessories
- Product Code: FDI
- Panel: Gastroenterology/Urology
- Regulation Number: 21 CFR 876.4300
- Device Class: Class II ●
- . Date Prepared: 11/23/2018
#### Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
- . K 172729, Polypectomy Snare manufactured by Hangzhou AGS MedTech Co., Ltd.
- K151197, Lariat Snare by United States Endoscopy Group, Inc. ●
- K172758, Rotatable Snares, Non-Rotatable Snares by Zhejiang Chuangxiang Medical Technology ● Co., LTD.
#### Indication for Use:
The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.
#### Device Description:
ClearGrasp Snare is a monopolar electrosurgical instrument intended for EMR which is Endoscopic Mucosal Resection by applying high-frequency current during endoscopic electrosurgical procedures. It is used for grasping and resecting the targeted lesion within the digestive tract by opening and closing the loop. Users can choose an oval or a crescent type based on their preference and the characteristic of the lesion. The loop sizes can be chosen in accordance with the size of the lesion and the rotational type helps approaching and grasping the lesion. There is a stiff type distinguished by the strength of loop.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "FINEMEDIX" in large, bold, purple letters. Above the "E" in "FINEMEDIX" is a small orange circle. The background of the image is white.
The device consists of a flexible Rope Wire and various size of loop, which can be extended and rotated from the flexible Catheter Tube using two handle control knobs such as Handle and Slider. When connected to an electrosurgical generator and activated, the loop delivers a monopolar electrical current to the surgical site. This device passes through the working channel of endoscope, and the average contact time of the product and the mucosa of the human digestive tract is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or re-sterilized.
| | Subject Device | Primary<br>Predicate Device | Reference Device | Reference Device | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Company | Finemedix Co.,<br>Ltd. | Hangzhou AGS<br>MedTech Co., Ltd. | United States<br>Endoscopy Group,<br>Inc. | Zhejiang Chuangxiang<br>Medical Technology<br>Co., LTD. | |
| Device Name | ClearGrasp Snare | Polypectomy Snare | Lariat Snare | Rotatable Snares,<br>Non-Rotatable Snares | |
| 510(k) Number | NA | K172729 | K151197 | K172758 | |
| Device<br>Classification<br>Name | Endoscopic<br>electrosurgical unit<br>and accessories | Endoscopic<br>electrosurgical unit<br>and accessories | Endoscopic<br>electrosurgical unit<br>and accessories | Endoscopic<br>electrosurgical unit<br>and accessories | |
| Product Code | FDI | FDI | FDI | FDI | |
| Regulation<br>Number | 876.4300 | 876.4300 | 876.4300 | 876.4300 | |
| Indications for Use | The ClearGrasp<br>Snare is used<br>endoscopically in<br>the removal of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated<br>polyps and tissue<br>from within the GI<br>tract. | The Polypectomy<br>Snare is used<br>endoscopically in<br>the removal of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated<br>polyps and tissue<br>from within the GI<br>tract. | The Lariat snare is<br>an electrosurgical<br>device designed to<br>be used to<br>endoscopically<br>grasp, dissect and<br>transect tissue<br>during GI<br>endoscopic<br>procedures. | The Polypectomy<br>Snare is used<br>endoscopically in the<br>removal and/or and<br>cauterization of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated polys<br>and tissue from within<br>the gastrointestinal<br>tract. | |
| Principle of<br>Operation | Monopolar<br>electrosurgical<br>instrument<br>intended for EMR | Monopolar<br>electrosurgical<br>instrument<br>intended for EMR | Monopolar<br>electrosurgical<br>instrument intended<br>for EMR | Monopolar<br>electrosurgical<br>instrument intended<br>for EMR | |
| | Loop<br>Catheter<br>Tube<br>Handle | SUS 304 | SUS 304 | Unknown | Unknown |
| Material | | PTFE | PTFE | | |
| | | ABS | ABS | | |
| | Components | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug | Loop, Catheter Tube,<br>Handle, Slider, Plug |
| | Shape of Loop | Oval and Crescent | Oval, Polygon,<br>Duck Bill,<br>Hexagonal and<br>Round | Oval, Hexagonal<br>and Diamond | Oval, Hexagonal,<br>Crescent, Round |
| | Working Length | 1600,1800,2200,<br>2400 mm | 1800, 2300 mm | 2300mm | 1600, 1800, 2400mm |
| | Loop Size | 6,10,13,15,<br>24,30 mm | 10,15,25,32 mm | 6,10,30mm | 10,15,22,25,32mm |
| | Rotational/Non-<br>Rotational | Both | Both | Non-Rotational | Both |
| | Sterile/Non-Sterile | Sterile | Sterile | Sterile | Sterile |
| | Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| | Single Use | Yes | Yes | Yes | Yes |
| | Used with<br>Electrosurgical<br>Unit | Yes | Yes | Yes | Yes |
| | Similarities | The subject device is substantially equivalent to the predicate device in Indications for Use,<br>materials, components, shape of loop, rotationality, sterilization methods and principle of<br>operation. | | | |
| | Differences | The difference between the subject and primary predicate are shape of loop, working<br>length, and loop size. To support the difference of the shape of loop such as the crescent<br>shape and working length such as 1600 and 2400mm, we added the K172758 as the<br>reference device. To support the difference of the loop size such as 6mm, we added the<br>K151197 as the reference device. These differences do not raise any questions of<br>substantial equivalence to the declared predicates. | | | |
### Summary of Technological Characteristics:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in a dark blue font, with a small orange circle above the letters 'NE'. The font is sans-serif and appears to be bolded.
## FINEMEDIX CO., LTD.
60, Maeyeo-ro, Dong-gu, Daegu, Republic of Korea (41065)
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in a dark blue color, with an orange circle above the letters 'ME'. The text is in all capital letters and is a sans-serif font. The logo is simple and modern.
#### Non-clinical testing data:
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Biocompatibility testing according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 ● and ISO 10993-11:2006
- . Performance testing such as appearance, dimension, continuity, tensile strength test of electrode cable, withstand voltage, high frequency leakage current test of electrode cable according to IEC 60601-2-2:2017 and ANSI/AAMI HF18:2001
- EO Sterilization residuals testing according to ISO 10993-7:2008
Below tests were performed on our own device and they can be leveraged for the subject device:
- EO Sterilization Validation Testing according to ISO 11737-1:2006 and ISO 11737-2:2009 ●
- . Shelf Life Testing according to ASTM F1980
The biocompatibility evaluation for ClearGrasp Snare was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing within a Risk Management Process " and FDA's biocompatibility guidance, G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (May 1, 1995). The following tests were completed: Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute Systemic Toxicity and Pyrogen Test.
Performance testing such as appearance, dimension, and continuity were performed as per Finemedix's design control system.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
#### Conclusion:
ClearGrasp Snare constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate device. Therefore, ClearGrasp Snare and its predicate are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.