K172758 · Zhejiang Chuangxiang Medical Technology Co., Ltd. · FDI · Mar 7, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K172758
Device Name
Rotatable Snares, Non-Rotatable Snares
Applicant
Zhejiang Chuangxiang Medical Technology Co., Ltd.
Product Code
FDI · Gastroenterology, Urology
Decision Date
Mar 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.
Device Story
Polypectomy snare consists of flexible wire cable and loop; extended/retracted from flexible outer sheath via three-ring handle; rotation actuator allows 360° loop rotation. Used endoscopically by physicians in clinical settings. Device passed through endoscope; monopolar electrical current delivered to loop to cut and cauterize gastrointestinal tissue. PTFE sheath lining minimizes friction during operation. Provides mechanical capture and electrosurgical treatment of polyps; benefits patient by enabling minimally invasive tissue removal.
Clinical Evidence
No clinical data. Bench testing only: includes working length, tortuous path functionality, rotation capability, tensile strength, electrical resistance, electrical safety, and snare actuation. Biocompatibility (ISO 10993-1), sterilization (ISO 11135), and shelf-life (ASTM F1980-16) testing performed.
Technological Characteristics
Flexible wire cable and loop; PTFE-lined sheath; three-ring handle with rotation actuator. Monopolar electrosurgical energy source. Dimensions: 1.8mm or 2.4mm OD; 1600-2400mm length. Sterile, single-use. Standards: ISO 11135 (sterilization), ISO 10993 (biocompatibility), ASTM F1980-16 (shelf-life), AAMI/ANSI ES 60601-1/IEC 60601-2-2 (electrical safety).
Indications for Use
Indicated for endoscopic removal and/or cauterization of diminutive, sessile, and pedunculated polyps and tissue from the gastrointestinal tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 7, 2018
Zhejiang Chuangxiang Medical Technology Co., LTD. Lucius Long Quality Supervisor 301B, No.22, XinYan Road Hangzhou, Zhejiang 311100 China
Re: K172758
Trade/Device Name: Rotatable Snares, Non-Rotatable Snares Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FDI Dated: January 4, 2018 Received: January 4, 2018
Dear Lucius Long:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172758
Device Name Rotatable Snares Non-Rotatable Snares
#### Indications for Use (Describe)
The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains a logo with a geometric shape on the left, transitioning from blue to green. To the right of the shape, there is text in blue that appears to be Chinese characters. Below the Chinese characters, the word "MEDNOVA" is written, with "MED" in blue and "NOVA" in green, matching the color gradient of the geometric shape.
Section 2 510(k) Summary( 21CFR 807.92)
# 1. Submitter's information
Name: Zhejiang Chuangxiang Medical Technology Co., LTD.
Address: 301B, No.22, XinYan Road Yuhang Economic And Technological Development Zone Hangzhou Zhejiang China
Contact person: Lucius.Long Telephone: 86-571-89167088 Fax: 86-571-89167086
## 2. Device
Trade Name: Rotatable Snares, Non-Rotatable Snares Common Name: Polypectomy Snare Classification name: Endoscopic electrosurgical unit and accessories Regulation class: 2 Regulation number:876.4300 Panel: Gastroenterology/Urology Product code: FDI
## 3. Predicative device
510(k) Number: K160637, Product Name: Rotatable Snares
## 4. Device description
The Polypectomy Snare consists of a flexible wire cable (core wire) and loop which can be extended and retracted from the Snare's flexible outer sheath using a threering handle. The flexible cable and loop can also be rotated 360° using the rotation actuator on the handle. The inner diameter of the sheath is PTFE to provide minimal friction during extension, rotation, and retraction of the loop from the sheath. When passed through an endoscope and activated, the Snare delivers a monopolar electrical current to cut and cauterize tissue with the loop.
For All models information in this submission, please refer to below table 1.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains a logo with text. On the left side, there is a geometric shape with a gradient of green and blue. To the right of the shape, there is the text "MEDNOVA" in green and blue. Above the text, there are three Chinese characters in blue.
# Table 1 List of models in this submission
| Model | Part Number | Max. OD<br>(mm)<br>≤ D (mm) | Snare Loop<br>Opening Width<br>(mm)<br>B +10%<br>-20% | Effective Length<br>(mm)<br>L (mm) ±100 | |
|-------------------|----------------|-----------------------------|-------------------------------------------------------|-----------------------------------------|------|
| E<br>Oval | MD-E-SNC181024 | Φ1.8 | 10 | 2400 | |
| | MD-E-SNC181524 | Φ1.8 | 15 | 2400 | |
| | MD-E-SNC181018 | Φ1.8 | 10 | 1800 | |
| | MD-E-SNC181518 | Φ1.8 | 15 | 1800 | |
| E<br>Oval | MD-E-SNC241024 | Φ2.4 | 10 | 2400 | |
| | MD-E-SNC241524 | Φ2.4 | 15 | 2400 | |
| | MD-E-SNC241018 | Φ2.4 | 10 | 1800 | |
| | MD-E-SNC241518 | Φ2.4 | 15 | 1800 | |
| | MD-E-SNC242516 | Φ2.4 | 25 | 1600 | |
| | MD-E-SNC242524 | Φ2.4 | 25 | 2400 | |
| | MD-E-SNC243224 | Φ2.4 | 32 | 2400 | |
| | MD-E-SNC242518 | Φ2.4 | 25 | 1800 | |
| Standard<br>(C) | MD-E-SNC243218 | Φ2.4 | 32 | 1800 | |
| | MD-H-SNC241024 | Φ2.4 | 10 | 2400 | |
| | MD-H-SNC241524 | Φ2.4 | 15 | 2400 | |
| H<br>Hexagonal | MD-H-SNC242224 | Φ2.4 | 22 | 2400 | |
| | MD-H-SNC243224 | Φ2.4 | 32 | 2400 | |
| | MD-H-SNC241018 | Φ2.4 | 10 | 1800 | |
| | MD-H-SNC241518 | Φ2.4 | 15 | 1800 | |
| H<br>Hexagonal | MD-H-SNC242218 | Φ2.4 | 22 | 1800 | |
| | MD-H-SNC243218 | Φ2.4 | 32 | 1800 | |
| | MD-C-SNC241524 | Φ2.4 | 15 | 2400 | |
| | MD-C-SNC242524 | Φ2.4 | 25 | 2400 | |
| C<br>Crescent | MD-C-SNC243224 | Φ2.4 | 32 | 2400 | |
| | MD-C-SNC241518 | Φ2.4 | 15 | 1800 | |
| | MD-C-SNC242518 | Φ2.4 | 25 | 1800 | |
| MD-C-SNC243218 | Φ2.4 | 32 | 1800 | | |
| | R | MD-R-SNC243224 | $Φ$ 2.4 | 32 | 2400 |
| | Round | MD-R-SNC243218 | $Φ$ 2.4 | 32 | 1800 |
| | E<br>Oval | MD-E-SNR181024 | $Φ$ 1.8 | 10 | 2400 |
| | | MD-E-SNR181524 | $Φ$ 1.8 | 15 | 2400 |
| | | MD-E-SNR181018 | $Φ$ 1.8 | 10 | 1800 |
| | | MD-E-SNR181518 | $Φ$ 1.8 | 15 | 1800 |
| | | MD-E-SNR241024 | $Φ$ 2.4 | 10 | 2400 |
| | | MD-E-SNR241524 | $Φ$ 2.4 | 15 | 2400 |
| | | MD-E-SNR241018 | $Φ$ 2.4 | 10 | 1800 |
| | | MD-E-SNR241518 | $Φ$ 2.4 | 15 | 1800 |
| | | MD-E-SNR242516 | $Φ$ 2.4 | 25 | 1600 |
| | | MD-E-SNR242524 | $Φ$ 2.4 | 25 | 2400 |
| | | MD-E-SNR243224 | $Φ$ 2.4 | 32 | 2400 |
| | | MD-E-SNR242518 | $Φ$ 2.4 | 25 | 1800 |
| | | MD-E-SNR243218 | $Φ$ 2.4 | 32 | 1800 |
| Rotational<br>(R) | H<br>Hexagonal | MD-H-SNR241024 | $Φ$ 2.4 | 10 | 2400 |
| | | | MD-H-SNR241524 | $Φ$ 2.4 | 15 |
| | | | MD-H-SNR242224 | $Φ$ 2.4 | 22 |
| | | | MD-H-SNR243224 | $Φ$ 2.4 | 32 |
| | | | MD-H-SNR241018 | $Φ$ 2.4 | 10 |
| | | | MD-H-SNR241518 | $Φ$ 2.4 | 15 |
| | | | MD-H-SNR242218 | $Φ$ 2.4 | 22 |
| | | | MD-H-SNR243218 | $Φ$ 2.4 | 32 |
| | C<br>Crescent | MD-C-SNR241524 | $Φ$ 2.4 | 15 | 2400 |
| | | | MD-C-SNR242524 | $Φ$ 2.4 | 25 |
| | | | MD-C-SNR243224 | $Φ$ 2.4 | 32 |
| | | | MD-C-SNR241518 | $Φ$ 2.4 | 15 |
| | | | MD-C-SNR242518 | $Φ$ 2.4 | 25 |
| | | | MD-C-SNR243218 | $Φ$ 2.4 | 32 |
| | R | MD-R-SNR243224 | $Φ$ 2.4 | 32 | 2400 |
| | Round | MD-R-SNR243218 | $Φ$ 2.4 | 32 | 1800 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo for a company called MEDNOVA. The logo consists of a green and blue shield with a white geometric shape inside. To the right of the shield is the company name, with "MED" in green and "NOVA" in blue. Above the company name are two Chinese characters in blue.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a green and blue shield-like shape on the left, with the text "MEDNOVA" in green below it. To the right of the shield is the company name in Chinese, written in blue.
# 5. Indication for use
The Polypectomy Snare is used endoscopically in the removal and/or and cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the gastrointestinal tract.
# 6. Technological Characteristics
The Polypectomy Snare incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the Boston Scientific Corporation predicate devices.
#### Comparison to predicate Devices:
| Item | Proposed device | Predicate device | Comparison to<br>Predicate<br>Devices |
|-------------------------------|---------------------------------------------------------|-------------------------------------|---------------------------------------|
| Device name | Polypectomy<br>Snare | Rotatable Snares | Similiar |
| 510(k)<br>number | NA, new<br>submission | K160637 | Different 510(k)<br>number |
| Manufacture<br>r | Zhejiang<br>Chuangxiang medical<br>technology Co., Ltd. | Boston Scientific<br>Corporation | Different<br>manufacturer |
| Product<br>Code | FDI | FDI, FGX | Same |
| Regulation<br>No. | 876.4300 | 876.4300 | Same |
| Class | 2 | 2 | Same |
| Supplied<br>Sterile | Yes | Yes | Same |
| Configuratio<br>n | Oval, Hexagonal,<br>Crescent, Round | Oval, Hexagonal,<br>Crescent, Round | Same |
| Minimal<br>working<br>channel | 2.8mm | 2.8mm | Same |
| Sheath OD | 2.4mm | 2.4mm | Same |
| Working<br>Length | 1600mm,1800m<br>m,2400mm | 2400mm | similar |
| Loop width | 10mm, 15mm, 22mm | 10mm, 11mm, 13mm, 15mm | similar |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a geometric shape in shades of green on the left, followed by the text "MEDNOVA" in green. To the right of the logo is Chinese text in blue.
| | mm,25mm,32mm, | mm,20mm,25mm,27mm,30<br>mm,33mm | |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Indications<br>for Use | The Polypectomy<br>Snare is used<br>endoscopically in the<br>removal and/or<br>cauterization of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated polyps<br>and tissue from within<br>the gastrointestinal<br>tract. | The Rotatable Snare is<br>used endoscopically in the<br>removal and/or<br>cauterization of diminutive<br>polyps, sessile polyps,<br>pedunculated polyps and<br>tissue from within the<br>gastrointestinal tract. | Same |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO<br>sterilized pouch with<br>one device per pouch | Single-use EO sterilized<br>pouch with one device per<br>pouch | Same |
# 7. Performance data
In-vitro Testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests. A summary of the test results has been provided for Working Length, Tortuous Path Functionality, Rotation Capability, Handle to Core Wire Tensile Strength, Electrical Resistance, Electrical Safety and Snare Actuation.
The EO residual and ECH residual were measured after sterilization of the device to meet the criteria defined in ISO 11135 Second edition 2014 and ISO 10993-7 Second Edition 2008-10-15.
The shelf-life for three years had been validated in accelerated testing according to ASTM F1980-16 (2016) and the requirements on packaging for terminally sterilized medical device per ISO 11607-1 First Edition 2006-04-15 and ISO 11607-2 First Edition 2006-04-15 are also met. The testing successfully demonstrated essential performance is achieved before and after the shelf life test.
Biocompatibility testing was performed in accordance with the FDA Guidance," Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on June 26, 2016. The cytotoxicity, sensitization, intracutaneous irritation, system toxicity pyrogen and endotoxin tests were performed to demonstrate the biocompatibility of the device.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for MEDNOVA. The logo consists of a blue and green hexagon shape on the left, with the text "MEDNOVA" in green and blue to the right of the shape. Above the text "MEDNOVA" are three blue Chinese characters.
Tests on Electromagnetic Compatibility and Electrical Safety were performed in accordance to requirements per AAMI/ANSI ES 60601-1:2005/(R)2012 and AAMI/ANSI/IEC 60601-2-2:2009.
# 8. Clinical Test Conclusion
No clinical study is included in this submission.
## 9. Conclusions
Chuangxiang medical has demonstrated that the proposed Polypectomy Snare are substantially equivalent to Boston Scientific Corporation's currently marketed Rotatable Snares(K160637).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.