← Product Code [FDI](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI) · K172729

# Polypectomy Snare (K172729)

_Hangzhou AGS MedTech Co., Ltd. · FDI · May 21, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K172729

## Device Facts

- **Applicant:** Hangzhou AGS MedTech Co., Ltd.
- **Product Code:** [FDI](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI.md)
- **Decision Date:** May 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Polypectomy Snare is used endoscopically in the removal of diminutive polyps, pedunculated polyps and tissue from within the GI tract.

## Device Story

Polypectomy snare is a single-use endoscopic instrument; consists of handle, sheath, flexible wire, and electrode. Used by clinicians during endoscopic procedures to remove GI tract polyps. Device passed through endoscope; electrode extends/retracts from sheath to capture tissue. Monopolar RF energy delivered via electrode to cut and cauterize tissue. Rotary models feature handle-based rotation actuator for cable/snare positioning. Available in five electrode shapes: Oval, Polygon, Duck Bill, Hexagonal, and Round. Sterile, single-use device; no animal/human origin substances.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, ISO 10993-10) and performance testing (appearance, size, material, usability, electrical properties).

## Technological Characteristics

Components: handle, sheath, flexible wire, electrode. Energy: Monopolar RF. Electrode shapes: Oval, Polygon, Duck Bill, Hexagonal, Round. Features: Rotary and non-rotary specifications. Biocompatibility: ISO 10993-1, ISO 10993-5, ISO 10993-10. Sterilization: Sterile, single-use.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Captivator II, Single-Use Polypectomy Snares ([K133987](/device/K133987.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 21, 2018

Hangzhou AGS MedTech CO., Ltd % Ethan Liu RA Specialist Shanghai Thinkwell Consulting Co., Ltd. Xin Ling Road, 211/6F Shanghai, 201100 China

Re: K172729

Trade/Device Name: Polypectomy Snare Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FDI Dated: April 13, 2018 Received: April 20, 2018

Dear Ethan Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K172729

Device Name Polypectomy Snare

Indications for Use (Describe)

Polypectomy Snare is used endoscopically in the removal of diminutive polyps, pedunculated polyps and tissue from within the GI tract.

| Type of Use (Select one or both, as applicable) | <span> <span style="margin-right: 5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> <span> <span style="margin-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

*PRAStaff@fda.hhs.gov*

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

d a person is espon دائل الملك ال ી । ઉપ information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a logo for AGS MEOTECH. The logo consists of a circular design with three curved shapes in blue and green. To the right of the circular design is the text "AGS MEOTECH" in English, followed by the same name in Chinese characters. The logo appears to be for a medical technology company.

#### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

# 5.1 Submitter

| Submitted by:                         | Hangzhou AGS MedTech CO., Ltd.<br>Building 6, Kangxin Road No.597, Qianjiang<br>Economic Development Area , Hangzhou ,<br>Zhejiang , 311106, China |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 3010288205                                                                                                                                         |
| Owner/Operator<br>Number:             | 10052959                                                                                                                                           |
| Contact Person:                       | Ethan Liu<br>Phone:0086-15216699240<br>Fax: 0086-21-60732022<br>Email:xtdeepwater@126.com                                                          |
| Date Prepared:                        | Jan 18, 2017                                                                                                                                       |

#### 5.2 Device

| Device Name:       | Polypectomy Snare                               |
|--------------------|-------------------------------------------------|
| Common Name:       | Polypectomy Snare                               |
| Regulatory Class:  | Class II                                        |
| Regulation Number  | 876.4300                                        |
| Product Code Name: | Snare, Flexible                                 |
| Regulation Name:   | Endoscopic electrosurgical unit and accessories |
| Product Code:      | FDI                                             |

# 5.3 Predicate Device

| Device Name:       | Single Use Polypectomy Snares<br>Captivator II, Single-Use Polypectomy Snares<br>K133987 |
|--------------------|------------------------------------------------------------------------------------------|
| Common Name:       | Polypectomy Snare                                                                        |
| Regulatory Class:  | Class II                                                                                 |
| Regulation Number  | 876.4300                                                                                 |
| Product Code Name: | Snare, Flexible                                                                          |
| Regulation Name:   | Endoscopic electrosurgical unit and accessories                                          |
| Product Code:      | FDI                                                                                      |

# 5.4 Device Description

The polypectomy snare consists of handle, sheath, flexible wire and electrode that is used to remove polys utilizing monopolar RF energy under endoscopic visualization.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains a logo with a circular design on the left and text on the right. The circular design consists of three curved shapes in blue and green, arranged in a triangular pattern. To the right of the logo, there is text in two lines, with the first line reading "AGS MEDTECH" in English. The second line contains Chinese characters, which translate to "An Jie Si Medical".

K172729 Page 2 of 3

The rotary snares' cable and polypectomy can be rotated using the rotation actuator on the handle.

Polypectomy snare is sterile for single use with no delayed hypersensitivity and no intracutaneous reactivity, have no animal or human origin substance.

The electrode of polypectomy snare has five shapes to be used in clinical practice: Oval, Polygon, Duck Bill, Hexagonal and Round, and the electrode can extend, retract from the snare's flexible outer sheath.

#### 5.5 Indication for Use:

Polypectomy Snare is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.

#### 5.6 Technology Characteristics

The polypectomy snare consists of handle, sheath, flexible wire and electrode, which are used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract. When passed through an endoscope and activated, the snare delivers a monopolar electrical current to cut and cauterize tissue with the electrode.

The proposed device contain both rotary and non-rotary specifications, the rotary snares' cable and polypectomy can be rotated using the rotation actuator on the handle, the main differences of these specifications are dimension.

# 5.7 Applicable Guidance Document

Guidance for the Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology

# 5.8 Substantial Equivalence

Captivator II, Single-Use Polypectomy Snare(Boston Scientific Corporation, K133987) is used as predicate device compared to proposed device Polypectomy Snare manufactured by Hangzhou MedTech Co., Ltd.

#### 5.9 Performance Data

# 5.9.1 Biocompatibility Testing

The biocompatibility evaluation for this device was conducted in accordance with the FDA Bluebook Memorandum G95-1 "Use Of International Standard ISO-10993, 'Biological Evaluation Of Medical Devices Part 1: Evaluation And Testing.'" and the International Standard ISO 10993-1 Fourth Edition 2009-10-15, Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process [Including: Technical Corrigendum 1 (2010)], as recognized by FDA.

| Biocompatibility<br>Test | Standards | Test Result | Verdict |
|--------------------------|-----------|-------------|---------|
|--------------------------|-----------|-------------|---------|

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo with a blue circular design on the left and text on the right. The circular design consists of three curved shapes that form a triangular pattern with a green center. To the right of the logo, the text "AGS MEDTECH" is written in a simple, sans-serif font. Below "AGS MEDTECH", there are three Chinese characters, which appear to be a translation or alternative name for the company.

Section 5 510(k) Summary Polynectomy Snari

| In<br>Cytotoxicity               | Vitro ISO 10993-5:2009 | No cytotoxicity observed.    | Pass |
|----------------------------------|------------------------|------------------------------|------|
| Delayed-type<br>Hypersensitivity | ISO 10993-10:2010      | No hypersensitivity observed | Pass |
| Irritation                       | ISO 10993-10:2010      | No irritation observed       | Pass |

# 5.8.2 Performance Testing

The following bench tests were performed on Polypectomy snare: Appearance, Size, Material, Usability, Electric property. The results of all testing were passing.

# 5.9 Clinical Test Conclusion

No Clinical Study is included in this submission.

#### 5.11 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS Medtech Co., Ltd has demonstrated that proposed device Polypectomy Snare is substantially equivalent to Boston Scientific Corporation's currently marketed Captivator II, Single-Use Polypectomy Snares(K133987).

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K172729](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K172729)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
