K152580 · Endochoice, Inc. · FDI · Nov 3, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K152580
Device Name
rescuenet
Applicant
Endochoice, Inc.
Product Code
FDI · Gastroenterology, Urology
Decision Date
Nov 3, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Indications for Use
The retrieval net is intended for use in endoscopic retrieval of foreign objects, food bolus, tissue fragments and excised tissue such as polyps.
Device Story
Disposable, single-use endoscopic retrieval device; consists of flexible wire cable, snare loop, and net housed within a flexible outer sheath. Operated via three-ring handle; advancing rings extends net from sheath; retracting rings closes net around target. Used by physicians during endoscopic procedures to capture and remove foreign bodies, food bolus, or tissue samples. Facilitates safe extraction through endoscope channel; minimizes risk of tissue damage or loss during retrieval.
Clinical Evidence
Bench testing only. Includes sterilization validation (ISO 11135, ISO 11737-2, ISO 11138-1), biocompatibility testing (ISO 10993-1), and performance testing to confirm functional operation as intended.
Technological Characteristics
Disposable, single-use endoscopic accessory. Length 230cm; catheter diameter 2.5mm. Compatible with 2.8mm endoscope channels. Oval net configuration. Sterilized via Ethylene Oxide. Biocompatibility per ISO 10993-1. Mechanical actuation via manual three-ring handle.
Indications for Use
Indicated for endoscopic retrieval of foreign objects, food bolus, tissue fragments, and excised tissue (e.g., polyps) in patients undergoing flexible or rigid endoscopy.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 3, 2015
Endochoice, Inc. Daniel Hoefer Regulatory Affairs Manager 11810 Wills Rd. Alpharetta, GA 30009
Re: K152580 Trade/Device Name: Rescuenet™ Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic and electrosurgical unit and accessories Regulatory Class: II Product Code: FDI, GCJ Dated: September 9, 2015 Received: September 10, 2015
Dear Daniel Hoefer,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152580
Device Name EndoChoice Retrieval Net
Indications for Use (Describe)
The retrieval net is intended for use in endoscopic retrieval of foreign objects, food bolus, tissue fragments and excised tissue such as polyps.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510(k) Summary
# 1. Company Identification
EndoChoice, Inc. 11810 Wills Road Alpharetta, GA 30009 Telephone (678) 708 4743 FAX (678) 878 3373 Establishment Registration: 3007591333
# 2. Contact Person
Daniel Hoefer Regulatory Affairs Manager
#### 3. Device Name
rescuenet™ Trade name: Common/Usual Name: Retrieval Net
#### 4. Device Classification
| Common Name: | Retrieval Net |
|-----------------|-------------------------------------------------------------------|
| Classification: | Endoscopic electrosurgical unit and accessories, 21CFR § 876.4300 |
| Product Code: | FDI, GCJ |
| Committee: | Gastroenterology/Urology |
# 5. Intended Use
The Retrieval Net is intended for use in endoscopic retrieval of foreign objects, food bolus, tissue fragments and excised tissue such as polyps.
# 6. Device Description
The EndoChoice Retrieval Net is a disposable, single-use sterile device comprised of a flexible wire cable and snare loop with net which can be extended and retracted form the flexible outer sheath using a three ring handle. Objects are retrieved by advancing the finger rings of the handle to open the retracted net. The net is then endoscopically manipulated over the foreign object, food bolus, tissue fragment or excised tissue, in which the finger rings of the handle are retracted to close the net once the article has been retrieved.
{4}------------------------------------------------
# 7. Substantial Equivalence
# 7.1 Predicate devices
The EndoChoice retrieval net is substantially equivalent to the predicate device, the Roth Net retriever product line by United States Endoscopy Group, Inc. The intended use, design and materials and labeling are all substantially equivalent.
# 7.2 Intended Use
The indications for use of the EndoChoice retrieval net is substantially equivalent to the predicate device, the Roth Net® retrieval product line (K122462) manufactured United States Endoscopy Group, Inc. In each case, the nets are intended for use in endoscopic retrieval of objects.
# 7.3 Technical Characteristics
The device length is 230cm and the device catheter diameter is 2.5mm.
Image /page/4/Figure/8 description: The image shows a medical device with measurements labeled. The diameter of the circular part on the left is 30 mm. The length of the device is 230 cm and the diameter of the shaft is 2.5 mm.
# 7.4 Performance Characteristics
The steps for operator use of each of the devices are equivalent. The instructions for use describe how to use the device with a standard endoscope, in which the stepwise instructions are similar to that of the predicate device.
{5}------------------------------------------------
| Substantial Equivalence Comparison | | | |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | EndoChoice Retrieval Net | Roth Net® retriever | Substantial |
| | | product line (K122462) | Equivalence |
| Product Code | FDI, GCJ | FDI, GCJ | Identical |
| Regulation No. | 21CFR § 876.4300 | 21CFR § 876.4300 | Identical |
| Classification | Endoscopic electrosurgical unit<br>and accessories | Endoscopic electrosurgical unit<br>and accessories | Identical |
| Manufacturer | EndoChoice Inc. | United States Endoscopy<br>Group, Inc. | N/A |
| Supplied Sterile | Yes | Yes | Identical |
| Sheath length | 230cm | 160cm, 230cm | Identical |
| Single use | Yes | Yes | Identical |
| Sheath diameter | 2.5mm | 2.5mm, 3.0mm | Identical |
| Compatibility | 2.8mm endoscope channel | 2.8mm, 3.2mm endoscope<br>channel | Equivalent |
| Net Size | 3 x 6 cm | 3 x 6 cm<br>2 x 4.5 cm<br>4 x 5.5 cm<br>4 x 8 cm | Identical |
| Configuration | Oval | Oval, Octagonal, Hexagonal | Identical |
| Indications for | This device is intended for use in<br>endoscopic retrieval of foreign<br>objects, food boluses, tissue<br>fragments and excised tissue<br>such as polyps. | The Roth Net® retriever<br>product line is intended to be<br>used to retrieve excised polyps,<br>tissue samples, foreign bodies<br>and calculi during flexible and<br>rigid endoscopy procedures. | Equivalent |
| use | | | |
| Packaging | Single-use pouch | Single-use pouch | Identical |
# 7.5 Substantial Equivalence Table
# 8. Non-Clinical testing
# 8.1 Sterilization and Shelf-Life
The EndoChoice Retrieval Net is sterilized by Ethylene Oxide.
- Validation has been completed in accordance with the following standards:
- ISO 11135: 2014 Sterilization of Health-Care Products Ethylene Oxide Requirements for the ● Development, Validation and Routine Control of a Sterilization Process for Medical Devices
- . AAMI ANSI ISO 11737-2: 2009 Sterilization of Medical Devices - Microbiological Methods -Part 2: Tests of Sterility Performed in the Definition, Validation and Maintenance of a Sterilization Process
- . AAMI ANSI ISO 11138-1: 2006 Sterilization of Health Care Products - Biological Indicators -Part 1: General Requirements
The shelf life of the EndoChoice Retrieval Net is 1 year from date of manufacturing.
{6}------------------------------------------------
# 8.2 Biocompatibility
Results of Biocompatibility testing in accordance with AAMI ANSI ISO 10993-1: 2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing within a Risk Management Process demonstrates that the device meets the requirements.
# 8.3 Performance Testing
Results from various performance testing indicates that the EndoChoice Retrieval Net functions as intended.
#### 9. Conclusion
Retrieval nets have been in use for over 20 years. A variety of retrieval net devices are available on the market from different manufacturers with various diameters and lengths. Based on the above information, the EndoChoice Retrieval Net is substantially equivalent to the predicate device listed. The information contained in this submission supports the fact that the EndoChoice Retrieval Net is as safe and effective as its predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.