← Product Code [FDG](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDG) · K964561

# ENDOLUMINA II TRANSILLUMINATION SYSTEM 56FR E-5630, 50FR E-5030 AND 40FR E-4020, ENDOLUMINA II DETACHABLE TIPS, 56FR, TA (K964561)

_Bioenterics Corp. · FDG · Feb 11, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDG/K964561

## Device Facts

- **Applicant:** Bioenterics Corp.
- **Product Code:** [FDG](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDG.md)
- **Decision Date:** Feb 11, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4530
- **Device Class:** Class 1
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The ENDOLUMINA II Transillumination System (EIB II) is a medical device designed to aid in the identification of the esophagus, rectum, and other structures by transillumination during laparoscopy, thoracoscopy, or open procedures.

## Device Story

Device aids anatomical identification via transillumination during surgery. Consists of glass fiberoptic bundle inside silicone elastomer tubing; proximal end features pop-fit connection for detachable, clear silicone tips (blunt or tapered); distal end connects to external light source. Light transmits through fiberoptics to tip, where it is deflected radially to illuminate structures. Used in OR by surgeons during laparoscopy, thoracoscopy, or open procedures. Cable is reusable; tips are single-use and sterile. Heat build-up at tip is controlled to prevent tissue damage. Provides visual guidance for surgeon to locate internal structures.

## Clinical Evidence

Bench testing only. Biocompatibility evaluation performed. Thermal testing conducted to ensure tip temperature does not cause tissue damage during high-intensity light source use. Sterilization validation performed for single-use tips to achieve 10⁻⁶ sterility assurance level.

## Technological Characteristics

Materials: glass fiberoptic bundle, silicone elastomer tubing/tips. Form factor: slender, flexible, cylindrical cable (96.1 inches) with pop-fit detachable tips (4-60 Fr). Energy: external light source. Connectivity: direct connection to commercially available light sources. Sterilization: reusable cable (cleaning/sterilization instructions provided); single-use tips (validated sterilization cycle, SAL 10⁻⁶).

## Regulatory Identification

A gastroenterology-urology fiberoptic retractor is a device that consists of a mechanical retractor with a fiberoptic light system that is used to illuminate deep surgical sites.

## Predicate Devices

- ENDOLUMINA® I Illuminated Bougie

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

BioEnterics CORPORATION
FEB 11 1997
K964561 p.1/2
1035 A Cindy Lane
Carpinteria, CA 93013
Tel 805 684-3045
Fax 805 684-0812

Contact person: Keith Lowrey, Regulatory Affairs Manager

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS:

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

## Substantially Equivalent Device which is legally marketed:

BioEnterics Corporation (BEC) considers the ENDOLUMINA® II Transillumination System to be substantially equivalent in design and use to the ENDOLUMINA® I Illuminated Bougie.

## Description of Premarket Notification Device:

The ENDOLUMINA II Transillumination System (EIB II) is a medical device designed to aid in the identification of the esophagus, rectum, and other structures by transillumination during laparoscopy, thoracoscopy, or open procedures. The EIB II consists of a glass fiberoptic bundle inserted into a length of silicone elastomer tubing in a configuration similar to a fiberoptic cable.

Attached to the cable’s proximal end is a ball-shaped or barb-shaped fitting which allows a clear, flexible, and detachable silicone tip to be attached or removed by means of a “pop-fit” connection. This tip is available in both blunt and tapered configurations. The EIB II is available in sizes ranging from 4 to 60 Fr.

A light source connector/fitting is permanently attached to the distal end of the cable and is designed to be inserted into a commercially-available light source. Light generated from the light source is transmitted through the cable’s fiberoptic bundle and down into the clear silicone tip, where it is deflected radially to enhance transillumination.

The EIB II is supplied with one (1) cable assembly, which is supplied nonsterile and is designed to be reusable by the customer. Instructions for methods for cleaning and sterilizing the EIB II cable assembly are provided. The EIB II also contains three (3) detachable tips which are individually packaged in a double pouch system and labeled as sterile and for single use only. The sterilization method for the tips has been qualified and validated to provide a sterility assurance level (SAL) of 10⁻⁶.

## Comparison to the Legally Marketed Device (Substantially Equivalent):

In comparison to the ENDOLUMINA Bougie (EIB I), BioEnterics Corporation considers the ENDOLUMINA Bougie II (EIB II) to be substantially equivalent in design, indications, and intended use. Both devices are slender, flexible, and cylindrical instruments for introduction into intact natural channels as well as surgically created pockets. Both devices consist of a glass fiberoptic bundle inserted into a length of silicone elastomer tubing in a configuration similar to a fiberoptic cable. One end of the cable is attached to a clear silicone elastomer tip and the other

Leading the Way in Surgical Innovation Worldwide
219

{1}

K964561
p. 2/2

# 510(k) Summary of Safety and Effectiveness of the ENDOLUMINA II Transillumination System

end connects to commercially-available light sources. Materials used to fabricate both devices have been evaluated and have passed requirements for biocompatibility.

The EIB I consists of a clear silicone elastomer tip which is *permanently* attached to the proximal end of an approximately 25.6 inch long cable. A connector/fitting is attached to the cable’s distal end to facilitate connection to a commercially-available light source cable. Like the EIB II cable, the EIB I can be cleaned, sterilized, and reused.

In comparison, the EIB II has a cable which is approximately 96.1 inches long and is designed to connect directly to a commercially available light source. However, when connected to a high intensity light source and left at maximum intensity until the temperature stabilizes, the “heat build-up” or maximum temperature of the tips is not hot enough to cause damage to surrounding tissues. The EIB II is designed to have *detachable* tips which are packaged, sterilized by a validated cycle, and provided separate from the cable assembly.

220

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDG/K964561](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDG/K964561)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).