The InfraVision Ureteral Kit with the InfraVision IR Illuminator is intended to transilluminate the ureter during surgical procedures. The Catheter accepts up to a 0.038"-guidewire. The Kit can be used for either open surgical (with the InfraVision Detector Probe) or laparoscopic (with the InfraVision Camera) access.
Device Story
Stryker Ureteral Illumination System IV consists of open-ended ureteral catheter; Pellethane construction with hydrophilic coating; incorporates fiberoptic light guide. Used with InfraVision IR Illuminator to transilluminate ureter during surgery. Compatible with up to 0.038" guidewire. Open surgical access utilizes InfraVision Detector Probe; laparoscopic access utilizes InfraVision Camera. Sterile, single-use disposable. Enables visualization of ureter to assist surgeons during procedures; helps prevent accidental injury to ureter.
Clinical Evidence
Bench testing only. Device validated for EtO sterilization to a SAL of 10^-6. Conformance to voluntary safety and performance standards including ISO 10993, ANSI/AAMI/ISO 11135, and various ASTM packaging/mechanical standards.
Technological Characteristics
Pellethane catheter with hydrophilic coating; fiberoptic light guide. Compatible with 0.038" guidewire. Sterile, single-use. EtO sterilization (SAL 10^-6). Standards: ISO 10993, G95-1, ANSI/AAMI/ISO 11135, EN 550, ASTM D4169, D999, D1140, D4332, D4728, D5276, F88, EN 868-1, ISO 11607, FPA SPMC 005-96.
Indications for Use
Indicated for patients undergoing surgical procedures requiring ureteral transillumination during open or laparoscopic access.
Regulatory Classification
Identification
A fiberoptic light ureteral catheter is a device that consists of a fiberoptic bundle that emits light throughout its length and is shaped so that it can be inserted into the ureter to enable the path of the ureter to be seen during lower abdominal or pelvic surgery.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Stryker Ureteral Illumination System III (K982542)
Submission Summary (Full Text)
{0}------------------------------------------------
San Jose, CA 95138 t: 408 754 2000 f: 408 754 2521
SEP 28 2006
K061548
**stryker**®
Endoscopy
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## Device Name:
| Proprietary Name: | Stryker Ureteral Illumination System IV |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common and Usual Name: | Stryker Universal Ureteral Kit or<br>Stryker Universal U-Kit or<br>Stryker Ureteral Illumination System IV or<br>Fiberoptic light ureteral catheters |
| Classification Name: | Fiberoptic light ureteral catheter (Class II, 876.4020, Product<br>Code FCS, Gastroenterology-Urology Devices Panel) |
The Stryker Ureteral Illumination System IV with the InfraVision IR Illuminator is intended to transluminate the ureter during surgical procedures. The Catheter accepts up to a 0.038" guidewire. The Kit can be used for either open surgical (with the InfraVision Detector Probe) or laparoscopic (with the InfraVision Camera) access.
The Stryker Ureteral IIIumination System IV consists of an open-ended ureteral catheter, which allows the use of a guidewire (up to 0.038" dia), is composed of Pellethane™ with a hydrophilic coating to ease insertion into the ureters and incorporates a fiberoptic light guide that is used with the InfraVision IR llluminator to transluminate the ureter during surgical procedures. It can be used for either open surgical or laparoscopic access and is distributed as a sterile, single use disposable product. The device is validated for EtO sterilization to a SAL of 10°.
The Stryker Ureteral Illumination System IV is substantially equivalent in terms of safety and effectiveness to the currently marketed Stryker Ureteral III (ref: K982542). The Styker Universal U-Kit will conform to the following voluntary safety and performance standards: ISO 10993, Blue Book Memorandum G95-1, ANSI/AAMI/ISO 11135, EN 550, ASTM D4169, ASTM D999, ASTM D1140, ASTM D4332, ASTM D4728, ASTM D5276, ASTM F88, EN 868-1, ISO 11607, and FPA SPMC 005-96.
The technological differences between the Stryker Ureteral Illumination System IV and the identified predicate devices (Stryker Ureteral Illuminator System III, K982542) do not affect the safety and efficacy of the product; therefore, the Stryker Ureteral Illumination System IV is substantially equivalent to the identified predicate devices.
Date:
By:
Erica A. Walters Sr. Requlatory Representative Stryker Endoscopy
Stryker Ureteral Illumination System IV
Page 6 of 48
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff and two snakes intertwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
、
SEP 2 8 2006
Ms. Erica A. Walters Sr. Regulatory Representative Stryker Endoscopy 5900 Optical Court SAN JOSE CA 95138
K061548 Re:
Trade/Device Name: Stryker Ureteral Illumination System IV Regulation Number: 21 CFR §876.4020 Regulation Name: Fiberoptic light ureteral catheter Regulatory Class: II Product Code: FCS Dated: August 28, 2006 Received: August 30, 2006
Dear Ms. Walters:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) presidentially equivalent (for the indications for referenced above and have determined the devices marketed in interstate commerce
use stated in the enclosure) to legally marketed predicated in interstate on to devices that use stated in the enclosure) to legally man Riccal Device Amendments, or to devices that prior to May 28, 1976, the enactment uith the provisions of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Febral Food, have been reclassified in accordance with the provisions and application (PMA). You may,
Act (Act) that do not require approval of a premarket approval application of the Ac Act (Act) that do not require apploval of a prematice approvisions of the Act. The general
therefore, market the device, subject to the general someration, listing of device therefore, market the device, subject to the generalion, fish registration, listing of devices, good
controls provisions of the Act include requirements for annon registratio controls provisions of the Act metade requirements see
If your device is classified (see above) into cither class II (Special Controls) or class III (Prematice)
President (see and to Microsoft of the street - Evigine major reaul If your device is classified (see above) microls. Existing major regulations affecting your Approval), it may be subject to such additions. Title 21, Pars 800 to 898. In addition, FDA
device can be found in the Code of Federal Regulations, Title 21, Paris 800 lo 89 device can be found in the Code of Pederal Regarities in the Federal Register.
Image /page/1/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top. Inside the circle, the letters "FDA" are prominently displayed above the word "Centennial". Three stars are located at the bottom of the circle. The logo appears to be a commemorative emblem for the FDA's centennial anniversary.
romoting Public 9
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (31 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C Hodgon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
Device Name: Stryker Ureteral Illumination System IV
Indications for Use:
The InfraVision Ureteral Kit with the InfraVision IR Illuminator is intended to transilluminate the ureter during surgical procedures. The Catheter accepts up to a 0.038"-guidewire. The Kit can be used for either open surgical (with the InfraVision Detector Probe) or laparoscopic (with the InfraVision Camera) access.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
text
Nancyc Broadon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices,
510(k) Number X061548
Stryker Ureteral Illumination System IV
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.