ELECTROSUGRICAL RESECTION AND VAPORIZATION ELECTRODE SERIES
Applicant
Olympus Winter & Ibe GmbH
Product Code
FAS · Gastroenterology, Urology
Decision Date
Jun 20, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HF-Resection Electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation or removal of soft tissue and where hemostasis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedures for which the devices can be used are transurethral resection in saline (TURis), transurethral prostatectomy, transurethral resection of the prostate (TURP) for benign prostatic hyperplasia, transurethral incision of the prostate (TUIP) or bladder neck, transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used to treating cancer of the prostate. The HF-Resection Electrode for Plasma Vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving the vaporization, ablation, coagulation, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include use in the prostate, bladder and bladder neck. The specific treatment indications include benign prostate hyperplasia BPH, bladder cancer, tumors, lesions and neoplasms. The specific urological indications include transurethral electrovaporization (TUVP, TVP, TUEVP), also known as transurethral vapor resection of the prostate (TUVRP) or transurethral vaporization in saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.
Device Story
Bipolar electrosurgical instrument series; includes active tips (loops, bands, rollers, needles, buttons), insulator, guiding tube, telescope clip, and shaft. Used in endoscopic urological procedures (TURP, TURBT, TUVP) in irrigated environments. Connects to electrosurgical generator (ESG-400) via specialized bipolar cable (WA00014A) to deliver high-frequency energy for tissue resection, vaporization, and hemostasis. Operated by urologists in clinical/surgical settings. Output is thermal energy for tissue modification; enables precise tissue removal while maintaining hemostasis. Benefits include reduced bleeding and efficient tissue management during urological surgeries.
Clinical Evidence
No clinical tests were performed or referenced. Substantial equivalence is supported by bench testing and design verification.
Technological Characteristics
Bipolar electrosurgical electrodes; materials include PTFE insulation. Energy source: external electrosurgical generator (ESG-400). Form factor: endoscopic instrument with various tip geometries (loops, buttons, etc.). Connectivity: wired via bipolar cable. Sterilization: not specified. No software/algorithm components.
Indications for Use
Indicated for patients requiring endoscopic urological soft tissue resection, ablation, vaporization, or coagulation in the prostate, bladder, or bladder neck. Applicable for benign prostatic hyperplasia (BPH), bladder tumors, lesions, and neoplasms. Contraindicated for prostate cancer treatment.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
HF Electrosurgical Resection and Vaporization Electrode Series (K100275)
{0}------------------------------------------------
# K120418 PAGE I OF 3
# JUN 2 0 2012
# 510(k) Summary of Safety and Effectiveness
#### June 20, 2012
## 1. General Information: Applicant:
Official Correspondent:
Manufacturer:
#### 2. Device(s) Identification:
Device Trade Name:
Common Name:
## Electrosurgical Resection and Vaporization Electrode Series Electrosurgical Cutting & Coagulation Device & Accessories Electrode, Electrosurgical, Active, Urological
#### Classification of the device:
Device Classification Name:
Electrosurgical, cutting & coagulation & accessories Endoscopic electrosurgical unit and accessories Resectoscope Resectoscope Working Element
Product Code: Device Classification No .:
Panel: Regulatory Status: FAS, GEI, FJL, FJC, FDC Part 876.4400, 21 CFR 876.4300, 21 CFR 876.1500
General & Plastic Surgery / Gastroenterology and Urology Class II
Olympus Winter & Ibe GmbH Kuehnstrasse 61 22045 Hamburg, Germany Registration Number: 9610773 Owner/Operator Number: 8010313
Sheri L. Musgnung Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway, PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-3147 Facsimile: (484) 896-7128 Email: Sheri.Musgnung@Olympus.com Registration No: 2429304
Olympus Winter & Ibe GmbH Kuehnstrasse 61 22045 Hamburg, Germany Registration Number: 9610773
{1}------------------------------------------------
#### 3. Legally Marketed Predicate Device to which Substantial Equivalence is Claimed:
| Device Trade Name: | HF Electrosurgical Resection and Vaporization Electrode Series |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Olympus Winter & Ibe GmbH |
| 510(k) No.: | K100275 |
| Model Nos.: | WA22301D, WA22302D, WA22503D, WA22305D, WA22306D,<br>WA22507D, WA22537D, WA22521C, WA22523C, WA22331D,<br>WA22332D, WA22538C, WA22539D, WA22351C, WA22438C,<br>WA22355C, WA 22351A, WA22355A, WA22558C, WA22557C |
#### 4. Device Description:
The HF-Resection Electrodes and HF Resection Button Electrode for Plasma Vaporization consist of an active tip, PTFE color code identification, an insulator between the electrode and electrode tube, a guiding tube, telescope clip and arm (shaft).
The design and dimensions of the electrodes vary to accommodate various procedural conditions. The active tips of the various electrodes may consist of loops, bands, rollers, needles or buttons.
The system includes Electrodes, working elements, and an HF bipolar cable that can be connected to an electrical surgical unit.
HF Resection Electrodes (K100275) can now be used with the newly compatible electrosurgical generator, the ESG-400 (K103032). For the HF Resection Electrodes to be used with the ESG-400, an additional HF cable is needed (WA00014A). The WA00014A bipolar reusable cable is an electrosurgical accessory designed to transfer electrosurgical power to electrosurgical working elements from the electrosurgical generator ESG-400 (K103032). The cable is designed to connect the working elements WA22366A and WA22367A (K100275).
#### 5. Indications for Use:
The HF-Resection Electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation or removal of soft tissue and where hemostasis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedures for which the devices can be used are transurethral resection in saline (TURis), transurethral prostatectomy, transurethral resection of the prostate (TURP) for benign prostatic hyperplasia, transurethral incision of the prostate (TUIP) or bladder neck, transurethral resection of bladder turnors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used to treating cancer of the prostate.
The HF-Resection Electrode for Plasma Vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving the vaporization, ablation, coagulation, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include use in the prostate, bladder and bladder neck. The
{2}------------------------------------------------
specific treatment indications include benign prostate hyperplasia BPH, bladder cancer, tumors, lesions and neoplasms. The specific urological indications include transurethral electrovaporization (TUVP, TVP, TUEVP), also known as transurethral vapor resection of the prostate (TUVRP) or transurethral vaporization in saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.
#### 6. Comparison of Technological Characteristics:
The HF Resection Electrodes and the HF Resection Button Electrode for Plasma vaporization are identical to the predicate devices in intended use, design and material specification. The only difference is the new cable (WA00014A) that now allows for the electrodes and working elements to be connected to the ESG-400 (K103032).
The cable is equivalent to WA00013A that allows for the electrodes and working elements to be connected to the UES-40 (K030194 and K100275). The only difference is with the plug.
### 7. Summary of Non-Clinical Performance Testing:
Design verification was performed to ensure the device functions according to its intended use and the results met their acceptance criteria.
Bench Testing according to the FDA-recognized consensus standards IEC 60601-1 and IEC 60601-2-2 was conducted to demonstrate that the design change does not raise any new concerns regarding electrical safety.
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007.
#### 8. Summary of Clinical Tests Submitted or Referenced:
No clinical tests were performed or referenced in support of this Special 510(k).
#### 9. Conclusion:
Olympus Winter & Ibe GmbH believes that the HF Electrosurgical Resection and Vaporization Electrode Series is substantially equivalent to the currently legally marketed devices. It does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Olympus Winter & Ibe GmbH % Olympus America Inc. Ms. Sheri Musgnung Associate Manager RA and Clinical Monitor 3500 Corporate Parkway P.O. Box 610 Center Valley, Pennsylvania 18034-0610
JUN 2 0 2012
Re: K120418
Trade/Device Name: Electrosurgical Resection and Vaporization Electrode Series HF-Resection Electrode for Plasma Vaporization, Model WA22557C HF-Resection Electrode Series, Models WA22301D, WA22302D, WA22503D, WA22305D, WA22306D, WA22507D, WA22537D, WA22521C, WA22523C, WA22331D, WA22332D, WA22538C, WA22539D, WA22351C, WA22438C, WA22355C, WA22351A, WA22355A, WA22558C Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FAS, GEI Dated: May 18, 2012 Received: May 21, 2012
Dear Ms. Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{4}------------------------------------------------
Page 2 - Ms. Sheri Musgnung
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eunice Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ·
Device Name: HF- Resection Electrode Series
Model Numbers:
WA22301D, WA22302D, WA22503D, WA22305D, WA22306D, WA22507D, WA22537D, WA22521C, WA22523C, WA22331D, WA22332D, WA22538C, WA22539D, WA22351C, WA22438C, WA22355C, WA22351A, WA22355A, WA22558C
Indications For Use:
The HF-Resection Electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation or removal of soft tissue and where hemostatis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedures for which the devices can be used for are transurethral resection in saline (TURis), transurethral prostatectorny, transurethral resection of the prostate (TURP) for benign prostatic hypertplasia, transurethral incision of the prostate (TUIP) or bladder neck, transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment.
These devices are not intended to be used for treating cancer of the prostate.
Prescription Use_
Over-The-Counter Use AND/OR
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Page 1 of 2
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K120418
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): _
Device Name: HF- Resection Electrode for Plasma Vaporization
Model Number: WA22557C
Indications For Use:
The HF-Resection Electrode for Plasma Vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving the vaporization, ablation, coagulation, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include use in the prostate, bladder neck. The specific treatment indications include benign prostate hyperplasia BPH, bladder cancer, tumors, The specific urological indications include transurethral and neoplasms. lesions electrovaporization (TUVP, TVP, TUEVP), also known as transurethral vapor resection of the prostate (TUVRP) or transurethral vaporization in saline (TUVis). These devices are intended to be used in an irrigated environment.
These devices are not intended to be used in treating cancer of the prostate.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
***_*** and*_* (Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
Page 2 of 2
510(k) Number K120418
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.