ArcTV Transvaginal Sling System

K183134 · Urocure, LLC · OTN · Feb 7, 2019 · General, Plastic Surgery

Device Facts

Record IDK183134
Device NameArcTV Transvaginal Sling System
ApplicantUrocure, LLC
Product CodeOTN · General, Plastic Surgery
Decision DateFeb 7, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ArcTV Transvaginal Sling System is a retropubic sling indicated for a transvaginal (TV) placement of a midurethral sling for the treatment of adult female stress urinary inconting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Device Story

Sterile, single-use retropubic sling system; includes detachable handle, two delivery needles, and sling assembly. Sling assembly consists of loosely knitted polymeric mesh with integrated bioresorbable suture, two removable plastic insertion sheaths, and two sling connectors. Physician uses handle to direct needles through tissue for transvaginal placement of mid-urethral sling. Integrated suture allows for sling adjustment post-placement without additional tools. Mesh and suture remain permanently implanted; needles, handle, sheaths, and connectors are not implanted. Used in clinical setting to treat female stress urinary incontinence; provides suburethral support to improve patient continence.

Clinical Evidence

Bench testing only. Testing included dimensional, fatigue, tensile, packaging, usability, and distribution testing. Biocompatibility testing performed per ISO 10993-1:2009 for limited duration contact (<24 hours) for needles/handle and permanent implant (>30 days) for mesh/suture.

Technological Characteristics

System components: handle, delivery needles, sling assembly (polymeric mesh, bioresorbable suture, plastic sheaths, sling connectors). Biocompatibility per ISO 10993-1:2009. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Single-use. Mechanical delivery system.

Indications for Use

Indicated for adult females with stress urinary incontinence (SUI) caused by urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line. February 7, 2019 UroCure LLC % Ming-Cheng Chew Regulatory Consultant Libra Medical, Inc. 8401 73rd Ave No., Suite 63 Brooklyn Park, MN 55428 Re: K183134 Trade/Device Name: ArcTV Transvaginal Sling System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OTN Dated: January 9, 2019 Received: January 11, 2019 Dear Ming-Cheng Chew: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Glenn B. Bell -S for Benjamin R. Fisher. Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K183134 Device Name ArcTV Transvaginal Sling System Indications for Use (Describe) The ArcTV Transvaginal Sling System is a retropubic sling indicated for a transvaginal (TV) placement of a midurethral sling for the treatment of adult female stress urinary inconting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD). Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(K) SUMMARY 1 ### 1.1 ADMINISTRATIVE INFORMATION Date of Summary Preparation: 1/9/2018 #### CONTACT INFORMATION 1.2 | Primary Submission Contact | Ming-Cheng Chew | |------------------------------|----------------------------| | | Regulatory Consultant, | | | Libra Medical Inc. | | | Phone: 763-232-3701 | | | Email: mcchew@libramed.com | | Secondary Submission Contact | Sew-Wah Tay, PhD | | | Regulatory Consultant, | ### DEVICE INFORMATION 1.3 | Trade Name | ArcTV Transvaginal Sling System | |---------------------------|-------------------------------------------------------------------------------------------------------------| | Common Name | Urinary sling | | Classification Name | Surgical Mesh | | Classification Regulation | 878.3300 | | Class | II | | Panel | Gastroenterology/Urology | | Product Code | OTN | | Product Code Name | Mesh, Surgical, Synthetic, Urogynecologic, For Stress<br>Urinary Incontinence, Retropubic Or Transobturator | Libra Medical Inc. Phone: 612-801-6782 Email: swtay@libramed.com ### 1.4 510(K) TYPE AND REASON FOR SUBMISSION This 510(k) is a traditional 510(k) and is submitted to obtain marketing clearance for a new device - the ArcTV Transvaginal Sling System. #### PREDICATE DEVICE 1.5 RetroArc Repubic Sling System (K132655) #### DEVICE DESCRIPTION 1.6 The ArcTV Transvaginal Sling System is a sterile, single-use system, consisting of a Handle, two delivery Needles and a Sling assembly. The tip portion of each delivery Needle is configured to allow for passage through tissue. The opposite end of the Needle is configured to connect with {4}------------------------------------------------ the Handle. The Handle is detachable and is used to direct both delivery Needles through tissue. The Sling assembly includes one piece of loosely knitted polymeric Mesh with an integrated bioresorbable Suture, two removable plastic insertion Sheaths, and two Sling Connectors. The Suture is an integral feature of the Mesh. The Suture and Mesh allow for adjustment of the Sling after initial placement in the patient without the use of additional tools. The two plastic Sheaths facilitate the placement of the Sling. The Sling Connectors, Sheaths, Needles, and Handle are used to facilitate placement of the Sling assembly, and are not implanted. The Mesh and the bioresorbable Suture are permanently implanted. ### 1.7 INDICATIONS FOR USE The ArcTV Transvaginal Sling System is a retropubic Sling indicated for a transvaginal (TV) placement of a mid-urethral Sling for the treatment of adult female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD) ### 1.8 PERFORMANCE DATA The ArcTV Transvaginal Sling System has been tested to meet the device's intended use and to ensure conformance to the product specifications. The ArcTV meets all its physical and performance specifications including: - Dimensional ● - Fatigue - Tensile - Packaging Testing - Usability Testing ● - Distribution Testing In addition, the ArcTV was tested for biocompatibility per ISO 10993-1:2009 for external communicating device with tissue for limited duration contact with (<24 hours) for the Needle and Handle, and for permanent implant in tissue (>30 days) for the Mesh and Suture. ### 1.9 SUBSTANTIAL EQUIVALENCE | Device<br>Characteristics | ArcTV Transvaginal Sling System | AMS RetroArc Retropubic Sling<br>System (K132655) | |---------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------| | Intended Use | The ArcTV Transvaginal Sling System<br>is intended to treat female stress urinary<br>incontinence. | Same | {5}------------------------------------------------ | Device<br>Characteristics | ArcTV Transvaginal Sling System | AMS RetroArc Retropubic Sling<br>System (K132655) | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication for Use | The ArcTV Transvaginal Sling System<br>is a retropubic Sling indicated for a<br>transvaginal (TV) placement of a mid-<br>urethral Sling for the treatment of adult<br>female stress urinary incontinence<br>(SUI) resulting from urethral<br>hypermobility and/or intrinsic sphincter<br>deficiency (ISD) | The RetroArc Retropubic Sling<br>System is intended for the placement<br>of a suburethral Sling for the<br>treatment of female stress urinary<br>incontinence (SUI) resulting from<br>urethral hypermobility and/or<br>intrinsic sphincter deficiency (ISD). | | Entry Method | Transvaginal Retropubic | Same | | Delivery System | Handle with detachable curved Needle | Same | | Sling | Sheathed polypropylene mesh with<br>adjustment bioresorbable suture | Same | | Sterilization | EO, SAL 10⁻⁶, single-use | Same | | Packaging | Molded tray with Tyvek lid and<br>separate pouch for Sling | Same | # 1.10 CONCLUSION Based on the test data and the same intended use, the ArcTV is substantially equivalent to its predicate. The minor differences do not raise any new questions of safety or efficacy.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...