K210087 · Athena Surgical, LLC · OTN · Apr 12, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210087
Device Name
Athena Surgical RMUS System
Applicant
Athena Surgical, LLC
Product Code
OTN · General, Plastic Surgery
Decision Date
Apr 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Athena Surgical RMUS System is indicated for use as a pubourethral sling for treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
Device Story
System comprises sterile, single-use retropubic mid-urethral sling (RMUS) implant assembly and non-sterile, reusable handle assembly. Implant consists of blue polypropylene mesh (1.1cm x 46cm) with clear plastic sheath and stainless steel needles. Used in surgical setting by physicians; needles attach to handle assembly to position mesh via transvaginal retropubic approach. Mesh provides sub-urethral support to treat stress urinary incontinence. Device is mechanical; no software or algorithms involved.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including mechanical performance (tensile strength, elongation, stiffness, burst strength, dimensional analysis), sterilization validation (ISO 11135:2014), reprocessing validation, shelf-life testing, and biocompatibility (ISO 10993-1).
Indicated for female patients with stress urinary incontinence caused by urethral hypermobility and/or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 9, 2021
Athena Surgical, LLC % Kellen Hills Biomedical Engineer Medavise Consulting 8725 Columbine Rd. #44952 Eden Prairie, MN 55344
Re: K210087
Trade/Device Name: Athena Surgical RMUS System Regulation Number: 21 CFR& 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTN Dated: January 11, 2021 Received: January 13, 2021
Dear Kellen Hills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
For
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210087
Device Name Athena Surgical RMUS System
Indications for Use (Describe)
The Athena Surgical RMUS System is indicated for use as a pubourethral sling for treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Athena Surgical. The word "ATHENA" is written in large, yellow, sans-serif font. Below the word "ATHENA" is the word "SURGICAL" written in a smaller, blue, sans-serif font.
**510(k) Summary**
[As Required by 21 CFR 807.92]
| (a)(1) Date Prepared: | | January 11, 2021 |
|---------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Submitted By: | Athena Surgical, LLC<br>6110 Blue Circle Drive, Suite 280<br>Minnetonka, MN 55343 |
| | Phone: | 612.888.9394 |
| | Contact: | Greg Slusser |
| | Prepared By: | Medavise, LLC<br>8725 Columbine Rd. #44952<br>Eden Prairie, MN 55344 |
| | Phone: | 612.405.4059 |
| (a)(2) Proprietary Name: | | Athena Surgical RMUS System |
| | Common Name: | Urinary sling |
| | Classification Regulation:<br>Class:<br>Product Code: | 21 CFR 878.3300 - Surgical Mesh<br>ll<br>OTN - Mesh, Surgical, Synthetic,<br>Urogynecologic, For Stress Urinary<br>Incontinence, Retropubic Or<br>Transobturator |
| (a)(3) Predicate Devices: | | |
| | Primary: | Ethicon GYNECARE TVT EXACT™ |
*No predicate has been subject to a design related recall.
### (a)(4) Device Description:
The Athena Surgical RMUS System consists of a sterile, single-use Retropubic Mid-Urethral Sling (RMUS) Implant Assembly and a non-sterile, reusable RMUS Handle Assembly. Each RMUS Implant Assembly contains one blue polypropylene mesh sling implant (1.1cm x 46cm), covered by a non-implantable clear plastic sheath and attached on each end to nonimplantable stainless steel needles. The mesh implant is constructed of knitted filaments of extruded polypropylene strands and is approximately 0.63mm thick.
(K132054);
The Athena Surgical RMUS Implant Assembly is designed to be used with the non-sterile, reusable Athena Surgical RMUS Handle Assembly, which consists of two parts: the Handle Body and the Handle Insert. The stainless steel needles from the RMUS Implant Assembly
6110 Blue Circle Drive | Suite 280 | Minnetonka, MN 55343
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Athena Surgical. The word "ATHENA" is written in large, gold, sans-serif font. Below the word "ATHENA" and slightly to the left is the word "SURGICAL" in a smaller, sans-serif font.
are designed to fit inside the RMUS Handle Assembly, which is used to position the implantable mesh in the patient from a vaginal incision up through the abdominal wall.
The purpose of this submission is to gain initial marketing authorization in the United States.
- (a)(5) Indications for Use:
The Athena Surgical RMUS System is indicated for use as a pubourethral sling for treatment of female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
- (a)(6) Comparison of Technological Characteristics:
The Athena Surgical RMUS System is substantially equivalent to the identified predicate based on similarities in intended use, design, materials, sterilization and performance. The technological characteristics do not raise any new questions of safety and efficacy.
| | Subject Athena Surgical RMUS System | Predicate Ethicon GYNECARE TVT™ |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | TBD | K132054 |
| Regulation and<br>Product Code | 21 CFR 878.3300 - OTN | Same |
| Intended Use | The Athena Surgical RMUS System is<br>intended to treat female stress urinary<br>incontinence. | Same |
| Indications | The Athena Surgical RMUS System is<br>indicated for use as a pubourethral sling<br>for treatment of female stress urinary<br>incontinence resulting from urethral<br>hypermobility and/or intrinsic sphincter<br>deficiency. | The GYNECARE TVT EXACT Continence<br>System is intended to be used as a pubo-<br>urethral sling for treatment of female<br>Stress Urinary Incontinence, resulting<br>from urethral hypermobility and/or<br>intrinsic sphincter deficiency. |
| Materials | Mesh - Polypropylene<br>Sheath - Low Density Polyethylene<br>Needle - Stainless Steel | Same |
| Delivery System<br>Design | Handle with detachable curved Needle | Handle with fixed curved Needle |
| Entry Method | Transvaginal Retropubic | Same |
| Sterility | EtO, SAL 10-6, single-use | Same |
| Performance | Ultimate Tensile Strength<br>Elongation at Break<br>Stiffness Bending Length | Same |
{5}------------------------------------------------
K210087 Page 3 of 3
Image /page/5/Picture/1 description: The image shows the word "ATHENA" in a stylized, sans-serif font. The letters are a golden color. Below the word "ATHENA" is the word "SURGICAL" in a smaller, sans-serif font. The word "SURGICAL" is a light blue color.
(b)(1) Non-clinical testing:
i
The subject device was successfully evaluated according to the following:
- Mechanical performance testing
- Ultimate tensile strength .
- . Elongation at break
- . Stiffness bending strength
- . Burst strength
- . Dimensional analysis (monofilament diameter, device width, pore size, mesh thickness, mesh density, mesh weave characteristics)
- 트 Sterilization validation per ISO 11135:2014
- Reprocessing validation for reusable instrumentation ■
- I Shelf life testing evaluating packaging integrity and mechanical performance
- 트 Biocompatibility per ISO 10993-1
- Implant Sheath and Needles (external communicating device contacting tissue for less than 24 . hours)
- o Cytotoxicity
- Sensitization O
- Irritation O
- Acute systemic toxicity O
- Mesh (permanent implant contacting tissue for greater than 30 days) ●
- Cytotoxicity O
- Sensitization O
- Irritation O
- Acute systemic toxicity O
- Implantation O
- Subacute toxicity, subchronic toxicity, genotoxicity, and carcinogenicity through O chemical characterization and toxicological risk assessment
- . Pyrogenicity testing
(b)(2) Clinical testing:
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
## (b)(3) Conclusions:
Based on the information provided in this premarket notification the subject Athena Surgical RMUS System demonstrates substantial equivalence to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.