← Product Code [OTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTM) · K182169

# AdVance XP Male Sling (K182169)

_Boston Scientific Corporation · OTM · Nov 27, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/OTM/K182169

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [OTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OTM.md)
- **Decision Date:** Nov 27, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The AdVance™ XP Male Sling System is intended for the treatment of male stress urinary incontinence (SUI) by the placement of a suburethral sling.

## Device Story

AdVance™ XP Male Sling System treats male stress urinary incontinence; repositions bulbar urethra 2-4 cm. System components: permanently implanted monofilament polypropylene mesh sling; two needle passers; SKW retractor; percutaneous needle. Procedure: sling connected to needle passers via keyed connectors; connectors removed after positioning. Used in clinical setting by surgeons. Device provides mechanical support to urethra to restore continence.

## Clinical Evidence

Evidence includes bench testing, biocompatibility testing, real-world complaint data, and clinical literature searches. No new clinical trials were conducted; data supports substantial equivalence to predicate.

## Technological Characteristics

Monofilament polypropylene mesh; transobturator surgical approach; keyed connector system for implantation; includes needle passers, SKW retractor, and percutaneous needle. Mechanical device; no software or energy source.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- AdVance™ Male Sling System ([K053371](/device/K053371.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 27, 2018

Boston Scientific Corporation Kaitlyn Rainbow Senior Regulatory Affairs Specialist 10700 Bren Road West Minnetonka, MN 55343

Re: K182169

> Trade/Device Name: AdVance™ XP Male Sling System Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTM Dated: August 16, 2018 Received: August 17, 2018

Dear Kaitlyn Rainbow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the name "Glenn B. Bell -S" in a large, bold font. The text is black and is set against a white background. The letters are evenly spaced and easy to read.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below.                                                                                         |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                                     |                                                                                                                                                                                     |
| 510(k) Number (if known)                                                | K182169                                                                                                                                                                             |
| Device Name                                                             | AdVance™ XP Male Sling System                                                                                                                                                       |
| Indications for Use (Describe)                                          | The AdVance™ XP Male Sling System is intended for the treatment of male stress urinary incontinence (SUI) by the             placement of a suburethral sling.                      |
| Type of Use (Select one or both, as applicable)                         | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Boston Scientific. The text "Boston" is on the top line, and the text "Scientific" is on the bottom line. The text is in a dark blue font.

Boston Scientific Corporation 300 Boston Scientific Way Marlborough, MA 01752 508-683-4000 www.bostonscientific.com

## 510(k) Summary

| Submitter                                                                      | Boston Scientific Corporation<br>10700 Bren Road West<br>Minnetonka, MN 55343<br>USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                | Phone: 952-930-6000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                | Contact Person: Kaitlyn Rainbow<br>Date Prepared: November 26, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device:                                                                        | Device Name: AdVance™ XP Male Sling System<br>Model: 720163-01<br>Common Name: Sub-Urethral Sling System; Surgical Mesh<br>Classification Name: Surgical Mesh<br>Classification Number: 21 CFR 878.3300<br>Regulatory Class: II<br>Product Code: OTM<br>Product Code Name: Mesh, Surgical, For Stress Urinary Incontinence, Male                                                                                                                                                                                                                                       |
| Predicate Device                                                               | AdVance™ Male Sling System, K053371<br>No reference devices were used for this submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description                                                             | AdVance™ XP is a sling system which treats male stress urinary incontinence<br>by repositioning the urethra with a suburethral mesh. The AdVance XP<br>system is comprised of a permanently implanted monofilament polypropylene<br>mesh sling, two needle passers used to implant the sling, SKW retractor and<br>percutaneous needle. The sling is connected during the procedure to the<br>needle passers through keyed connectors at each end of the sling which are<br>removed after the sling is positioned. The sling repositions the bulbar urethra<br>2-4cms. |
| Indications for Use                                                            | The AdVance™ XP Male Sling System is intended for the treatment of male<br>stress urinary incontinence (SUI) by the placement of a suburethral sling.                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                | The indications for use of the AdVance™ XP Male Sling System are the same<br>as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device | AdVance™ XP and the predicate device achieve their mechanism of action<br>by repositioning the bulbar urethra 2-4 cm. Both devices have the same mesh<br>material, pore size, sling arm width, sheath and connectors, and transobturator<br>surgical approach.                                                                                                                                                                                                                                                                                                         |

{4}------------------------------------------------

|                  | The technological differences that exist between the two devices are sling<br>length, mesh weave, anchoring mechanism, and needle passer shape.                                                                                                                                                                  |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data | To demonstrate substantial equivalence of the AdVance™ XP to the predicate<br>device, technological characteristics and performance criteria were evaluated<br>using bench testing, biocompatibility testing, real world complaint data,<br>clinical literature searches, and clinical evaluations.              |
|                  | The results from this testing and other data demonstrate that the technological<br>characteristics and performance criteria of the AdVance™ XP are comparable<br>to the predicate device and can perform in a manner equivalent to the devices<br>currently on the market for the same intended use.             |
| Conclusion       | As the indications for use and fundamental scientific technology have not<br>changed, non-clinical and clinical performance data supports a determination<br>that the subject device is substantially equivalent to the predicate device, and<br>that it is at least as safe and effective for its intended use. |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/OTM/K182169](https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/OTM/K182169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
