← Product Code [ESW](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/ESW) · K030559

# RUSCH POLYFLEX STENT FOR THE ESOPHAGUS W/INTRODUCER/DELIVERY SYSTEM (K030559)

_Rusch Intl. · ESW · Jun 30, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/ESW/K030559

## Device Facts

- **Applicant:** Rusch Intl.
- **Product Code:** [ESW](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/ESW.md)
- **Decision Date:** Jun 30, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3610
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Rüsch Polyflex Stent for the Esophagus with Introducer System intended for use with Esophageal Stenoses, indications are: Stenting: Esophageal stenoses (strictures), such as stenting refractory benign and/or malignant strictures. Esophageal-respiratory-fistula. Maintaining esophagus lumen patency in esophageal strictures caused by intrinsic or extrinsic tumors.

## Device Story

Esophageal stent system; used for treatment of esophageal strictures and fistulas. Components include: Polyflex stent, introducer sleeve, stent loader, insertion tube with dilator, soft positioner, stopper, and stent clamp. Stent consists of integral polyester braid encapsulated in medical-grade silicone. Delivered orally via introducer system to maintain esophageal lumen patency. Used by clinicians to manage refractory benign or malignant strictures and esophageal-respiratory-fistulas. Provides mechanical support to keep esophagus open; improves patient ability to swallow/maintain nutrition.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design similarities to predicate devices.

## Technological Characteristics

Stent composed of integral polyester braid encapsulated in medical-grade silicone. Delivery via oral insertion tube/introducer system. Mechanical device; no energy source or software.

## Regulatory Identification

An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

## Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

## Predicate Devices

- Hood Esophageal Prosthesis
- Boston Scientific Ultraflex™
- Boston Scientific Wall Stent
- Cook, Inc. Esophageal Z-Stent
- Wilson-Cook Esophageal Prosthesis Set

## Reference Devices

- Rüsch Polyflex Stent Kit ([K982614](/device/K982614.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 3 0 2003

Image /page/0/Picture/1 description: The image shows a document with the text "K030559" at the top, which appears to be a document or reference number. Below that, it says "page 1 of 2", indicating that this is the first page of a two-page document. The word "RUSCH" is prominently displayed in a bold, stylized font, followed by "INTERNATIONAL" and "Group Regulatory Affairs" in smaller fonts. The text "A Subsidiary of Teleflex incorporated (USA)" is at the bottom of the image.

Tall Pines Park Jaffrey, NH 03452 (603) 532-7706 FAX (603) 532-8211 or 6108

# 510 (k) Summary

#### Submitter Name, Address, and Date of Submission. 1.

Karenann J. Brozowski Group Regulatory Affairs Director Rüsch International Tall Pines Park Jaffrey, New Hampshire 03452

(603) 532-7706 Telephone: Facsimile: (603) 532-6207

Contact: Same as above

### 2 . Name of the Device, Common, Proprietary (if Known), and Classification.

## Classification:

Class II, Product Code 79ESW, 21 CFR 878.3610

### Common Name:

Esophageal Stent

## Proprietary Name:

Rüsch Polyflex Stent for the Esophaqus with Introducer/Delivery System

### 3. Identification of the legally marketed device to which the submitter claims equivalence.

The Rüsch Polyflex Stent for the Esophagus with Introducer System is substantially equivalent to the Hood Esophageal Prosthesis, Boston Scientific Ultraflex™, Boston Scientific Wall Stent, Cook, Inc. Esophageal Z-Stent, and the Wilson-Cook Esophageal Prosthesis Set.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K030559", and the text below reads "page 2 of 2". The handwriting is in black ink and appears to be clear and legible.

page 2

#### Description of the Device. 4 .

The Rüsch Polyflex Stent for the Esophagus with Introducer System consists of the Polyflex Stent, Introducer Sleeve, Stent Loader, Insertion Tube with Dilator, Soft Positioner, Stopper and Stent Clamp.

#### Intended Use of the Device. 5.

The Rüsch Polyflex Stent for the Esophagus with Introducer System intended for use with Esophageal Stenoses, indications are: Stenting: Esophageal stenoses (strictures), such as stenting refractory benign and/or malignant strictures. Esophageal-respiratory-fistula. Maintaining esophagus lumen patency in esophageal strictures caused by intrinsic or extrinsic tumors.

### Summary of Technological Characteristics. 6.

The following technological characteristics are the same as or equivalent to predicate devices:

The stent is constructed with an integral polyester braid, which is surrounded by medical grade silicone. This is the same construction as Rüsch Polyflex Stent Kit, which was cleared under K 982614.

The product is delivered through the mouth via a delivery system as is the Wilson Cook Esophageal Prosthesis, the Boston Scientific Ultraflex or Hood Prosthesis.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human figures.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 3 0 2003

Ms. Karenann J. Brozowski Group Regulatory Affairs Director Rüsch International Tall Pines Park JAFFREY NH 03452

Re: K030559

Trade/Device Name: Rüsch Polyflex Stent for the Esophagus with Introducer/Delivery System Regulation Number: 21 CFR §878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: 79 ESW Dated: June 6, 2003 Received: June 10, 2003

Dear Ms. Brozowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrf/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. B rogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K030559

Device Name: Rüsch Polyflex Stent for the Esophagus with Introducer System

Indications for Use:

Stenting:

Esophaqeal stenoses (strictures), such as stenting refractory beniqn strictures and/or malignant strictures

Esophageal-respiratory-fistula

Maintaining esophagus lumen patency in esophageal strictures caused by intrinsic or extrinsic tumors

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

V Prescription Use (Per 21 CFR 801.109)

OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

David A. Seagram

510k) N

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/ESW/K030559](https://fda.innolitics.com/submissions/GU/subpart-d%E2%80%94prosthetic-devices/ESW/K030559)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
