K083838 · Fook Tin Technologies Limited · MNW · May 12, 2009 · Cardiovascular
Device Facts
Record ID
K083838
Device Name
SCALEMAN BODY FAT SCALES, MODEL: FS-148BW1 SERIES
Applicant
Fook Tin Technologies Limited
Product Code
MNW · Cardiovascular
Decision Date
May 12, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Scaleman Body Fat Scales - Models in "Family Model List 1A" is a series of body composition analyzers that measure body weight and impedance and estimate percentage of body fat and body water using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment. The Scaleman Body Fat Scales - Models in "Family Model List 1B" is a series of body composition analyzers that measure body weight and impedance and estimate percentage of body fat and body water, bone mass and muscle mass using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment.
Device Story
Scaleman Body Fat Scales are body composition analyzers using foot-to-foot bioelectrical impedance analysis (BIA). Device measures body weight and electrical impedance; estimates body fat, body water, bone mass, and muscle mass. Used in home environment by consumers. User stands on scale; device applies 50 KHz signal; internal electronics calculate composition metrics based on impedance. Output displayed to user for personal health monitoring. Benefits include tracking body composition changes over time.
Clinical Evidence
Comparative clinical study evaluated Scaleman scales against predicate device. Sample size not specified. Primary endpoints included body fat (%), FFM (kg), body water (%), muscle mass (kg), bone mass (kg), and weight (kg). Results showed significant correlation (P<.05) between test and predicate devices in both normal and athlete modes.
Technological Characteristics
Foot-to-foot bioelectrical impedance analysis (BIA) at 50 KHz. 2 or 4 electrode configuration. Powered by replaceable 9V, CR2032, or AAA batteries. Form factor is a digital scale. No specific material standards or software architecture details provided.
Indications for Use
Indicated for healthy children 10-17 years old and healthy adults with active, moderately active, or inactive lifestyles for body composition assessment in home environments.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Tanita Innerscan Body Composition Monitor (K040778)
Submission Summary (Full Text)
{1}------------------------------------------------
| 510(k) Summary | |
|--------------------------|--|
| Scaleman Body Fat Scales | |
CONFIDENTIAL
| Submitter Name: | Fook Tin Technologies Ltd. | MAY 12 2009 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Submitter Address: | 4/F Eastern Center, 1065 King's Road, Quarry Bay, Hong Kong | |
| Contact Person: | Dr. John Chai, Managing Director | |
| Phone Number: | (852) 2960 7288 | |
| Fax Number: | (852) 2565 9672 | |
| Date: | October 30, 2008 | |
| Device Trade Name: | Scaleman Body Fat Scales | |
| Device Common Name: | Impedance plethysmograph (body composition analyzer/scales) | |
| Classification Name: | Impedance plethysmograph | |
| Classification Number: | 21 CFR 870.2770 | |
| Product code: | MNW | |
| Predicate Device: | Tanita Innerscan Body Composition Monitor, K040778 | |
| Statement of Intended Use: | The Scaleman Body Fat Scales - Models in "Family Model List 1A" is a series of body composition analyzers that measure body weight and impedance and estimate percentage of body fat and body water using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment. | |
The Scaleman Body Fat Scales - Models in "Family Model List 1B" is a series of body composition analyzers that measure body weight and
{2}------------------------------------------------
# CONFIDENTIA
impedance and estimate percentage of body fat and body water, bone mass and muscle mass using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment.
Device Description, Summary of Technologica! Characteristics, and Comparison to the Predicate Device
The models in Scaleman body fat scales' Family Model List 1A and 1B submitted in this 510(k) are body composition analyzers. The devices estimate percentages of body fat, body water, bone mass and muscle mass (as noted in the indications statement) based on bioelectrical impedance analysis (BIA). The following table illustrates the substantial equivalence in technology and intended use to the predicate device.
| Feature | Proposed Device:<br>Scaleman Body Fat Scales<br>(Model FS-148BW1) | Predicate Device:<br>Tanita InnerScan Body Composition<br>Monitor (Model BC-533) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | NIL | K040778 |
| Manufacturer | Fook Tin Technologies Ltd. | Tanita Corp of America |
| Classification # | 21 CFR 870.2770 | 21 CFR 870.2770 |
| Product code | MNW | MNW |
| Indication for use | The Scaleman Body Fat Scales -<br>Models in "Family Model List 1A" is a<br>series of body composition analyzers<br>that measure body weight and<br>impedance and estimate percentage<br>of body fat and body water using BIA<br>(bioelectrical impedance analysis).<br>They are intended for use by healthy<br>children 10-17 years old and healthy<br>adults with active, moderately active,<br>to inactive lifestyles for body<br>composition assessment in the home<br>environment.<br><br>The Scaleman Body Fat Scales -<br>Models in "Family Model List 1B" is a<br>series of body composition analyzers | The Tanita family of InnerScan Body<br>Composition Monitors measure body<br>weight and impedance and estimate<br>percentage of body fat and body water,<br>visceral fat rating, bone mass, muscle<br>mass, physique rating, daily calorie<br>intake (DCI) and metabolic age using<br>BIA. They are intended for use by<br>healthy children 7-17 years old and<br>healthy adults with active, moderately<br>active, to inactive lifestyles for body<br>composition assessment in the home<br>environment. |
{3}------------------------------------------------
## UUINFIDEN I IAL
| | that measure body weight and<br>impedance and estimate percentage<br>of body fat and body water, bone<br>mass and muscle mass using BIA<br>(bioelectrical impedance analysis).<br>They are intended for use by healthy<br>children 10-17 years old and healthy<br>adults with active, moderately active,<br>to inactive lifestyles for body<br>composition assessment in the home<br>environment. | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description | Body composition analyzer/scale that utilizes a “foot-to-foot” bioelectrical impedance (BIA) technology to determine internal body composition. | Body composition analyzer/scale that utilizes a “foot-to-foot” bioelectrical impedance (BIA) technology to determine internal body composition. |
| Analysis method | BIA | BIA |
| Operating parameters | 50 KHz | 50 KHz |
| Power source | Replaceable 9V or CR2032 or AAA batteries, depending on the model | AA batteries |
| Operating keys | Range of 3 to 6, depending on the model | 13 |
| Number of electrodes | 4 or 2 | 4 |
| Clinical performance | A comparative clinical study, submitted in this notification (Section 20), showed<br>that body fat (%) and FFM (kg) in normal and athlete modes, body water (%)<br>muscle mass (kg), bone mass (kg) and weight (kg) measured or estimated by<br>test and predicate devices were significantly correlated (P<. 05). | |
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
MAY 12 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Public Health Service
Fook Tin Technologies Ltd c/o Dr. John Chai Managing Director 4/F Eastern Center 1065 King's Road Quarry Bay, Hong Kong
Re: K083838
Trade/Device Name: Scaleman Body Fat Scales Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II (two) Product Codes: MNW Dated: April 18, 2009 Received: May 5, 2009
Dear Dr. Chai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Dr. John Chai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Dima R. Willing
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
### 510(k) Number (if known): KO83838
Device Name: Scaleman
Indications For Use:
Family Model List 1A
A scries of body composition analyzers that measure body weight and impedance and estimate percentage of body fat and body water, using BIA (bioelectrical impedance analysis).
They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment.
Family Model List 1B
A series of body composition analyzers that measure hody weight and impedance and estimate percentage of body fat and body water, bone mass and muscle mass using BIA (bioelectrical impedance analysis). They are intended for use by healthy children 10-17 years old and healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment.
NO Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
YES Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Holmes
n of Cardiovascular Devices
510/k) Numbe
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.