← Product Code [MNW](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW) · K072325

# BODY COMPOSITION ANALYZER, EASYBODY 202, 203, 205 (K072325)

_Jawon Medical Co., Ltd. · MNW · Nov 30, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K072325

## Device Facts

- **Applicant:** Jawon Medical Co., Ltd.
- **Product Code:** [MNW](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW.md)
- **Decision Date:** Nov 30, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2770
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

Body Composition Analyzer Models EasyBody 202 and EasyBody 203 are intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method. Body Composition Analyzer Model EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake. daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method. The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.

## Device Story

Non-invasive body composition analyzers (EasyBody 202, 203, 205) measure user weight and bio-electrical impedance (BIA) at 50kHz. Devices use 4 electrodes (202/203) or 8 electrodes (205) in a tetra-polar configuration. Inputs (impedance, weight) processed via experimentally derived algorithm to estimate body fat percentage, lean body mass, BMR, and exercise recommendations. Operated by user in upright position; results displayed to user. Provides health-related metrics to assist in fitness and lifestyle management. Powered by AA batteries or AC adapter.

## Clinical Evidence

Clinical comparison tests performed against predicate device ZEUS 9.9. Results demonstrate no significant difference between subject devices and predicate. Bench testing confirmed compliance with IEC 60601-1, EN 60601-1-2, and ISO 10993 standards.

## Technological Characteristics

BIA-based body composition analyzer. 50kHz operating frequency. Tetra-polar electrode method (4 electrodes for 202/203; 8 electrodes for 205). Powered by 6V DC (AA batteries) or AC adapter. Measures impedance (100-950 ohms) and weight (1-150 kg). Upright patient position.

## Regulatory Identification

An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Body Composition Analyzer Model ZEUS 9.9 ([K053556](/device/K053556.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072325
Page 1 of 4

# Section 5

# 510(k) Summary

NOV 3 0 2007

[As Required by 21 CFR 807.92]

| Date Prepared:       | August 03, 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:           | Jawon Medical Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                      | #1208-12, Sinsang-Li, Jinryang-Myun, Kyungsan-city, Kyungsang-Bukdo,                                                                                                                                                                                                                                                                                                                                                                                                               |
|                      | 712-830, Republic of Korea                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                      | Tel: 82-53-856-0993                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                      | Establishment Registration Number: 9616164                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Contact Person:      | MI Consulting Co., Ltd. (Mr. HL Jung)                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                      | RM 624, Life Officetel, 61-3, Yoido-dong, Youngdeungpo-gu, Seoul, 150-731,                                                                                                                                                                                                                                                                                                                                                                                                         |
|                      | Republic of Korea                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                      | Tel: 82-2-6264-9001                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                      | Fax: 82-2-6269-9001                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                      | Email: jung_halyong@yahoo.com                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:          | Body Composition Analyzer                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                      | Models EasyBody 202, EasyBody 203 and EasyBody 205                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Common Name:         | Body Fat Analyzer / Body Fat Monitor / Body Composition Monitor                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification Name: | Body Composition Analyzer (Impedance plethysmograph) / MNW                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device:    | Body Composition Analyzer Model ZEUS 9.9 (K053556)                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Description:  | The EasyBody 202, EasyBody 203 and EasyBody 205 are non-invasive body<br>composition analyzers intended for use only in healthy subjects between the<br>age of 5-89. The devices employ BIA(Bio-electrical Impedance Analysis)<br>method, 4 electrodes for EasyBody 202 and EasyBody 203, 8 electrodes for<br>EasyBody 205, and then measure body composition using an experimentally<br>derived algorithm. The devices are powered by four AA batteries (DC 6V) or<br>AC adapter. |
| Intended use:        | Body Composition Analyzer Models EasyBody 202 and EasyBody 203 are<br>intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat),                                                                                                                                                                                                                                                                                                                                     |
| Jawon Medical        | EasyBody                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

Section 5: Page 1 of 4

{1}------------------------------------------------

LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.

Body Composition Analyzer Model EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake. daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.

The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.

- Technologic characteristics: The devices EasyBody 202, EasyBody 203 and EasyBody 205 have been compared to the predicate device ZEUS 9.9 such as a comparing table of technology characteristics. The differences in this submission don't raise new questions concerning either safety or effectiveness.
- Non-clinical and clinical tests: The subject devices EasyBody 203 and EasyBody 203 meet the requirements of IEC 60601-1, EN 60601-1-2 and ISO 10993 series. The results of clinical comparison tests with predicate device ZEUS 9.9 demonstrate that there is no significant difference between the subjected devises and the predicate device.
- Conclusions: Based on non-clinical and clinical tests, the subject devices EasyBody 202, EasyBody 203 and EasyBody 205 are as safe, as effective, and perform as well as the predicate device ZEUS 9.9. Accordingly, the subject devices are substantially equivalent to the predicate device.

{2}------------------------------------------------

| ltems                                                     | Zeus 9.9                                                                                                                                                                                                                                                                                                                                                                                                                                 | EasyBody 202                                                                                                                                                                                                                                                                                                     | EasyBody 203                                                                                                                                                                                                                                                                                                     | EasyBody 205                                                                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) #                                                  | K053556                                                                                                                                                                                                                                                                                                                                                                                                                                  | New                                                                                                                                                                                                                                                                                                              | New                                                                                                                                                                                                                                                                                                              | New                                                                                                                                                                                                                                                                                                                             |
| Intended Use                                              | Body composition<br>Analyzer                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Indications<br>for use                                    | Estimated of:<br>PBF(Percentage of<br>Body Fat), MBF(Mass<br>of Body Fat),<br>LBM(Lean Body<br>Mass), TBW(Total<br>Body Water,<br>BMI(Body Mass<br>Index), BMR(Basic<br>Metabolic Rate),<br>Segmental LBM,<br>ICW(Intra-Cellular<br>Water), ECW(Extra-<br>Cellular Water), and<br>Ratio of ECW/TBW<br>Actual :<br>Impedance,<br>BMI(Body Mass<br>Index), weight,<br>WHR(Waist to Hip<br>Ratio)<br>Healthy subjects of<br>the age of 7-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean<br>Body Mass),<br>BMR(Basic<br>Metabolic Rate),<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean<br>Body Mass),<br>BMR(Basic<br>Metabolic Rate),<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean Body<br>Mass), BMR(Basic<br>Metabolic Rate),<br>Segmental LBM,<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 |
| Analysis<br>Method                                        | Bio-electrical<br>Impedance<br>Analysis (BIA)                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Operating<br>parameters                                   | Frequency:<br>1,5,50,<br>250,550,1000kHz                                                                                                                                                                                                                                                                                                                                                                                                 | Frequency:<br>50kHz                                                                                                                                                                                                                                                                                              | Frequency:<br>50kHz                                                                                                                                                                                                                                                                                              | Frequency:<br>50kHz                                                                                                                                                                                                                                                                                                             |
| Electrode<br>type                                         | Tactile                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Number/<br>Placement of<br>Electrodes                     | 8 electrodes<br>Placed on Hands and<br>feet(or ankle)<br>(Tetra-polar Electrode<br>method)                                                                                                                                                                                                                                                                                                                                               | 4 electrodes<br>Placed on feet<br>(Tetra-polar<br>Electrode method)                                                                                                                                                                                                                                              | 4 electrodes<br>Placed on feet<br>(Tetra-polar<br>Electrode method)                                                                                                                                                                                                                                              | 8 electrodes<br>Placed on Hands<br>and feet<br>(Tetra-polar<br>Electrode method)                                                                                                                                                                                                                                                |
| Items                                                     | Zeus 9.9                                                                                                                                                                                                                                                                                                                                                                                                                                 | EasyBody 202                                                                                                                                                                                                                                                                                                     | EasyBody 203                                                                                                                                                                                                                                                                                                     | EasyBody 205                                                                                                                                                                                                                                                                                                                    |
| Patient<br>Position                                       | Upright                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Impedance<br>Measuring<br>Site                            | Whole body,<br>Right Arm, Right Leg,<br>Left Arm, Left Leg<br>Trunk                                                                                                                                                                                                                                                                                                                                                                      | Foot to Foot                                                                                                                                                                                                                                                                                                     | Foot to Foot                                                                                                                                                                                                                                                                                                     | Whole body,<br>Right Arm, Right Leg,<br>Left Arm, Left Leg,<br>Trunk                                                                                                                                                                                                                                                            |
| Impedance<br>Measurement<br>Range                         | 100 ~ 950 ohms                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Measuring<br>height                                       | 110 ~ 200 cm                                                                                                                                                                                                                                                                                                                                                                                                                             | 100 ~ 200 cm                                                                                                                                                                                                                                                                                                     | 100 ~ 200 cm                                                                                                                                                                                                                                                                                                     | 100 ~ 200 cm                                                                                                                                                                                                                                                                                                                    |
| Measuring<br>weight                                       | 10 ~ 250 kg                                                                                                                                                                                                                                                                                                                                                                                                                              | 1~150 kg                                                                                                                                                                                                                                                                                                         | 1~150 kg                                                                                                                                                                                                                                                                                                         | 1~150 kg                                                                                                                                                                                                                                                                                                                        |
| Measuring<br>PBF                                          | 3 ~ 50%                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Measuring<br>Time                                         | Within 1 minute                                                                                                                                                                                                                                                                                                                                                                                                                          | About 15sec                                                                                                                                                                                                                                                                                                      | About 15sec                                                                                                                                                                                                                                                                                                      | About 15sec                                                                                                                                                                                                                                                                                                                     |
| Electrical<br>current<br>applied<br>during<br>measurement | Maximum current<br>360uA                                                                                                                                                                                                                                                                                                                                                                                                                 | Maximum current<br>180uA                                                                                                                                                                                                                                                                                         | Maximum current<br>180uA                                                                                                                                                                                                                                                                                         | Maximum current<br>180uA                                                                                                                                                                                                                                                                                                        |
| Power<br>consumption                                      | 50VA                                                                                                                                                                                                                                                                                                                                                                                                                                     | 0.36VA                                                                                                                                                                                                                                                                                                           | 0.36VA                                                                                                                                                                                                                                                                                                           | 0.36VA                                                                                                                                                                                                                                                                                                                          |
| Power<br>Supply                                           | AC120V 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                           | AA Battery DC6V<br>or AC Adapter                                                                                                                                                                                                                                                                                 | AA Battery DC6V<br>or AC Adapter                                                                                                                                                                                                                                                                                 | AA Battery DC6V<br>or AC Adapter                                                                                                                                                                                                                                                                                                |
| Operating<br>ambient                                      | Temp: 10 ~ 40 ℃<br>RH: 30 ~ 75%                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Storage<br>ambient                                        | Temp: -20 ~ 60 ℃<br>RH: lower than 95%                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                            |
| Dimensions<br>(W x L x H)                                 | 470 x 655 x 1220mm                                                                                                                                                                                                                                                                                                                                                                                                                       | 310 x 311 x 38.5mm                                                                                                                                                                                                                                                                                               | 300 x 356 x 79mm                                                                                                                                                                                                                                                                                                 | 300 x 300 x 369mm                                                                                                                                                                                                                                                                                                               |
| Weight                                                    | 45Kg                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1.7kg                                                                                                                                                                                                                                                                                                            | 2.2kg                                                                                                                                                                                                                                                                                                            | 2.7kg                                                                                                                                                                                                                                                                                                                           |

## < Comparison Table of technical characteristics >

Jawon Medical

EasyBody

Section 5: Page 3 of 4

ે છે. રે

{3}------------------------------------------------

{4}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stripes representing its wings.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### NOV 3 0 2007

Jawon Medical Co., Ltd. c/o H. L. Jung MI Consulting Co., Ltd. Room 624, Life Officetel, 61-3 Yoido-dong, Youngdeungpo-gu Seoul. 150-731 REPUBLIC OF KOREA

K072325 Re:

> Trade/Device Name: JAWON MEDICAL BODY COMPOSITION ANALYZER MODELS EasyBody 202, EasyBody 203, and EasyBody 205 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: November 19, 2007 Received: November 21, 2007

Dear Mr. Jung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA' may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Section 4

## Indications for Use

510(k) Number (if known): _ KO72325

Device Name: JAWON MEDICAL BODY COMPOSITION ANALYZER MODELS EasyBody 202, EasyBody 203 and EasyBody 205

Indications For Use:

The EasyBody 202 and EasyBody 203 are intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.

The EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.

The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Per 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IN NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert Lewin

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

EasyBody 202/203/205

Section 4: Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K072325](https://fda.innolitics.com/submissions/GU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MNW/K072325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
