TANITA BODY COMPOSITION ANALYZER; TANITA BODY FAT MONITOR; TANITA ULTIMATE SCALE
K014009 · Tanita Corp. of America · MNW · Jun 24, 2002 · Cardiovascular
Device Facts
Record ID
K014009
Device Name
TANITA BODY COMPOSITION ANALYZER; TANITA BODY FAT MONITOR; TANITA ULTIMATE SCALE
Applicant
Tanita Corp. of America
Product Code
MNW · Cardiovascular
Decision Date
Jun 24, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
Intended to be used as a body fat analyzer that determines body weight and estimates body fat composition with the use of BIA (bioelectrical impedance analysis).
Device Story
Device consists of platform scale with four stainless steel footpad electrodes; subject stands on scale to measure body weight and lower extremity impedance. System utilizes bioelectrical impedance analysis (BIA) to estimate body composition parameters. Used in professional and consumer settings; operated by clinicians or patients. Output displayed to user to assist in monitoring body composition and weight management. Benefits include non-invasive assessment of body fat and related metrics.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of BIA methodology to predicate.
Indicated for individuals requiring body composition analysis, including measurement of body weight, impedance, body fat percentage, BMI, fat mass, fat-free mass, basal metabolic rate, and total body water (for professional models) or body fat percentage (for consumer models) using bioelectrical impedance analysis.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
| Trade Name: | TANITA Body Fat Analyzer TBF/BF/ Ultimate Scale |
|----------------------|------------------------------------------------------------------------------------------------|
| Common Name: | Body Fat Analyzer/ Body Fat Monitors/ Body Composition Analyzers/<br>Body Composition Monitors |
| Classification Name: | ANALYZER, BODY COMPOSITION<br>21 CFR § 870.2770 |
## Description of Applicant Device:
The TANITA Body Fat Monitor/Scales (multiple models) are designed to determine body weight and body fat composition. Each model will offer a range of features. The system consists of two subdivided stainless steel footpad electrodes mounted on a platform scale. The four-foot electrodes are designed where each foot makes contact with a heel and toe electrode. Impedance of the lower extremities and body weight is measured simultaneously while the subject stands on the scale.
## Intended Uses of Applicant Device:
Intended to be used as a body fat analyzer that determines body weight and estimates body fat composition with the use of BIA (bioelectrical impedance analysis).
#### Predicate Device:
TANITA Body Fat Analyzer Model TBF-105 K930599
## Scientific Concepts and Significant Performance Characteristics:
| | Body Fat Analyzer Model TBF-<br>105<br>K930599 | TANITA Body Fat Analyzer<br>TBF/BF/ Ultimate Scale |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED USE: | A combination non-invasive<br>device, which determines weight<br>and estimates body fat<br>composition using BIA<br>(bioelectrical impedance<br>analysis). | A combination non-invasive<br>device, which determines weight<br>and estimates body fat<br>composition using BIA<br>(bioelectrical impedance<br>analysis). |
| PRODUCT DESCRIPTION: | Body composition analyzer/scales<br>utilize a “foot-to-foot” BIA (bio-<br>electrical impedance) technology -<br>to determine internal body<br>composition. | Body composition analyzer/scales<br>utilize a “foot-to-foot” BIA (bio-<br>electrical impedance) technology<br>to determine internal body<br>composition. |
| ANALYTICAL<br>METHOD/MEASUREMENT | • “Foot-to-Foot” BIA<br>• Patented in-house BIA and<br>DEXA reference methods | • “Foot-to-Foot” BIA<br>• Patented in-house BIA and<br>DEXA reference methods |
{1}------------------------------------------------
# 510(k) SUMMARY, continued
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
Side by side comparisons of the TANITA Body Fat Analyzer TBF/ BF/ Ultimate Scale to the predicate device Body Fat Analyzer Model TBF-105 clearly demonstrate that the applicant device is substantially equivalent to the legally marketed device.
Based on the results of using the previously approved "Foot-to-Foot" BIA methodology with our patented in-house BIA it was concluded that the TANITA Body Fat Analyzer(s) TBF/BF/Ultimate Scales perform as well as the predicate device and therefore have proven its safety and efficacy.
Carol Alloian (Benson) TANITA Corporation of America Manager Requiatory Affairs Manager Phone: (847) 640-9251 ext. 116 (847) 640-7978 Fax:
November 30, 2001
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name around the perimeter of the circle. Inside the circle is a stylized emblem with three wave-like shapes.
JUN 2 4 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Carol Alloian (Benson) Regulatory Affairs Manager TANITA Corporation of America 2625 South Clearbrook Drive ARLINGTON HEIGHTS IL 60005 Re: K014009 Trade/Device Name: Body Fat Analyzer, Professional Series: Models TBF-300A. TBF-310. TBF-410, TBF-215, and BF-350; Consumer Body Fat Monitor/Scale Series: Models TBF-611, TBF-612, TBF-621, TBF-622, BF-623, BF-625, BF-626, BF-541, BF-555, BF-542, TBF-551, BF-556, BF-558, BF-559, BF-572, TBF-521 Ultimate Scales 2000, 2001, and 2001T Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: 74 MNW Dated: March 25, 2002 Received: March 26, 2002
Dear Ms. Alloian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C.Bogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K014009 510(k) Number (if known):
Tanita Body Fat Analyzer Device Name:
Indications for Use:
MODELS TBF-300A, TBF-310, TBF-410, TBF-215, BF-350:
A body composition analyzer that measures body weight and impedance and estimates body fat percent. BMI, fat mass, fat free mass, basal metabolic rate, and total body water using BIA (bioelectrical impedance analysis).
MODELS TBF-611, TBF-612, TBF-621, TBF-622, BF-623, BF-625, BF-626, BF-541, BF-555, BF-542, TBF-551, BF-556, BF-558, BF-559, BF-572, TBF-521, Ultimate Scale 2001:
A body composition analyzer that measures body weight and impedance and estimates body fat percent using BIA (bioelectrical impedance analysis).
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| | | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|------------------------------------------|----|------------------------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use <span style="text-decoration: overline;">✓</span> |
| | | (Optional Format 1-2-96) |
(Division Sign-Off)
Division of Reproductive, Abdominal,
Revised: 6/24/02
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.