C-MOUNT OBJECTIVE LENS F = 17 MM, STEAM-STERILIZABLE: MODEL 85 261.172, C-MOUNT OBJECTIVE LENS F = 21 MM, STEAM-STERILIZ
Device Facts
| Record ID | K964173 |
|---|---|
| Device Name | C-MOUNT OBJECTIVE LENS F = 17 MM, STEAM-STERILIZABLE: MODEL 85 261.172, C-MOUNT OBJECTIVE LENS F = 21 MM, STEAM-STERILIZ |
| Applicant | Richard Wolf Medical Instruments Corp. |
| Product Code | FET · Gastroenterology, Urology |
| Decision Date | Dec 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The C-Mount objective lenses serve to connect endoscope to video cameras. The lenses with clamping mechanism are used for video-controlled diagnostics and therapy and allow quick adaptation and fixation of the standard telescopes with eyepiece. The focus ring serves to focus an endoscopic image on the monitor.
Device Story
C-Mount objective lenses facilitate coupling of endoscopes to video cameras; feature rotatable focus ring for image adjustment on monitor; include clamping mechanism for fixation to standard telescopes. Used in clinical settings by physicians for video-controlled diagnostics and therapy. Device redesigned to support steam sterilization; maintains original production method and function.
Clinical Evidence
No clinical tests performed. Bench testing confirmed no leakage or reduction in optical quality following steam sterilization.
Technological Characteristics
C-Mount objective lenses with integrated clamping mechanism and rotatable focus ring. Redesigned for steam sterilization compatibility. Mechanical optical coupling device; no electronic or software components.
Indications for Use
Indicated for use in video-controlled endoscopic diagnostics and therapy to connect standard telescopes with eyepieces to video cameras.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- RIWO lenses f = 15 mm, 30 mm, 60 mm (pre-enact.)
- C-mount lenses f = 25 mm, 38 mm (Karl Storz)
- Coupler 28 mm, 35 mm (Snowden Pencer)
Related Devices
- K994090 — KARL STORZ AUTOCLAVABLE CAMERA HEAD · Karl Storz Imaging, Inc. · Feb 4, 2000
- K990154 — KARL STORZ DIRECT COUPLED INTERFACE (DCI 5) CAMERA SYSTEM · Karl Storz Imaging, Inc. · Jun 1, 1999
- K020336 — ENDOSCOPIC CAMERA MC404/C3 · World of Medicine Lemke GmbH · Mar 14, 2002
- K102059 — AUTOCLAVABLE CAMER HEAD MODEL OTV-Y0017 · Olympus Medical Systems Corporation · Oct 8, 2010
- K973252 — KARL STORZ DIRECT COUPLED CAMERA HEAD · Karl Storz Imaging, Inc. · Nov 24, 1997