K182696 · KARL STORZ Endoscopy-America, Inc. · FET · Nov 20, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K182696
Device Name
Telepack X LED
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
FET · Gastroenterology, Urology
Decision Date
Nov 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The TELE PACK X LED is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.
Device Story
Portable, all-in-one imaging system; includes 15-inch monitor, camera control unit (CCU), and internal LED light source. Connects to compatible camera heads or video endoscopes. Used for visualization and documentation of endoscopic/microscopic procedures and stroboscopy. Operated by clinicians in clinical settings. Inputs: optical signals from endoscopes. Processing: image processing (capture, flip/mirror, zoom, brightness control, white balance). Outputs: visual display on integrated monitor; storage of image/video/audio files to SD card or network server. Benefits: provides integrated illumination and visualization for surgical/diagnostic procedures.
Clinical Evidence
No clinical data required for substantial equivalence. Evidence based on non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), and photobiological safety (IEC 62471). Performance testing verified minimum illumination, spatial resolution, color performance, latency, white balance, and AE step response.
Technological Characteristics
Portable all-in-one system (450x350x150 mm, 7kg). Features: LED light source, 15-inch display, CCU. Connectivity: SD card, network server. Electrical safety: Class I, Type BF (stroboscopy/camera), Type CF (light). Moisture protection: IPX1. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471.
Indications for Use
Indicated for visualization of endoscopic and microscopic procedures in patients requiring such interventions.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K193235 — Telepack + · Karl Storz Endoscopy America, Inc. · Dec 20, 2019
K071292 — VIEW MD OFFICE SYSTEM, MODEL 41066.5540 · Richard Wolf Medical Instruments Corp. · Jul 19, 2007
K152513 — 8X-10-XXXX Camera System · Santa Barbara Imaging Systems · Dec 15, 2015
K022490 — KSEA MEDI PACK · KARL STORZ Endoscopy-America, Inc. · Sep 29, 2003
K221174 — Telepack + · Karl Storz Endoscopy America, Inc. · May 13, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 20, 2018
KARL STORZ Endoscopy-America, Inc. Winkie Wong Manager, Regulatory Affairs 2151 E. Grand Avenue EI Segundo, CA 90245
Re: K182696 Trade/Device Name: Telepack X LED Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FET Dated: September 25, 2018 Received: September 27, 2018
Dear Winkie Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeffrey W. Cooper -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182696
Device Name Telepack X LED
#### Indications for Use (Describe)
The TELE PACK X LED is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-size: 12px;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 12px;"></span> |
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
K182696
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | KARL STORZ Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>El Segundo, CA 90245 |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Winkie Wong<br>Regulatory Affairs Manager<br>424-218-8379 (phone) |
| Date of Preparation: | September 24th, 2018 |
| Type of 510(k)<br>Submission: | Traditional |
| Device Identification: | Trade Name: Telepack X LED<br>Classification Name: Endoscopic Video Imaging<br>System/Component |
| Product Code: | FET |
| Regulation: | 21 CFR 876.1500 (Endoscope and Accessories) |
| Predicate Device(s): | Image 1 Video Imaging System (K070716) – Primary<br>KARL STORZ ICG Imaging System (K180146) – Reference<br>**The above predicate and reference device have not been subject to any recall** |
| Device Description: | The Telepack X LED is a portable and compact all-in-one imaging<br>system that includes a 15 inch screen display, a camera control<br>unit and internal LED light source, that is intended to be<br>connected to a compatible device (camera head or<br>videonendoscope) for the purpose of visualization and<br>documentation of endoscopic and microscopic procedures as well<br>as stroboscopy. |
| | The Telepack X LED includes a LED illumination light source to<br>illuminate the intended area and a 15 inch monitor for display. It<br>also allows the users to redefine the functions that take place<br>when a button is pressed. The Telepack X LED is a non-patient<br>contacting and require only wipe down as needed. |
| Intended Use and<br>Indications for use: | The TELE PACK X LED is an all-in-one Imaging System, which<br>comprises a light source for illumination, Camera Control Unit<br>(CCU) for use with compatible camera heads or video endoscopes<br>for image processing, as well as a monitor for image display,<br>intended for the visualization of endoscopic and microscopic<br>procedures. |
| Technological<br>Characteristics: | The Telepack X LED is a portable and compact all-in-one imaging<br>system that includes a 15 inch screen display, a camera control<br>unit and internal LED light source, that is intended to be<br>connected to a compatible device (camera head or<br>videonendoscope) for the purpose of visualization and<br>documentation of endoscopic and microscopic procedures as well<br>as stroboscopy.<br>The Telepack X LED includes, but not limited to, the following<br>features:<br>Image capture Flip or mirror image Zoom Brightness control Light source control White Balance The dimension of the subject device is 450 x 350 x 150 mm and<br>weighs 7kg. It is not intended to be soiled and is non-patient<br>contacting. It includes moderate level of concern software. The<br>device has been tested and passed the electrical safety and EMC<br>testing, which is certified to be Class I protection against<br>electrical shock, Type BF protection against electrical shock from<br>stroboscopy and camera applied parts, Type CF protection<br>against electrical shock from light and lastly drip-water<br>protection against moisture per IPX1. |
| Non-Clinical<br>Performance Data: | There are no performance standards or special controls<br>developed under Section 514 of the FD&C Act for endoscopes.<br>However, the KARL STORZ ICG Imaging System follows the FDA<br>recognized consensus standards and is tested according to the<br>following standards and FDA Guidance:<br>Electrical Safety and EMC IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18 IEC 52471 Software Verification and Validation Testing Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical<br>Device Level of concern: Moderate Performance Testing Minimum Illumination Spatial Resolution Color Performance Latency White Balance AE Step Response Head Button Functionality Additional bench testing was performed to ensure the device met<br>its design specifications. The bench testing performed verified<br>and validated that the Telepack X LED has met all its design<br>specification and is substantially equivalent to its predicate<br>devices. |
| Clinical<br>Performance Data: | Clinical published literatures were provided to support the<br>effectiveness of NIR imaging in the neuro- and endonasal skull<br>base surgeries as well as the use of the KARL STORZ ICG Imaging<br>System in pediatrics. |
| Substantial<br>Equivalence: | The intended use, operating principles, technological<br>characteristics and features are similar, if not identical, between<br>that subject device and the Image 1 Video Imaging System<br>(K070716). The minor difference between the subject and<br>predicate devices that does not raise new or different questions<br>or safety and effectiveness are: |
| | The subject device seeks clearance on the CCU, internal light source and internal display to be used with compatible camera heads and videoendoscopes, whereas the predicate device sought clearance on the CCU and camera heads with use of an external standard light source and external display as part of the submission. The subject device allows image, video and audio files to be stored via an external SD memory card or the hospital's/office's network server, whereas the predicate allows image, video and audio files to be stored via an external device. (Function not under review as this function is 510k exempt per product code, LMD, and does not have impact on the performance of the device) The subject device does not offer interoperability (the ability to control external devices), whereas the predicate does. The subject device does not offer split screen enhancement, whereas the predicate does. The subject device is slightly larger and heavier than the predicate device Due to the internal light source the subject device offer, the subject device is compliant to IEC 62471, whereas for the predicate, the compliance to IEC 62471 is achieved by the external light source. As proven by the comparisons, the above differences do not raise different questions of safety and effectiveness because the intended use, operating principles, technological characteristics, and features are similar, if not identical. Both systems also comply with identical standards and safety testing, where applicable.<br>Substantial equivalence on the effectiveness of the subject device is supported by the comparison of the images and standard image quality characteristics including, but not limited to, resolution, latency, white balance and AE step response between subject and predicate devices. |
| Clinical Performance<br>Data: | Clinical performance is not required to demonstrate substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish substantial equivalence. |
| Conclusion: | The Telepack X LED is substantially equivalent to its predicate device. The non-clinical bench and comparative testing demonstrate that the device is as safe and effective as the legally marketed devices. |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.