K123013 · Novadaq Technologies, Inc. · FED · Mar 22, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K123013
Device Name
ENDOSCOPIC INTRODUCER
Applicant
Novadaq Technologies, Inc.
Product Code
FED · Gastroenterology, Urology
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The Novadaq Disposable Rigid Scope Introducer is intended to facilitate passage of a laparoscope through the anus allowing visual examination of the rectum and distal portions of the colon.
Device Story
Single-use rigid scope introducer; used with laparoscope to facilitate access to rectum and distal colon. Device provides pathway for laparoscope and includes insufflation port for rectal/colonic insufflation. Maximum insertion length 25cm. Operated by clinician; used in clinical setting. Provides visual access for examination; insufflation enables visualization. Benefits patient by providing controlled, disposable access for diagnostic examination.
Clinical Evidence
Clinical evaluation based on exhaustive literature search and review of adverse events database. Data supports safety and performance claims. No prospective clinical trial data reported.
Technological Characteristics
Single-use rigid introducer; includes insufflation port. Materials verified per ISO 10993-1. Testing performed per ISO 8600-1 and ISO 8600-4. Dimensions similar to predicates with 25cm insertion length. Non-powered, mechanical device.
Indications for Use
Indicated for patients requiring visual examination of the rectum and distal colon via laparoscope; facilitates passage of the laparoscope through the anus.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K242668 — Gastro Concepts Air Assist · Gastro Concepts, LLC · Dec 13, 2024
K110792 — GELPOINT PATH · Applied Medical Resources Corp. · May 23, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Summary Date:
R123013
pg. 1 of 2
Endoscopic Introducer
## MAR 2 2 2013
### 510(k) Summary For the Novadag Disposable Riqid Scope Introducer
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
Applicant: Novadaq Technologies Inc. 13155 Delf Place, Unit 250 Richmond, British Columbia Canada V6V 2A2 Tel: (604) 232 9861
- Contact Name: Tim Verspagen Sr. Manager, Requlatory Compliance Tel: (604) 232-9861 Ext. 240 Fax: (604) 232-9641
March 14, 2013
- Trade Name: Disposable Rigid Scope Introducer
- Classification Name: Endoscopic Access Overtube
- Classification: 21 CFR § 876.1500
- Product Code: 78 (FED)
- Classification: Class II
- Predicate Devices: K092221 Colonic Splinting Overtube (US Endoscopy)
K033954 Shape Locking Endoscopic Overtube (USGI Medical Inc.)
K770291 Disposable Rigid Sigmoidoscope (Welch Allyn)
#### Device Description:
The single-use Disposable Rigid Scope Introducer is used in conjunction with a laparoscope to allow visual examination of the rectum and distal portions of the colon. When used with the laparoscope, the Disposable Rigid Scope Introducer and laparoscope are advanced into the rectum under visual guidance to a maximum insertion length of 25cm. An insufflation port is provided as part of the Endoscopic Introducer to allow insufflation of the rectum and lower colon.
{1}------------------------------------------------
K123013
Pg. 2 of 2
#### Indications for Use:
The Novadag Disposable Rigid Scope Introducer is intended to facilitate passage of a laparoscope through the anus allowing visual examination of the rectum and distal portions of the colon.
#### Predicate Device Technological Comparison Summary:
Considered in combination, the predicate devices are used to provide an introduction pathway to enable visual examination of the rectum and distal colon by also providing a channel for insufflation.
The predicate devices and the Disposable Rigid Scope Introducer are similar in shape and materials. The effect of use of the Disposable Rigid Scope Introducer and the predicate devices on the patient is the same. With respect to mechanical dimensions, the diameters of the devices are similar, while the length of the Disposable Rigid Scope Introducer is less than the predicate device as it is intended to be inserted a shorter distance.
Similar to one of the predicate devices, the Disposable Rigid Scope Introducer provides a channel to insufflate the rectum and distal colon which enables the visual examination.
Any additional indications for use claimed by the predicate device do not affect the safe and effective use of the Disposable Rigid Scope Introducer when used according to its instructions for use.
#### Nonclinical Tests:
The device safety and performance were verified by tests conducted by Novadaq Technologies and accredited third party laboratories. Standards used / relied upon for testing are:
- ISO 10993-1:2009 .
- ISO 8600-1 Second Ed. .
- . ISO 8600-4 First Ed.
#### Clinical Performance Data:
For clinical evaluation, an exhausive literature search was conducted according to written protocol. The clinical evaluation report is based on clinical data obtained from review of clinical literature and clinical experience from adverse events database.
Data presented in the clinical evaluation report provide evidence that supports the safety and performance claims made for the Endoscopic Introducer.
#### Safety & Effectiveness:
It has been shown in this 510(k) submission that any differences in intended use or technological characteristics between the Disposable Rigid Scope Introducer and the referenced predicate device do not raise any questions regarding its safety and effectiveness. The Endoscopic Introducer, as designed, manufactured and used according to its labeled intended use is determined to be substantially equivalent to the referenced predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three bars representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2013
Novadaq Technologies, Inc. % Mr. Tim Verspagen Sr. Manager Regulatory Compliance 13155 Delf Place, Unit 250 RICHMOND BRITISH COLUMBIA CANADA V6V 2A2
Re: K123013
Trade/Device Name: Disposable Rigid Scope Introducer Regulation Number: 21 CFR8 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: February 19, 2013 Received: February 20, 2013
Dear Mr. Verspagen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Herbert P. Verner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): -K123013
Device Name: Disposable Rigid Scope Introducer
#### Indications for Use:
The Novadaq Disposable Rigid Scope Introducer is intended to facilitate passage of a laparoscope through the anus allowing visual examination of the rectum and distal portions of the colon.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
HerbertLerner -S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123013 510(k) Number _
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.