SPYSCOPE ACCESS AND DELIVERY CATHETER

K051504 · Boston Scientific Corp · FED · Jun 16, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051504
Device NameSPYSCOPE ACCESS AND DELIVERY CATHETER
ApplicantBoston Scientific Corp
Product CodeFED · Gastroenterology, Urology
Decision DateJun 16, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The SpyScope Access and Delivery Catheter will be intended to guide both optical and accessory type devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts.

Device Story

The SpyScope Access and Delivery Catheter is a sterile, single-use device consisting of a flexible delivery catheter and a handle. It is used during endoscopic retrograde cholangiopancreatography (ERCP) procedures. The device is introduced through a duodenoscope (minimum 4.2mm working channel) to reach the biliary system. The distal tip features four-way articulation controlled by the handle. The device serves as a conduit to guide visualization probes (e.g., SpyGlass Direct Visualization Probe) and various accessory devices such as biopsy forceps, cytology brushes, and stone retrieval baskets. It is operated by a physician in a clinical setting. By providing a stable, steerable pathway, it facilitates targeted diagnostic and therapeutic interventions within the biliary tree.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Single-use, sterile catheter with four-way articulating distal tip, handle with control knobs, and a working channel. Mechanical operation; no energy source or software. Designed for use through a duodenoscope with a minimum 4.2mm working channel.

Indications for Use

Indicated for patients undergoing endoscopic procedures in the biliary system, including hepatic ducts, requiring guidance of optical or accessory devices for diagnostic or therapeutic applications.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K05150f Image /page/0/Picture/1 description: The image shows the words "Boston Scientific" stacked on top of each other. The font is a serif font, and the text is black. The word "Boston" is on top, and the word "Scientific" is below it. # JUN 1 6 2005 SECTION 10 510(K) SUMMARY #### 510(K) SUMMARY #### 1. Submitter: Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760 Owner/operator #9912058 Contact: Jennifer Johnson Sr. Regulatory Specialist Boston Scientific Marlborough 100 Boston Scientific Way Marlborough, MA 01752 Phone: 508-683-4178, Fax: 508-683-5939 Date Prepared: May 16, 2005 #### 2. Device: Trade Name: SpyScope Access and Delivery Catheter Common Name: Catheter Classification Name: Endoscopes and Accessories #### 3. Predicate Device: Olympus SwingTip Cannula, K011149 Olympus CHF Type BP30 Choledochofiberscope, K944473 #### 4. Device Description: The SpyScope Access and Delivery Catheter is a sterile, single-use device comprised of two main components: a flexible delivery catheter and a handle. The device is intended to be used to guide both the SpyGlass Direct Visualization Probe (K050403) or other visualization devices and accessory devices, (such as biopsy forceps, cytology brushes, stone retrieval baskets, etc.) during endoscopic retrograde cholangiopancreatography (ERCP) procedures. The SpyScope Access and Delivery Catheter is introduced to the desired anatomical location through a duodenoscope with a minimum working channel diameter of 4.2mm. The distal tip of the SpyScope Access and Delivery Catheter is designed to articulate in four directions. #### 5. Intended Use: The SpyScope Access and Delivery Catheter will be intended to guide both optical and accessory type devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Boston Scientific. The logo is in a serif font and is black. The word "Boston" is on top of the word "Scientific". #### 6. Technological Characteristics: Technological similarities between the SpyScope Access and Delivery Catheter and the Olympus SwingTip Cannula include the articulating tip and between the SpyScope Access and Delivery Catheter and the Olympus Choledochofiberscope includes the working channel, control knobs, and locking mechanism to hold tip in position. In instances where the technological characteristics may differ, it has been demonstrated that there are no new questions raised regarding safety and effectiveness of the SpyScope Access and Delivery Catheter. ### 7. Performance Data: Bench testing was conducted to evaluate the design features of the SpyScope Access and Delivery Catheter and establish specifications of the descriptive characteristics for the proposed device. These specifications were used to demonstrate substantial equivalence of the proposed device to the predicate devices. ### 8. Conclusion: BSC has demonstrated that the SpyScope Access and Delivery Catheter is substantially equivalent to Olympus's currently marketed SwingTip Cannula and Olympus's currently marketed CHF Type BP30 Choledochofiberscope. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo with a stylized bird figure inside a circle. The bird is depicted with three curved lines representing its body and wings. The text "university" is written along the left side of the circle, oriented vertically. The logo appears to be a simple, symbolic representation of a university or academic institution. r பு JUN 1 6 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MI) 20050 Boston Scientific Corporation c/o Mr. Neil E. Devine Sr. Staff Engineer Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719 Re: K051504 Trade/Device Name: SpyScope™ Access and Delivery Catheter Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: KOG Dated: June 2, 2005 Received: June 7, 2005 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Aniendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Hogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## SECTION 3 INDICATIONS FOR USE # Indications for Use | 510(k) Number (if known): | To be determined | |---------------------------|---------------------------------------| | Device Name: | SpyScope Access and Delivery Catheter | Indications For Use: The SpyScope Access and Delivery Catheter is intended to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices K051504 Page ! of Premarket Notification, SpyScope Access and Delivery Catheter, May 16, 2005 Proprietary and Confidential Information of Boston Scientific Corporation
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...