CELLVIZIO 100 SERIES SYSTEM WITH CONFOCAL MINIPROBES (TM)
Applicant
Mauna Kea Technologies
Product Code
OWN · Gastroenterology, Urology
Decision Date
Jul 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Cellvizio® 100 Series F700-v2 system with Confocal Miniprobes™ is a confocal laser system with fiber optic probes that is intended to allow imaging of the internal microstructure of tissues in anatomical tracts, i.e., gastrointestinal or respiratory, accessed by an endoscopic accessories.
Device Story
Cellvizio 100 Series F700-v2 is a confocal laser imaging system using fiber optic probes. Device inputs laser light at 785nm to illuminate internal tissue microstructures within gastrointestinal or respiratory tracts. System transforms reflected light into real-time confocal images via adapted optical filters. Used in clinical settings by physicians during endoscopic procedures. Output displayed to clinician to visualize tissue microstructure, aiding in clinical assessment. Benefits include real-time, in-vivo visualization of tissue morphology.
Clinical Evidence
Bench testing only. Evidence includes laser safety testing (IEC 60825-1:2007), 21 CFR Part 1040 compliance, spectral sensitivity evaluation, electrical safety testing (IEC 60601-1-1:2000, IEC 60601-1-2:2001, IEC 60601-1-4, IEC 60601-2-18), and software validation. No clinical data presented.
Technological Characteristics
Confocal laser imaging system; 785nm laser source; fiber optic probes; adapted optical filters. Electrical safety per IEC 60601-1 series; laser safety per IEC 60825-1. Materials and packaging identical to predicate. Software-controlled image acquisition.
Indications for Use
Indicated for imaging internal microstructure of gastrointestinal or respiratory tract tissues accessed via endoscope or endoscopic accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Cellvizio® 100 Series System with Confocal Miniprobes™ (K111047)
Cellvizio® (-GI, -Lung) with Confocal Miniprobe™ (K061666)