K073507 · Epitek, Inc. · ODC · Mar 5, 2008 · Gastroenterology, Urology
Device Facts
Record ID
K073507
Device Name
ANCHORAGE ACCESS KIT
Applicant
Epitek, Inc.
Product Code
ODC · Gastroenterology, Urology
Decision Date
Mar 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The Anchorage™ Access Kit facilitates percutaneous access in order to establish a guide through which surgical instruments or devices can be intubated.
Device Story
Anchorage Access Kit is a system of components designed to gain percutaneous access and provide a conduit for surgical instruments, specifically the Anchorage Closure Device or other compatible tools. Used by clinicians in a surgical setting to facilitate the placement of instruments through a guide. The device functions as an access tool to enable subsequent surgical procedures. It does not involve electronic processing or AI/ML algorithms.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of in-vitro testing, biocompatibility testing, and conformance to consensus and voluntary standards.
Technological Characteristics
Percutaneous access system components. Materials and specifications are compliant with consensus and voluntary standards. Non-electronic, mechanical device. No software or energy source.
Indications for Use
Indicated for patients requiring percutaneous access to establish a guide for the intubation of surgical instruments or devices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
USGI Medical Transport™ Endoscopic Access Device (K061216)
K251490 — 8 mm Assist Cannula; 12 mm Assist Cannula · Intuitive Surgical, Inc. · Jul 2, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
Epitek, Inc. Werner Hampl President, RA/OA 4801 West 81st Street, Suite 105 Bloomington, MN 55437
JUL 2 7 2015
Re: K073507 Trade/Device Name: Anchorage™ Access Kit Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC Dated (Date on orig SE ltr): December 10, 2007 Received (Date on orig SE ltr): December 14, 2007
Dear Werner Hampl,
This letter corrects our substantially equivalent letter of March 5, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): _ K073507
Device Name: Anchorage™ Access Kit
Indications for Use:
The Anchorage™ Access Kit facilitates percutaneous access in order to establish a guide through which surgical instruments or devices can be intubated.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
T
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nuthp. Olen for mxm Page 1 of 1
(Division Sign-C Division of General, Restorative, and Neurological Devices
**510(k) Number** K073507
{3}------------------------------------------------
K073507
### - 5 2008
# 510(k) Summary Epitek Anchorage™ Access Kit
#### APPLICANT:
Epitek, Inc. 4801 W. 81st St., Suite 105 Bloomington, MN 55437 USA
| Contact Person: | Werner Hampl |
|-----------------|----------------------|
| Telephone: | (952) 230-9886 |
| e-mail: | whampl@epitekinc.com |
| Date Prepared: | December 10, 2007 |
#### DEVICE:
Proprietary Name: Classification:
Anchorage Access Kit Class II; 21 CFR 876.1500; Endoscope and accessories; Product Code: KOG
#### PREDICATE DEVICE:
The subject device is substantially equivalent (i.e., has the same intended use and technological characteristics) to the USGI Medical Transport™ Endoscopic Access Device (K061216).
#### DEVICE DESCRIPTION:
The Anchorage Access Kit is a system of components used to gain percutaneous access and provide a conduit for the Anchorage Closure Device (K073096) or other surgical instruments.
#### INDICATIONS FOR USE:
The Anchorage Access Kit facilitates percutaneous access in order to establish a guide through which surgical instruments or devices can be intubated.
{4}------------------------------------------------
## TEST RESULTS:
Results of the in-vitro testing, biocompatibility testing, and conformance to consensus and voluntary standards conducted prior to introduction into commerce demonstrate that the Anchorage Access Kit is substantially equivalent to the specified predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.