← Product Code [OCX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX) · K111274 # JETPREP FLUSHING DEVICE OL-103037 / OL-133037 / OL-168037 / PE-105038 / PE-150038 / PE-170038 / FU-169038 (K111274) _Jetprep, Ltd. · OCX · Apr 4, 2012 · Gastroenterology, Urology · SESE_ **Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K111274 ## Device Facts - **Applicant:** Jetprep, Ltd. - **Product Code:** [OCX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX.md) - **Decision Date:** Apr 4, 2012 - **Decision:** SESE - **Submission Type:** Traditional - **Regulation:** 21 CFR 876.1500 - **Device Class:** Class 2 - **Review Panel:** Gastroenterology, Urology ## Intended Use The JetPrep Flushing Device is intended for use as a flexible endoscopic accessory to apply legally marketed solutions for washing mucosal tissue in the gastrointestinal tract. ## Device Story The JetPrep Flushing Device is a sterile, single-use, disposable flexible endoscopic accessory. It consists of a catheter with a spray tip at the distal end. The device is inserted into the working channel of an endoscope. The user manually positions the spray tip at the distal end of the endoscope to deliver a 360-degree irrigation spray pattern. The device is connected to external, legally marketed irrigation pumps or manual syringes to supply irrigation fluids. It is designed to allow for the simultaneous aspiration of debris and fluids through the endoscope working channel while the device is in place. The device is used by clinicians during gastrointestinal endoscopic procedures to wash mucosal tissue, potentially improving visualization and cleaning of the site. ## Clinical Evidence Bench testing only. The device underwent safety and performance validations, including biocompatibility testing for all patient-contacting and fluid-contacting components. ## Technological Characteristics Catheter-based device with a distal spray tip. Provides 360-degree irrigation spray pattern. Compatible with standard endoscope working channels. Does not obstruct aspiration. Connects to external irrigation pumps or manual syringes. Biocompatible materials used for patient and fluid-contacting parts. Sterile, single-use, disposable. ## Regulatory Identification An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations. ## Special Controls *Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9. ## Predicate Devices - Hobbs Mistifier Spray Catheter ([K030765](/device/K030765.md)) ## Related Devices - [K122740](/device/K122740.md) — JETPREP FLUSHING DEVICE OL-103037, JETPREP FLUSHING DEVICE OL-133037, JETPREP FLUSHING DEVICE OL-168037, JETPREP FLUSHIN · Jetprep, Ltd. · Dec 10, 2012 - [K101094](/device/K101094.md) — CLEARJET · Easyglide , Ltd. · May 17, 2010 - [K113166](/device/K113166.md) — CLEARPATH UPPER GI · Easyglide , Ltd. · May 16, 2012 - [K182319](/device/K182319.md) — Glo-Tip Spray Catheter · Wilson-Cook Medical, Inc. · Feb 1, 2019 - [K251986](/device/K251986.md) — AUXILIARY WATER TUBE MAJ-855 · Olympus Medical Systems Corporation · Sep 26, 2025 ## Submission Summary (Full Text) {0}------------------------------------------------ # 510(k) SUMMARY APR - 4 2012 ۽ : JetPrep, Ltd. JetPrep Flushing Device K111274 | Manufacturer: | JetPrep, Ltd.<br>71 Ha'Nadiv St.<br>Herzliya 46485<br>lsrael<br>Phone: +972-9-950-6712<br>Fax: +972-9-950-6710<br>Contact: David Nitsan, CEO | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Summary Prepared: | June 6, 2011 | | Representative/Consultant: | John Smith, MD, JD<br>Hogan Lovells US LLP<br>Columbia Square<br>555 Thirteenth Street, NW<br>Washington, DC 20004<br>Phone: +1-202-637-5600<br>Fax: +1-202-637-5910 | | Device: | Trade Name: JetPrep Flushing Device<br>Common/Classification Name: Endoscope and accessories<br>Product Code: KOG<br>Classification Regulation: 21 C.F.R. § 876.1500 | | Predicate Devices: | Hobbs Mistifier Spray Catheter<br>Hobbs Medical, Inc.<br>K030765 | | Device Description: | The JetPrep Flushing Device is a sterile, disposable, single<br>use device, intended for use as flexible endoscopic<br>accessory to apply legally marketed solutions for washing<br>mucosal tissue in the gastrointestinal tract. | {1}------------------------------------------------ Intended Use: The JetPrep Flushing Device is intended for use as a flexible endoscopic accessory to apply leqally marketed solutions for washing mucosal tissue in the gastrointestinal tract. Technological Characteristics: The device is composed of a catheter with a spray tip on its distal tip. During operation, the catheter should be inserted into the endoscope working channel. The spray tip location can be manually controlled by the user to be positioned on the distal end of the endoscope working channel, and thus apply 360 degree irrigation spray pattern. The device does not impede aspiration of debris and fluids through the endoscope working channel while it remains within the endoscope. For providing irrigation fluids, the device should be connected to legally marketed irrigation pumps or manual syringe for endoscopy. Performance Data: The JetPrep Flushing Device has been subjected to extensive safety and performance validations prior to release. The device parts that come in contact with the irrigation fluids and/or the patient's tissue are composed of materials that were tested for biocompatibility. JetPrep, Ltd., demonstrated that, for the purposes of FDA's Substantial Equivalence: requlation of medical devices, JetPrep Flushing Device is substantially equivalent in indications for use, design and operation principles to a legally marketed predicate device. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter of the circle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JetPrep, Ltd. % Mr. John J. Smith, MD, JD Partner Hogan Lovells US LLP Columbia Square 555 13th Street, NW WASHINGTON DC 20004 Re: K111274 Trade/Device Name: JetPrep Flushing Device Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCX Dated: March 26, 2012 Received: March 26, 2012 Dear Mr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r four of a how a determination that your device complies with other requirements of the Act that I Drinan statutes and regulations administered by other Federal agencies. You must or urf I cather the Act's requirements, including, but not limited to: registration and listing Configity with an the 110 b ong (21 CFR Part 801); medical device reporting (reporting of medical APR - 4 2012 {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerély vours, Sincerely yours, Benjamin R. Twichell Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K111274 Device Name: JetPrep Flushing Device Indications for Use: The JetPrep Flushing Device is intended for use as a flexible endoscopic accessory to apply legally marketed solutions for washing mucosal tissue in the gastrointestinal tract. Prescription Use _ X AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Virological Devices --- **Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K111274](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K111274) **Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact). **Cite as:** FDA Device Explorer (Innolitics), https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K111274
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...