The HD-500 Video Endoscope System, which includes a video gastroscope /video colonoscope, image processor, light source, monitor, accessories and other peripheral devices, is intended for endoscopies examination, diagnosis and treatment of the disease of the upper and lower gastrointestinal tract.
Device Story
HD-500 Video Endoscope System is a flexible endoscopic imaging platform for GI tract visualization and intervention. System comprises video gastroscopes (EG-500 series), video colonoscopes (EC-500 series), image processor (HD-500 series), and light source (HDL-500 series). Endoscopes capture visual data via distal CMOS sensors; image processor performs signal processing, enhancement (edge, structure, contrast, color), and output to external monitors; light source provides illumination (LED or Xenon). Used in clinical settings by physicians for observation, diagnosis, and treatment. VIST observation mode uses optical filtering to enhance mucosal vessel visibility, analogous to NBI technology. Output includes real-time video, still images, and video recording. System aids clinical decision-making by providing high-definition visualization of GI mucosa, facilitating biopsy, and therapeutic procedures.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), endoscopic equipment safety (IEC 60601-2-18), optical performance (resolution, depth of field, magnification, distortion, intensity uniformity), and physical/functional performance (angulation, sealing, body model testing). Biocompatibility verified per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
System includes flexible video endoscopes (CMOS sensor), image processor, and light source (LED or Xenon). Materials meet ISO 10993-5/10 standards. Connectivity includes DVI, VGA, SDI, Y/C, and VBS video outputs. Features include automatic white balance, automatic gain control, and multiple image enhancement modes (edge, structure, contrast, color). System provides automatic and manual brightness control and automatic exposure (19 steps).
Indications for Use
Indicated for endoscopic examination, diagnosis, and treatment of upper GI tract (esophagus, stomach, duodenum) and lower GI tract (anus, rectum, colon, ileocecal segment) in patients requiring endoscopic intervention.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 5, 2018
Sonoscape Medical Corp. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K173921
> Trade/Device Name: HD-500 Video Endoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NWB, FDF, FDS Dated: July 13, 2018 Received: July 24, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173921
Device Name HD-500 Video Endoscope System
#### Indications for Use (Describe)
HD-500 Video Endoscope System
The HD-500 Video Endoscope System, which includes a video colonoscope, image processor, light source, monitor, accessories and other peripheral devices, is intended for endoscopies examination, diagnosis and treatment of the disease of the upper and lower gastrointestinal tract.
HDL-500 Series Light Source, HDL-500E, HDL-500X
The HDL-500 Series Light Source has been designed to be used with the endoscope, image processor and other peripheral devices for endoscopic observation, diagnosis and treatment.
HD-500 Series Image Processor, HD-500, HD-500S, HD-330Plus
The HD-500 Series Image Processor has been designed to be used with the endoscope, light source, monitor and other peripheral devices for endoscopic observation, diagnosis, treatment, and video recording.
EG-500 Series Video Gastroscope, EG-500, EG-500L
The EG-500 Series Video Gastroscope has been designed to be used with the image processor, light source, monitor and other peripheral devices for endoscopic observation, diagnosis and treatment of the upper digestive tract (including the esophagus, stomach and duodenum).
EC-500 Series Video Colonoscope, EC-500, EC-500T, EC-500L, EC-500L/T
The EC-500 Series Video Colonoscope has been designed to be used with the image processor, light source, monitor and other peripheral devices for endoscopic observation, diagnosis and treatment of the lower digestive tract (including the anus, rectum, colon and ileocecal segment).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K173921
- 1. Date of Preparation: 11/30/2017
- 2. Sponsor Identification
### SONOSCAPE MEDICAL CORP.
4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen, 518051, Guangdong, China
Establishment Registration Number: 3004705634
Contact Person: Toki Wu Position: Regulatory Affairs Manager Tel: 0755-26722890 Fax: 0755-26722850 Email: ra@sonoscape.net
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 001-3609253199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: HD-500 Video Endoscope System
Common Name: Endoscopic Video Imaging System
Components and Component Models:
| EG-500 Series Video Gastroscope: | EG-500, EG-500L |
|----------------------------------|-------------------------------------|
| EC-500 Series Video Colonoscope: | EC-500, EC-500T, EC-500L, EC-500L/T |
| HD-500 Series Image Processor: | HD-500, HD-500S, HD-330Plus |
| HDL-500 Series Light Source: | HDL-500E, HDL-500X |
Regulatory Information
Classification Name: Endoscope and accessories Classification: II Product Code: NWB, FDF and FDS Regulation Number: 21 CFR 876.1500 Endoscope and accessories Review Panel: Gastroenterology/Urology
Intended Use Statement:
HD-500 Video Endoscope System
The HD-500 Video Endoscope System, which includes a video gastroscope /video colonoscope, image processor, light source, monitor, accessories and other peripheral devices, is intended for endoscopies examination, diagnosis and treatment of the disease of the upper and lower gastrointestinal tract.
#### EG-500 Series Video Gastroscope, EG-500, EG-500L
The EG-500 Series Video Gastroscope has been designed to be used with the image processor, light source, monitor and other peripheral devices for endoscopic observation, diagnosis and treatment of the upper digestive tract (including the esophagus, stomach and duodenum).
#### EC-500 Series Video Colonoscope, EC-500, EC-500T, EC-500L, EC-500L/T
The EC-500 Series Video Colonoscope has been designed to be used with the image processor, light source, monitor and other peripheral devices for endoscopic observation, diagnosis and treatment of the lower digestive tract (including the anus, rectum, colon and ileocecal segment).
#### HD-500 Series Image Processor, HD-500, HD-500S, HD-330Plus
The HD-500 Series Image Processor has been designed to be used with the endoscope, light source, monitor and other peripheral devices for endoscopic observation, diagnosis, treatment, and video recording.
#### HDL-500 Series Light Source, HDL-500E, HDL-500X
The HDL-500 Series Light Source has been designed to be used with the endoscope, image processor and other peripheral devices for endoscopic observation, diagnosis and treatment.
{5}------------------------------------------------
#### Device Description
The proposed device, HD-500 Video Endoscope System, which includes a video gastroscope /video colonoscope, image processor, light source, monitor, accessories and other peripheral devices.
HD-500 Video Endoscope System can be offered in several configurations with the options of different models of primary components
The EG-500 Series Video Gastroscope/ EC-500 Series Video Colonoscope is the hand-held, direct-viewing flexible endoscope used for endoscopy and endoscopic surgery within the upper and lower gastrointestinal tract.
The HD-500 Series Image Processor is a video processing system which is designed to be used with endoscopes, light source, monitor of the proposed system. Apart from the image processing functions, it also provides power supply for the endoscopes.
The HDL-500 Series Light Source provides illumination for endoscopic diagnosis, treatment and video observation.
- Identification of Predicate Device న.
510(k) Number: K100584 Product Name: EVIS EXERA II 180 System Primary components and Component Models:
| EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE | OLYMPUS GIF TYPE Q180,<br>OLYMPUS GIF TYPE H180 |
|-------------------------------------------|--------------------------------------------------|
| EVIS EXERA II COLONOVIDEOSCOPE | OLYMPUS CF TYPE Q180AL<br>OLYMPUS CF TYPE Q180AI |
| EVIS EXERA II VIDEO SYSTEM CENTER | OLYMPUS CV-180 |
| EVIS EXERA II XENON LIGHT SOURCE | OLYMPUS CLV-180 |
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-Part 1: General requirements for basic safety and essential performance, including the US National Differences
- > IEC 60601-1-2:2007, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
{6}------------------------------------------------
- > IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- > ISO 8600-1:2015 Endoscopes--Medical endoscopes and endotherapy devices - part 1: General requirements
- > ISO 8600-7:2012 Endoscopes --Medical endoscopes and endotherapy devices - part 7: Basic requirements for medical endoscopes of water-resistant type
- > ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
#### Optical performance testing
The following testing was conducted to evaluate the optical performance characteristics for each endoscopic system mode.
- A Photobiological safety
- A Resolution (test results comparison between the proposed device and predicate device)
- A Depth of field (test results comparison between the proposed device and predicate device)
- > Optical magnification and distortion (test results comparison between the proposed device and predicate device)
- > Image intensity uniformity (test results comparison between the proposed device and predicate device)
#### Physical/functional performance testing
The endoscope performance test was conducted to evaluate the ability of proposed endoscope to maintain the maximum angulation/deflection when in use, the physical/functional performance of the proposed device, including a) Appearance visual inspection and handle strength inspection, b) Image function visual inspection, c) Sealing performance, and d) Maximum bending angle measurement and body model testing.
#### Imaging performance testing
The imaging performance testing was conducted to demonstrate that the imaging quality of the proposed device is still in a better condition when the device is over its lifetime of clinical use. The degradations of imaging performance are very little which will not affect the normal use of the endoscope.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
#### Substantially Equivalent (SE) Comparison 8.
The whole system and components of the proposed device is basically identical to its predicate device in indication for use, and similar in specification. Comparisons between the proposed device and predicate device are shown in Table 1 to Table 6.
| Table 1 General Comparison |
|------------------------------------------------|
| Proposed device: HD-500 Video Endoscope System |
| ITEM | Proposed Device | Predicate Device |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | NWB, FDF and FDS | K100584 |
| Regulation<br>Number | 21 CFR 876.1500 | 21 CFR 876.1500 |
| Class | II | II |
| Indications<br>for Use<br>(SE Analysis 1) | HD-500 Video Endoscope System<br>The HD-500 Video Endoscope System, which<br>includes a video gastroscope /video<br>colonoscope, image processor, light source,<br>monitor, accessories and other peripheral /<br>devices, is intended for endoscopies<br>examination, diagnosis and treatment of the<br>disease of the upper and lower gastrointestinal<br>tract. | EVIS EXERA II XENON LIGHT<br>SOURCE OLYMPUS CLV-180<br>This light source has been designed<br>to be used with Olympus<br>endoscopes, video system center, and<br>other ancillary equipment for<br>endoscopic diagnosis, treatment and<br>video observation. |
| | HDL-500 Series Light Source, HDL-500E,<br>HDL-500X<br>The HDL-500 Series Light Source has been<br>designed to be used with the endoscope,<br>image processor and other peripheral devices<br>for endoscopic observation, diagnosis and<br>treatment. | |
| | HD-500 Series Image Processor, HD-500,<br>HD-500S, HD-330Plus<br>The HD-500 Series Image Processor has been<br>designed to be used with the endoscope, light<br>source, monitor and other peripheral devices<br>for endoscopic observation, diagnosis,<br>treatment, and video recording. | EVIS EXERA II VIDEO SYSTEM<br>CENTER OLYMPUS CV-180<br>This video system center has been<br>designed to be used with OLYMPUS<br>camera heads, endoscopes, light<br>sources, monitors, endo-therapy<br>accessories and other ancillary<br>equipment for endoscopic diagnosis,<br>treatment and video observation. |
| | EG-500 Series Video Gastroscope, EG-500,<br>EG-500I | EVIS EXERA II<br>GASTROINTESTINAL |
| Configuration<br>(primary<br>components) | The EG-500 Series Video Gastroscope has<br>been designed to be used with the image<br>processor, light source, monitor and other<br>peripheral devices for endoscopic<br>observation, diagnosis and treatment of the<br>upper digestive tract (including the<br>esophagus, stomach and duodenum). | VIDEOSCOPE OLYMPUS GIF<br>TYPE Q180, OLYMPUS GIF TYPE<br>H180<br>These instruments have been<br>designed to be used with an Olympus<br>video system center, light source,<br>documentation equipment, monitor,<br>EndoTherapy accessories (such as a<br>biopsy) and other ancillary<br>equipment for endoscopy and<br>endoscopic surgery within the upper<br>digestive tract (including the<br>esophagus, stomach, and duodenum). |
| | EC-500 Series Video Colonoscope, EC-500,<br>EC-500T, EC-500L, EC-500L/T | EVIS EXERA II<br>COLONOVIDEOSCOPE<br>OLYMPUS CF TYPE Q180AL.<br>OLYMPUS CF TYPE Q180AI. |
| | The EC-500 Series Video Colonoscope has<br>been designed to be used with the image<br>processor, light source, monitor and other<br>peripheral devices for endoscopic<br>observation, diagnosis and treatment of the<br>lower digestive tract (including the anus,<br>rectum, colon and ileocecal segment). | These instruments have been<br>designed to be used with an Olympus<br>video system center, light source,<br>documentation equipment, monitor,<br>EndoTherapy accessories (such as a<br>biopsy forceps), and other ancillary<br>equipment for endoscopy and<br>endoscopic surgery within the lower<br>digestive tract (including the anus,<br>rectum, sigmoid colon, colon, and<br>ileocecal valve). |
| | Light Source | Light Source |
| | Image processor | Image processor |
| | Video Gastroscope | Video Gastroscope |
| | Video Colonoscope | Video Colonoscope |
| | Accessories and peripheral devices | Accessories and peripheral devices |
### Predicate device: EVIS EXERA II 180 SYSTEM
{8}------------------------------------------------
SE Analysis 1-Indications for Use
The indications for use of proposed device is summarized for the whole video endoscope system and the primary components, the indications for use of predicate device is only summarized for the primary components. It is only the different expressions on indications for use, actually the proposed device and the predicate device have the same primary components, each of the primary component has the same intended use. Therefore, this item is considered as substantial equivalence.
{9}------------------------------------------------
### Table 2 Specifications Comparison of Image Processor Proposed device: HD-500 Series Image Processor (HD-500, HD-500S, HD-330Plus)
Predicate device: EVIS EXERA II VIDEO SYSTEM CENTER (OLYMPUS CV-180)
| ITEM | | Proposed Device | Predicate Device | | |
|-------------------------------------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | | HD-330Plus | HD-500 | HD-500S | OLYMPUS CV-180 |
| Power supply | | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz |
| Over-current protection | | Fuse type | Fuse type | Fuse type | Fuse type |
| Input current<br>(SE Analysis 2) | | 160VA | 160VA | 160VA | 150VA |
| Size<br>(SE Analysis 3) | | 370(W)×124(H)×455(D)mm | 370(W)×124(H)×455(D)mm | 370(W)×124(H)×455(D)mm | 383(W)×162(H)×536(D)mm |
| Weight<br>(SE Analysis 3) | | 9.5 Kg | 9.5 Kg | 9.5 Kg | 15.4kg |
| compatible endoscope | | Videoscope | Videoscope | Videoscope | Fiber/rigid scope with camera head<br>Videoscope |
| | Video signal<br>output<br>(SE Analysis 4) | DVI:1<br>VGA:1<br>Y/C:1<br>VBS:1 | DVI:1<br>VGA:1<br>SDI:1<br>Y/C:1<br>VBS:1 | DVI:1<br>VGA:1<br>SDI:1<br>Y/C:1<br>VBS:1 | RGB:3<br>Y/C:4<br>VBS:4<br>HDTV:1 |
| | Observation | Auto<br>white<br>balance | Automatically adjusted using<br>the white balance switch. At the<br>time of connection with the<br>scope in which the scope ID is<br>provide, compensation is<br>performed automatically | Automatically adjusted using<br>the white balance switch. At the<br>time of connection with the<br>scope in which the scope ID is<br>provide, compensation is<br>performed automatically | Automatically adjusted using<br>the white balance switch. At the<br>time of connection with the<br>scope in which the scope ID is<br>provide, compensation is<br>performed automatically |
| | Standard color<br>char output | Color bar image | Color bar image | Color bar image | Color bar image |
| | Color<br>tone<br>adjustment | R: ±8 steps B:±8 steps<br>Chroma:±8 steps | R: ±8 steps<br>B:±8 steps<br>Chroma:±8 steps | R: ±8 steps<br>B:±8 steps<br>Chroma:±8 steps | R:±8 steps<br>B:±8 steps<br>Chroma:±8 steps |
| | Automatic<br>gain control | Provided | Provided | Provided | Provided |
| | Image<br>enhancement<br>(SE Analysis 5) | Edge enhancement: [1][2][3] 3-mode selectable and each mode have 9 levels adjustable by the user<br>Structure enhancement: [1][2][3] 3-mode selectable and each mode have 9 levels adjustable by the<br>user<br>Contrast enhancement: [low][med][high] 3 levels available<br>Color enhancement: [1][2][3] 3-mode selectable and each mode have 9 levels adjustable by the user | Edge enhancement:<br>[off][low][med][high] 4 levels<br>available<br>structure enhancement:<br>[off][low][med][high] 4 levels<br>available | | |
| IRIS mode<br>selection<br>(SE Analysis 6) | | Peak/AVE/Auto<br>photometry mode | Peak/AVE/Auto<br>photometry mode | Peak/AVE/Auto<br>photometry mode | Peak/Auto photometry mode |
| | Zoom<br>(SE Analysis 7) | ×1.4 /×1.6/×1.8 | ×1.4 /×1.6/×1.8 | ×1.4 /×1.6/×1.8 | ×1/×1.2/×1.5 3-mode |
| | Imaging<br>modes<br>(SE Analysis 8) | Common observation mode;<br>VIST observation mode | Common observation mode;<br>VIST observation mode | Common observation mode;<br>VIST observation mode | White-light imaging mode<br>Narrow Band Imaging (NBI) modes |
| Foot switch connector | | Provided | Provided | Provided | Provided |
| record to memory card | | Provided | Provided | Provided | Provided |
{10}------------------------------------------------
#### SE Analysis 2-Input Current
The input current of proposed devices is similar as that of the input current proposed device comples with IEC 60601-1 standard. Therefore, this difference on input current is considered not effectiveness of the proposed device.
{11}------------------------------------------------
#### SE Analysis 3-Size and Weight
The physical specifications are different between the proposed device; the difference on physical specification will not affect the safety and effectiveness of the proposed device.
#### SE Analysis 4-Video Signal Output
The types of video signal output are different between the difference on video signal output will not affect the safety and effectiveness of the proposed device based on the following reasons:
- a. Both the proposed system and predicate system are the high-definition endoscope system; they all have the standard -definition interface and high-definition interface. The proposed system has of the predicate system, these interfaces improve the system's versatility.
- Standard -definition interface: Y/C interface and VBC interface belong to the standard -definition interface. b.
- High-definition interface DVI interface is substantially equivalent to the HDTV interface is substantially equivalent to the RGB C. interface.
- d. SDI interface belongs to the direct output interface. The display of the monitor can be switched to SDI signal output in case of the image processor suddenly doesn't work or the system suddenly crashes.
#### SE Analysis 5- Image enhancement
The proposed devices provide more selection about the image enhancement; it will offer the physician to adjust the observation. This difference will not affect the safety and effectiveness of the proposed device.
#### SE Analysis 6- IRIS mode selection
The proposed devices provide more selection about the IRIS mode; it will offer the physician more selection. This difference will not affect the safety and effectiveness of the proposed device.
#### SE Analysis 7- Zoom
The zoom of image processor is similar as that of the processor. This difference will not affect the safety and effectiveness of the proposed device.
{12}------------------------------------------------
#### SE Analysis 8- Imaging modes
The predicate device has both white-light inaging (NB), modes. The NBI mode selectively processes green and blue light images by filtering the white light spectum with an optical filter, and thus enhances visualization of superficial vacce also has two modes, the common observation modes. In the VIST observation mode, a dedicated light filter is insetted into the light path to filter out the light of a specific color, theress. The common observation mode of proposed device is same as the white-light imaging mode of the predicate device) and NBI technology (proposed device) and NBI technology (predicate device) are special light imaging technologies taking advantage of the extinction spectra between mucosal vessels and mucosa. Both of them utilize optical filtering and digital image processing to enhance the visibility of mucosal vesses. Although there are digital filtering parameters, leading to slight difference in the corresponding images, they are not the essential differences. VIST can still achieve the same clinical application purposes as NBI does. Therefore, the VIST observation mode of proposed device is substantially equivalent to the NBI imaging mode of the predicate device.
{13}------------------------------------------------
#### Table 3 Specifications Comparison of Video Gastroscope
Proposed device: EG-500 Series Video Gastroscope
| Predicate device: EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS | |
|---------------------------------------------------------------------|--|
|---------------------------------------------------------------------|--|
| ITEM | Proposed device | | Predicate device | |
|------------------------------------------------------|------------------------------------------------|------------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| Model | EG-500 | EG-500L | GIF-Q180 | GIF-H180 |
| Field of view | 140° | 140° | 140° | 140° |
| Depth of focus | 3-100mm | 3-100mm | 3-100mm | 2-100mm |
| Front view | 0° | 0° | 0° | \ |
| Sensor type<br>(SE Analysis 9) | color CMOS | color CMOS | color CCD | color CCD |
| Distal end outer diameter<br>(SE Analysis 10) | 9.3mm | 9.8mm | 8.8mm | 9.8mm |
| Insertion section outer diameter<br>(SE Analysis 10) | 9.3mm | 9.8mm | 8.8mm | 9.8mm |
| Bend angle | UP:210°<br>DOWN:90°<br>RIGHT:100°<br>LEFT:100° | UP:210°<br>DOWN:90°<br>RIGHT:100°<br>LEFT:100° | UP:210°<br>DOWN:90°<br>RIGHT:100°<br>LEFT: 100° | UP:210°<br>DOWN:90°<br>RIGHT:100°<br>LEFT: 100° |
| Insertion section length<br>(SE Analysis 11) | 1050mm | 1050mm | 1030mm | 1030mm |
| Biopsy channel inner diameter<br>(SE Analysis 10) | 2.8mm | 3.2mm | 2.8mm | 2.8mm |
SE Analysis 9- Sensor type
Although the CMOS sensor and CCD sensor have logic, but both they can provide high quality image. And the CMOS sensor has been commonly used in the endoscope systems. This difference on sensor type does not affect the safety and effectiveness of the proposed device.
{14}------------------------------------------------
#### SE Analysis 10- Diameters
The Distal end outer diameter, Insertion outer diameter and Biopsy chamel inner diameter of proposed device are similar as those of the predicate device. Those diameters meet the industrial standard. Therefore, this item is considered as substantial equivalence.
#### SE Analysis 11- Insertion section length
The insertion length of proposed device is similar as that of the predicate device. Therefore this item is considered as substantial equivalence.
{15}------------------------------------------------
## Table 4 Specifications Comparison of Video Colonoscope
Proposed device: EC-500 Series Video Colonoscope
Predicate device: EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE
| ITEM | Proposed device | | | | Predicate device |
|------------------------------------------------------|--------------------------------------------------|--------------------------------------------------|--------------------------------------------------|--------------------------------------------------|--------------------------------------------------|
| Model | EC-500 | EC-500T | EC-500L | EC-500L/T | CF-Q180AL/ CF-Q180AI |
| Field of view (SE Analysis 12) | 140° | 140° | 140° | 140° | 170° |
| Depth of focus | 3-100mm | 3-100mm | 3-100mm | 3-100mm | 3-100mm |
| Front view | 0° | 0° | 0° | 0° | 0° |
| Sensor type (SE Analysis 9) | color CMOS | color CMOS | color CMOS | color CMOS | color CCD |
| Distal end outer diameter (SE Analysis 10) | 12mm | 12mm | 12.9mm | 12.9mm | 13.2mm |
| Insertion section outer diameter<br>(SE Analysis 10) | 12.5mm | 12.5mm | 12.9mm | 12.9mm | 12.8mm |
| | | | | | |
| Bend angle | UP:180°<br>DOWN:180°<br>RIGHT:160°<br>LEFT: 160° | UP:180°<br>DOWN:180°<br>RIGHT:160°<br>LEFT: 160° | UP:180°<br>DOWN:180°<br>RIGHT:160°<br>LEFT: 160° | UP:180°<br>DOWN:180°<br>RIGHT:160°<br>LEFT: 160° | UP:180°<br>DOWN:180°<br>RIGHT:160°<br>LEFT: 160° |
| Insertion section length (SE Analysis 11) | 1350mm | 1700mm | 1350mm | 1700mm | 1680mm |
| | | | | | |
SE Analysis 12- Field of view
Although the field of view of the proposed video colonoscope is different than that of the predicate video colonoscope, the field of view of the proposed video colonoscope meets the requirements of ISO 8600-1 standard. Therefore, this difference on field of view does not affect the safety and effectiveness of the proposed device.
{16}------------------------------------------------
### Table 5 Specifications Comparison of Light Source Proposed device: HDL-500E Light Source/ HDL-500X Light Source Predicate device: EVIS EXERA II XENON LIGHT SOURCE (OLYMPUS CLV-180)
| ITEM | HDL-500E Light Source | HDL-500X Light Source | Predicate device |
|----------------------------------------------|---------------------------------------------|---------------------------------------------|----------------------------------------------|
| Power supply (SE Analysis 13) | AC 100-240V<br>50Hz/60Hz | AC 100-240V<br>50Hz/60Hz | AC 100-120V,<br>50/60HZ ±1Hz |
| Over-current protection | Fuse type | Fuse type | Fuse type |
| Input current (SE Analysis 14) | 160VA | 500VA | 500VA |
| Examination lamp (SE Analysis 15) | 50W LED | 300W xenon lamp | 300W Xenon lamp |
| Average lamp life (SE Analysis 16) | 50000 hours | 500 hours | Approximately 500 hours of<br>continuous use |
| Emergency lamp (SE Analysis 15) | 14W LED | 12V 60W halogen lamp | 12V 35W Halogen lamp |
| Average emergency lamp life (SE Analysis 17) | 50000 hours | 4000 hours | Approximately 500 hours |
| Brightness control | Automatic and manual | Automatic and manual | Automatic and manual |
| Automatic exposure (SE Analysis 18) | 19 steps | 19 steps | 17 steps |
| Air feeding | 4 levels available (off, low,<br>mid, high) | 4 levels available (off, low,<br>mid, high) | 4 levels available (off, low,<br>mid, high) |
| System connector | Provided | Provided | Provided |
| Foot switch connector | Provided | Provided | Provided |
| CV connector | Provided | Provided | Provided |
#### SE Analysis 13-Power supply
Although the power supply of proposed light source is different from that of the electrical safety testing of proposed device demonstrates that the power supply of the proposed device meets the requirements of IEC 60601-1 Standard. Therefore, on power supply does not affect the safety and effectiveness of the proposed device.
#### SE Analysis 14-Input current
{17}------------------------------------------------
The input current of proposed HDL-500X Light Source is same as that of the input current of proposed HDL-500E Light Source is different from that of the predicate device, the input current of proposed HDL-500E Light Standard), Therefore this difference on input current is considered not affectiveness of the proposed device.
#### SE Analysis 15- Examination lamp and Emergency lamp
The proposed HDL-500X Light Source has that of the predicate device. The proposed HDL-500E has different light source from that of the predicate device. The LED light source of HDL-500E is also a common light source used in endoscope system, therefore this difference on examination lamp and emergency lamp does not affect the safety and effectiveness of the proposed device.
#### SE Analysis 16- Average lamp life
The proposed HDL-500X Light Source has that of the predicate device. The proposed HDL-500E has longer lamp life than that of the predicate device. The LED ight source of HDL-500E is also a common light source used in endoxope system. Therefore on lamp life does not affect the safety and effectiveness of the proposed device.
#### SE Analysis 17- Average emergency lamp life
The proposed Light Sources have longency lamp life than that of the predicate device. This difference on average emergency lamp life does not affect the safety and effectiveness of the proposed device.
#### SE Analysis 18- Automatic exposure
Both the proposed device and predicate device provide the fination of automatic exposure. Although the adjustable levels the proposed device and predicate device, the purpose of image brightness adjustment. Therefore this item is considered as substantial equivalence.
{18}------------------------------------------------
| Predicate device: EVIS EXERA II 180 SYSTEM | | |
|------------------------------------------------------------------|------------------------------------------|------------------------------------------|
| ITEM | Proposed Device | Predicate Device |
| | HD-500 Video Endoscope System | K100584 |
| | | EVIS EXERA II 180 SYSTEM |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Particular<br>requirements | Comply with IEC 60601-2-18 | Comply with IEC 60601-2-18 |
| Product<br>Performance | Comply with ISO 8600-1 and ISO<br>8600-7 | Comply with ISO 8600-1 and ISO<br>8600-7 |
| Patient-contact<br>component and<br>material<br>(SE Analysis 19) | Insertion section | Unknown |
| | Distal end | Unknown |
| | Adhesive | Unknown |
| Biocompatibility<br>(SE Analysis 19) | Cytotoxicity, ISO 10993-5 | |
| | Sensitization, ISO 10993-10 | |
| | Irritation, ISO 10993-10 | |
## Table 6 Safety Comparison Proposed device: HD-500 Video Endoscope System
#### SE Analysis 19- Patient-contact component and material /Biocompatibility
The patient-contact components of the proposed device are same as those of the predicate device. Although the patient-contacting materials of the predicated device are unknown, the proposed device is biocompatible and conforms to ISO 10993 series standards. Therefore, the proposed device is claimed to be substantially equivalent to the predicate devices.
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis in section 8 and the side-by-side optical performance tests, the proposed device and the predicate device have the same intended use, comparable product specification and optical performance. Therefore, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.