1788 4K Camera System with Advanced Imaging Modality
Applicant
Stryker
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Mar 31, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Endoscopic white light and near-infrared illumination and imaging during endoscopic procedures.
Device Story
1788 4K Camera System provides live surgical video in visible and infrared spectrums. System components: Camera Control Unit (CCU), camera head with cable, and coupler for endoscope attachment. Principle of operation: light projected from source through endoscope onto CMOS image sensors; sensors acquire continuous image data; CCU processes data into video stream for display. Used in OR by surgeons (general, pediatric, gynecological, cardiac, thoracic, plastic, orthopedic, ENT, neuro, urology) for endoscopic visualization. Output allows real-time surgical field monitoring, aiding clinical decision-making and procedural guidance.
Clinical Evidence
Bench testing only. No clinical data required. Performance testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-6, IEC 60601-2-18), and software validation (IEC 62304). Mechanical bench testing verified tether pull force, connector insertion/removal force, and activation timing.
Technological Characteristics
System utilizes CMOS image sensors for 4K (3840 x 2160) video at 60 fps. Imaging modes include white light, near-infrared fluorescence, and near-infrared transillumination. Connectivity via CCU and camera head/coupler. Complies with IEC 60601-1, 60601-1-6, 60601-2-18, and 60601-1-2. Operates via alternate or simultaneous frame processing. Not patient-contacting.
Indications for Use
Indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, neurosurgery, and plastic surgery in adults and pediatric patients whenever a laparoscope/endoscope/arthroscope/sinuscope is indicated. Common procedures include cholecystectomy, hernia repair, appendectomy, pelvic lymph node detection, hysterectomy, spinal fusion, ACL reconstruction, knee/small joint arthroscopy, decompression fixation, wedge resection, lung/pleural biopsy, sympathectomy, pleurodesis, coronary artery bypass grafting, and cardiac chamber examination.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
1688 4K Camera System with Advanced Imaging Modality (K230216)
1688 4K Camera System with Advanced Imaging Modality (K214046)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
March 31, 2023
Stryker April Malmborg Senior Director, Regulatory Affairs 5900 Optical Court San Jose, California 95138
Re: K230605
Trade/Device Name: 1788 4K Camera System with Advanced Imaging Modality Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, GWG Dated: March 2, 2023 Received: March 3, 2023
Dear April Malmborg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K230605
#### Device Name
1788 4K Camera System with Advanced Imaging Modality
#### Indications for Use (Describe)
The 1788 Video Camera is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, sinuscopy, neurosurgery and plastic surgery whenever a laparoscope/ arthroscope/ sinuscope is indicated for use. The 1788 Video Camera is indicated for use in adults and pediatric patients.
A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repar, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the 1788 Video Camera are general and pediatric surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT/neurosurgeons and urologists.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# stryke
## 510(k) Summary
### Submitter:
| Applicant: | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg<br>Senior Director, Regulatory Affairs<br>Email: April.malmborg@stryker.com<br>Phone: (408) 754-2473 |
| Date Prepared: | March 3, 2023 |
#### Subject Device:
| Name of Device: | 1788 4K Camera System with Advanced Imaging Modality |
|-------------------------|-----------------------------------------------------------------------------------------------------------------|
| Common or<br>Usual Name | 3-chip Video Camera |
| Classification<br>Name: | Laparoscope, General and Plastic Surgery (21 C.F.R. §876.1500)<br>Endoscope, Neurological (21 C.F.R. §882.1480) |
| Regulatory Class: | II |
| Product Code: | GCJ<br>GWG |
| 510(k) Review<br>Panel: | General & Plastic Surgery<br>Neurology |
#### Predicate Device(s):
K230216*, K214046 1688 4K Camera System with Advanced Imaging Modality *Primary predicate
#### Device Description:
The 1788 4K Camera System with Advanced Imaging Modality is an endoscopic camera system that produces live video in the surgical field during surgical endoscopic procedures. The system is sensitive in the visible and infrared spectrums. The optical image is transferred from the surgical site to the camera head by a variety of rigid and flexible endoscopes, which are attached to the camera head. The 1788 4K Camera System consists of three main components: (1) a camera control unit (CCU); (2) a camera head with an integral cable that connects to the CCU; and (3) a coupler for attaching an endoscope to the camera head.
{4}------------------------------------------------
# K230605
# stryker
## Indications for Use:
| Item | Subject Device | Predicate Device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| | 1788 4K Camera System | 1688 4K Camera System |
| Intended Use | Endoscopic white light and near-infrared illumination<br>and imaging during endoscopic procedures. | Same as subject device |
| Indications<br>for Use | The 1788 4K Camera System is indicated for use in<br>general laparoscopy, nasopharyngoscopy, ear endoscopy,<br>sinuscopy, neurosurgery and plastic surgery whenever a<br>laparoscope/ endoscope/ arthroscope/ sinuscope is<br>indicated for use. The 1788 4K Camera System is<br>indicated for adults and pediatric patients.<br><br>A few examples of the more common endoscope<br>surgeries are Laparoscopic cholecystectomy,<br>Laparoscopic hernia repair, Laparoscopic appendectomy,<br>Laparoscopic pelvic lymph node detection,<br>Laparoscopically assisted hysterectomy, Laparoscopic<br>and thorascopic anterior spinal fusion, Anterior cruciate<br>ligament reconstruction, Knee arthroscopy, Small joint<br>arthroscopy, Decompression fixation, Wedge resection,<br>Lung biopsy, Pleural biopsy, Dorsal sympathectomy,<br>Pleurodesis, Internal mammary artery dissection for<br>coronary artery bypass, Coronary artery bypass grafting<br>where endoscopic visualization is indicated and<br>Examination of the evacuated cardiac chamber during<br>performance of valve replacement.<br><br>The users of the 1788 Video Camera are general and<br>pediatric surgeons, gynecologists, cardiac surgeons,<br>thoracic surgeons, plastic surgeons, orthopedic surgeons,<br>ENT surgeons, neurosurgeons and urologists. | Same as subject device. |
{5}------------------------------------------------
## K230605
# stryker
### Comparison of Technological Characteristics with the Predicate Device:
| Item | Subject Device | Predicate Device |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | 1788 4K Camera System | 1688 4K Camera System |
| Manufacturer | Stryker | Same as subject device. |
| Submission Reference(s) | Current Submission | K230216, K214046 |
| Imaging Modes | White Light<br>Near-infrared fluorescence<br>Near-infrared transillumination | Same as subject device. |
| Principles of Operations | Via an optical endoscope and<br>coupler, light is projected from a<br>light source onto one or more<br>complementary metal oxide<br>semiconductor image sensors<br>which acquire a continuous stream<br>of image data. The image data is<br>processed to provide a video<br>stream that is then sent to a display<br>for viewing. | Same as subject device. |
| Camera System<br>Components | Camera Control Unit<br><br>Camera Head(s) – Standard,<br>Inline, Integrated, Pendulum<br><br>Coupler(s) – AIM 4K | Camera Control Unit<br><br>Camera Head(s) – Standard,<br>Inline, Pendulum, Autoclavable<br><br>Coupler(s) – AIM 4K, AIM 4K<br>Autoclavable |
| Safety Standards | IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-2-18<br>IEC 60601-1-2 | Same as subject device. |
| Mode of Operation | Alternate Frame Processing<br>Simultaneous Frame Processing | Same as subject device. |
| Image Sensor | CMOS image sensor | Same as subject device. |
| Image Processing/Video<br>Output | Digital | Same as subject device. |
| Resolution | 4K (up to 3840 x 2160) | Same as subject device. |
| Frame Rate | 60 frames per second | Same as subject device. |
{6}------------------------------------------------
# stryke
### Performance Data:
The following performance data were provided in support of the substantial equivalence determination:
| Test | Method | Results |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electromagnetic<br>Compatibility | In accordance with IEC 60601-1-2:2014 | Pass |
| Electrical Safety | In accordance with:<br>• ANSI/AAMI ES60601-1:2005 + A1:2012<br>• IEC 60601-1-6:2010 + A1:2013<br>• IEC 60601-2-18:2009 | Pass |
| Software | In accordance with IEC 62304:2015 | Pass |
| Performance<br>Testing - Bench | Camera Head Soaker Cap Tether Pull Force<br>Camera Head to Camera Control Unit Connector Insertion/ Removal Force<br>Camera Head Button to Camera Control Unit Activation Timing | Pass |
NOTE: The 1788 4K Camera System is not patient contacting; therefore, biocompatibility testing was not required to support the determination of substantial equivalence.
NOTE: The 1788 4K Camera System does not require clinical studies to support the determination of substantial equivalence.
#### Conclusions:
The 1788 4K Camera System is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. In summary, the 1788 4K Camera System is substantially equivalent with respect to safety and effectiveness to the legally marketed predicate device.