ACMI AUTOCLAVABLE URETEROSCOPE, MODELS MR-6A AND MR-6AL
Applicant
Acmi Corporation
Product Code
FGB · Gastroenterology, Urology
Decision Date
Aug 16, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes (MR-6A) are intended for The examination/operation of the urinary tract, and using additional accessories, to the claimmation operation of the exapentic procedures. These devices are marketed as perform various diagnostic and mering, disinfecting and sterilization are included in reusable devices, member for creaming, excition 7.0, "Cleaning, Disinfection, and Sterilization".
Device Story
Semi-rigid ureteroscope; two working channels for flexible accessories (stone baskets, retrievers, forceps, lithotripter probes, laser fibers). Components: metal shaft, eyepiece, optical imaging fibers, light guide connector, distal lens, light carrier fibers. Used by physicians in clinical settings for urinary tract visualization and intervention. Modifications from predicate include MP35N alloy shaft, hermetic sealing via micro-TIG/laser welding, and validated autoclaving sterilization. Device provides direct visualization to assist clinical decision-making during urological procedures; benefits include improved durability and sterilization compatibility.
Clinical Evidence
Bench testing only.
Technological Characteristics
Semi-rigid endoscope; MP35N alloy shaft; micro-TIG and laser welding for hermetic sealing; optical imaging and light carrier fibers; two working channels; autoclavable.
Indications for Use
Indicated for examination and operation of the urinary tract in patients requiring diagnostic or therapeutic procedures via ureteroscope.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K200369 — OES Elite Ureteroscopes and Accessories · Olympus Winter & Ibe GmbH · Aug 28, 2020
K233372 — KARL STORZ Fiber Telescopes for Urology · Karl Storz SE & CO. KG · Jul 25, 2024
K960323 — KARL STORZ GELET URETERO-RENOSCOPES/KARL STORZ ALKEN MOTION CONTROL DEVICE · KARL STORZ Endoscopy-America, Inc. · Apr 22, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes ACMI Corporation Turnpike Road AUG 1 6 2005 uthborough, MA 01772
Special 510(k) Notification Summary of Safety and July 26, 2005
## 510(k) Summary of Safety and Effectiveness ACMI Corporation ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes
#### General Information
Manufacturer:
K052044
pg 1 of 2
ACMI Corporation 136 Turnpike Rd. Southborough, MA 01772-2104
Establishment Registration Number:
Contact Person:
Director, Regulatory Affairs
Endoscope and accessories
(21 CFR 876.1500), Class II Gastroenterology/Urology Panel
Endoscope and accessories
ACMI® MR-6A/MR-6LA Autoclavable
Terrence E. Sullivan
2020483
July 26, 2005
Date Prepared:
Device Description
Classification Name:
Trade Name:
Generic/Common Name:
#### Predicate Device
ACMI® MR-Series Ureteroscopes
K011849
Ureteroscopes
### Intended Uses
The ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes (MR-6A) are intended for The examination/operation of the urinary tract, and using additional accessories, to the claimmation operation of the exapentic procedures. These devices are marketed as perform various diagnostic and mering, disinfecting and sterilization are included in reusable devices, member for creaming, excition 7.0, "Cleaning, Disinfection, and Sterilization".
{1}------------------------------------------------
Special 510(k) Notification Summary of Safety and Effectiveness July 26, 2005
#### Product Description
K052044
py r of 2
Like the predicate ACMI® MR-Series Ureteroscopes, the ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes are semi-rigid endoscopes with two working channels. Flexible accessories such as stone baskets, retrievers, forceps, electrohydraulic lithotripter probes and laser fibers may be used through either working channel.
The main component parts of each device include:
- The metal shaft 1.
- 2. Two working channels
- An evepiece 3.
- 4. Optical imaging fibers
- Light guide connector post 5.
- A distal lens 6.
- 7. Light carrier fibers
This Special 510(k) proposes the addition of autoclaving processes as a sterilization method, a change in shaft material from stainless steel to MP35N alloy, as well as a change in the manufacturing process to allow the shaft to be hermetically sealed using a micro-TIG and laser welding process in place of epoxy seals. The indications for use, principles of operation, overall length, working channel length and working channel inner diameters of the ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes remain the same or essentially the same as the predicate device.
#### Summary of Safety and Effectiveness
The proposed modifications for the ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes, as described in this submission, are substantially equivalent to the predicate device. The proposed addition of autoclaving as a recommended sterilization process and change in shaft modification in design specifications, performance specifications, and dimensional specifications are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the devices.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a black and white logo. The logo features a stylized bird with three lines representing its wings or feathers. The bird is facing to the right. There is some text in a circular arrangement to the left of the bird, but the text is not legible.
AUG 1 6 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Terrence E. Sullivan Director, Regulatory Affairs ACMI Corporation 136 Turnpike Road SOUTHBOROUGH MA 01772
Re: K052044
> Trade/Device Name: ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB Dated: July 26, 2005 Received: July 28, 2005
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Device Name: ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes
510(k) Number: K 052044
#### Indications for use:
The ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes are intended for the examination/operation of the urinary tract, and using additional accessories, to perform various diagnostic and therapeutic procedures.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: __ X ___
OR Over-the-Counter Use: ________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
. 1
Nancy C brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.