ULTRASOUND GASTROSCOPE GF TYPE UM30P
Device Facts
| Record ID | K963023 |
|---|---|
| Device Name | ULTRASOUND GASTROSCOPE GF TYPE UM30P |
| Applicant | Olympus America, Inc. |
| Product Code | FDS · Gastroenterology, Urology |
| Decision Date | Feb 7, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Indications for Use
The Olympus GF-UM30P Ultrasonic Endoscope is intended to be used in combination with the Ultrasound System, EU-M20 or EU-M30, for endoscopic ultrasound imaging of the gastrointestinal tract wall, biliary and pancreatic duct, and surrounding organs. This instrument also provide for Endoscopic Ultrasound (EUS) guided fine needle aspiration (FNA) to acquire submucosal tissue.
Device Story
Mechanically scanning ultrasonic endoscope; operates in B-mode; utilizes de-aerated water immersion or balloon contact method; provides 250° scan. Enables ultrasonic guidance of endoscopic accessories (biopsy forceps, aspiration needles) by aligning sonographic scan plane with accessory entry path. Used in GI tract by physicians; provides real-time ultrasound imaging to facilitate tissue sampling and diagnostic visualization. Benefits include improved targeting for submucosal tissue acquisition compared to conventional radial scanners.
Clinical Evidence
Bench testing only.
Technological Characteristics
Insertion tube diameter 11.7 mm; working length 1265 mm; 250° scan angle; B-mode imaging; mechanical scanning mechanism. Ancillary MAJ-213 balloon provided clean, requires EtO sterilization. Device components marketed non-sterile, requiring manual reprocessing.
Indications for Use
Indicated for patients requiring endoscopic ultrasound imaging of the gastrointestinal tract, biliary and pancreatic ducts, and surrounding organs, and for those requiring EUS-guided fine needle aspiration for submucosal tissue acquisition.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- GF/JF/CF-UM20 (K926514)