Flyte Mechanotherapy System (aka Flyte System) (MTI-1.0)
K233362 · Pelvital USA, Inc. · HIR · Dec 29, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K233362
Device Name
Flyte Mechanotherapy System (aka Flyte System) (MTI-1.0)
Applicant
Pelvital USA, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Dec 29, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flyte System is intended for strengthening of the pelvic floor muscles to treat women with stress urinary incontinence.
Device Story
Flyte Mechanotherapy System is a reusable, non-sterile, vaginal device for in-home pelvic floor muscle (PFM) strengthening. It consists of a vaginal wand and a handheld controller. The wand, covered in medical-grade silicone, contains an electric motor with an eccentric weight to generate mechanical vibrations and sensors (accelerometer/gyroscope) to track movement. During Kegel exercises, the wand provides a resistive surface while delivering vibrations; the controller processes sensor data to provide real-time visual feedback to the user. This feedback confirms correct muscle contraction, guiding the user through treatment sessions. The device is intended for patient self-use to treat stress urinary incontinence by conditioning the PFM system.
Clinical Evidence
No clinical or non-clinical performance data was submitted; the application was limited to minor labeling changes for an identical device.
Technological Characteristics
Reusable vaginal wand and handheld controller. Wand: ABS plastic housing, medical-grade silicone sheath, electric motor with eccentric weight, accelerometer, and gyroscope. Controller: ABS plastic housing, 3.7V Lithium-ion Polymer battery (700-750 mAh), PCB assembly for motor control and visual feedback. Connectivity: Wired cable between wand and controller. Power: Rechargeable via AC adapter. Non-sterile.
Indications for Use
Indicated for strengthening pelvic floor muscles in women with stress urinary incontinence.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 29, 2023
Pelvital USA, Inc. Pamela Snyder Correspondent & Consultant 860 Blue Gentian Rd. Suite 200 Eagan, Minnesota 55121
Re: K233362
Trade/Device Name: Flyte Mechanotherapy System (aka Flyte System) (MTI-1.0) Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: December 11, 2023 Received: December 11, 2023
Dear Pamela Snyder:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jessica K. Nguyen -S
Jessica K.Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K233362
Device Name
Flyte System
Indications for Use (Describe)
The Flyte System is intended for strengthening of the pelvic floor muscles to treat women with stress urinary incontinence.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
### Special 510(k) Submission: Labeling Modification Flyte System
#### Applicant & Manufacturer
Pelvital USA, Inc. 860 Blue Gentian Rd. Suite 200 Eagan, MN 55121 866-735-8482
#### Applicant Contact
Lydia Zeller President & CEO lydia.zeller@pelvital.com 612-643-9724
#### Date Prepared: December 29, 2023
#### Device Identification
Trade names: The Flyte System or Flyte or MTI-1.0. Common Name: Perineometer Product Code: HIR Classification Panel: Obstetrics/Gynecology Regulation Number: 21 CFR § 884.14251 Classification Name: Perineometer
#### Legally Marketed Predicate
K212655 Flyte System The predicate devices have not been subject to a design-related recall.
#### Device Description
The Flyte System is a repeat use, non-sterile, vaginal device intended to condition and strengthen the pelvic floor muscle (PFM) system. The Flyte System is designed for in-home use to deliver treatment to the pelvic floor muscles during normal Kegel exercises. The product consists of a wand and a hand-held controller. The Flyte wand is placed in the vagina and delivers a series of mechanical vibrations while the pelvic floor muscles are contracting.
The hand-held controller controls the wand, guides the user through the treatment session and provides a visual feedback function to inform the user when the correct muscles are being contracted during a Kegel exercise. The wand provides a resistive surface against which the user can contract the PFM, and houses an electric motor containing an eccentric weight that
{5}------------------------------------------------
generates mechanical oscillations when the motor is running. A cable connects the wand and the controller.
Controller: The controller consists of a 3.7V Lithium-ion Polymer rechargeable battery with a capacity of 700-750 mAh and built-in safety protection. The controller also contains a PCB assembly used to control the motor speed and frequency and to provide the user with visual feedback information. The controller housing is made of ABS plastic (Acrylonitrile-butadienestyrene copolymer).
Wand: The wand contains an accelerometer and gyroscope which enable the Controller to generate the visual feedback information. It also houses the motor and weight used to generate the mechanical vibrations. The wand is available in Large and Small sizes. The wand housing is cylindrical and is made of ABS plastic. The wand is the only part of the device that directly contacts the user's vaginal cavity (mucosal membrane contact, ≤24hr) and is covered entirely with a biocompatible medical-grade silicone sheath.
#### Additional Components:
Charging Cable: Connects the Controller to the Charging Block Charging Block: AC adapter that allows you to plug the Controller into a wall outlet for charging.
#### Indications for Use
The Flyte System is intended for strengthening of the pelvic floor muscles to treat women with stress urinary incontinence.
#### Substantial Equivalence Summary
The Special 510(k) is a Labeling Modification to the Flyte System, predicate K212655. There are no changes to the Flye System device or principles of operation. The predicate and subject device are identical. A summary of the technological feature comparison is included in the SE table for completeness.
| | K233362 Under review | Predicate K212655 | Comparison |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Trade/Device Name | Flyte or Flyte System or Flyte<br>MTI-1.0.System | Flyte or Flyte System or<br>Flyte MTI-1.0. System | Identical |
| Manufacturer | Pelvital USA, Inc | Pelvital USA, Inc | Same Manufacturer |
| 510(k) Number | K233362 | K212655 | Same device |
| Regulation Number | 21 CFR § 884.1425 | 21 CFR § 884.1425 | Identical |
| Regulation Name | Perineometer | Perineometer | Identical |
| Regulatory Class | II | II | Identical |
| Product Code | HIR | HIR | Identical |
| | K233362 Under review | Predicate K212655 | Comparison |
| Over the Counter<br>(OTC) | Yes | Yes | Identical<br>Rx use was removed |
| Intended Use | Intended for strengthening of<br>the pelvic floor muscles | Intended for strengthening of<br>the pelvic floor muscles | Identical |
| Indications for Use | The Flyte System is intended<br>for strengthening of the pelvic<br>floor muscles to treat women<br>with stress urinary incontinence | The Pelvital System is<br>intended for the<br>strengthening of the pelvic<br>floor muscles, which has<br>been found to help women<br>with stress urinary<br>incontinence. | Similar<br>Same population<br>Same disease or<br>condition it treats.<br>Same intended use |
| Target Population | Women with urinary<br>incontinence | Women with urinary<br>incontinence | Identical |
| Anatomical Site | Vagina-Pelvic floor muscles | Vagina-Pelvic floor muscles | Identical |
| Single Patient Device | Yes | Yes | Identical |
| Single Use or<br>Reusable | Reusable | Reusable | Identical |
| Sterility | Non-Sterile; Clean | Non-sterile; Clean | Identical |
| Device Design | A handheld Controller and a<br>vaginally inserted Wand | A handheld control unit<br>and a vaginally inserted<br>Wand | Identical<br>Component name<br>changes from Control<br>unit/ Biofeedback unit<br>to Controller |
| Flyte Components | Wand<br>Controller<br>Charging Cable<br>Charging Block | Wand<br>Control Unit/ Biofeedback Unit<br>Charging Cable<br>Wall Charger | Identical Device<br>Component name<br>change from Control<br>unit to Controller |
| Visual feedback<br>Display | Yes | Yes | Identical |
| | K233362 Under review | Predicate K212655 | Comparison |
| Wand Dimensions | SMALL:<br>Max Shaft Diameter: 29 mm<br>Length: 97.85 mm<br>Weight: 89 +/- 1g<br>LARGE:<br>Max Shaft Diameter: 35 mm<br>Length:101 mm<br>Weight: 99+/- 2g | SMALL:<br>Max Shaft OD: 29 mm<br>Length: 97.85 mm<br>Weight: 89 +/- 1g<br>LARGE:<br>Max Shaft Diameter: 35 mm<br>Length:101 mm<br>Weight: 99+/- 2g | Identical |
| Technological Characteristics: The Devices are the Identical. There are No changes to the device. | | | |
{6}------------------------------------------------
{7}------------------------------------------------
The changes that are the subject of this premarket notification are limited to minor changes to the indication for use statement and labeling. The Flyte System is substantially equivalent to the predicate, Flyte System K212655. The devices are manufactured by Pelvital USA, Inc. There are no changes to the device's technological characteristics, materials, or principles of operation. The devices are identical.
The Flyte System has the same intended use and the same technological characteristics as the previously cleared predicate device, K212655. The Indication for use clarifies the indication without affecting the substance or meaning of the indications or intended use. It does not alter the original intended use or therapeutic effect. There is no change to patient populations, demographics, diagnosis, prognosis, comorbidities, and/or potential for complications. The riskbased assessment included an analysis of both safety and effectiveness. The modifications do not increase the risk profile of the device. There are no changes to contraindications, warnings, or precautions. There are no new risks are modifications of risks.
#### Clinical and Non-clinical Performance Testing
Given the changes proposed in this submission were limited to minor labeling changes, no clinical and non-clinical performance data was submitted in this submission to support the changes.
#### Conclusion
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The differences do not raise any different questions regarding safety and effectiveness. The Flyte System, as designed and manufactured, is substantially equivalent to the predicate device, K212655 Flyte System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.