Post-market user survey; Post-market surveillance report
The sponsor provided post-market data to demonstrate that the device is well-understood and easy to use without professional supervision, supporting the change in availability from prescription to over-the-counter.
Post-market surveillance; User survey; OTC transition
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-market user survey; Post-market survey
Users of the PeriCoach device
Not applicable for this study
Ease of use, understanding of device operation without supervision
Post Market Surveillance Report; Post-market surveillance
Users of the PeriCoach device
Not applicable for this study
Safety and efficacy in an OTC setting
Indications for Use
The PeriCoach® OTC is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.
Device Story
PeriCoach OTC is a perineometer for pelvic floor muscle training. Device consists of a rigid probe with a medical-grade silicone sheath; inserted vaginally by the user. Internal sensors measure contraction strength of pubococcygeus and puborectalis muscles. Data is transmitted wirelessly to a smartphone application. App provides real-time biofeedback, displaying relative contraction magnitudes or graphical representations of sensor output to guide user through Kegel exercises. Intended for over-the-counter, single-patient, reusable use. Benefits include improved pelvic floor muscle strength and management of urinary incontinence symptoms. Healthcare providers may use the output to monitor patient progress.
Clinical Evidence
No clinical trials were required. Evidence consists of a post-market user survey demonstrating that the device is well-understood and easy to use without supervision, supporting the OTC classification. Bench testing included biocompatibility (ISO 10993: cytotoxicity, sensitization, irritation, systemic toxicity), electrical safety/EMC (IEC 60601-1, 60601-1-2, 60601-1-11, 62133), and mechanical testing (drop, durability, immersion/cleaning, sensor behavior).
Technological Characteristics
Rigid polymer probe with medical-grade silicone sheath. Sensing via force-sensing resistors. Wireless data transmission to smartphone. Non-sterile, reusable, single-patient device. Electrical safety per IEC 60601-1, 60601-1-2, 60601-1-11, 62133. Biocompatibility per ISO 10993.
Indications for Use
Indicated for women with stress, mild-moderate urge, and mixed urinary incontinence to strengthen pelvic floor muscles via exercise.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other, with a wing-like shape extending above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 11, 2016
Analytica Ltd. % Tammy Ahmadzada Senior Consultant Brandwood Biomedical Suite 5, Level 9, 1 Chandos Street St. Leonards, NSW 2065 Australia
Re: K160758
> Trade/Device Name: PeriCoach OTC Regulation Number: 21 CFR§ 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: March 29, 2016 Received: April 12, 2016
Dear Tammy Ahmadzada:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160758
Device Name PeriCoach OTC
Indications for Use (Describe)
The PeriCoach® OTC is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Date Prepared: | 29 February 2016 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Analytica Pty Ltd<br>320 Adelaide Street,<br>Brisbane, QLD 4000,<br>Australia<br>Tel: +61 (0) 732781950 |
| Application Contact: | Tammy Ahmadzada<br>Suite 5, 1 Chandos St<br>St Leonards NSW 2065<br>Australia<br>Tel: +61(0) 299062984<br>Fax: +61(0) 285804613<br>tammy@brandwoodbiomedical.com |
| Trade Name: | PeriCoach® OTC |
| Common Name: | Perineometer |
| Classification Name: | (21CFR 884.1425) Perineometer |
| Product Code: | HIR |
| Predicate: | PeriCoach® (K143580) |
| Intended Use: | The PeriCoach® OTC is a perineometer designed to treat stress, mild-<br>moderate urge and mixed urinary incontinence in women, by<br>strengthening of the pelvic floor muscles through exercise. This device<br>provides biofeedback via smart phone technology. |
| Device Description: | The PeriCoach® OTC device consists of a rigid probe covered in a<br>silicone sheath that is temporarily inserted into the vagina. Sensors<br>located under the sheath measure the strength of contraction of the<br>user's pelvic floor muscles. This information is then transmitted<br>wirelessly to a smartphone application in order to provide real-time<br>feedback to the user. It is a single patient, reusable device to be<br>supplied over-the-counter. |
| Non-clinical Testing: | Biocompatibility<br>The patient contacting material in the PeriCoach® OTC have been<br>tested in accordance with ISO 10993 standards and found to be safe<br>for the intended purpose. Biocompatibility testing included |
5—1
{4}------------------------------------------------
Cytotoxicity, Sensitization, Vaginal Irritation, and Systemic Toxicity.
#### Software
The software used in the PeriCoach OTC was evaluated in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
## Electrical Safety and EMC
The PeriCoach OTC was evaluated in accordance with the following standards IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 62133. The PeriCoach OTC meets the respective standards for electrical safety and EMC to establish the safety of the device.
### Bench testing
Various mechanical tests have also been conducted to establish the safety of the device. The following bench tests were completed:
- Drop test
- Durability test
- Immersion/long term cleaning exposure
- Sensor behavior test
The bench testing established that the PeriCoach OTC met all of the acceptance criteria needed to demonstrate substantial equivalence.
### Clinical Testing: A post market user survey to substantiate the safety and effectiveness of the subject device for OTC use was provided. Although this type of information is not typically required for the review of OTC perineometer devices, the survey information was reviewed to further substantiate the substantial equivalence determination. The post market user survey demonstrated that the subject device is generally well understood and easy to use without supervision. The survey does not raise any concerns with OTC use of the device.
{5}------------------------------------------------
# Summary of Basis for Substantial Equivalence:
| Parameters | Predicate: PeriCoach®<br>(K143580) | PeriCoach® OTC<br>(Proposed device) |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Mode of Use | Reusable for single<br>patient | Same |
| Target Population | Adult female urinary<br>incontinence patients | Same |
| Principle of<br>Operation | A probe inserted into the<br>vagina to determine the<br>strength of the pelvic<br>floor muscles. Probe<br>sends signals to external<br>device to indicate muscle<br>contraction strength to<br>encourage and assist<br>user with voluntary kegel<br>exercises. | Same |
| Sensing method | Output from force<br>sensing resistors. | Same |
| Parameter<br>monitored | Analogue to digital output<br>of uncalibrated force<br>exerted against external<br>walls of device by<br>pubococcygeus and<br>puborectalismuscles. | Same |
| User Feedback | Provides real-time<br>feedback, via an<br>application on the user's<br>Android or iOS<br>smartphone. The<br>smartphone application<br>displays the relative<br>magnitudes of pelvic<br>muscle contraction or<br>graphically displays the<br>normalized analogue to<br>digital sensor output<br>depending on which<br>option is selected. | Same |
| Anatomical Sites | Female<br>Pubococcygeusmuscle<br>area | Same |
| Over the Counter | No, prescription only | Yes |
| Energy used<br>and/or delivered | The device is not<br>intended to deliver<br>energy to the patient.<br>Energy is used to operate<br>the device and<br>communicate the results. | Same |
| Compatibility with<br>environment and<br>other devices | Probe is not known to<br>conflict with other devices<br>or cause environmental<br>hazards and tested in<br>accordance with<br>IEC60601-1-2 (2007) | Same |
| Sterility | Non-sterile device | Same |
| Body Materials | Medical grade silicone | Same |
| Biocompatibility of<br>body material | Biocompatible in<br>accordance with<br>ISO10993 | Same |
| Electrical Safety | Tested in accordance<br>withIEC60601-1-2 (2007)<br>andIEC60601-1(2005) | Same |
| Chemical Safety | Probe outer surface<br>constructed of chemically<br>inert materials and tested<br>in accordance with<br>ISO10993 | Same |
| Construction | Rigid polymer structure<br>enclosed within a medical<br>grade silicone outer layer | Same |
5—3
{6}------------------------------------------------
- Conclusion: The predicate and proposed devices share indications for use, usage environments, outer construction materials and general principle of operation. The devices are all single patient, reusable and non-sterile.
The primary differences between the predicate device and the proposed device are:
- . the original PeriCoach® predicate device was prescription use only and the new PeriCoach® OTC device will be available as an over the counter device;
- . Software and firmware upgrades have been made to improve functionality
A Post Market Surveillance Report supports the decision to change the availability of the PeriCoach® device from prescription to over-the-counter. Non-clinical testing demonstrates that the PeriCoach® OTC device raises no
5—4
{7}------------------------------------------------
new safety or efficacy concerns and is therefore substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.