K202376 · Gi Supply, Inc. · PLL · Sep 30, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K202376
Device Name
EverLift Submucosal Lifting Agent
Applicant
Gi Supply, Inc.
Product Code
PLL · Gastroenterology, Urology
Decision Date
Sep 30, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
EverLift™ Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions prior to excision with a snare or other appropriate endoscopic device.
Device Story
EverLift is a sterile, injectable liquid composition supplied in a prefilled 10mL plastic syringe with a luer lock connection. It is used by endoscopists in hospital or clinic settings during endoscopic mucosal resection (EMR), endoscopic mucosal dissection (ESD), and polypectomy procedures. The agent is injected into the submucosal layer via a standard endoscopic injection needle inserted through the endoscope's working channel. The injection creates an in situ cushion, lifting the gastrointestinal mucosa from the submucosal layer to facilitate safe and easy resection of lesions in the esophagus, stomach, small intestine, sigmoid colon, and rectum. The device is for single-use only.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included syringe tip cap removal force, product color, injection flow rate, graduation marking tolerance (ISO 7886-1:2017), container closure integrity, tray lid peel force, sterilization validation (ISO 17665-1), LAL testing (USP <85>, USP <161>), distribution testing (ASTM D7386-16), and accelerated aging (ASTM F1980-16).
Technological Characteristics
Prefilled 10mL COC syringe; contains water, hydroxyethyl cellulose, glycerin, methylene blue, benzyl alcohol, sodium phosphate, and potassium phosphate. Sterilized by moist-heat (steam). Shelf life: 2 years. Single-use. Luer lock interface for standard endoscopic needles. Packaging: thermoformed tray with lid stock.
Indications for Use
Indicated for patients undergoing gastrointestinal endoscopic procedures requiring submucosal lift of polyps, adenomas, early-stage cancers, or other lesions prior to excision.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K241591 — Blue Eye · The Standard Co., Ltd. · Jul 3, 2024
K180068 — ORISE Gel · Boston Scientific · Sep 28, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 30, 2020
Gi Supply, Inc. Erika Parry Manager, Quality and Regulatory 5069 Ritter Road Suite 104 Mechanicsburg, Pennsylvania 17055
Re: K202376
Trade/Device Name: EverLift Submucosal Lifting Agent Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: PLL Dated: September 1, 2020 Received: September 2, 2020
Dear Erika Parry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202376
Device Name EverLift Submucosal Lifting Agent
Indications for Use (Describe)
EverLift Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions with a snare or other appropriate endoscopic device.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 14 M2 14 L14 2" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over The Counter Use (21 CFR 801 Subpart C) </div> |
||X| Prescription Use (Part 21 CFR 801 Subpart D)
_ | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER [Per 807.92(a)(1)]
Sponsor/Manufacturer GI Supply 5069 Ritter Road Suite 104 Mechanicsburg, PA 17055 USA Phone: (800)-451-5797
Contact Person Erika Parry Manager, Quality and Regulatory Phone: (717)-562-7580 Email: e.parry@gi-supply.com
### Date Prepared
August 19, 2020 Amended September 29, 2020
#### II. DEVICE [Per 807.92(a)(2)]
| Device Trade/Proprietary Name | EverLift™ Submucosal Lifting Agent |
|-------------------------------|------------------------------------|
| Device Common or Usual Name | Submucosal Injection Agent |
| Regulation Number | 21 CFR 876.1500 |
| Regulation Name | Endoscope and Accessories |
| Regulatory Class | Class II |
| Submission Type | Special 510(k) |
| Product Code | PLL |
| Classification Panel | Gastroenterology/Urology |
### III. PREDICATE DEVICE [Per 807.92(a)(3)]
The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent in terms of its intended use to the claimed predicate device, the 5mL EverLift™ Submucosal Lifting Agent (5mL) (K191923) with respect to device design, fundamental technology, physical characteristics, performance and intended use.
The device design and fundamental technology of the subject device are nearly identical to that of the predicate device. The formulation and composition of the primary packaging materials are identical. There are slight differences in the dimensions of the subject device and predicate device which do not impact safety or effectiveness. Both the subject and predicate devices are delivered sterile and are indicated for single-use only. The Sterilization Method and Shelf Life for the subject device are identical as that for the predicate device.
| Predicate<br>Device | The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent to<br>the following predicate device manufactured by GI Supply:<br>• EverLift™ Submucosal Lifting Agent (5mL) (K191923) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION [Per 807.92(a)(4)]
The GI Supply EverLift™ Submucosal Lifting Agent (10mL) is a prefilled plastic syringe with attached plunger rod containing 10mL of lifting agent. The syringe has a luer lock connection capable of interfacing with a standard, commercially available endoscopic injection needle.
EverLift™ Submucosal Lifting Agent is an injectable liguid composition for use as a submucosal injection agent during endoscopic mucosal resection (EMR), endoscopic mucosal dissection (ESD), and polypectomy procedures in the gastrointestinal tract. The device is intended for use in endoscopic resection procedures in the upper and the lower gastrointestinal tract, including the esophagus, the stomach, the small intestine, the sigmoid colon, and the rectum, as a submucosal injectable agent during the removal of polyps, adenomas, early-stage cancers, and other pathological lesions by EMR, ESD, or polypectomy.
EverLift™ Submucosal Lifting Agent is injected into the submucosal layer by means of a standard, commercially available endoscopic injection needle, which is inserted into the working channel of the endoscope. The composition, when injected, creates a cushion in situ by lifting the gastrointestinal mucosa from the submucosal layer, allowing the endoscopist to perform an easy and safe resection procedure (EMR, ESD, or polypectomy).
# V. INTENDED USE / INDICATIONS FOR USE [Per 807.92(a)(5)]
### Intended Use / Indications for Use
EverLift™ Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions prior to excision with a snare or other appropriate endoscopic device.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [Per 807.92(a)(6)]
The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent to the Gl Supply EverLift™ Submucosal Lifting Agent (5mL) (K191923) [predicate device] based on the same indication for use as the predicate device and the similar or identical functional and performance characteristics of the subject device when compared to the predicate device. The differences between the subject device and predicate device do not raise different issues of safety and effectiveness.
The detailed substantial equivalence comparison of the similarities and differences between the Gl Supply EverLift™ Submucosal Lifting Agent (10ml) [subject device] and the GI Supply EverLift™ Submucosal Lifting Agent (5mL) (K191923) [predicate device] is provided in the table below.
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | GI Supply | GI Supply | Same |
| Device Trade<br>or<br>Proprietary<br>Name | EverLift™ Submucosal Lifting Agent<br>(10mL) | EverLift™ Submucosal Lifting Agent<br>(5mL) | --- |
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| 510(k)<br>Number | K202376 | K191923 | --- |
| Device Class | Class II | Class II | Same |
| Device<br>Classification<br>Name | Endoscope and Accessories | Endoscope and Accessories | Same |
| Device<br>Common<br>Name | Submucosal Injection Agent | Submucosal Injection Agent | Same |
| Product Code | PLL | PLL | Same |
| Regulation<br>Number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same |
| Design Features and Capabilities of the Device | | | |
| Indications<br>for Use | EverLift™ Submucosal Lifting Agent<br>is indicated for use in<br>gastrointestinal endoscopic<br>procedures for submucosal lift of<br>polyps, adenomas, early-stage<br>cancers, or other gastrointestinal<br>lesions prior to excision with a<br>snare or other appropriate<br>endoscopic device. | EverLift™ Submucosal Lifting Agent<br>is indicated for use in<br>gastrointestinal endoscopic<br>procedures for submucosal lift of<br>polyps, adenomas, early-stage<br>cancers, or other gastrointestinal<br>lesions prior to excision with a<br>snare or other appropriate<br>endoscopic device. | Same |
| Intended Use | Same as Indications for Use | Same as Indications for Use | Same |
| Prescription<br>or Over-the-<br>Counter<br>(OTC) Use | Prescription Use | Prescription Use | Same |
| Use<br>Environment | Hospital / Clinic | Hospital / Clinic | Same |
| Sterile | Sterile | Sterile | Same |
| Single-Use | Single-Use Only | Single-Use Only | Same |
| Design Features | | | |
| Packaging | Each individual 10mL COC syringe is<br>packed inside a thermoformed tray.<br>The tray is sealed with a labeled lid<br>stock with three (3) patient labels<br>affixed to the lid stock. Ten (10) of<br>the sealed trays are packed into a<br>shelf box with a copy of the IFU<br>placed between the trays and the<br>shelf box wall. The shelf box is<br>shipped in a regular die cut and<br>scored 200-pound B-flute kraft<br>shipper box sealed with a shipper<br>box label. | Each individual 5mL COC syringe is<br>packed along with three (3) patient<br>labels inside a propionate tube with<br>a friction fit press on closure. Ten<br>(10) tubes are packed into a shelf<br>box on top of a copy of the IFU. The<br>shelf box is shipped in a regular die<br>cut and scored 200-pound B-flute<br>kraft shipper box sealed with a<br>shipper box label. | Different. The<br>differences in<br>packaging do<br>not raise<br>different<br>questions of<br>safety or<br>effectiveness.<br>A Distribution<br>Study was<br>performed on<br>the new<br>packaging<br>configuration. |
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| Images<br>of Device | Image: Subject Device | Image: Predicate Device | --- |
| Sterilization<br>Method | Sterilized by moist-heat (steam)<br>sterilization | Sterilized by moist-heat (steam)<br>sterilization | Same |
| Shelf Life | 2-Year Shelf Life | 2-Year Shelf Life | Same |
| Composition | Each steam sterilized syringe<br>contains 10mL of lifting agent with<br>the following ingredients:<br>• Water<br>• Hydroxyethyl Cellulose<br>• Glycerin<br>• Methylene Blue<br>• Benzyl Alcohol<br>• Sodium Phosphate<br>• Potassium phosphate | Each steam sterilized syringe<br>contains 5mL of lifting agent with<br>the following ingredients:<br>• Water<br>• Hydroxyethyl Cellulose<br>• Glycerin<br>• Methylene Blue<br>• Benzyl Alcohol<br>• Sodium Phosphate<br>• Potassium phosphate | Same<br>concentration;<br>however,<br>volume of<br>lifting agent<br>provided<br>differs. |
{5}------------------------------------------------
{6}------------------------------------------------
# VII. SUMMARY OF PERFORMANCE DATA AND PERFORMANCE TEST CONCLUSIONS [Per 807.92(b)(1)(2)(3)]
The determination of substantial equivalence is based on an assessment of non-clinical performance data. To verify that the device design meets its functional and performance requirements, the EverLift™ Submucosal Lifting Agent (10mL) underwent performance testing. Biocompatibility testing was performed only on the 5mL variant as it represented worst-case surface area to volume ratio.
The EverLift™ Submucosal Lifting Agent was developed under GI Supply's risk management process in accordance with ISO 14971:2012 - Medical Devices-Application of Risk Management to Medical Devices. The identified risks were adequately mitigated and verified by means of non-clinical performance testing.
### Summary of Performance Testing
A series of non-clinical performance testing was conducted on the subject device. Please refer to the table below for a summary of all non-clinical performance testing conducted on the subject device in support of substantial equivalence.
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe Tip Cap<br>Removal Force<br>Testing | None | Syringe tip cap removal force testing was performed on<br>the subject device using the same method and<br>acceptance criteria as the predicate device. Results of<br>the testing confirmed that the syringe tip cap removal<br>force is equivalent between the subject and predicate<br>devices, thereby demonstrating substantial equivalence. |
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
| Product Color<br>Testing | None | Product color testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that product color is identical between the two devices, thereby demonstrating substantial equivalence. |
| Injection Flow<br>Rate Testing | None | Injection flow rate testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the injection flow rate is equivalent between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Graduation<br>Marking<br>Tolerance | ISO 7886-1:2017, Sterile<br>hypodermic needles for single<br>use – Part 1: Syringes for manual<br>use | Graduation mark tolerance testing was performed on the subject device using the same method as the predicate device, and acceptance criteria as defined in ISO 7886-1. Results of the testing confirmed that the marking tolerance meets the acceptance criteria defined in ISO 7886-1, thereby demonstrating substantial equivalence. |
| Container<br>Closure<br>Integrity Testing | None | Container closure integrity testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the container closure integrity is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Tray Lid Peel<br>Force Testing | None | The secondary packaging of the subject device required a minor design change to accommodate the larger 10mL syringe. Tray lid peel force test results post-distribution simulation met acceptance criteria, confirming that the differences in secondary packaging do not raise different questions of safety or effectiveness. |
| Sterility<br>Assurance Level | ISO 17665-1, Sterilization of<br>healthcare products – Moist<br>heat – Part 1: Requirements for<br>the development, validation and<br>routine control of a sterilization<br>process for medical devices | A sterilization validation was performed on the subject device using the same sterilization process and sterilization parameters as the predicate device. Validation results confirmed that the Sterility Assurance Level (SAL) is identical for both devices (10-6), thereby demonstrating substantial equivalence. |
| LAL Testing | USP <85>, Bacterial Endotoxins<br><br>USP <161>, Transfusion and<br>Infusion Assemblies and Similar<br>Medical Devices | LAL testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the LAL level is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. Additionally, LAL testing will be performed on each lot prior to final product release. |
| Distribution and<br>Post-<br>Distribution<br>Testing | D7386-16, ASTM Standard<br>Practice for Performance Testing<br>of Packages for Single Parcel<br>Delivery Systems | Post-distribution testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that package integrity is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
| Accelerated and<br>Post-<br>Accelerated<br>Aging Studies | F1980-16, ASTM Standard Guide<br>for Accelerated Aging of Sterile<br>Barrier Systems for Medical<br>Devices | Post-accelerated aging testing was performed on the<br>subject device using the same method and acceptance<br>criteria as the predicate device. Results of the testing<br>was identical between the subject and predicate devices,<br>thereby demonstrating substantial equivalence. |
{7}------------------------------------------------
{8}------------------------------------------------
# VIII. CONCLUSIONS
Based on the results of the non-clinical performance testing conducted on the subject device, it has been concluded that the proposed EverLift™ Submucosal Lifting Agent is as safe and effective and performs as well as the legally marketed predicate device (K191923). The similar indications for use, intended use, technological characteristics, and performance characteristics for the proposed EverLift™ Submucosal Lifting Agent have been assessed to be substantially equivalent to the predicate device, and any differences do not raise different issues of safety and effectiveness when compared to the predicate device. Therefore, the EverLift™ Submucosal Lifting Agent is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.