K242163 · SmartsurgN, Inc. · OWN · Apr 20, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K242163
Device Name
SmartSurgN Visualization System
Applicant
SmartsurgN, Inc.
Product Code
OWN · Gastroenterology, Urology
Decision Date
Apr 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The SmartSurgN Visualization System is intended to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging. Upon intravenous administration and use of an ICG consistent with its approved label the SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Device Story
Endoscopic video imaging system providing real-time visible (VIS) and near-infrared (NIR) fluorescence imaging; used during minimally invasive surgery. System components: EyeRSurgN console, PrizMCurgN camera head, IRLightSurgN light source, and OptIcRSurgN 10mm ICG laparoscope. Light source provides VIS and NIR illumination; user selects imaging mode (Regular, IRMax, IRFlo, or IRTrue). Camera head acquires images; console processes signals; output displayed on 4K medical-grade monitor. Optional Rec4KSurgN recorder captures procedure. Operated by surgeons in clinical settings. NIR imaging enables visual assessment of vessels, blood flow, tissue perfusion, and biliary ducts when used with ICG. Benefits include enhanced intraoperative visualization of anatomy and perfusion, aiding surgical decision-making.
Clinical Evidence
No clinical investigations were required. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (EN IEC 60825-1), optical performance (ISO 8600-3, ISO 8600-5), software validation (IEC 62304), and sterilization validation (SAL 10^-6).
Technological Characteristics
4K, 4CMOS video processing; LED illumination and NIR laser excitation. Laparoscope: 10mm diameter, 31cm length, 0°/30° DOV, 70° FOV. Materials: Stainless steel, epoxy, sapphire, PEEK. Sterilization: Ethylene Oxide (console/camera/light source) and Steam (laparoscope). Connectivity: Standalone system with optional 4K video recorder. Software: IEC 62304 compliant.
Indications for Use
Indicated for patients undergoing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for visual assessment of vessels, blood flow, tissue perfusion, and major extra-hepatic bile ducts (cystic, common bile, and common hepatic ducts) using ICG. Not for standalone biliary duct visualization.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 20, 2026
SmartSurgN Inc.
Jocelyn Long
Chief Compliance Officer
1038 Leigh Avenue
Suite 101B
San Jose, California 95126
Re: K242163
Trade/Device Name: SmartSurgN Visualization System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: OWN
Dated: July 10, 2024
Received: July 24, 2024
Dear Jocelyn Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242163 - Jocelyn Long
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242163 - Jocelyn Long
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**YAN FU-S**
Digitally signed by YAN FU-S
Date: 2026.04.20 13:33:08
-04'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242163 | |
| Device Name SmartSurgN Visualizaton System | |
| Indications for Use (Describe) The SmartSurgN Visualization System is intended to provide realtime endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging. Upon intravenous administration and use of an ICG consistent with its approved label the SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 5
510(k) Summary
K242163
This summary statement is submitted in accordance with the requirements of 21 CFR § 807.92
| 510 (k) Sponsor: | | SmartSurgN Incorporated |
| --- | --- | --- |
| Address | | 1038 Leigh Ave, Suite 101B
San Jose, CA 95126 USA |
| Correspondence Person: | Jocelyn Long |
| --- | --- |
| Contact Information: | jocelyn.long@smartsurgn.com
408-802-8586 |
| Date of Preparation: | | April 15, 2026 |
| --- | --- | --- |
| Device Identification: | | Trade Name:
SmartSurgN Visualization System |
| Common Name: | | Endoscopic Video Imaging System |
| Classification Name: | | Endoscope and Accessories |
| Regulation Number: | | 21 CFR 876.1500 |
| Product Code: | | OWN |
| Regulatory Class | | II |
| Indications for Use: | The SmartSurgN Visualization System is intended to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging. Upon intravenous administration and use of an ICG consistent with its approved label, the SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. |
| --- | --- |
| Device Description | The SmartSurgN Visualization System is designed to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging during minimally invasive surgery.
The SmartSurgN Visualization System is comprised of the following main components:
• EyeRSurgN Console with PrizMCurgN Camera Head
• IRLightSurgN Light Source |
{5}
Page 2 of 5
- OptIcRSurgN 10mm ICG Laparoscope, 0° or 30°
The SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
During use, the IRLightSurgN Light Source provides visible and near-infrared illumination of the surgical site. The user selects the imaging mode (Regular, IRMax, IRFlo, or IRTrue), which determines the light spectrum used for image capture.
The IRLightSurgN is connected to the SmartSurgN Laparoscope using a fiber optic light cable. Additionally, the SmartSurgN OptIcRSurgN 10mm ICG Laparoscope connects to the PrizMCurgN Camera Head.
Images are acquired by the PrizMCurgN Camera Head and transmitted to the EyeRSurgN Console. Images are processed by the EyeRSurgN Console and outputted to the View4SurgN 4K medical grade monitor. If connecting a Rec4KSurgN 4K video recorder to the EyeRSurgN Console, a record of the surgical procedure can be saved.
| Predicate Device: | SmartSurgN Visualization System
(K213943)
Regulation: 876.1500
Product Code: OWN |
| --- | --- |
# Technological Comparison
The SmartSurgN Visualization System was previously cleared under K213943 and is the primary predicate device; the predicate device has not been subject to any design-related recalls. The subject device has the same intended use, indications for use, intended users, and use environment as the predicate device.
| Item/Feature | Proposed Device
K242163 | Predicate Device
K213943 |
| --- | --- | --- |
| | SmartSurgN Visualization System | SmartSurgN
Visualization System |
| Intended Use | The SmartSurgN Visualization System is intended to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence imaging during minimally invasive surgery. | Same as proposed device |
| Indications for Use | The SmartSurgN Visualization System is intended to provide real-time endoscopic visible (VIS) and real-time near-infrared (NIR) fluorescence | Same as proposed device |
{6}
| Item/Feature | Proposed Device K242163 | Predicate Device K213943 |
| --- | --- | --- |
| | imaging. Upon intravenous administration and use of an ICG consistent with its approved label, the SmartSurgN Visualization System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct) using near-infrared imaging. Fluorescence imaging of biliary ducts with the SmartSurgN Visualization System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization | |
| Device Design | 4K, 4CMOS video signal processing and projection with LED illumination and NIR laser excitation | Same as proposed device |
| Component Device: | IRLightSurgN Light Source and EyeRSurgN | |
| Packaging: Physical Dimensions | EyeRSurgN, IRLightSurgN = 12”x12”x4.5” (30cm x30cmx11cm) | Same as proposed device |
| Packaging: Weight | EyeRSurgN = 9 lb. (4kg) IRLightSurgN = 10 lb. (4.5kg) | Same as proposed device |
| Device Design | 4K, 4CMOS video signal processing and projection with LED illumination and NIR laser excitation | Same as proposed device |
| Component Device: | PrizMCurgN | |
| Reprocessing: Cleaning | Manual | Same as proposed device |
| Device Design | 4K, 4CMOS video signal processing and projection with LED illumination and NIR laser excitation | Same as proposed device |
| Reprocessing: Sterilization Methods | Ethylene Oxide | Same as proposed device |
| Component Device: | OptIcRSurgN Laparoscope | |
| Outer diameter | 10mm | Same as proposed device |
| Working Length | 31 cm | Same as proposed device |
| DOV | 0°, 30° | Same as proposed device |
| FOV | 70° | Same as proposed device |
| Patient-Contacting Materials | Stainless Steel, Epoxy, Sapphire, PEEK | Same as proposed device |
| Single or Reusable | Reusable | Same as proposed device |
| Cleaning | Manual | Same as proposed device |
| Sterilization Methods | Steam Sterilization | Same as proposed device |
| Sterility Assurance Level | 10-6 | Same as proposed device |
The SmartSurgN Visualization System was previously cleared under K213943 and is the primary predicate device for this submission. The predicate device has not been subject to any design-related
{7}
recalls. The SmartSurgN Visualization System provides real-time endoscopic visible (VIS) and near-infrared (NIR) fluorescence imaging to support minimally invasive surgical procedures.
The subject SmartSurgN Visualization System has the same intended use, indications for use, intended users, and use environment as the predicate device. It continues to enable surgeons to perform minimally invasive surgery using standard endoscopic visible light and NIR fluorescence imaging when used with indocyanine green (ICG), consistent with the cleared indications. No new clinical claims or standalone use claims are introduced.
There are no design changes to the imaging console, camera head, surgical light source, or system operation compared to the predicate device. The major technological characteristics, including image acquisition, optical signal transmission, image processing, illumination, and display output, remain unchanged. Updates to software that improve device reliability and operator usability were tested and released.
The subject device includes the addition of laparoscope suppliers. The laparoscope maintains the same diameter, working length range, field of view, materials, and patient-contacting profile as the predicate laparoscope.
# Performance Testing
Non-Clinical Testing:
The following non-clinical testing was performed to support the safety and substantial equivalence.
| Item/Feature | Verification/Validation Method(s) | Acceptance Criteria | Summary of Results |
| --- | --- | --- | --- |
| Electrical Safety and EMC | in accordance with IEC60601-1-1:2005+AMD1:2012 CSV Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Meet IEC60601-1-1:2005+AMD1:2012 CSV | Pass |
| | in accordance with IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Meet IEC 60601-1-2:2014 | Pass |
| | in accordance with EN IEC60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements | Meet EN IEC60825-1:2014 | Pass |
| ISO 8600-5 Second Edition 2020-10 | in accordance with ISO 8600-5 Second Edition 2020-10, Optics and photonics - Medical endoscopes and endotherapy devices - Part 5: Determination of optical resolution of rigid endoscopes with optics | Meet ISO 8600-5 Second Edition 2020-10 | Pass |
| | In accordance with ISO 8600-3 Second Edition 2019-8 Endoscopes - Medical endoscopes and endotherapy devices Part 3: Determination of field of view and direction of view of endoscopes with optics. | Meet ISO 8600-3Second Edition 2019-8 | Pass |
| Software | In accordance with IEC62304:2006 Medical device software - Software life cycle processes | Meet IEC62304:2006 | Pass |
| | ISO/IEC 27000:2016 Information Technology - Security techniques - Information security | ISO/IEC 27000:2016 | Pass |
{8}
| | management systems - Overview and vocabulary | | | |
| --- | --- | --- | --- | --- |
| Packaging | in accordance with ISTA 2A-2011 Packaged-Products Weighing 150 Lb (68 Kg) Or Less | | Meet ISTA 2A-2011 | Pass |
| Device Change | Risks | Verification/Validation Method(s) | Acceptance Criteria | Summary of Results |
| Additional laparoscope suppliers added as a device component alternate source | Patient Infection
Device failure causes patient injury or delay in procedure. | Performance Specifications and Testing: ICG Fluorescence Sensitivity Test | Substantially equivalent to Predicate Device | Pass |
| | | Performance Specifications and Testing: Brightness Uniformity and Illumination Test | Substantially equivalent to predicate device | Pass |
| | | Performance Specifications and Testing | ISO 8600-1:2015-10-15, ISO 8600-3:2019-08, ISO 8600-5:2020 | Pass |
| | | Performance Specifications and Testing: Image Quality Comparison | Substantially equivalent to predicate device | Pass |
| | | Sterilization validation was conducted using FDA-recognized methods to demonstrate sterility assurance. | Devices shall maintain package integrity and have SAL of 10-6. | Sterilization validation testing results passed (n=3).
Sterilization validation established SAL of 10-6. |
| Rec4K SurgN 4K video recorder added as an accessory | Malfunction during procedure
Device failure causes delay in procedure | Compatibility testing was completed using TP-32 Test Protocol for TEAC Video Recording Accessory Compatibility | Device shall record video image without error. | Compatibility Verification testing results passed (n=1). |
Note: No clinical investigations were required for a determination of substantial equivalence.
## Conclusion
The SmartSurgN Visualization System has the same intended use, indications for use, technological characteristics, and principles of operation as the predicate device cleared under K213943. Non-clinical testing demonstrates that the subject device is as safe, as effective, and performs as well as or better than the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.