KARL STORZ ICG Imaging System

K212695 · Karl Storz Endoscopy America · OWN · Nov 15, 2021 · Gastroenterology, Urology

Device Facts

Record IDK212695
Device NameKARL STORZ ICG Imaging System
ApplicantKarl Storz Endoscopy America
Product CodeOWN · Gastroenterology, Urology
Decision DateNov 15, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesPediatric

Intended Use

The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging. The KARL STORZ VITOM II ICG System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light. Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Device Story

System provides real-time visible (VIS) and near-infrared (NIR) fluorescence imaging for minimally invasive surgery. Inputs: endoscopic video signals from TIPCAM1 Rubina or other compatible telescopes. Processing: Image1 S Camera Control Unit processes VIS and NIR signals; offers three NIR display modes (Overlay, Intensity Map, Monochromatic). Output: 4K video displayed on monitors for surgeon assessment of blood flow, tissue perfusion, and anatomical structures (biliary ducts, lymphatic system). Used in OR by surgeons. Benefits: intraoperative visual assessment of perfusion and anatomy to guide surgical decision-making.

Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing, including electrical safety (IEC 60601-1, -1-2, -2-18), ISO endoscopic standards (ISO 8600-1, 3-6), and software verification/validation.

Technological Characteristics

System includes Image1 S 4U Rubina camera head, Power LED Rubina light source, and TIPCAM1 Rubina Videoendoscope. Supports 4K output via 12G/3G-SDI, DVI-D, DisplayPort. Features 1x-2.5x zoom. Software level of concern: Moderate. Sterilization/cleaning validated for camera head.

Indications for Use

Indicated for minimally invasive surgery in adults and pediatrics (cranial neurosurgery) and adults and pediatrics > 6 years (endonasal skull base surgery) for visualization of vessels, blood flow, tissue perfusion, and lymphatic systems using ICG fluorescence. Also indicated for plastic, micro-, and reconstructive surgery for tissue perfusion assessment.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION". KARL STORZ Endoscopy America Mario Trujillo Regulatoary Affairs Specialist 2151 E. Grand Avenue El Segundo, California 90245 November 15, 2021 Re: K212695 Trade/Device Name: KARL STORZ ICG Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OWN. GWG Dated: October 28, 2021 Received: October 29, 2021 Dear Mario Trujillo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Neil R.P. Ogden Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212695 #### Device Name KARL STORZ ICG Imaging System ### Indications for Use (Describe) The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging. Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for biliary duct visualization. Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels. blood flow and related tissue perfusion using near infrared imaging. The KARL STORZ VITOM II ICG System is intended for capturing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light. Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> | |----------------------------------------------|---------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters. The logo is simple and professional, and it is likely used to represent the company on its products and marketing materials. # 510(k) Summary This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92 and the FDA guidance document titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued on July 28, 2014. All data included in this document is accurate and complete to the best of KARL STORZ SE & Co. KG knowledge. | Submitter: | KARL STORZ SE & Co. KG<br>Dr.-Karl-Storz-Straße 34<br>78532 Tuttlingen, Germany | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Mario Trujillo<br>Regulatory Affairs Specialist<br>Tel.: (424) 218-8481<br>Email: Mario.Trujillo@karlstorz.com | | Date of Preparation: | October 21, 2021 | | Type of 510(k) Submission: | Traditional | | Device Identification: | Trade Name: KARL STORZ ICG Imaging System<br>Classification Name: Confocal Optical Imaging<br>21 CFR 876.1500 (Endoscope and Accessories)<br>21 CFR 882.1480 (Neurological Endoscopes) | | Regulatory Class: | 2 | | Product Code: | OWN, GWG | | Guidance Document: | Not Applicable | | Predicate Device: | Primary predicate device KARL STORZ ICG Imaging System<br>(K202925). Secondary predicate device: KARL STORZ ICG Imaging<br>System (K180146). Reference device: TIPCAM1 Rubina Video<br>Endoscope System (K201526) | | Device Description: | The modified KARL STORZ ICG Imaging System is identical to the<br>KARL STORZ ICG Imaging System recently cleared under K202925.<br>The modified KARL STORZ ICG Imaging System now includes the<br>following components:<br>TIPCAM®1 Rubina Videoscope (0°, 30°): a 3D image capable<br>videoendoscope with 2D auto-leveling (auto-rotation) and 2D auto-<br>switch display modes. | | Intended Use: | The KARL STORZ ICG Imaging System is intended to provide real-<br>time visible and near-infrared fluorescence imaging. | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters on the top line. The bottom line says "KARL STORZ - ENDOSKOPE" in smaller, blue letters. | Indications For Use: | The KARL STORZ ICG Imaging System is intended to provide real-<br>time visible (VIS) and near-infrared (NIR) fluorescence imaging.<br><br>Upon intravenous administration and use of ICG consistent with its<br>approved label, the KARL STORZ Endoscopic ICG System enables<br>surgeons to perform minimally invasive surgery using standard<br>endoscopic visible light as well as visual assessment of vessels, blood<br>flow and related tissue perfusion, and at least one of the major extra-<br>hepatic bile ducts (cystic duct, common bile duct and common<br>hepatic duct), using near infrared imaging. Fluorescence imaging of<br>biliary ducts with the KARL STORZ Endoscopic ICG System is<br>intended for use with standard of care white light and, when<br>indicated, intraoperative cholangiography. The device is not<br>intended for standalone use for biliary duct visualization.<br>Additionally, the KARL STORZ Endoscopic ICG System enables<br>surgeon to perform minimally invasive cranial neurosurgery in<br>adults and pediatrics and endonasal skull base surgery in adults and<br>pediatrics > 6 years of age using standard endoscopic visible light as<br>well as visual assessment of vessels, blood flow and related tissue<br>perfusion using near infrared imaging.<br><br>The KARL STORZ VITOM II ICG System is intended for capturing and<br>viewing fluorescent images for the visual assessment of blood flow,<br>as an adjunctive method for the evaluation of tissue perfusion, and<br>related tissue-transfer circulation in tissue and free flaps used in<br>plastic, micro- and reconstructive surgical procedures. The VITOM II<br>ICG System is intended to provide a magnified view of the surgical<br>field in standard white light.<br><br>Upon interstitial administration and use of ICG consistent with its<br>approved label, the KARL STORZ Endoscopic ICG System is used to<br>perform intraoperative fluorescence imaging and visualization of<br>the lymphatic system, including lymphatic vessels and lymph nodes. | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters. | Technological<br>Characteristics: | The clinical application for the subject KARL STORZ ICG Imaging<br>System is identical to the cleared KARL STORZ ICG Imaging System,<br>K202925. | |-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The 4mm, 5mm & 10mm Endoscopes and VITOM ICG telescopes<br>connected to the optical coupler of the Image1 S 4U Rubina camera<br>head, and the TIPCAM®1 Rubina Videoendoscope which connects to<br>the Image1 S Camera Control Unit for image processing, as well as to<br>the Power LED Rubina light source light source via compatible light<br>cable as the source of illumination to allow visualization of internal<br>anatomy. Visualization and navigation is performed initially using<br>VIS imaging. NIR imaging is selected when visual assessment and/or<br>confirmation of vessels, blood flow or tissue perfusion is desired. | | | For the NIR image, the user has three presentations of the ICG<br>imagery to choose from: | | | a. Overlay: The white light image is overlaid with the NIR image.<br>The NIR image could either by blue or green | | | b. Intensity Map: The white light image is overlaid with color<br>transformed NIR image. | | | c. Monochromatic: The NIR image is indicated by the color white<br>against a dark background. | | | The KARL STORZ ICG Imaging System can output a 4K image to the<br>monitor 12G/3G-SDI, DVI-D and DisplayPort digital outputs and also<br>offers 7 increments of zoom ranging from 1x to 2.5x. | | | The reference device, TIPCAM1 Rubina Video Endoscope System, is<br>identical in design and materials of construction to the subject<br>device. The only difference is that the ICG functionality was not<br>included as part of K201526. | | Non-Clinical<br>Performance Data: | There are no performance standards or special controls developed<br>under Section 514 of the FD&C Act for endoscopes. | | | However, the KARL STORZ ICG Imaging System, as well as the<br>primary, secondary and reference devices follow the FDA<br>recognized consensus standards and is tested according to the<br>following standards and FDA Guidance: | | | Electrical Safety and EMC<br>•<br>o IEC 60601-1<br>o IEC 60601-1-2<br>o IEC 60601-2-18 | | | ISO Endoscopic Standards<br>•<br>o ISO 8600-1<br>o ISO 8600-3<br>o ISO 8600-4<br>o ISO 8600-5 | | | o ISO 8600-6 | | | •<br>Software Verification and Validation Testing<br>o Guidance for the Content of Premarket Submissions for | | | Software Contained in Medical Device | | | o Level of concern: Moderate | | | Cleaning and sterilization validations were conducted for the<br>Image1 S 4U Rubina camera head. | | | Additional bench testing was performed to ensure the device met its<br>design specifications. The bench testing performed verified and<br>validated that the KARL STORZ ICG Imaging System has met all its<br>design specification and is substantially equivalent to its predicate<br>device. | | | Conformance to these standards was not the basis upon which<br>substantial equivalence was determined. | | Clinical Performance<br>Data: | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate devices. Non-clinical bench testing was<br>sufficient to establish the substantial equivalence of the<br>modifications. | | Conclusion: | The conclusions drawn from the nonclinical tests demonstrate that<br>the subject device, the KARL STORZ ICG Imaging System performs as<br>well as or better than the predicate devices that are currently<br>marketed for the same intended use. | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...