The ASC TriPort and QuadPort Laparoscopic Access Devices are intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
Device Story
Laparoscopic access devices providing port for multiple instruments/camera during minimally invasive abdominal surgery. Device comprises introducer, retractor, and valve components. Operates by applying radial force to abdominal incision to create aperture; valve maintains pneumoperitoneum. QuadPort variant accommodates four instruments via larger boot. Used in OR by surgeons. Benefits include simultaneous instrument passage through single incision and specimen removal without full port removal. No clinical data; performance verified via bench testing and porcine model.
Clinical Evidence
No clinical testing submitted. Evidence consists of bench-top design verification and validation testing in a porcine model. Porcine study confirmed device functionality across various clinician experience levels and verified that formal user training was not required.
Technological Characteristics
Sterile, single-use laparoscopic access ports. Components: introducer, retractor, valve. Mechanism: radial force application to incision. Materials: consistent with predicate devices. QuadPort features increased boot diameter for fourth instrument valve. No software or electronic components.
Indications for Use
Indicated for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery in patients undergoing such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
K101794
Page 1 of 5
# 510(k) Summary For the Advanced Surgical Concepts (ASC) TriPort and QuadPort Laparoscopic Access Devices
(per 21 CFR 807.92 and http://www.fda.gov/MedicalDeviceRegulationandGuidance/default.htm)
## 1. SUBMITTER/510(K) HOLDER
Advanced Surgical Concepts NOV 2 9 2010 Unit 4 Sunnybank Centre Upper Dargle Road Bray, County Wicklow Ireland Establishment Registration Number: 9616720 Contact Person: Tanya Kavanagh
+353-1276-4413 Telephone:
Date Prepared: November 29, 2010
# 2. DEVICE NAME
| Proprietary Name: | ASC TriPort and QuadPort Laparoscopic Access Devices |
|----------------------|---------------------------------------------------------------|
| Common/Usual Name: | Laparoscopic Accessory |
| Classification Name: | Endoscopic Accessory and Surgical Retractor (21 CFR 876.1500) |
| Product Code: | OTJ |
# 3. PREDICATE DEVICES
- · TriPort Laparoscopic Access Device subject of K073719
- · R-Port II Laparoscopic Access Device subject of K073170
- · R-Port Laparoscopic Access Device subject of K070158
# 4. DEVICE DESCRIPTION
# Physical Description
# TriPort Laparoscopic Access Device:
The modified ASC TriPort Laparoscopic Access Device, like the parent TriPort cleared by the FDA under K073719, is still comprised of the following three components:
{1}------------------------------------------------
Kio1794
- · an introducer component which creates an abdominal incision (except in the case where the surgeon creates a Hasson cut-down incision) and delivers the Distal Ring of the ASC TriPort into the abdominal cavity
- · a retractor component which retracts an abdominal incision to allow the passage of laparoscopic instruments
- · a valve component which maintains the pneumoperitoneum established for the surgical procedure
The modified ASC TriPort Laparoscopic Access Device is still intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery. The modified ASC TriPort Laparoscopic Access Device is identical in function to the parent device, which has been cleared for marketing under K073719.
The modifications made to the parent device labeling to produce the proposed TriPort Laparoscopic Access device informs users and patients that the device does not contain Natural Rubber Latex. Additionally the labelling has been revised to extend the shelf life from three months to three years as notified in the original 510(k) and to make the device safer and easier to use. The changes made to the parent TriPort Laparoscopic Access Device to produce the proposed TriPort are minor and do not represent modifications to the indications for use, operating principles, or mechanism of action for the device. Therefore, the 510(k) Premarket Notification for the modified Tri-Port Laparoscopic Access Device is appropriate for review as a Special 510(k).
Since the submission of the original 510(k) K073719, a number of additional nonsignificant modifications, which were the subject of letters to file, were made to the parent TriPort device as follows:
- A second insufflation valve was added to facilitate evacuation of electrocautery . smoke.
- Retaining clips are no longer blue in color (colorant has been removed from the . material formulation).
- . Flange on insufflation valve was extended from a roughly circular shape of 6mm diameter to a roughly oval shape 7.5mm x 4.5mm to effect a more secure connection to the Boot.
November 29, 2010
- Modification to the labeling to include a three year shelf life. .
{2}------------------------------------------------
Kio1794
Page 3 of 5
In accordance with FDA Guidance, "Deciding When to Submit a 510(k) for a Change to an Existing Device," issued in 1997, these modifications are included in this Special 510(k).
## OuadPort Laparoscopic Access Device:
The TriPort has additionally been modified resulting in the QuadPort device. The modifications to the TriPort comprise the following:
- A fourth valve was added to accommodate an additional laparoscopic . instrument.
- The Boot is larger in diameter to accommodate the fourth instrument valve. .
- An Introducer and Retaining Clips are not required. .
- The Removal Ring is made from a different material (the same material as is . used for the TriPort distal ring).
These modifications fulfill the requirements for Special 510(k): Device Modification as defined in "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications - Final Guidance," issued in 1998.
## How the Device Functions
The ASC TriPort and QuadPort Laparoscopic Access Devices are laparoscopic multiinstrument ports which perform the following two functions:
- . They retract a small abdominal incision to allow multiple laparoscopic instruments to pass through a small incision to the abdomen at the same time during laparoscopic surgery.
- They ensure that pneumoperitoneum is maintained in the abdomen during the . surgical procedure, whether or not one or more laparoscopic instruments are passing through them.
#### Scientific Concepts That Form the Basis for the Device
The ASC TriPort and QuadPort Laparoscopic Access Devices apply radial force upon the incision creating an aperture though which the laparoscopic instruments pass. The ability to remove the boot, allows the clinician to deliver specimens without removing the entire port, and with the QuadPort, generally without the need to extend the incision.
{3}------------------------------------------------
Kio1794
Page 425
# Significant Physical and Performance Characteristics of the Device, such as Device Design, Material Used, and Physical Properties
The physical design and performance of the modified TriPort is unchanged. Materials changes relate to substitution with materials already used for the TriPort. The physical properties of the QuadPort differ from the TriPort in that the diameter of the boot is increased, allowing space for the addition of a fourth port.
## 5. INTENDED USE/INDICATION FOR USE
The ASC TriPort and QuadPort Laparoscopic Access Devices are intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
## 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE/S
The ASC TriPort and QuadPort Laparoscopic Access Devices provide an access path for laparoscopic instruments through a small incision in the abdominal wall. Their function is identical to that of the parent ASC TriPort Laparoscopic Access Device (K073719).
Both the proposed ASC TriPort/QuadPort and the parent device are laparoscopic instrument access ports that are used to perform the same function as a trocar. They retract a small abdominal incision to allow laparoscopic instruments to pass through to the abdomen and maintain pneumoperitoneum in the abdomen during the surgical procedure, whether or not laparoscopic instruments are passing through them.
Both the proposed ASC TriPort/QuadPort and the parent device, allow for the simultaneous introduction of up to 4 laparoscopic instruments through a single incision.
Like the predicate devices, the ASC TriPort and QuadPort are sterile, single-use (disposable) devices. The use of the ASC TriPort/QuadPort and the predicates are identical in that they facilitate the passage of laparoscopic instrumentation while maintaining pneumoperitoneum. The insertion of the ASC TriPort and the parent device are identical and are made consistently and safely during laparoscopic surgery.
# 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Biocompatibility and performance verification testing of the ASC TriPort demonstrates that the modifications that are the subject of this notification do not
{4}------------------------------------------------
K101794
Page 5 of 5
raise new issues of safety or effectiveness. Validation testing of the ASC QuadPort in a porcine model enrolled clinicians with various levels of experience and demonstrate that the QuadPort functioned as intended, the performance did not raise new issues of safety and effectiveness, and that formal user training was not required.
## 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
No clinical testing was submitted as the basis for determining substantial equivalence.
#### 9. SUMMARY OF OTHER INFORMATION
Other information submitted for review included revised Instructions for Use and promotional materials for use in the US.
## 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Based on design verification testing of both the TriPort and QuadPort, along with the validation testing of the QuadPort in the porcine animal model, the modified TriPort and QuadPort fulfill prospectively defined design and performance requirements.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three wing-like shapes stacked on top of each other. Encircling the bird symbol is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 2 9 2010
Advanced Surgical Concepts % Medical Device Consultants, Inc. Ms. Rosina Robinson 49 Plain Street North Attleboro, Massachusetts 02760
Re: K101794
Trade/Device Name: ASC TriPort and QuadPort Laparoscopic Access Devices Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II ﻌﺸ Product Code: OTJ Dated: November 09, 2010 Received: November 10, 2010
Dear Ms. Robinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{6}------------------------------------------------
Page 2 - Ms. Rosina Robinson
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours
Mark N. Melkerson Director Division of Surgical. Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K10 1794
# Indications for Use
NOV 2 9 2000
Special 510(k) Number (if known):
Device Name: ASC TriPort and QuadPort Laparoscopic Access Devices
Indications for Use:
The ASC TriPort and QuadPort Laparoscopic Access Devices are intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Butler. Farmer
(Division Sign of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101794
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.