EndoSound Vision System; EVS

K232518 · Endosound, Inc. · ODG · Dec 27, 2023 · Gastroenterology, Urology

Device Facts

Record IDK232518
Device NameEndoSound Vision System; EVS
ApplicantEndosound, Inc.
Product CodeODG · Gastroenterology, Urology
Decision DateDec 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EndoSound Vision System (EVS), when affixed to an endoscope, is intended to provide ultrasonic visualization of, and ultrasound guided therapeutic access to the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The EVS, mounted on an endoscope, is introduced orally when indications consistent with the requirement for a GI procedure are met. The EVS is a prescription only device to be used by a qualified physician. The clinical environments where the system can be used include clinics, hospitals, and ambulatory surgery centers.

Device Story

Add-on ultrasound system; attaches externally to standard upper GI endoscope (EGD) to convert it into an EUS endoscope. Components: EVSScanner (beamformer), Ultrasound Transducer Module (UTM), Transducer Extension Cable (TEC), and sterile disposable Ultrasound Disposable Kit (UDK-T). UDK-T provides needle/transducer articulation via external control knob. System connects to PC/tablet via USB 3.0 running EVSViewer software. Physician operates endoscope and EVS controls to perform real-time ultrasound imaging and needle aspiration. Enables visualization of upper GI organs/tissues; assists in therapeutic procedures. Benefits: provides EUS capabilities without requiring dedicated, expensive EUS endoscopes; maintains native gastroscope functions (video, suction, insufflation).

Clinical Evidence

Bench testing and animal study performed. GLP animal study (n=3) compared EVS to PENTAX EG-3870UTK. Results: no clinically relevant damage identified; imaging quality equivalent to control; needle targeting and maneuverability equivalent; no cable trapping or mucosal damage observed. Bench testing included biocompatibility (ISO 10993), sterilization validation, and electrical safety (IEC 60601 series).

Technological Characteristics

Convex curved-linear array transducer (6.4mm ROC, 96-element, 5-12 MHz). External attachment via disposable UDK-T. Connectivity: USB 3.0 to PC. Software: EVSViewer (customized Telemed Echo Wave II). Sterilization: ETO for UDK-T; high-level disinfection (Cidex OPA) for UTM. IPX7/IPX8 rated. Compliant with IEC 60601-1, 60601-2-18, 60601-2-37, ISO 10993, ISO 14971.

Indications for Use

Indicated for ultrasonic visualization and ultrasound-guided therapeutic access to the upper GI tract (esophagus, stomach, duodenum) in patients requiring GI procedures. For use by qualified physicians in clinics, hospitals, and ambulatory surgery centers.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 27, 2023 Endosound, Inc. Patrick Herriman VP Quality 4640 S. Macadam Ave Unit 200 Portland, OR 97239 Re: K232518 Trade/Device Name: EndoSound Vision System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODG, IYN, IYO, ITX Dated: November 21, 2023 Received: November 28, 2023 Dear Patrick Herriman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Shanil P. Haugen -S Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232518 Device Name EndoSound Vision System #### Indications for Use (Describe) The EndoSound Vision System (EVS), when affixed to an endoscope, is intended to provide ultrasonic visualization of, and ultrasound guided therapeutic access to the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The EVS, mounted on an endoscope, is introduced orally when indications consistent with the requirement for a GI procedure are met. The EVS is a prescription only device to be used by a qualified physician. The clinical environments where the system can be used include clinics, hospitals, and ambulatory surgery centers. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(K) Summary # ldentifying information | Manufacturer: | EndoSound, Inc. | |-------------------------------|--------------------------------------------------------------------------| | Device Trade Name: | EndoSound Vision System | | Address: | 4640 S Macadam Avenue<br>Suite 200<br>Portland, OR 97239 | | Telephone: | (971) 231-4791 | | Web: | www.endosound.com | | E-mail: | info@endosound.com | | Contact Person: | Patrick Herriman, VP of Quality, EndoSound, Inc., herriman@endosound.com | | Date 510(k) Summary Prepared: | August 7th, 2023 | #### Class and Predicate Information | Classification Name | Regulation Number | Product Code | |----------------------------------------------------------|-------------------|--------------| | Endoscopic Ultrasound System<br>Gastroenterology-Urology | 876.1500 | ODG | | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN | | Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO | | Diagnostic Ultrasonic Transducer | 892.1570 | ITX | Common Name: Endoscopic Ultrasound System # Proprietary Name: EndoSound Vision System (EVS) Regulatory Class: Class II Predicate Devices: Olympus GF Type UCT180, Olympus GF Type UST180, Evis Exera II Ultrasound Gastrovideoscope used with ALOKA SSD-a 10 Ultrasound System (K093395) > PENTAX Medical EG-J10U Endoscopic Ultrasound System with EG38-J10UT Ultrasound Upper GI Video Scope (Convex Array Type) (used with Hitachi ARIETTA 70 Ultrasound console (K134016)) Reference Predicate Device: Telemed ArtUs Pulsed Doppler ultrasound imaging system (K211248) {5}------------------------------------------------ # LIST OF CONSENSUS STANDARDS The EndoSound Vision System (EVS) was tested to be in conformance with the following consensus standards, with applicable test certifications included in this submission: - . IEC 60601-1 Medical electrical equipment –Part 1: General requirements for safety and essential performance. - . IEC 60601-1-2 Medical electrical equipment –Part 1-2 General requirements for safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests. - IEC 60601-1-6 Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance, Collateral Standard: Usability - . IEC 60601-2-18 Medical Electrical Equipment - Part 2-18: Particular Requirements For The Basic Safety And Essential Performance Of Endoscopic Equipment - . IEC 60601-2-37 Medical electrical equipment -Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. - . IEC 62304 Medical device software - Software life cycle processes. - IEC 62366-1 Medical devices Part 1: Application of usability engineering to medical devices ● - ISO 14971 Medical devices - Application of risk management to medical devices. - ISO 15223-1 Medical devices - Symbols to be used with medical device labelling and information to be supplied - Part 1: General requirements. - . ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. - . ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. - . ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. - . ISO 10993-11 Biological evaluation of medical devices - Part 11: Acute Cytotoxicity # Indication for Use The EndoSound Vision System (EVS), when affixed to an endoscope, is intended to provide ultrasonic visualization of, and ultrasound quided therapeutic access to the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The EVS, mounted on an endoscope, is introduced orally when indications consistent with the requirement for a GI procedure are met. The EVS is a prescription only device to be used by a qualified physician. The clinical environments where the system can be used include clinics, hospitals, and ambulatory surgery centers. {6}------------------------------------------------ # Device Description The EndoSound Vision System® (EVS) is an add-on ultrasound system designed to attach externally to an upper gastrointestinal (GI) endoscope (gastroscope/upper (EGD) endoscope). Once attached, it temporarily converts an EGD endoscope to a fully capable EUS endoscope. The EVS consists of: - an electronic ultrasound beamformer/image analyzer (EVSScanner), - a reusable transducer assembly (Ultrasound Transducer Module (UTM)), . - . a cable/connector to interface the transducer to the beamformer (Transducer Extension Cable (TEC)). - a sterile, disposable mounting kit with an operator control mechanism (Ultrasound Disposable Kit (UDK-T)) used to externally affix the EVS to a standard gastroscope/upper (EGD) endoscope and to provide needle and transducer angle control. When attached to an upper GI endoscope, the EVS enables real-time ultrasound imaging, ultrasound quided needle aspiration and other EUS quided procedures within the upper gastrointestinal tract and surrounding organs. Native functions of the gastroscope including video imaging, articulation insufflation and suctioning are available after attachment of the EVS device. The system works in tandem with a PC computer/monitor/tablet connected via a USB 3.0 cable that runs custom software (EVSViewer) and provides the user interface and ultrasound display. The EVS components and technical characteristics are described individually below. EVSScanner (Beamformer): The EVSScanner is a small, portable, self-contained module that contains the electronics that perform the ultrasound scanning and receiving functions (beamforming). The beamformer is the commercially available ArtUs ultrasonic system (K211248) manufactured by Telemed with a firmware modification that allows the system to recognize the EndoSound transducer. The device produces real-time ultrasound images in combination with the EVS curved linear array transducer and a computer/display running EVSViewer software. Ultrasound Transducer Module (UTM): The UTM is a 6.4mm radius-of-curvature (ROC) convex, 96element, curved-linear ultrasound arrav imaging over a frequency range from 5 to 12 MHz. The array produces a fixed ultrasound field-of-view of 150 degrees. The array is connected to a flat, multi-wire cable to a waterproof connector. The transducer is designed to be high-level disinfected using Cidex OPA as commonly employed for gastroscopes and EUS scopes. # Transducer Extension Cable (TEC) The TEC consists of a multiwire coax cable with connector interfaces on each end. One end plugs into the EVSScanner and the other end plugs into the UTM. It functions as an extension cable to extend the length of the UTM while in use and the UTM facing connector provides a waterproof seal with the UTM while in use. EVSViewer Software: The EVSViewer is the software run on a PC/monitor which processes the ultrasound data received from the EVSScanner module and displays the ultrasound image. It also functions to produce the user interface on the computer monitor. The software is equivalent to the Echo Wave II software manufactured by Telemed (K211248) with the following changes: - EVS Probe Integration: The EVS UTM probe was integrated to work with the base o software. - Probe Recognition Function: The probe recognition function was altered to read a singleo wire EEPROM (electronic chip) residing in the EVS UTM. {7}------------------------------------------------ - UTM Counter in the EVSViewer: A dialog box displays upon bootup requiring the user to O press a button to proceed, debiting the number of preprogrammed uses of the UTM attached array until the probe is locked out. - Ultrasound Field of View (FOV) lock: The EVSViewer removes the ability of the user to o change the ultrasound field-of-view (only the full 150-degree field of view is allowed). - EndoSound Splash Screen: Displays an EndoSound splash screen upon bootup. o - Insert EndoSound Icons: EndoSound branded small icons are displayed on the user O interface. All changes to the software have been verified and validated. # Ultrasound Disposable Kit (UDK-T) The UDK-T is a disposable, single-use kit provided sterile, that is used to attach the EVS to the compatible endoscope and provide transducer and needle articulation in addition to the native gastroscope controls. Sterility and shelf-life testing for six-month life was conducted and reports may be found in Appendix 4, Clinical Safety & Effectiveness. Silicone sleeves are included to protect the endoscope and help secure it to compatible gastroscopes. The control wire is terminated to a control knob which is affixed to the proximal end of the gastroscope. The control knob allows the transducer assembly as well as any needle inserted through the working channel to be articulated/deflected over a working range. # Comparison Table to Predicate Devices | Table 1 | Proposed Device<br>(this submission) | Predicate Device | Predicate Device | |-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | EndoSound Vision System<br>(EVS) | Evis Exera II Ultrasound<br>Gastrovideoscope<br>(Olympus GF-UCT180)<br>(used with ALOKA<br>SSD-a 10 Ultrasound<br>System) | PENTAX Medical EG-<br>J10U Endoscopic<br>Ultrasound System with<br>EG38-J10UT<br>Ultrasound Upper Gl<br>Video Scope (Convex<br>Array Type) (used with<br>Hitachi ARIETTA 70<br>Ultrasound console<br>(K134016)) | | Manufacturer | EndoSound, Inc. | Olympus | Pentax | | 510(k) Number | K232518 | K093395 | K200090 | | Product Code | IYN, IYO, ITX, ODG | IYN, IYO, ITX, ODG,<br>NWB | IYN, IYO, ITX, ODG | | Indications for<br>Use | The EndoSound Vision<br>System (EVS), when<br>affixed to an endoscope,<br>is intended to provide<br>ultrasonic visualization of,<br>and ultrasound guided<br>therapeutic access to the<br>Upper Gastrointestinal<br>Tract including but not<br>restricted to the organs,<br>tissues, and subsystems:<br>Esophagus, Stomach,<br>Duodenum and underlying<br>areas. The EVS, mounted<br>on an endoscope, is<br>introduced orally when<br>indications consistent with<br>the requirement for a Gl<br>procedure are met. | This instrument has<br>been designed to be<br>used with an Olympus<br>universal endoscopic<br>ultrasound center or a<br>diagnostic ultrasound<br>system (ALOKA CO.<br>LTD), video system<br>center, light source,<br>documentation<br>equipment, monitor,<br>EndoTherapy<br>accessories and other<br>ancillary equipment.<br>This instrument is<br>designed for<br>endoscopic real-time<br>ultrasound imaging,<br>ultrasound guided<br>needle aspiration and<br>other endoscopic<br>procedures within the<br>upper gastrointestinal<br>tract and surrounding<br>organs | The PENTAX Medical<br>Endoscopic Ultrasound<br>System is intended to<br>provide optical<br>visualization of<br>ultrasonic visualization<br>of, and therapeutic<br>access to, the Upper<br>Gastrointestinal Track<br>including but not<br>restricted to the organs,<br>tissues, and<br>subsystems:<br>Esophagus, Stomach,<br>Duodenum, Small<br>Bowel, and underlying<br>areas. The instrument<br>is introduced per orally<br>when indications<br>consistent with the<br>requirement for<br>procedure are observed<br>in adult and pediatric<br>patient populations. | | Endoscope<br>Compatibility | Gastroscopes/endoscopes<br>with bending section<br>diameters from 9.5-<br>11.5mm | NA | NA | | Distal Mechanical<br>Dimensions<br>(As attached to<br>compatible<br>endoscope) | Height in attachment<br>region (max): 19.4 mm<br>Width in attachment<br>region (max): 13.5mm | Diameter: 15.5 mm | Diameter 14.65 mm | | Working Length | 1167 mm | 1250 mm | 1250 mm | | Attachment<br>method | Disposable component | NA | NA | | Needle deflection | Disposable component<br>with wire control of<br>transducer assembly and<br>needle deflection | Integral mechanical<br>elevator | Integral mechanical<br>elevator | | Ultrasound<br>Transducer Type | Convex Curved Linear | Convex Curved Linear | Convex Curved Linear | | Transducer<br>Radius of<br>Curvature | 6.4 mm | 5 mm | 6 mm | | Imaging Modes | B, M, Pulse Wave Doppler<br>(PWD), Color Doppler<br>(CFM), Power Doppler<br>(PDI), Directional Power<br>Doppler (DPDI),<br>Combined modes (B+B,<br>B+M, 4B, B+PWD<br>(Duplex),<br>B+CFM/PDI/DPDI+PWD<br>(Triplex)), Tissue<br>Harmonic Imaging (THI)<br>and Inverted Tissue<br>Harmonic Imaging (ITHI) | B, M, Color Doppler,<br>Power Doppler, Pulse<br>Wave Doppler,<br>Combined modes: B/M,<br>B/PWD, M/CD,<br>B/CD/PWD, B/CWD,<br>B/CD/CWD B/CD/PWD | Bmode, Mmode, PW<br>mode (Pulsed Wave<br>Doppler), CW mode<br>(Continuous Wave<br>Doppler), Color<br>Doppler, Amplitude<br>Doppler (Color Flow<br>Angiography),<br>TDI (Tissue Doppler<br>Imaging), 3D Imaging,<br>Real-Time Tissue<br>Elastography, and<br>Real Time Virtual<br>Sonography. | | Scanning<br>Direction | Parallel to insertion<br>direction (adjustable to<br>forward viewing) | Parallel to insertion<br>direction | Parallel to insertion<br>direction | | Scanning<br>Frequencies | 5, 5.3, 6.7, 8, 9, 11 MHz | 5, 6, 7.5, 10, 12 MHz | 5.0, 6.5, 7.5, 10.0, 13.0<br>MHz | | Ultrasound Field<br>of View | 150 degrees | 180 degrees | 150 degrees | | Contacting<br>Method | Direct | Balloon, Direct | Balloon, Direct | | Transducer<br>Articulation<br>Range (using<br>UDK control)<br>(relative to<br>endoscope axis) | 5-85 degrees | Fixed | Fixed | | Bending portion<br>angulation range | UP 157°, DOWN 73°<br>RIGHT 87° LEFT 80°<br>(Typical, attached to<br>Pentax 29i10<br>gastroscope) | UP 130°, DOWN 90°<br>RIGHT 90° LEFT 90° | UP 160°, DOWN 130°<br>RIGHT 120° LEFT 120° | | Transducer<br>Surface<br>Temperature<br>(Maximum) | 41 degrees C | 43 degrees C | 43 degrees C | | Ingress Protection<br>Rating | IPX7/IPX8 | IPX7 | IPX7 | | Electrosurgery<br>compatible | Yes | Yes | Yes | | Acoustic Output | Track 3 Compliant | Track 3 Compliant | Track 3 Compliant | | Disinfection<br>Compatibility | Liquid Chemical High<br>Level Disinfection | Liquid Chemical High<br>Level Disinfection | Liquid Chemical High<br>Level Disinfection | {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ # General Safety and Effectiveness # Ultrasound Scanning The EVSScanner employs the same basic principles of operation including signal generation, digital acquisition, processing, and display as the predicate device. The transducer is of the same principal design and form and uses similar transduction processes to generate and receive the ultrasound signal. Phantom and in vivo study comparison images show similar scanning resolution, image quality and ultrasound image performance. The EVS has no ultrasound characteristics, modes, or features that are not previously evaluated or present in the predicate device, or that are not commonly used in traditional ultrasound systems. # Physical Characteristics The EVS is placed on an upper GI gastroscope to create a device similar in performance to a dedicated ultrasound endoscope. In the EVS the cabling and control wires are external to the endoscope, whereas in the predicates, the control wires and electronic cabling are interior to the endoscope. For efficient management of the external cabling, the UDK-T control wire is wrapped around the endoscope during attachment, which manages the external electronic cabling to the endoscope. Verification testing has shown that the external cabling does not impact the performance and safety of the EVS relative to the listed predicates. Once attached, the physical form, ultrasound image, and physical manipulations of the device are substantially equivalent to the listed predicates. The EVS varies from the predicate in the manner of deflection of the predicate devices use a built-in elevator mechanism to deflect the needle whereas the EVS uses an external, disposable attachable kit (UDK-T) to manipulate the needle to the intended target. The process entails using the UDK-T Control Knob to alter the articulation of the transducer assembly. This allows the needle to be inserted through a built-in slot in the EVS to place the needle in the ultrasound scan plane. Once through the aperture, the directed movement of the transducer assembly deflects the needle. Overall, the physical attribute differences do not impact safety or efficacy of the combined system or performance as compared to the predicate devices as reported in the body of this submission. # Biocompatibility A summary of the successfully completed biological safety testing on components of the EVS in contact with the patient is shown in Table 2. {11}------------------------------------------------ # Table 2. Summary of Completed ISO 10993-1 Tests Performed | Ultrasound Disposable kit (UDK-T) PN 18040<br>Ultrasound Transducer Module (UTM) PN 18010 | | |-------------------------------------------------------------------------------------------|------------------------------| | Cytotoxicity | ISO 10993-5 | | Sensitization | ISO 10993-10 | | Irritation | ISO 10993-23 or ISO 10993-10 | | Acute systemic toxicity | ISO 10993-11:2017 | # UTM Reprocessing Manual Cleaning, High-Level Disinfection and Rinsing Validation studies were performed on the EVS UTM by an outside GLP accredited laboratory with passing results. # UDK Sterilization and Packaging Integrity, ECH and ETO Residual Testing The UDK-T sterile kit (PN 18040) is sterilized using ETO sterilization. Packaging Verification including sterilization efficacy testing was performed at time TO (after manufacture) and time T1 (post 6 months accelerated aging). # Animal Testing A GLP lab study was performed comparing safety and performance of the EVS with a predicate endoscope (PENTAX EG-3870UTK EUS scope) (K130247, K182004). The study was conducted with an expert endoscopist as the user under the supervision of the GLP quality director and the study director. A detailed optical and endoscopic ultrasound regimen was utilized for both the control device. The regimen was implemented four times in each of 3 live animal models. Study results included: Overall Safety Assessment: Following the completion of six procedures, no clinically relevant damage was identified, either radiologically, endoscopically, or macroscopically. Optical and EUS imaging: The study incorporated both standard endoscopic (i.e., optical) assessment and complete EUS assessment. The comparison of images obtained from these tests revealed no significant differences from the control, suggesting that the EVS is substantially equivalent to the control device in terms of imaging quality. Behavior of EVS during Needle Targeting: No issues or remarkable differences were observed when angulating the EVS to the same extent as the predicate device. The clinical behavior of the EVS is substantially equivalent to the control device. The EVS successfully conducted procedures in the evaluated scenarios. Endoscope and Device Maneuverability: The study evaluated the manipulation of both the endoscope and the EVS device. The findings suggest that mounting the EVS on the endoscope does not have a significant impact on the bending and maneuverability of the endoscope compared to when the EVS is not mounted. Cable management: No cable trapping of mucosal tissue between the cable and the endoscope occurred during use, or any other damage to the upper GIT occurred throughout the study. {12}------------------------------------------------ Verification and validation testing for the EVS was completed in accordance with EndoSound's Design and Control process in compliance with 21 CFR Part 820.30. Successful results of testing are included in the body of this submission. Potential risks of the EVS were identified according to ISO 14971. Mitigations were implemented and tested as part of the performance testing described above, with all mitigations verified and validated as documented in the attached risk management file. Technological differences from the predicate were shown not to result in any new issues of safety or efficacy according to the performance data submitted. # Conclusion Based on a comparison of physical characteristics, safety testing and animal studies performed, and conformance to consensus standards cited above and information provided in this Premarket notification, EndoSound considers the EVS to be as safe and effective and the performance is substantially equivalent to the predicate devices described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...