# FUJIFILM Ultrasonic Endoscope (K183433)

_Fujifilm Corporation · ODG · Sep 5, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODG/K183433

## Device Facts

- **Applicant:** Fujifilm Corporation
- **Product Code:** [ODG](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODG.md)
- **Decision Date:** Sep 5, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. These products are intended to be used with a FUJIFILM ultrasonic processor. These products are not intended for use on children and infants.

## Device Story

Ultrasonic endoscopes (EG-580UT/EG-580UR) provide ultrasonic imaging of upper GI tract submucosal and peripheral organs; used for observation, diagnosis, and endoscopic treatment. Device comprises control portion, insertion portion, and umbilicus. Insertion portion contains CCD image sensor, glass fiber bundles for illumination, and channels for air/suction/accessories (e.g., forceps). Operates by connecting to FUJIFILM video processors, light sources, and ultrasonic processors. Physician operates device in clinical setting; visual output displayed on monitor. Enables real-time visualization of internal structures to guide diagnostic or therapeutic procedures; benefits patient through minimally invasive assessment of GI tract.

## Clinical Evidence

Bench testing only. Performance evaluated against consensus standards including electrical safety (ANSI/AAMI ES 60601-1, IEC 60601 series), biocompatibility (ISO 10993), and endoscope-specific standards (ISO 8600). Additional performance testing confirmed specifications for field of view, resolution, bending capability, suction/air/water supply rates, and ultrasonic penetration depth. Reprocessing validation performed per AAMI TIR12/TIR30 and FDA guidance.

## Technological Characteristics

Components: control portion (G7), insertion portion, umbilicus. Materials: biocompatibility validated per ISO 10993-1, -5, -10. Imaging: CCD sensor, fiber optic illumination. Connectivity: interfaces with FUJIFILM video/ultrasonic processors. Standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, IEC 60601-2-18:2009, IEC 60601-2-37:2015, ISO 8600-1:2015, ISO 8600-3:1997, ISO 8600-4:2014.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 ([K181763](/device/K181763.md))

## Reference Devices

- FUJIFILM Endoscope Model EG-760R ([K172916](/device/K172916.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.

September 5, 2019

Fujifilm Corporation % Jeffrey Wan Specialist, Regulatory Affairs Fujifilm Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

Re: K183433

Trade/Device Name: FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODG, ITX, FDS Dated: July 26, 2019 Received: July 29, 2019

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Martha W. Betz, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K183433

Device Name

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR

Indications for Use (Describe)

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. These products are intended to be used with a FUJIFILM ultrasonic processor. These products are not intended for use on children and infants.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY

### FUJIFILM Corporation's FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR

Date: July 26, 2019

#### Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

#### Contact Person:

Jeffrey Wan Senior Regulatory Affairs Specialist Telephone: (201) 675-8947 E-Mail: jeffrey.wan@fujifilm.com

#### ldentification of the Proposed Devices:

| Proprietary/Trade Name: | FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UF |
|-------------------------|------------------------------------------------------|
| Common Name:            | Endoscope                                            |
| Device Class:           | Class II                                             |
| Review Panel:           | Gastroenterology/Urology                             |

#### Classification Information:

| Regulation Description           | CFR Section     | Product Codes |
|----------------------------------|-----------------|---------------|
| Endoscope and accessories        | 21 CFR 876.1500 | ODG, FDS      |
| Diagnostic ultrasonic transducer | 21 CFR 892.1570 | ITX           |

#### Predicate Devices:

- . FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 (K181763)
### Reference Devices:

- FUJIFILM Endoscope Model EG-760R (K172916) ●

{4}------------------------------------------------

### Intended Use / Indications for Use

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. These products are intended to be used with a FUJIFILM ultrasonic processor. These products are not intended for use on children and infants.

### Device Description

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are comprised of three general sections: a control portion, an insertion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion contains glass fiber bundles, several channels and a chargecoupled device (CCD) image sensor in its distal end. The insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged into the video processor, light source, and ultrasonic processor. The endoscopes are used in combination with FUJIFILM's video processors, light sources, ultrasonic processors and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K181763.

### Comparison of Technological Characteristics

FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR differ from the predicate devices EG-530UT2 and EG-530UR2 in terms of technological characteristics and materials. The subject and predicate devices share the same mode of operation and intended use.

A summary of major differences between the subject devices EG-580UR and the predicate devices EG-530UT2 and EG-530UR2 is provided as follows:

- . The subject device uses a G7 control portion, whereas the primary predicate device contains a G5 control portion.
- Compatibility with different accessories ●
- Minor material changes to the insertion portion 0

# Performance Data

Electrical safety of the subject device was evaluated using following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, IEC 60601-2-18:2009, and IEC 60601-2-37:2015.

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010. Biocompatibility testing was conducted in accordance with the FDA guidance, Use of International Standard ISO 10993-1, "Biological

{5}------------------------------------------------

evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", issued June 16, 2016.

Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015, ISO 8600-3:1997, and ISO 8600-4:2014.

Cleaning, high-level disinfection, and sterilization of the subject device were evaluated according to the following consensus standards: AAMI TIR12:2010, AAMI TIR30:2011. Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015.

Rate of balloon suction

The subject device met performance specifications in the following additional testing:

.

- Field of view
supply

- Rate of suction
- Resolution ●

- o Bending capability
- Rate of air supply
- Working length ●
- . Diameter of forceps channel
- o Rate of water supply ● Rate of balloon water
	- o
		- Viewing direction
- LG output ●
- Axial resolution
- . Lateral resolution
- o Penetration depth

# Substantial Equivalence

The subject devices FUJIFILM Ultrasonic Endoscopes EG-580UR are substantially equivalent to the predicate devices, FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 (K181763). The subject devices have intended use/indications for use and substantially similar technological characteristics and principles of operation as that of the predicate devices. Material changes to the predicate devices have been validated through biocompatibility testing. Thus, the subject devices FUJIFILM Ultrasonic Endoscopes EG-580UR are substantially equivalent to the predicate devices.

# Conclusions

The subject devices FUJIFILM Ultrasonic Endoscopes EG-580UR are substantially equivalent to the predicate devices based on the same intended use, indications for use, similar technological characteristics and materials. The differences in technological characteristics and materials between the subject and predicate devices raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject devices are substantially equivalent in performance to the predicate devices. The difference in materials between subject and predicate devices has been validated through biocompatibility testing. Thus, the subject devices FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are substantially equivalent to the predicate devices, FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 (K181763).

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODG/K183433](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODG/K183433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).