← Product Code [ODD](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODD) · K202583

# BioShield biopsy valve EUS - Linear (K202583)

_STERIS Corporation · ODD · Nov 6, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODD/K202583

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [ODD](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODD.md)
- **Decision Date:** Nov 6, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Intended Use

The single use BioShield® biopsy valve EUS is used to cover the biopsy/suction channel of flexible echoendoscopes. It provides access for endoscopic device passage and exchange, helps maintain insufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure and provides access for irrigation.

## Device Story

BioShield biopsy valve EUS - Linear is a single-use, disposable accessory for linear echoendoscopes. Device consists of a valve body and cap with a tether. It is manually applied to the echoendoscope's accessory port. During endoscopic procedures, the valve covers the biopsy/suction channel, allowing the physician to insert and exchange endoscopic instruments (e.g., biopsy forceps, injection needles) while maintaining insufflation and minimizing leakage of biomaterial from the port. It also provides an access point for irrigation. The device is used in clinical settings by healthcare professionals performing GI endoscopy. It does not use energy and is provided sterile or non-sterile.

## Clinical Evidence

Bench testing only. Performance testing included retention force, leakage testing under 5 psi pressure, and device exchange testing with biopsy forceps and injection needles. Biocompatibility testing was performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). Shelf life was validated via accelerated aging per ASTM F1980-16.

## Technological Characteristics

Materials: Thermoplastic Elastomer. Dimensions: 0.52" x 0.52" x 0.77". Energy source: None. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ASTM F1980-16 (accelerated aging), ISO 10993-5/10 (biocompatibility), AAMI/ANSI/ISO 11135-1:2014 (sterilization).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- BioShield ERCP Biopsy Valve ([K070420](/device/K070420.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

November 6, 2020

Steris Corporation Jacqueline Oliver Regulatory Affairs Specialist 5976 Heisley Rd Mentor, OH 44060

Re: K202583 Trade/Device Name: BioShield biopsy valve EUS - Linear Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODD Dated: September 3, 2020 Received: September 8, 2020

Dear Jacqueline Oliver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls' provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K202583

Device Name BioShield® biopsy valve EUS - Linear

### Indications for Use (Describe)

The single use BioShield® biopsy valve EUS is used to cover the biopsy/suction channel of flexible echoendoscopes. It provides access for endoscopic device passage and exchange, helps maintain insufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure and provides access for irrigation.

| Type of Use (Select one or both, as applicable)                                                       |                                                                                                      |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> | <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary For the BioShield® biopsy valve EUS - Linear

STERIS Corporation 5960 Heisley Road Mentor, OH 44060

Contact: Jackie Oliver Regulatory Affairs Specialist Tel: 440-358-6289 Email: Jackie_oliver@steris.com

September 1, 2020 Summary Date:

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

#### 1. Device Name

| Trade/Device Name: | BioShield biopsy valve EUS - Linear |
|--------------------|-------------------------------------|
| Regulation Number: | 21 CFR 876.1500                     |
| Regulation Name:   | Endoscope and accessories           |
| Regulatory Class:  | Class II                            |
| Product Code:      | ODD                                 |

#### Predicate Device 2.

K070420 BioShield ERCP Biopsy Valve

#### 3. Description of Device

The BioShield® biopsy valve EUS - Linear is an accessory to a linear echoendoscope. The EUS biopsy valve allows the end user to cover the accessory port of a linear echoendoscope. The BioShield® biopsy valve EUS - Linear consists of tether and cap. The biopsy valves' cap can be removed from the accessory port while being held on the echoendoscope via the tether to the cap.

The biopsy valve is a single-use, disposable device that is supplied sterile or nonsterile.

#### 4. Indications for Use

The single use BioShield® biopsy valve EUS is used to cover the opening to the biopsy/suction channel of flexible echoendoscopes. It provides access for endoscopic device passage and exchange, helps maintain insufflation, minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure and provides access for irrigation.

{5}------------------------------------------------

#### Technological Characteristic Comparison Table 5.

The BioShield® biopsy valve EUS - Linear is similar in design to the predicate and has similar intended uses. The differences between the proposed and predicate devices are the design of the valves and the particular type of endoscope on which the devices are used. The design is different than the predicate (see table below) because the BioShield® biopsy valve EUS - Linear is used on linear echoendoscopes that have different dimensions than the endoscopes that are used with the predicate device. These differences do not raise any new concerns of safety and effectiveness when compared to the predicate device.

Table 1 summarizes the similarities/differences between the proposed device and the predicate.

| Feature                          | BioShield® biopsy valve EUS -<br>Linear (Proposed)                                                                                                                                                                                                                                                                                                                                              | BioShield ERCP Valve<br>(Predicate K070420)                                                                                                                                                                                                                                                                                                                                                                                            | Comparison |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for use              | The single use BioShield®<br>biopsy valve EUS is used to<br>cover the opening to the<br>biopsy/suction channel of<br>flexible echoendoscopes. It<br>provides access for endoscopic<br>device passage and exchange,<br>helps maintain insufflation,<br>minimizes leakage of<br>biomaterial from the biopsy<br>port throughout the endoscopic<br>procedure and provides access<br>for irrigation. | The single use BioShield®-<br>ERCP biopsy valve is used to<br>cover the opening to the biopsy/<br>suction channel of Olympus and<br>G5 and newer Fujinon<br>gastrointestinal endoscopes. It<br>provides access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation,<br>minimizes leakage of biomaterial<br>from the biopsy port throughout<br>the endoscopic procedure, and<br>provides access for irrigation. | Similar    |
| Construction                     | Valve Body<br>Valve Cap                                                                                                                                                                                                                                                                                                                                                                         | Valve Body<br>Valve Body Insert<br>Valve Cap                                                                                                                                                                                                                                                                                                                                                                                           | Similar    |
| Sterile/Non-sterile              | Sterile & Non-sterile                                                                                                                                                                                                                                                                                                                                                                           | Sterile & Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                  | Same       |
| Sterilization<br>Method          | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                  | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                         | Same       |
| Sterilization<br>Assurance Level | 10-6                                                                                                                                                                                                                                                                                                                                                                                            | 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same       |
| Usage                            | Single use                                                                                                                                                                                                                                                                                                                                                                                      | Single use                                                                                                                                                                                                                                                                                                                                                                                                                             | Same       |
| Materials                        | Thermoplastic Elastomer                                                                                                                                                                                                                                                                                                                                                                         | Thermoplastic Elastomer                                                                                                                                                                                                                                                                                                                                                                                                                | Same       |

### Table 1. Proposed/Predicate Device Technological Characteristics Comparison Table

{6}------------------------------------------------

| STERIS Traditional 510(k) PREMARKET NOTIFICATION<br>BioShield® biopsy valve EUS – Linear | K202583<br>Page 4 of 7 |
|------------------------------------------------------------------------------------------|------------------------|
|------------------------------------------------------------------------------------------|------------------------|

| Feature                  | BioShield® biopsy valve EUS -<br>Linear (Proposed) | BioShield ERCP Valve<br>(Predicate K070420)                      | Comparison |
|--------------------------|----------------------------------------------------|------------------------------------------------------------------|------------|
| Device<br>Dimensions     | Length: 0.52"<br>Width: 0.52"<br>Height: 0.77"     | Length: 0.87"<br>Width w/side-tabs 0.88"<br>Height: 0.86"        | Similar    |
| Target<br>Population     | GI Endoscopic procedures                           | GI Endoscopic procedures                                         | Same       |
| Energy<br>Used/Delivered | None                                               | None                                                             | Same       |
| Method of<br>Application | Manual application                                 | Manual application                                               | Same       |
| Compatible<br>Endoscopes | Linear Echoendoscopes                              | Olympus & G5 and newer<br>Fujinon gastrointestinal<br>endoscopes | Similar    |
| Packaging                | Polybag (non-sterile)<br>Peel Pouch (sterile)      | Polybag (non-sterile)<br>Peel Pouch (sterile)                    | Same       |

#### Description of Safety and Substantial Equivalence 6.

Testing for the proposed biopsy valves, both sterile and non-sterile showed they were comparable to the predicate devices in maintaining insufflation, allowing for device exchange and minimizing leakage from the biopsy port. The proposed devices met all acceptance criteria of the verification testing. Performance testing includes:

### Bench Testing Summary

Performance testing was conducted to demonstrate that the proposed BioShield biopsy valve EUS are comparable to the predicate K070402 devices.

A summary of the testing conducted is provided in the table below:

{7}------------------------------------------------

| Testing                       | Acceptance<br>Criteria                                                                                                                                                                                                |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Retention Force<br>Testing    | The retention force of the BioShield EUS biopsy valve -<br>Linear (sterile & non-sterile) must be equal to or greater than<br>the predicate device.                                                                   |
| Leakage Testing               | The BioShield biopsy valve EUS - Linear (sterile & non-<br>sterile) must be comparable to or better than the predicate<br>device. The device(s) shall not<br>spray or experience geyser type leaks while pressurized. |
| Device<br>Exchange<br>Testing | The BioShield biopsy valve EUS - Linear should allow<br>device passage and should not become detached during<br>instrument usage and exchange.                                                                        |
| Shelf Life Testing            | The proposed devices were accelerated aged for a period<br>equal to one year of real-time aging prior to conducting the<br>testing provided in this submission.                                                       |

Summary of Tests Conducted for the BioShield biopsy valve EUS - Linear

Provided below is a detailed summary of the testing activities.

### Retention Force Testing

Objective: This test was intended to determine the force required to remove the proposed devices, the BioShield Biopsy Valves EUS - Linear (sterile & nonsterile) from the echoendoscope and compare those results to the predicate device.

Sample: Fifteen BioShield biopsy valves EUS - Linear (non-sterile) from 3 different lots (5 from each lot) and 15 BioShield biopsy valves EUS - Linear (sterile) from 3 different lots (5 from each lot) were used for testing.

Procedure: The testing procedure included the use of a calibrated force gauge to measure tensile force in pounds. The BioShield biopsy valve EUS - Linear (nonsterile) was attached to the inlet port of an echoendoscope. The valve was pulled off the endoscope and the force required to remove the biopsy valve from the linear echoendoscope was measured. This procedure was repeated for the remaining non-sterile valves and the sterile valves.

Standards utilized: None

{8}------------------------------------------------

### Leakage Testing

Objective: To establish that the proposed devices, the BioShield Biopsy Valves EUS - Linear (sterile & non-sterile), do not leak when attached to a pressurized echoendoscope.

Sample: Fifteen BioShield biopsy valves EUS - Linear (non-sterile) from 3 different lots (5 from each lot) and 15 BioShield biopsy valves EUS - Linear (sterile) from 3 different lots (5 from each lot) were used for testing.

Procedure: Attach the BioShield biopsy valve EUS - Linear (non-sterile) to the inlet port of the echoendoscope. Affix a pressure tube assembly at the end of the linear echoendoscope and introduce 5 psi of water pressure up the accessory channel. Verify that leaks do not occur at the biopsy valve cap while the accessory channel is pressurized. Repeat this procedure for the remaining nonsterile valves and the sterile valves.

Standards Utilized: None

### Device Exchange Testing

Objective: To establish that the proposed devices, the BioShield Biopsy Valves EUS - Linear (sterile & non-sterile), allow for the passage of endoscopic devices of varying sizes.

Sample: Fifteen BioShield biopsy valves EUS - Linear (non-sterile) from 3 different lots (5 from each lot) and 15 BioShield biopsy valves EUS - Linear (sterile) from 3 different lots (5 from each lot) were used for testing.

Procedure: Secure a linear echoendoscope to the lab bench in a tortuous configuration. Attach the BioShield Biopsy Valve EUS – Linear (non-sterile) to the inlet port of the echoendoscope. Insert and remove the below listed devices. Test insertion and removal of the following device types: biopsy forceps and injection needles.

Standards Utilized: None

### Shelf Life Testing (Accelerated aging)

Tested in accordance with Standard: ASTM F1980-16 - Standard Guide For Accelerated Aging Of Sterile Barrier Systems for Medical devices.

### Biocompatibility Testing:

Cytotoxicity: Tested in accordance with ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity

Irritation: Tested in accordance with ISO 10993-10: 2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization

Sensitization: Tested in accordance with ISO 10993- 10: 2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization

{9}------------------------------------------------

## Sterilization

Tested in accordance with AAMI/ANSVISO 11135-1:2014, Sterilization of health care products - Ethylene Oxide - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical device

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODD/K202583](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/ODD/K202583)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
