Disposable Endoscope Valves Set A

K151714 · Wilson Instruments (Shanghai) Company Limited · ODC · Dec 3, 2015 · Gastroenterology, Urology

Device Facts

Record IDK151714
Device NameDisposable Endoscope Valves Set A
ApplicantWilson Instruments (Shanghai) Company Limited
Product CodeODC · Gastroenterology, Urology
Decision DateDec 3, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

A collection of sterile device intended to be fitted to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of fluids, gases, and other materials. It typically includes a suction valve, an air/water valve, and a biopsy valve. This is a single-use device.

Device Story

Disposable Endoscope Valves Set A comprises three sterile, single-use components: suction valve, air/water valve, and biopsy valve. Designed for attachment to endoscope working channels/ports; enables operator control of channel functions; prevents retrograde flow of fluids/gases. Used in clinical endoscopic procedures; operated by physicians/endoscopy staff. Provides mechanical control of endoscope ports; facilitates procedure workflow by allowing convenient replacement of valves. Benefits patient by maintaining sterile environment and ensuring proper endoscope function during diagnostic or therapeutic procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility, sterility, and performance testing to demonstrate substantial equivalence to predicate devices.

Technological Characteristics

Sterile, single-use endoscope accessory set. Includes suction, air/water, and biopsy valves. Designed for compatibility with standard endoscope working channels. Sterilization method: Ethylene Oxide (EO).

Indications for Use

Indicated for use by endoscope operators to control working channel/port functions and prevent retrograde flow of fluids, gases, and materials during endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 10, 2015 Wilson Instruments (Shanghai) Company Limited Lijuan Zhang Regulatory Affairs Manager Building 5, No.258 Shuangbang Road Xujing Town, Qingpu District Shanghai China K151714 Re: > Trade/Device Name: Disposable Endoscope Valves Set A Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODC Dated: November 6, 2015 Received: November 10, 2015 Dear Lijuan Zhang, This letter corrects our substantially equivalent letter of December 3, 2015. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the word "WILSON" in large, bold, purple letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. # 008_Indications for Use {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "WILSON" in large, bold, purple letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame, with a white background. ### Indications for Use K151714 510(k) Number (if known): Device Name: Indications for Use: Disposable Endoscope Valves Set A A collection of sterile device intended to be fitted to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of fluids, gases, and other materials. It typically includes a suction valve, an air/water valve, and a biopsy valve. AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) Prescription Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) This is a single-use device. Concurrence of CDRH, Office of In Vitro Diagnostic Devices and Radiological Health (OIR) > (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k) _ {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "WILSON" in large, bold, purple letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame. ## 510(k) Summary [As required by 21 CFR 807.92] #### 1. Date Prepared [21 CFR807.92 (a) (1)] Jun 23th, 2015 #### 2. Submitter's Information [21 CFR807.92 (a) (1)] | Name of Sponsor: | Wilson Instruments (Shanghai) Company Limited | |------------------|----------------------------------------------------------------------------------| | Address: | Building 5, No.258 Shuangbang Road, Xujing Town<br>Qingpu DistrictShanghai China | | Contact Name: | Xin HUANG | | Telephone No.: | +86 (021) 39290005 | | Fax No.: | +86 (021) 66311472 | | Email Address: | xhuang@heyinovo.com | #### 3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)] | Trade Name: | Disposable Endoscope Valves Set A | |----------------------------|------------------------------------------| | Model Name: | VO series, M series | | Common Name: | Endoscope Channel Accessory | | Regulatory Classification: | 21 CFR876.1500 Endoscope and Accessories | | Product Code: | ODC | | Classification Panel: | Gastroenterology/Urology | | Device Class: | II | #### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicates within this submission are as follows: OLYMPUS CORP. OLYMPUS DISPOSABLE SUCTION VALVE has been cleared by FDA through 510(k) No.K920025 (Decision Date -03/24/1992). OLYMPUS CORP. OLYMPUS DISPOSABLE BIOPSY VALVE has been cleared by FDA through 510(k) No.K911412 (Decision Date —06/28/1991). THE OLYMPUS OPTICAL CO. EVIS EXERA COLONOVIDEOSCOPE CF-Q160 AL/I AND PCF-160 AL/I(MH-438) have been cleared by FDA through 510(k) No.K001241 (Decision Date -05/09/2000). {5}------------------------------------------------ #### 5. Description of the Device [21 CFR 807.92(a)(4)] Wilson's Disposable Endoscope Valves Set A collects three types of valve products into one package unit. Disposable Endoscope Valves Set A can make it convenient for the replacement of Air/Water Valves, Suction Valves, Biopsy Valve during an endoscopic procedure. And these three types' valves can realize their intended uses respectively. #### 6. Intended Use [21 CFR 807.92(a)(5)] A collection of sterile device intended to be fitted to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of fluids, gases, and other materials. It typically includes a suction valve, an air/water valve, and a biopsy valve. This is a single-use device. #### 7. Technological Characteristics [21 CFR 807.92(a)(6)] As the reason that the working situation, environment and the intended use of Wilson's Disposable Endoscope Valves Set A are the same as the intended uses' combination from the similar Olympus Disposable Suction Valve, Olympus Disposable Biopsy Valve and EVIS EXERA COLONOVIDEOSCOPE CF-Q160 AL/I AND PCF-160 AL/I(MH-438), the technological characteristics of this product series are designed to make same as that of the equivalence product structure for compatible endoscope, and the application of materials over different parts of the product series are also be designed to be equal or better respectively and etc. It applies EO sterilization method, which is also same as that of SE product. #### 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92] The Disposable Endoscope Valves Set A of Wilson has taken the biocompatibility, sterility and performance testing into concern in accordance to Food and Drug Administration related guidance and recognized international standards. Test data and report information are included in this submission. #### 9. Conclusion [21 CFR 807.92(b) (3)] In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Wilson Instruments (Shanghai) Company Limited concludes that Disposable Endoscope Valves Set A is substantially equivalent to predicate devices with regard to safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...