ENDOCHOICE GRASPING INSULATED AND ALIGATOR JAW FORCEPS MODEL AMHGFA, ENDOCHOICE GRASPING INSULATED AND RAT TOOTH FORCEPS

K101298 · Endochoice, Inc. · OCZ · Aug 5, 2010 · Gastroenterology, Urology

Device Facts

Record IDK101298
Device NameENDOCHOICE GRASPING INSULATED AND ALIGATOR JAW FORCEPS MODEL AMHGFA, ENDOCHOICE GRASPING INSULATED AND RAT TOOTH FORCEPS
ApplicantEndochoice, Inc.
Product CodeOCZ · Gastroenterology, Urology
Decision DateAug 5, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

Grasping forceps are intended to be used to grasp tissue and / or retrieve foreign bodies, excised tissue, and stents from within the gastrointestinal tract.

Device Story

EndoChoice Grasping Forceps are endoscopic accessories used to grasp tissue or retrieve foreign bodies, excised tissue, and stents from the gastrointestinal tract. The device consists of a flexible shaft and a proximal control handle. The operator (physician) uses the proximal handle to actuate the distal tip grasping jaws under endoscopic visualization. The device is used in clinical settings during endoscopic procedures to facilitate tissue management or foreign body removal, potentially benefiting the patient by enabling minimally invasive retrieval or biopsy.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Flexible shaft with proximal control handle; distal tip grasping jaws; manual actuation; endoscopic accessory; non-powered.

Indications for Use

Indicated for use in the gastrointestinal tract to grasp tissue, retrieve foreign bodies, retrieve excised tissue, and retrieve stents.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 101298 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92. | Submitter Name and<br>Address: | EndoChoice, Inc.<br>11800 Wills Rd, Suite 100<br>Alpharetta, GA 30009<br>AUG - 5 2010 | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Theron Gober<br>Quality and Regulatory Manager | | Phone Number: | 678-534-6021 | | Fax Number: | 770-410-9008 | | Establishment<br>Registration Number: | 300759133 | | Date Prepared: | April 16, 2010 | | Device Trade Name(s): | EndoChoice Grasping Forceps: | | Device Common Name: | Grasping forceps | | Classification Name: | KOG - Endoscope and accessories | | Predicate Device(s): | Olympus FG – 4/14/20/21/22/26/32/38/40/##-1 | | General Device<br>Description: | The EndoChoice grasping forceps are designed for grasping<br>tissue and / or retrieving foreign body, and excised tissue<br>under endoscopic visualization. These forceps consist of a<br>flexible shaft and a proximal control handle. Operation of the<br>proximal control handle actuates the distal tip grasping jaws. | | Intended Use: | Grasping forceps are intended to be used to grasp tissue and<br>/ or retrieve foreign bodies, excised tissue, and stents from<br>within the gastrointestinal tract. | | Technological<br>Characteristics: | From a clinical perspective and comparing design<br>specifications, the EndoChoice grasping forceps and the<br>predicate device are substantially equivalent. Based on the<br>technological characteristics and overall performance of the<br>devices, EndoChoice, Inc. believes that no significant<br>differences exist between the proposed grasping forceps and<br>the predicate devices. | Description {1}------------------------------------------------ K 101298 pg 2 of 2 EndoChoice, Inc. believes the minor differences of the grasping forceps and its predicates device should not raise any concerns regarding the overall safety or effectiveness. Performance Data: No performance standards exist for this device. Conclusion: Based on the technological characteristics and overall performance of the devices, EndoChoice, Inc. believes that the grasping forceps and the predicate device selected are substantially equivalent and that the differences between the devices are minor which do not raise new issues of safety or effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission of protecting the health of all Americans and providing essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Mr. Theron Gober Quality and Regulatory Manager EndoChoice, Inc. 11800 Wills Road, Suite 100 ALPHARETTA GA 30009-2089 Re: K101298 Trade/Device Name: EndoChoice Grasping Forceps Regulation Number: 21 CFR 8876.1500 Regulation Name: Endoscopes and accessories Regulatory Class: II Product Code: OCZ Dated: May 7, 2010 Received: May 11, 2010 000 - 5 - 2010 ## Dear Mr.Gober: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {3}------------------------------------------------ Page 2 adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Huker Lemur MD Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health. ## Enclosure {4}------------------------------------------------ K101298 ## INDICATIONS FOR USE **510(k) Number (if known):** K101298 AUG - 5 2010 Device Name: Grasping Forceps Indications for Use Grasping forceps are intended to be used to grasp tissue and / or retrieve foreign bodies, excised tissue, and stents from within the gastrointestinal tract. . (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) X Prescription Use_ · (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...