ERBEFLO 2 Disposable Tubing System

K190469 · Erbe USA, Inc. · OCX · May 23, 2019 · Gastroenterology, Urology

Device Facts

Record IDK190469
Device NameERBEFLO 2 Disposable Tubing System
ApplicantErbe USA, Inc.
Product CodeOCX · Gastroenterology, Urology
Decision DateMay 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ERBEFLO® 2 Disposable Tubing System is intended to provide sterile water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures.

Device Story

The ERBEFLO® 2 Disposable Tubing System is a sterile, disposable accessory set used during endoscopic procedures. It consists of tubing, port connectors, channel adapters, and back-flow valves. Clinicians connect the system to a sterile water source and an irrigation pump to deliver water through an endoscope channel. The device facilitates irrigation to improve visualization or clear debris during endoscopy. The system includes redundant back-flow protection (check valves) to prevent cross-contamination between patients. It is intended for use by clinicians in a clinical setting. The device is provided sterile and is designed for single-patient or 24-hour use depending on the specific component configuration.

Clinical Evidence

No clinical data. Safety and efficacy were demonstrated through bench testing, including biological evaluation (biocompatibility), 2X sterilization functional testing (visual inspection, flow, back-flow pressure, tensile strength, durability, connection testing), packaging integrity, ethylene oxide residual testing, barrier integrity/simulated use testing, and human factors/usability engineering testing.

Technological Characteristics

Materials include medical-grade plastics, brass, and adhesives. Components include tubing sets, port connectors, channel adapters, and back-flow valves. Connectivity is mechanical (luer-style connections per ISO 594-1/2). Sterilization is via ethylene oxide. Standards applied: ISO 10993-1, ISO 594-1/2, ISO 15223-1, ISO 11607-1/2, ISO 11135-1, and ISO 10993-7.

Indications for Use

Indicated for use in endoscopic procedures to provide a sterile water source through an irrigation pump and/or endoscope for irrigation.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 23, 2019 Erbe USA, Inc. John Tartal Director of Quality and Regulatory Affairs 2225 Northwest Parkway Marietta. GA 30067 Re: K190469 Trade/Device Name: ERBEFLO® 2 Disposable Tubing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: February 23, 2019 Received: February 26, 2019 Dear John Tartal: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For: Shani Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190469 Device Name ERBEFLO® 2 Disposable Tubing System Indications for Use (Describe) The ERBEFLO® 2 Disposable Tubing System is intended to provide sterile water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div> <input checked="true" type="checkbox"/> Residential Use (CA STC 400 Solar Only) Please </div> | <div> <input type="checkbox"/> On-Site Consumption (CA STC 400 </div> | |------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------| |------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY | Submitted By: | Erbe USA, Inc.<br>2225 Northwest Parkway<br>Marietta, GA 30067<br>Tel: 770-955-4400<br>Fax: 770-955-2577 | |---------------------------------------|----------------------------------------------------------------------------------------------------------| | Contact Person: | John Tartal<br>Director of Quality and Regulatory Affairs | | Date Prepared: | February 23, 2019 | | Common Name: | Endoscopic Irrigation Tubing System | | Trade/Proprietary Names: | ERBEFLO® 2 Disposable Tubing System | | Classification Name: | Endoscopes and Accessories (21 CFR Part 876.1500) | | Regulatory Class: | II | | Product Code: | OCX | | Legally Marketed<br>Predicate Device: | ERBEFLO® 2 Disposable Tubing System, 510(k) Number K103235 | ### Device Description: The ERBEFLO® 2 Disposable Tubing System products are manufactured with medical grade materials or agents used in the medical device industry such as plastics, brass, adhesive, etc. The devices consist of tubing sets, port connectors, channel adapter, and back flow valve to deliver sterile water source through designated pumps and to/or through a scope channel of various endoscopes for irrigation in endoscopy procedures. Clinicians connect the associated products to a water source (i.e., a sterile water bottle) and then to a designated pump and endoscope. The ERBEFLO® 2 Disposable Tubing System are provided sterile and are disposable. ### Intended Use: The ERBEFLO® 2 Disposable Tubing System is intended to provide sterile water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures. ## Similarities and Differences of the Proposed Device (Predicate Comparison/Substantial Equivalence): ### Similarities The ERBEFLO® 2 Disposable Tubing System products in this 510(k) are essentially the same (i.e., materials, construction, dimensionally, etc.) as the products in the previously cleared 510(k). The intended use, packaging, sterilization, use conditions, disposability, etc. are the same. {4}------------------------------------------------ February 2019 ### Differences The ERBEFLO® 2 Disposable Tubing System Notes On Use's (i.e., the Labeling) Precautions and Warnings section is being updated to better inform users of potential hazards. Additionally, a back flow (check) valve is being added to the 24 hour use Endoscopy Port Connectors for Olympus® and Pentax® Scopes [Note: When Erbe scope accessories (i.e., connectors, channel adapter, and back flow valve) are used with Erbe tubing sets there will be two (2) back flow (check) valves. The redundant back flow protection features with Erbe products will provide an additional layer of protection to prevent back flow from occurring that could cause possible patient cross contamination and infection.]. These measures are being taken to warn against the use of Erbe products with other manufacturer's tubing products that are not specified in Erbe's Notes On Use (NOU), provide more clarity involving the importance of priming the scope channel and maintaining the water barrier in the scope channel, as well as mitigate risk if Connectors are used "off label" (i.e., use of non-Erbe tubing products not specified in Erbe's NOU). Fundamentally from a product design prospective, the single use Endoscopy Port Connector w/Valve for Olympus® Scopes submitted/cleared in the original 510(k) is also being packaged and labeled as the 24 hour use Endoscopy Port Connector for Olympus® Scopes and the single use Endoscopy Port Connector w/Valve for Pentax® Scopes submitted/cleared in the original 510(k) is also being packaged and labeled as the 24 hour use Endoscopy Port Connector for Pentax® Scopes (i.e., there is no product design change that hasn't already been reviewed/cleared. The changes are a designation/labeling change for both Connectors.). ### Evaluations and Testing: The following evaluations and tests were performed to demonstrate safety and efficacy. Biological Evaluation The evaluation was performed per the current recognized standard and demonstrated that there were no biocompatibility issues with the materials used for the ERBEFLO® 2 Disposable Tubing System. 2X Sterilization Functional Testing Visual inspection, flow testing, back flow pressure testing, tensile strength testing, durability testing, and connections testing demonstrated that the devices upon 2X sterilization met established performance specifications. ERBEFLO® 2 Endoscopy Port Connectors for Pentax® Scopes 2X Sterilization Functional Testing Visual inspection and connections testing upon aging demonstrated that the devices upon 2X sterilization met established performance specifications. Packaqing Evaluation The evaluation demonstrated the adequacy and integrity of the packaging for the products. Sterilization Evaluation The evaluation was performed using current recognized standards to demonstrate product sterility as well as that the products met ethylene oxide residual requirements. ### Barrier Integrity and Simulated Use Testing The testing showed that Erbe port connectors when used with an Erbe tubing set provides sufficient back flow protection and verified 24 hour multi-patient use. Human Factors Engineering/Usability Engineering Testing The evaluation testing demonstrated that Erbe's port connectors can be used as intended by following ERBEFLO® 2 Disposable Tubing System's Notes On Use. Applied Standards AAMI / ANSI / ISO 10993-1, ISO 594-1, ISO 594-2, AAMI / ANSI / ISO 15223-1, AAMI / ANSI / ISO 11607-1, AAMI / ANSI / ISO 11607-2, AAMI / ANSI / ISO 11135(-1), AAMI / ANSI / ISO 10993-7 {5}------------------------------------------------ February 2019 ### Conclusion: The ERBEFLO® 2 Disposable Tubing System is safe and efficacious.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...