← Product Code [OCX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX) · K161167

# Seal Single-Use Biopsy Valve (K161167)

_Endochoice, Inc. · OCX · May 26, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K161167

## Device Facts

- **Applicant:** Endochoice, Inc.
- **Product Code:** [OCX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX.md)
- **Decision Date:** May 26, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation.

## Device Story

The Seal Single-Use Biopsy Valve is a mechanical accessory for endoscopes. It consists of a cylindrical base, a pre-perforated cap, and a flexible internal diaphragm. The device attaches to the endoscope's working channel port. During procedures, clinicians insert surgical instruments through the cap and diaphragm; the elastomer diaphragm deforms around the instrument to create a seal, preventing fluid leakage and maintaining insufflation. The cap allows for pressure adjustment by venting gases. The device is used in clinical settings by physicians or trained staff to facilitate instrument exchange and irrigation while maintaining a closed system. It benefits patients by preventing leakage of biomaterials and maintaining necessary insufflation during endoscopic interventions.

## Clinical Evidence

Bench testing only. Performance testing included pressure tests (holding 10 PSI for 1 minute) with the valve closed, with an instrument inserted, and following instrument removal. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous injection, and systemic injection tests in accordance with 21 CFR Part 58 and ISO 10993-1.

## Technological Characteristics

Thermoplastic elastomer construction. Mechanical diaphragm seal. Dimensions: 7.1 mm inner diameter, accommodates instruments up to 3.2 mm. Compatible with Olympus (160, 180, 190), Fujinon (530, 590, 600), and EndoChoice Fuse endoscopes. Non-sterile, single-use. No energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Single-Use Biopsy Valve ([K133734](/device/K133734.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, which is a symbol often associated with healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 26, 2016

Endochoice, Inc. Daniel Hoefer Manager, Regulatory Affairs 11810 Wills Road Alpharetta, GA 30009

Re: K161167 Trade/Device Name: Seal Single-use Biopsy Valve Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: April 22, 2015 Received: April 26, 2016

Dear Daniel Hoefer,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

**Herbert P. Lerner -S**

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161167

Device Name Seal Biopsy Valve

Indications for Use (Describe)

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

#### Seal Single-Use Biopsy Valve

#### May 26, 2016

#### 1. Company Identification

EndoChoice, Inc. 11810 Wills Road Alpharetta, GA 30009 Telephone: 678-708-4743 Fax: 678-567-8218 Establishment Registration: 3007591333

#### 2. Contact Person

Daniel Hoefer Regulatory Affairs Manager

#### 3. Device Name

| Commercial Name:     | Seal® Single-Use Biopsy Valve         |
|----------------------|---------------------------------------|
| Common/Usual Name:   | Biopsy Valve                          |
| Classification Name: | Endoscopic, irrigation/suction system |

## 4. Device Classification

| Product Code:      | OCX                      |
|--------------------|--------------------------|
| Regulation Number: | 876.1500                 |
| Class:             | II                       |
| Review Panel:      | Gastroenterology/Urology |

#### 5. Intended Use:

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation.

#### 6. Device Description:

The Seal "Single-Use Biopsy valve is provided non-sterile, and is composed of a cylindrical base and a connected cap. The Single-Use Biopsy Valve provides sealable access via the working channel port of an endoscope for surgical instruments. The valve consists of an attached cap and valve body with a circular opening. The cap is pre-perforated, and ensures instrument access and removal is tight and leak-free.

{4}------------------------------------------------

The valve is designed with a hollow body with a distal end that releasably attaches to the inlet port of the working channel. The valve body provides a flexible diaphragm seal that separates the body into a proximal chamber and distal chamber. The distal chamber secures onto the endoscope, while the proximal chamber secures the cap. Instruments are inserted through the cap and chambers, through the diaphragm seal, and into the endoscope. The diaphragm seal is configured such that a seal is formed due to deformation of the elastomer around an instrument when it is inserted through the opening, preventing fluids from passing through the biopsy valve.

The cap serves two primary functions. One function of the cap is to provide an additional seal around surgical instruments to prevent leakage. The second function is to allow modification of pressure during insufflation by opening the cap, allowing gasses to flow from the endoscope, and out of the biopsy valve.

# 7. Substantial Equivalence:

The device submitted for review is a modification of the Biopsy Valve (K133734). Changes to the device include a modification in material. Both the modified and unmodified device is composed of thermoplastic elastomers. The modified device also has a reduction in overall height of the biopsy valve to provide an improved fit between the cap and body, as well as the valve to the instrument port.

The modified device is identical in terms of intended use, operating principle, performance, technology, energy used, and packaging.

| Characteristic             | Single-Use Biopsy Valve<br>(Unmodified)                                                                                                                                                                                                           | Single-Use Biopsy Valve<br>(Modified)                                                                                                                                                                                                             |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number              | K133734                                                                                                                                                                                                                                           | Pending                                                                                                                                                                                                                                           |
| Indications for Use        | The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation. | The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation. |
| Material                   | Thermoplastic elastomer                                                                                                                                                                                                                           | Thermoplastic elastomer                                                                                                                                                                                                                           |
| Valve Inner<br>Diameter    | 7.1 mm                                                                                                                                                                                                                                            | 7.1 mm                                                                                                                                                                                                                                            |
| Biocompatiblity            | Tested in accordance with ISO<br>10993-1                                                                                                                                                                                                          | Tested in accordance with ISO<br>10993-1                                                                                                                                                                                                          |
| Endoscope<br>compatibility | Olympus series 160, 180, and 190.<br>Fujinon series 530, 590, and 600.<br>EndoChoice Fuse gastrointestinal                                                                                                                                        | Olympus series 160, 180, and 190.<br>Fujinon series 530, 590, and 600.<br>EndoChoice Fuse gastrointestinal                                                                                                                                        |

{5}------------------------------------------------

| Characteristic                 | Single-Use Biopsy Valve<br>(Unmodified)                                                                                | Single-Use Biopsy Valve<br>(Modified)                   |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
|                                | endoscopes                                                                                                             | endoscopes.                                             |
| Removable cap<br>with slit     | Yes                                                                                                                    | Yes                                                     |
| Slit<br>accommodate<br>devices | Up to 3.2 mm                                                                                                           | Up to 3.2 mm                                            |
| Packaging                      | Individually packaged plastic pouch<br>or sterile pouch                                                                | Individually packaged plastic pouch<br>or sterile pouch |
| Performance<br>testing:        | Design verification Pressure test with<br>valve closed - the device holds pressure<br>for 1 Min at 10 PSI              | Same                                                    |
|                                | Design verification Pressure test with<br>instrument inserted - the device holds<br>pressure for 1 Min at 10 PSI       | Same                                                    |
|                                | Design verification Pressure test:<br>Following removal of instrument the<br>device holds pressure for 1 Min at 10 PSI | Same                                                    |

# 8. Non-Clinical Testing:

The modified device has undergone both bench testing of performance and laboratory biocompatibility testing for cytotoxicity, sensitization, intracutaneous injection test, and system injection test, in accordance with 21 CFR, Part 58. The modified Single-Use Biopsy Valve results in no safety or efficacy concerns regarding biocompatibility or performance. Likewise, in conformance with 21 CFR 807.92(b)(3), the modified device performs as well as the predicate in all testing performed.

## 9. Conclusion:

The modified Single-Use Biopsy Valve is substantially equivalent to the unmodified predicate device listed above in performance, technical characteristics, biocompatibility, and intended use.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K161167](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/OCX/K161167)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
